K141496 · Reflectance Medical, Inc. · MUD · Aug 4, 2014 · Cardiovascular
Device Facts
Record ID
K141496
Device Name
MULTI-PARAMETER MOBILE CAREGUIDE 4100 OXIMETER
Applicant
Reflectance Medical, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Aug 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Multi-Parameter Mobile CareGuide™ 4100 Oximeter is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor. The sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 4100 Oximeter is intended to allow for display of SmO2 and pHm data on a third party or Reflectance supplied device, which would interface with the Multi-Parameter Mobile CareGuide 4100 Oximeter via a powered USB connection. The Multi-Parameter Mobile CareGuide 4100 Oximeter is intended for prescriptive use (Rx only) by a trained healthcare professional inside and outside a hospital. The Multi-Parameter Mobile CareGuide 4100 Oximeter provides output of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 4100 Oximeter should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 4100 Oximeter has not been demonstrated in disease states.
Device Story
Non-invasive optical sensor; utilizes near-infrared (NIR) spectroscopy to monitor skeletal muscle tissue. Sensor illuminates skin with LEDs; reflected light analyzed by internal microprocessor. Device calculates SmO2 and pHm using stored algorithms. System comprises reusable sensor and disposable adhesive pad; attaches to deltoid, calf, or thigh. Communicates via powered USB to third-party display or RMI-supplied tablet using proprietary protocol. Provides real-time numeric values and trends of SmO2 and pHm; includes device status/error reporting. Used by trained healthcare professionals in clinical or out-of-hospital settings. Assists clinicians by providing physiological data; not intended as sole diagnostic tool. Benefits include non-invasive, real-time monitoring of microvascular blood status.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing included device performance, sensor optimization, background checks, and algorithm verification. Compliance with IEC 60825-1, IEC 60601-1-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10, AAMI TIR 12, AAMI TIR 30, and ISTA 1A standards confirmed.
Technological Characteristics
NIR spectroscopy (700-900 nm). Components: monitor, reusable sensor, disposable pad. Light source: LEDs. Parameters: SmO2, pHm. Connectivity: USB interface to external display. Standards: IEC 60825-1, IEC 60601-1-1, IEC 60601-1-2, ISO 10993-5/10, AAMI TIR 12/30. Software: embedded algorithm for spectral analysis.
Indications for Use
Indicated for use as an adjunct, non-invasive monitor of hemoglobin oxygen saturation (SmO2) and pH (pHm) of microvascular blood in skeletal muscle tissue. Intended for prescriptive use by trained healthcare professionals in or out of hospital settings. No specific patient population age or gender restrictions; sensor may be used on pigmented skin. Not for use as sole basis for diagnosis or therapy.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Multi-Parameter Mobile CareGuide Oximeter (K130079)
Multi-Parameter Mobile CareGuide Oximeter with Tablet (K133923)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure embracing a bird, which is a common symbol associated with the department.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2014
Reflectance Medical, Inc. % Nandini Murthy Regulatory Consultant 116 Flanders Road, Suite 1000 Westborough, Massachusetts 01581
Re: K141496
Trade/Device Name: Multi-parameter Mobile Careguide 4100 Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD. CBZ Dated: June 4, 2014 Received: June 6, 2014
Dear Nandini Murthy,
This letter corrects our substantially equivalent letter of August 4, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
sincerely yours,
[signature]
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 4100 Oximeter
# Indications for Use Form
# Indications for Use
510(k) Number (if known):
Multi-Parameter Mobile CareGuide™ 4100 Oximeter Device Name:
Indications for Use:
The Multi-Parameter Mobile CareGuide™ 4100 Oximeter is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor. The sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 4100 Oximeter is intended to allow for display of SmO2 and pHm data on a third party or Reflectance supplied device, which would interface with the Multi-Parameter Mobile CareGuide 4100 Oximeter via a powered USB connection. The Multi-Parameter Mobile CareGuide 4100 Oximeter is intended for prescriptive use (Rx only) by a trained healthcare professional inside and outside a hospital. The Multi-Parameter Mobile CareGuide 4100 Oximeter provides output of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 4100 Oximeter should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 4100 Oximeter has not been demonstrated in disease states.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of of
{3}------------------------------------------------
510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 4100 Oximeter
