K141440 · Oehm Und Rehbein GmbH · MQB · Aug 29, 2014 · Radiology
Device Facts
Record ID
K141440
Device Name
DICOMPACS DX-R WITH FLAT PANEL
Applicant
Oehm Und Rehbein GmbH
Product Code
MQB · Radiology
Decision Date
Aug 29, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The dicompacs dx-r with flat panel digital imaging system is intended for use in generating radiographic images of human anatomy. This device is intended to replace film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. This device is intended for use by qualified medical persomel and is contraindicated when, in the judgment of the physician, procedures would be contrary to the patient.
Device Story
System captures radiographic images using solid-state flat panel detectors (Thales Pixium Portable 2430 EZ, 3543 EZ, or Perkin Elmer 4336 XRpad). Detectors transmit raw image data to a Windows-based console PC via LAN or WLAN (Wi-Fi). Software performs image acquisition, processing, and optimization for display. Used in clinical settings by radiographers/X-ray assistants. Output provides high-quality diagnostic images to assist physicians in clinical decision-making. Benefits include rapid image availability (under 5 seconds) and replacement of traditional film/screen systems.
Clinical Evidence
Clinical image review performed by a US Board Certified Radiologist. Images from all three panels evaluated as clinically acceptable, high resolution, and comparable or superior to predicate devices. Clinical study conducted per FDA guidance 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices' (1999).
Indicated for generating radiographic images of human anatomy for general purpose diagnostic procedures. Intended for use by qualified medical personnel. Contraindicated when procedures are contrary to the patient's best interest. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are stacked vertically, creating a sense of unity and collaboration.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2014
Oehm und Rehbein GmbH % Daniel Kamm, P.E. Principal Engineer ( Kamm & Associates) 8870 Ravello Ct NAPLES, FL 34114
Re: K141440
Trade/Device Name: dicomPACS DX-R with flat panel Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-ray System Regulatory Class: II Product Code: MOB Dated: May 29, 2014 Received: June 2, 2014
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M Morris Division Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141440
Device Name dicomPACS DX-R with flat panel
#### Indications for Use (Describe)
The dicompacs dx-r with flat panel digital imaging system is intended for use in generating radiographic images of human anatomy. This device is intended to replace film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. This device is intended for use by qualified medical persomel and is contraindicated when, in the judgment of the physician, procedures would be contrary to the patient.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for "OR Technology". The logo consists of a 2x2 grid of squares on the left, with the bottom left square being blue and containing a smiley face. The other three squares are gray. To the right of the squares is the text "OR Technology" in blue, followed by the text "| Digital X-ray and | Imaging Solutions" in a smaller font.
# 510(k) Summary K140551
| Submission<br>Correspondent: | Oehm und Rehbein GmbH<br>Contact: Franziska Günther<br>Phone: (49) 381 36 600 500<br>Email: franziska.guenther@or-technology.com | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 3006542593 | |
| Submission Sponsor: | Oehm und Rehbein GmbH<br>Neptunallee 7c<br>18057 Rostock<br>GERMANY<br>Phone: (49) 381 36 600 500<br>Fax: (49) 381 36 600 555 | |
| Date summary prepared: | May 23, 2014 | |
| Device trade name: | dicomPACS®DX-R with flat panel | |
| Device common name: | Solid State X-ray Imager (Flat Panel/Digital Imager) | |
| Classification: | Class II | |
| Product codes: | MQB at 21 CFR Part 892.1680 LLZ at 21 CFR Part 892.2050 | |
| Predicate devices: | K131211, dicomPACS®DX-R; Oehm und Rehbein GmbH. | |
