The EasyDrill Autostop Cranial Perforator is a sterile, single use cutting device intended for performing cranial bone trephination.
Device Story
EasyDrill Autostop Cranial Perforator is a sterile, single-use mechanical bone-cutting instrument; used for cranial trephination. Device comprises inner/outer drill and Hudson shank; attaches to pneumatic or electric surgical drill motors via Hudson chuck. Principle of operation: mechanical torque/rotational speed from motor drives drill; integrated clutch mechanism automatically disengages drill upon perforation when resistance to bone ceases. Used in clinical/surgical settings by neurosurgeons. Output is a created access hole in the skull. Benefits include controlled, automated cessation of drilling to prevent unintended penetration of underlying structures.
Clinical Evidence
Bench testing and cadaveric study provided. Cadaveric study on human skull specimens evaluated performance and suitability for cranial trephination. Side-by-side bench testing performed against predicate devices using bovine scapula. Endotoxin testing conducted via LAL Gel-Clot procedure, confirming compliance with 2.15 EU/Device limit.
Technological Characteristics
Mechanically powered cranial perforator; stainless steel patient-contacting components; Hudson shank interface for pneumatic/electric drill motors. Features integrated clutch mechanism for automatic disengagement. Gamma sterilized. Available in multiple diameters (9mm, 11mm, 14mm) and shelf depths (1.5mm, 3mm).
Indications for Use
Indicated for patients requiring cranial bone trephination. No specific age or gender restrictions provided.
Regulatory Classification
Identification
Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2015
Micromar Industria e Comercio Ltda. % Deep Pal US Agent- Principal Regulatory Affairs Specialist Medtronic Powered Surgical Solutions 4620 North Beach Street Fort Worth, Texas 76137
Re: K141455
Trade/Device Name: EasyDrill Autostop Cranial Perforator Regulation Number: 21 CFR 882.4305 Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBF Dated: February 12, 2015 Received: February 13, 2015
Dear Deep Pal.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Carlos L. Pena -SD/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141455
Device Name EasyDrill Autostop Cranial Perforator
Indications for Use (Describe)
The EasyDrill Autostop Cranial Perforator is a sterile, single use cutting device intended for performing cranial bone trephination.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary is submitted in accordance with the requirements of 21CFR807.92.
# DATE PREPARED
March 16, 2015
# NAME/ADDRESS OF MANUFACTURER
Micromar Industria e Comercio Ltda CNPJ: 53.168.142/0001-29 Av. Marginal ao Corrego da Serraria, nº 168 CEP 09980-390 - Serraria, Diadema, Sao Paulo, Brazil
#### SUBMITTER'S NAME AND ADDRESS
Micromar Industria e Comercio Ltda CNPJ: 53.168.142/0001-29 Av. Marginal ao Corrego da Serraria, nº 168 CEP 09980-390 - Serraria, Diadema, Sao Paulo, Brazil
# CONTACT PERSON
Deep Pal US Agent for Micromar Telephone: 817.788.6685 Facsimile: 817.788.6225 E-Mail: deep.pal@medtronic.com
#### PROPRIETARY NAME
EasyDrill Autostop Cranial Perforator
# COMMON/USUAL NAME
Cranial Perforator
#### CLASSIFICATION NAME
Powered compound cranial drills, burrs, trephines, and their accessories, HBF, 21CFR882.4305
#### PREDICATE DEVICE IDENTIFICATION
K071931: Codman Disposable Perforator, Codman & Shurtleff, Inc. K082010: Zyphr Disposable Cranial Perforator, Stryker Instruments K892866: Acra-Cut Disposable Cranial Perforator, Acra-Cut Inc.
{4}------------------------------------------------
# DEVICE DESCRIPTION
The EasyDrill Autostop Cranial Perforator comprises of an Inner and Outer Drill, and a Hudson Shank that is designed to lock onto a Pneumatic or an Electric Drill Motors.