# SECTION 5
# 510(k) SUMMARY
# SUMMARY OF SAFETY AND EFFECTIVENESS FOR Multi-Parameter Mobile CareGuide™ 4100 Oximeter
# Submitter Information
| Name: | Reflectance Medical, Inc. (RMI) |
|----------------------|--------------------------------------------------------------------------------------------------------------------------|
| Address: | 116 Flanders Road, Suite 1000<br>Westborough, MA 01581 USA |
| Telephone Number: | 508.366.4700 |
| Registration Number: | NA (RMI will apply for registration number following 510(k)<br>clearance, prior to commencement of commercial shipment.) |
| Contact Person: | Dr. Babs Soller |
| Telephone Number: | 508.366.4700, Ext 223 |
| Fax Number: | 508.366.4770 |
| Email: | babs.soller@reflectancemedical.com |
| Date Prepared: | August 1, 2014 |
Device Name
| Trade name of New Device: | Multi-Parameter CareGuide™ Oximeter |
|---------------------------|------------------------------------------------------------------------------------------|
| Model Number: | 4100 |
| 510(k) Holder/Submitter: | Reflectance Medical, Inc. (RMI) |
| 510(k) Number: | N/A |
| Applicable product codes: | MUD, 21 CFR 870.2700 Oximeter and<br>CBZ, 21 CFR 868.1170 Analyzer, Hydrogen-Ion<br>(Ph) |
| Classification Panel: | Cardiovascular |
Predicate Devices
| Predicate Device #1: | CareGuide Oximeter |
|--------------------------|---------------------------|
| Trade name of Device: | CareGuide™ Oximeter |
| Model #: | 1100 |
| 510(k) holder/Submitter: | Reflectance Medical, Inc. |
{4}------------------------------------------------
510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 4100 Oximeter
| 510(k) Number: | K113656 |
|--------------------------|--------------------------------------------|
| Predicate Device #2: | CareGuide Oximeter |
| Trade name of Device: | Mobile CareGuide™ Oximeter |
| Model #: | 2100 |
| 510(k) holder/Submitter: | Reflectance Medical, Inc. |
| 510(k) Number: | K122645 |
| Predicate Device #3: | CareGuide Oximeter |
| Trade name of Device: | Multi-Parameter Mobile CareGuide™ Oximeter |
| Model #: | 3100 |
| 510(k) holder/Submitter: | Reflectance Medical, Inc. |
| 510(k) Number: | K130079 |
| Predicate Device #4: | CareGuide Oximeter |
| Trade name of Device: | Multi-Parameter Mobile CareGuide™ Oximeter |
| Model #: | 3100 with Tablet |
| 510(k) holder/Submitter: | Reflectance Medical, Inc. |
| 510(k) Number: | K133923 |
## Device Description
The Multi-Parameter Mobile CareGuide 4100 has the following technical characteristics in common with the CareGuide family of predicate devices.
| Characteristics | Reflectance Medical Multi-Parameter Mobile CareGuide 4100<br>Oximeter |
|------------------------|-----------------------------------------------------------------------|
| Principle of Operation | NIR spectroscopy |
| Components | Monitor with reusable sensor and disposable pad |
| Light Source | LEDs |
| Parameters Measured | Tissue oxygen saturation (SmO2) and muscle pH (pHm) |
The Multi-Parameter Mobile CareGuide 4100 is a noninvasive optical sensor that determines two separate medical parameters and reports them out for display to the user in real-time. Light sources in the sensor illuminate the skin with near infrared (NIR) light passes through the skin and fat with only some loss to be primarily absorbed by small blood vessels in the muscle tissue. Light which is not absorbed is scattered back and analyzed by the spectroscopic
{5}------------------------------------------------
510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 4100 Oximeter
detector, also contained in the sensor. The microprocessor in the sensor converts the reflected light to an absorbance spectrum which is then analyzed by two separate algorithms, also stored in the sensor's microprocessor. The two algorithms calculate muscle oxygen saturation (Sm22), and muscle pH (pHm).
The sensor is attached to the patient using the CareGuide Disposable. One Disposable is used per patient, but the CareGuide Sensor is reusable. To use, the adhesive liner is removed from the Disposable and, with the sensor clipped into the Disposable, is adhered to the patient's skin over either the deltoid, calf or thigh muscle.
When the sensor is first placed on the patient, software checks to make sure the LEDs are functioning properly and then the sensor automatically performs an optimization routine which sets up the spectral data collection parameters for the individual patient. Once conditions are established the sensor begins collecting spectra and reporting parameter values.
The Multi-Parameter Mobile CareGuide 4100 measures and provides for output of SmO2 and pHm data. The Multi-Parameter Mobile CareGuide 4100 communicates with either a 3ª party display or monitoring device, or RMI-supplied tablet, which is compliant with the Mobile CareGuide communications protocol. Sm02 and pHm values suitable for display and trending are sent via the Mobile CareGuide communications protocol to the display device as well as error information and device states.
