← Product Code [GEI](/productcode/GEI) · K141507

# TWO SKIN TREATMENT SYSTEM (K141507)

_Syneron Medical, Ltd. · GEI · Dec 8, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141507

## Device Facts

- **Applicant:** Syneron Medical, Ltd.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Dec 8, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The eTwo Skin Treatment System is intended for dermatological procedures. The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin. the Sublative RF applicator is limited to skin types I-IV. The Sublime applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm3, the Sublime applicator is limited to skin types I-IV.

## Device Story

Mobile dermatological system; utilizes two applicators for skin treatment. Sublative RF applicator delivers bipolar radiofrequency (RF) via multi-electrode pin array to heat skin, causing ablation and resurfacing. Sublime applicator combines infrared (IR) light and bipolar RF to bulk-heat dermis, stimulating collagen for wrinkle reduction. Operated by clinicians in professional settings. Output is thermal energy delivery to target tissue. Healthcare providers use device to improve skin appearance and reduce wrinkles. Benefits include non-invasive or minimally invasive skin rejuvenation with demonstrated safety and efficacy profiles.

## Clinical Evidence

Prospective multi-center study of 72 subjects (mean age 49, mostly female/Caucasian, Fitzpatrick types II-III). Primary endpoint: wrinkle reduction at 12 weeks post-treatment using Fitzpatrick Wrinkle and Elastosis Scale. Results: 71% of subjects achieved wrinkle reduction; 77% showed at least 1-unit improvement on Fitzpatrick Scale; 91% showed overall facial appearance improvement. Safety: 6.9% (5/72) device-related adverse events, all mild/moderate and resolved. No serious adverse events reported.

## Technological Characteristics

Mobile system; bipolar RF and IR light sources. Sublative RF uses multi-electrode pin arrays (including 44-pin tip). Sublime uses IR light and bipolar RF. Electrical safety/EMC per IEC 60601-1 and 60601-1-2. Sterilization provides SAL 10^-6. Software controls energy delivery parameters.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Syneron eTwo Skin Treatment System ([K110672](/device/K110672.md))
- EndyMed Medical, Ltd.'s Glow ([K120513](/device/K120513.md), [K083461](/device/K083461.md), [K101510](/device/K101510.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 8, 2014

Syneron Medical Ltd. c/o Ms. Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103

Re: K141507

Trade/Device Name: eTwo Skin Treatment System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical, Cutting & Coagulation & Accessories Regulatory Class: Class II Product Code: GEI Dated: November 4, 2014 Received: November 4, 2014

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

510(k) Number (if known) K141507

Device Name

eTwo Skin Treatment System

Indications for Use (Describe)

The eTwo Skin Treatment System is intended for dermatological procedures.

The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin. the Sublative RF applicator is limited to skin types I-IV.

The Sublime applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm3, the Sublime applicator is limited to skin types I-IV.

Type of Use (Select one or both, as applicable)

Z Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number. "

FORM FDA 3881 (1/14)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EF

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

□ Over-The-Counter Use (21 CFR 801

{3}------------------------------------------------

#### 510(k) SUMMARY Syneron Medical Ltd.'s eTwo Skin Treatment System

#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Syneron Medical Ltd. P.O.B 550 Industrial Zone Tavor Building, Yokneam Illit 20692, Israel

Tel: +972 (0) 732-442-200 Fax: +972 (0) 732-442-202

Contact Person: Ruthie Amir, MD, Global Vice President of Clinical and Regulatory Affairs

Date Prepared: December 5, 2014

#### Name of Device

eTwo Skin Treatment System

#### Common or Usual Name

Electrosurgical cutting and coagulation device and accessories

### Classification Name/Product Code

Classification: 21 CFR 878.4400, Electrosurgical cutting and coagulation device and accessories

Product Codes: GEI (Electrosurgical, cutting & coagulation & accessories), OUH (Skin resurfacing RF applicator)

#### Predicate Devices

Syneron eTwo Skin Treatment System (K110672) EndyMed Medical, Ltd.'s Glow (K120513, K083461, K101510)

#### Intended Use / Indications for Use

The eTwo Skin Treatment System is intended for dermatological procedures.

The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin, the Sublative RF applicator is limited to skin types I-IV.

{4}------------------------------------------------

The Sublime applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm², the Sublime applicator is limited to skin types I-IV.

# Device Description

The eTwo Skin Treatment System is a mobile system combining two existing technologies and applications for dermatological procedures.

The Sublative RF applicator delivers bipolar radiofrequency (RF) to the skin surface via an array of multi-electrode pin tips. The applicator delivers bipolar RF energy to the skin, which results in heating of demarcated spots of skin to temperatures that result in ablation and resurfacing of the skin that comes in contact with the multi-electrode pins.

The Sublime applicator uses a combination of an infrared (IR) light source and bipolar radiofrequency (RF) to bulk heat the dermis and affect dermal collagen in order to treat wrinkles. The bulk heating of the dermal layers refers to the gradual and gentle accumulation of heat with each additional pass performed.

# Technological Characteristics

The eTwo System has generally the same technological characteristics as cleared in the predicate eTwo System, including device design, material, and energy source, except the subject eTwo System includes an additional 44-pin electrode tip and allows for higher RF energy levels to be delivered. The Sublime applicators of the eTwo Systems provide IR light and RF energy to the treatment area, whereas the non-ablative applicator of the EndyMed predicate provides RF energy only. The ablative applicators of the eTwo System and its predicates deliver RF energy to the treatment area. The hardware of the subject eTwo remains the same as previously cleared in the prior iteration of the device system.