The EasyDrill Autostop Cranial Perforator device is a bone cutting and drilling instrument used in conjunction with a surgical motor and a Hudson Chuck - speed reducer attachment, to drill access holes through a patient's skull. When properly used, the EasyDrill Autostop Cranial Perforator employs a clutch mechanism to automatically disengage once perforation is accomplished and as the drill ceases to find resistance to bone.
The EasyDrill Autostop Cranial Perforator is a mechanically powered tool, designed to create an access hole through the skull. The subject Perforator device derives its mechanical torque and rotational speed from various legally marketed Electric and/or Pneumatic Drill Motors. The EasyDrill Autostop Cranial Perforator device is latched onto a Hudson Chuck device, which is in turn attached to the Drill Motors.
The EasyDrill Autostop Cranial Perforator is provided Gamma Sterilized, in a sealed packaging and is designed for single patient use only.
| Model | External Diameter | Internal Diameter | Shelf | Color | Minimum Bone Thickness |
|-------------------|-------------------|-------------------|-------|--------|------------------------|
| Ø9 x Ø6 x 3mm | 9 mm | 6 mm | 3 mm | Yellow | 3.25 mm |
| Ø11 x Ø7 x 3mm | 11 | 7 | 3 | Blue | 3.25 mm |
| Ø14 x Ø11 x 3mm | 14 | 11 | 3 | White | 3.25 mm |
| Ø14 x Ø11 x 1.5mm | 14 | 11 | 1.5 | Green | 1.75 mm |
# INDICATIONS FOR USE
The EasyDrill Autostop Cranial Perforator is a sterile, single use cutting device intended for performing cranial bone trephination.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Mechanically powered tool, designed to create an access hole through the skull is the technological principle for both the subject and predicate devices. It is based on the use of powered instrumentation for deriving the needed mechanical torque and rotational speed for cranial bone trephination. At a high level, the subject and predicate devices are based on the following same technological elements:
- > Cranial Perforator Application: Used for cranial bone trephination;
- > Used in conjunction with Electric and/or Pneumatic Drill System and a Hudson Chuck Speed Reducer Attachment: Comprises of an Inner and Outer Drill, and a Hudson Shank that is designed to lock onto a Pneumatic or an Electric driving tool. This allows the device the needed mechanical torque and rotational speed to create an access hole through the skull;
- A Clutch Mechanism: Automatically disengages once perforation is accomplished and as the drill ceases to find resistance to bone.
In terms of the materials used in manufacturing of the patient contacting components of the subject EasyDrill Autostop Cranial Perforator devices, the subject EasyDrill Autostop Cranial Perforator device is similar to the predicate devices. Like the predicate devices, the patient contacting Central and the Outer Drill components of the EasyDrill Autostop Cranial Perforator devices are made from Stainless Steel materials.
{5}------------------------------------------------
#### PERFORMANCE TESTING
The following performance data were provided in support of the substantial equivalence determination:
#### CADAVER STUDY
In order to evaluate the performance, its application and suitability of the subject EasyDrill Autostop Cranial Perforator devices for its intended use, the EasyDrill Autostop Cranial Perforator devices were tested under simulated use conditions on human cadaveric specimen - skull. Results of cadaveric testing demonstrate that the EasyDrill Autostop Cranial Perforator does not present any new issues of safety or effectiveness, and that the devices perform as intended during cranial bone trephination.
# SIDE BY SIDE BENCH TOP TESTING
In order to evaluate the performance of the subject EasyDrill Autostop Cranial Perforator devices against the predicate devices, for its intended use, the EasyDrill Autostop Cranial Perforator devices were tested alongside the predicate devices in a Bovine Scapula.
#### ENDOTOXIN TESTING
The subject EasyDrill Autostop Cranial Perforator devices were tested for Endotoxins using the Limulus Amebocyte Lysate (LAL) Gel-Clot test procedure. The data from the study show that the subject EasyDrill Autostop Cranial Perforators meet the required endotoxin specification limit of 2.15 EU/Device.
# CONCLUSIONS
The performance data from the side-by-side testing, of the subject EasyDrill Autostop Cranial Perforator device against the predicate perforator devices, were compared to support the subject device and the cadaver study demonstrates that the subject EasyDrill Autostop Cranial Perforator should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.