| Description of the<br>device: | The dicomPACS®DX-R with flat panel digital imaging system consists of two<br>components, the dicomPACS®DX-R software for viewing captured images on<br>a Windows based computer, and one of three solid state X-ray imaging<br>devices: Thales Pixium Portable 2430 EZ (wireless), Thales Pixium Portable<br>3543 EZ (wireless), or the Perkin Elmer 4336 XRpad (also wireless). The<br>system will display high quality images in less than five seconds over a wide<br>range of X-ray dose settings. The software has the following characteristics:<br>The dicomPACS® DX-R software runs on an off-the shelf PC which forms the<br>operator console. Images captured with the flat panel digital detector are<br>communicated to the operator console via LAN or WLAN connection,<br>depending on the model and the user's choice. dicomPACS®DX-R software<br>uses the software API of the panel manufacturers to control the flat panels and<br>to receive and to calibrate image data. The dicomPACS® DX-R software is an<br>independent product for the acquisition, processing and optimization of X-ray<br>images (raw images) provided by flat panel (DR) systems or CR systems. In<br>general, such software is also called ",console software" as it is installed on the<br>so-called ",console PC" of the imaging device. dicomPACS® DX-R carries out<br>the image processing of the raw images provided by the particular device and<br>provides the radiographer / X-ray assistant with optimum workflow for their<br>work. The key difference between the modified device and our predicate device<br>is the panels are wireless, Wi-Fi. | |
| X-Ray Generators<br>Supported | Manufacturer<br>Sedecal, Spain<br>Stadler, Swizerland<br>Röntgenwerk Bochum, Germany<br>CPI, Canada<br>EMD Technologies<br>Control-X, Hungary<br>Shimadzu, Japan | Product Name<br>SHF and RST types<br>8X Series<br>Editor HFe401 - 801<br>Indico100, CMP 200 DR<br>EPS 45-80 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for OR Technology. The logo consists of a 2x2 grid of rounded squares on the left, with the bottom left square being blue and containing a smiley face. To the right of the squares is the text "OR Technology" in a sans-serif font, followed by "| Digital X-ray and | Imaging Solutions" in a smaller font.
| Indications for Use | The dicomPACS® DX-R with flat panel digital imaging system is intended for<br>use in generating radiographic images of human anatomy. This device is<br>intended to replace film/screen systems in all general purpose diagnostic<br>procedures. This device is not intended for mammography applications.<br>This device is intended for use by qualified medical personnel and is<br>contraindicated when, in the judgment of the physician, procedures would be<br>contrary to the best interest of the patient. |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>equivalence to<br>predicate devices: | All three receptor panels are identical in structure and share nearly<br>identical Indications for Use. The included software offers the same<br>functions as previously cleared in our software K131211. All panels share<br>similar technical parameters as well as the same materials and conversion<br>technique. All panels have been previously cleared by FDA.<br>The key technical difference is that the new panels are wireless, using<br>standard Wi-Fi technology. |
| Summary of<br>Technological<br>Characteristics | The three new panels have the following specifications;<br>Thales Pixium Portable 2430 EZ (wireless) 24 x 30 cm, 148 $\mu$ m pixels<br>Thales Pixium Portable 3543 EZ (wireless) 35 x 43 cm, 148 $\mu$ m pixels<br>Perkin Elmer 4336 XRpad (wireless) : 35 x 43 cm, 100 $\mu$ m pixels<br>All use standard Wi-Fi communications and have wired Ethernet available. |
| Summary of Findings<br>from Non-Clinical<br>Testing | Electrical safety and EMC testing was conducted by the respective panel<br>manufacturers in compliance with IEC 60601-1 and IEC 60601-1-2.<br>Software validation and risk analysis was conducted in compliance with<br>FDA guidance documents. The suppliers of the respective panels<br>performed MTF, DQE, linearity, and resolution measurements. Wireless<br>coexistence is assured via use of standard 802.11(n). Wireless<br>communication technology was chosen with the FDA wireless guidance<br>document taken into consideration. |
| Summary of Findings<br>from Review of Clinical<br>Images | Images from all three panels were reviewed by a US Board Certified<br>Radiologist The panels produce images that are clinically acceptable. The<br>images are of excellent quality, high resolution and are comparable to or<br>better than the images from the predicate devices. A clinical study as<br>described in the FDA Guidance Document Guidance for the Submission of<br>510(k)'s for Solid State X-ray Imaging Devices Document issued on: August<br>6, 1999 was conducted successfully. |
| Conclusion: | There are no significant differences between the dicomPACS® DX-R with flat<br>panel digital imaging system and the predicate devices and therefore,<br>dicomPACS® DX-R with flat panel does not raise any questions regarding<br>safety and effectiveness. The dicomPACS® DX-R with flat panel, as<br>designed, is as safe and effective as the predicate devices, and the device is<br>determined to be substantially equivalent to the referenced predicate devices<br>currently on the market |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.