### Performance Data

### Software

Software verification and validation was performed to evaluate the minor software changes made to support the increase in RF energy for the Sublime and Sublative RF applicators and to support the additional tip configuration (44 pin). Results showed that the Sublime and Sublative RF applicators with the eTwo Skin Treatment System function as intended.

# Electrical Safety and Electromagnetic Compatibility

The eTwo System has been evaluated for electromagnetic compatibility and electrical safety testing per applicable standards of IEC 60601-1-2 and 60601-1, and all results were passing.

{5}------------------------------------------------

# Histological Evaluations

Histological evaluation of skin tissues treated with the Sublime and Sublative applicators examined the impact of each applicator on skin tissue under the treated area, and demonstrated that the device performs as intended.

# Sterilization Validation

Sterilization validation study results demonstrated that the sterilization method for the eTwo System continues to provide a minimum sterility assurance level of 10°, including for the additional 44-pin tip.

# Clinical Study

In addition, the eTwo System treatment was evaluated in a prospective multi-center clinical study of 72 subjects through 24 weeks of follow up after the end of treatment. Four follow up visits (1 week, 6 weeks, 12 weeks after treatment) were scheduled for subjects receiving 1 treatment session, and five follow up visits (1 week, 6 weeks, 12 weeks, 18 weeks, and 30 weeks after first treatment) were scheduled for subjects receiving 2 treatment sessions. The study evaluated the higher energy levels and the additional 44-pin tip of the subject eTwo System through assessments by blinded evaluators, investigators, and subjects. Subjects' results at each follow up visit were compared to baseline to determine the amount of improvement with treatment. The primary endpoint of the study was to evaluate the reduction in wrinkles in the treated areas at 12 weeks following the last treatment session compared to baseline, based on photographic assessments by three independent blinded evaluators using the Fitzpatrick Wrinkle and Elastosis Scale (Degree of Elastosis categorized from class I (mild) to class The study also evaluated the safety of the treatment and recorded the III (severe)). number and type of any adverse event during the study. Additional assessments included the Global Aesthetic Improvement scale and patient satisfaction.

The majority of the enrolled population was female, Caucasian, with Fitzpatrick skin type II or III. Mean age was 49 years. The study population presented to the study with a range in Fitzpatrick Wrinkle and Elastosis scores and skin quality.

The pivotal study of the eTwo System demonstrated that device treatment achieves wrinkle reduction under the proposed parameters for a majority (71%) of the study subjects, meeting the primary study endpoint for effectiveness (50%). Results of secondary endpoint analyses showed improvement in wrinkles as observed in blinded evaluation, by investigators, and as perceived by subjects themselves over time. At the primary endpoint visit (12 weeks post treatment end), investigator assessments determined that 77% of the subjects were rated as improved from baseline (i.e., at least 1 unit improvement in Fitzpatrick Scale) and 91% of the subjects showed an overall improvement in facial appearance, further confirming the favorable results reported in 

{6}------------------------------------------------

blinded evaluation. In addition, high levels of patient satisfaction were reported by the subjects, where the majority of subjects (≥63%) reported being satisfied or very satisfied with treatment throughout follow up. Treatment-associated discomfort or pain immediately following treatment generally resolved within a week.

At the proposed higher RF energy levels and with the additional 44-pin tip, the eTwo treatment continued to demonstrate a favorable safety profile with 6.9% (5/72 subjects) device-related adverse events throughout the study. All of the device-related adverse events were mild to moderate, and each event completely resolved during the study (most resolved within a few hours or days). There were or serious adverse events in the study and no subjects discontinued from the study due to adverse events. A few subjects reported hyperpigmentation or hypopigmentation following treatment, which were also recognized as treatment-associated responses. In total, 11 subjects reported any changes in pigmentation or texture. Of these subjects, the events resolved for 6 subjects (most within a few days). For the remaining 5 subjects, results showed that the noted event was not associated with a negative clinical impact on the patient. Therefore, the clinical study data demonstrate that the eTwo System performs as intended with a favorable safety and effectiveness profile.

### Substantial Equivalence

The eTwo System has the same intended use as the predicate devices. The eTwo System indications for use are the same as the predicate eTwo device, and similar to those of the EndyMed Glow predicate. The eTwo System presents similar technological characteristics and principles of operation as the predicate devices. The key technological differences between the current and predicate eTwo Systems are in terms of the increased RF energy level and the addition of an electrode pin tip size. The proposed eTwo treatment parameters have been evaluated in clinical testing, and results demonstrate that the device continues to perform as intended. The EndyMed predicate similarly presents higher levels of RF energy. Although the EndyMed device differs in terms of the energy type delivered (RF only) compared to the eTwo Systems (RF only and IR combined with RF), this difference does not raise new types of safety or effectiveness questions given that the predicate eTwo utilizes the same combination of energies. Further, the underlying mechanisms of all three devices are consistent to achieve the same intended therapeutic effect for wrinkles treatment. Therefore, the eTwo System has satisfied the criteria for substantial equivalence.

# Conclusions

Syneron's eTwo System has been evaluated in nonclinical testing, and results demonstrated the device performance is appropriate for its intended use. The eTwo System presents the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate devices. Therefore, the eTwo System is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K141507](https://fda-staging.innolitics.com/device/K141507)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com