# Indications for Use
The Multi-Parameter Mobile CareGuide™ 4100 Oximeter is intended for use as an adjunct, noninvasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 4100 Oximeter is intended to allow for display of SmO2 and pHm data on a third party or Reflectance supplied device, which would interface with the Multi-Parameter Mobile CareGuide 4100 Oximeter via a powered USB connection. The Multi-Parameter Mobile CareGuide 4100 Oximeter is intended for prescriptive use (Rx only) by a trained healthcare professional inside and outside a hospital. The Multi-Parameter Mobile CareGuide 4100 Oximeter provides output of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 4100 Oximeter should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 4100 Oximeter has not been demonstrated in disease states.
# Rationale for Substantial Equivalence
The Multi-Parameter Mobile CareGuide 4100 Oximeter is substantially equivalent to the Reflectance Medical CareGuide predicate devices by intended use and design.
{6}------------------------------------------------
510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 4100 Oximeter
- . The Mobile CareGuide 4100 Oximeter is identical to the predicate family in technological optical principles. They all use the identical NIR Spectroscopy hardware and software, using multiple wavelengths in the range 700-900 nm to measure tissue oxygen saturation and pH.
- The Multi-Parameter Mobile CareGuide 4100 Oximeter has a reusable sensor, disposable . and sensor check functionality equivalent to the predicates.
- The Multi-Parameter Mobile CareGuide 4100 Oximeter provides identical data to be . displayed as numeric values or trend.
- . The Intended Use is the same as the predicate.
- . Therefore, the Multi-Parameter Mobile CareGuide 4100 Oximeter is substantially equivalent to the predicates in terms of the proposed intended use and technology.
- There are no new or open questions of safety and effectiveness. ●
## Summary of Safety and Effectiveness Data
Testing demonstrates that the Multi-Parameter Mobile CareGuide 4100 Oximeter is a safe and effective oximeter meeting all relevant FDA recognized standards and internal testing as listed in Table 1.
| Standard #<br>or Internal<br>Test | Description |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC60825-1<br>Ed 2.0 (2007) | Safety of laser products - Part 1: Equipment classification, and<br>requirements |
| IEC 60601-1-<br>1: 2005 | Medical electrical equipment -- Part 1-1: General requirements for safety –<br>Collateral standards: Safety requirements for medical electrical systems |
| IEC 60601-1-<br>2: 2007-03 | Medical Electrical Equipment - Part 1-2: General Requirements for basic<br>safety and essential performance – Collateral standard: Electromagnetic<br>Compatibility – Requirements and tests |
| ISO 10993-5<br>2009 | Biological evaluation of medical devices-Part 5: Test for in vitro<br>cytotoxicity |
| ISO 10993-10<br>2010 | Biological evaluation of medical devices-Part 10: Tests for irritation and<br>delay-type hypersensitivity |
| AAMI TIR<br>12:2010 | Designing, testing and labeling reusable medical devices for reprocessing<br>in health care facilities: A guide for medical device manufacturers |
| AAMI TIR<br>30:2003 | A compendium of processes, materials, test methods, and acceptance<br>criteria for cleaning reusable medical devices |
#### Table 1 Standards and Internal Testing
{7}------------------------------------------------
510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 4100 Oximeter
| Standard #<br>or Internal<br>Test | Description |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISTA 1A | Series Non-Simulation Integrity Performance Test Procedure: Packaged-<br>Products 150 lb (68 kg) or Less |
| Internal Tests | Device Verification and Validation tests such as Vibration and shock,<br>Auto-start, Sensor optimization and background checks and verification of<br>minor updates to the CareGuide 4100 algorithm |
The test results in this submission demonstrate that the Multi-Parameter Mobile CareGuide 4100 Oximeter meets the expected performance requirements for an Oximeter. The accuracy, functionality and safety of the Multi-Parameter Mobile CareGuide 4100 Oximeter is the equivalent to the predicate device.
#### Conclusion
The Multi-Parameter Mobile CareGuide 4100 Oximeter is equivalent to predicate devices in terms of technology (NIR Spectroscopy) and intended use. The Multi-Parameter Mobile CareGuide 4100 Oximeter does not raise new questions of safety or effectiveness, as compared to the predicate. Therefore, the Multi-Parameter Mobile CareGuide 4100 Oximeter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.