The venous hardshell cardiotomy reservoir is used to collect, store and filter blood in extracorporeal circulation, in cardiopulmonary bypass operations on pediatric patients for up to 6 hours. The reservoir can also be employed postoperatively as drainage and autotransfusion reservoir (e.g., for thorax drainage) to return the autologous blood to the patient which was removed from the thorax for the volume exchange.
Device Story
Device functions as blood buffer in extracorporeal circuits; collects, stores, and filters blood during pediatric cardiopulmonary bypass; serves as postoperative drainage and autotransfusion reservoir for thorax drainage. Input consists of sucked blood; device performs defoaming and filtration; output is processed autologous blood returned to patient. Used in clinical settings (OR/post-op) by medical professionals. Modification involves changing defoamer coating from ASC to MACE and updating specifications for maximum blood flow, dynamic hold-up volume, and breakthrough volume.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by non-clinical design verification testing, including air handling, defoaming, breakthrough volume, dynamic priming volume, functional testing, bioburden, LAL, and biocompatibility.
Technological Characteristics
Hardshell reservoir; MACE defoamer coating; SOFTLINE coating; sterile, non-pyrogenic. Modifications include updated specifications for maximum blood flow, dynamic hold-up volume, and breakthrough volume. Mechanical device; no software or energy source.
Indications for Use
Indicated for pediatric patients undergoing cardiopulmonary bypass procedures requiring blood collection, storage, and filtration for up to 6 hours, or for postoperative drainage and autotransfusion.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating (K102919)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 0 7 2014 Premarket Notification Special 510(k) - Device Modification; Modified Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating
## 510(k) Summary
# [As required by 21 CFR 807.92(c)]
Submitter Information
MAQUET Cardiopulmonary AG
Kehler Strasse 31 76437 Rastatt
Germany
Contact Person:
Katrin Schwenkglenks
Phone: 011 49 7471 9973 151
011 49 7471 9973 8667 Fax:
e-mail: Katrin.Schwenkglenks@maquet.com
1
Summary Date:
May 23, 2014
#### Device Name
Proprietary name: Neontal Venous Hardshell Cardiotomy Reservoir with and without SOFTLINE Coting Pediatric Venous Hardshell Cardiotomy Reservoir with and without SOFTLINE Coating
Common/Usual name:Venous Hardshell Reservoir
Classification: Class II
CFR Sections: 21 CFR 870.4400; 21 CFR 870.4230; 21 CFR 870.4270
Predicate Device: Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating (K102919).
{1}------------------------------------------------
Premarket Notification Special 510(k) - Device Modification;
Modified Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating
#### Device Modification
ﺘﺮﺗ
The modifications subject of this submission comprise the material change of the defoamer coating from ASC to MACE coating combined with a change of the specifications - the increase of the maximum blood flow of the cardiotomy part and thus an increase in the dynamic hold-up volume of the cardiotomy part, as well as a change of the specifications of the break through volume of the venous and of the cardiotomy part.
#### Device Description
The Neonatal and Pediatric Venous Harshell Cardiotomy Reservoirs with and without SOFTLINE Coating are developed for surgical procedures requiring cardiopulmonary bypass for pediatric patients. They are used as a blood buffer in the extracorporeal circuit and are used as a collecting and defoaming device for sucked blood. The device is supplied sterileand non-pryogenic.
#### Intended Use
The venous hardshell cardiotomy reservoir is used to collect, store and filter blood in extracorporeal circulation, in cardiopulmonary bypass operations on pediatric for up to 6 hours.
The reservoir can also be employed postoperatively as drainàge and autotransfusion reservoir (e.g., for thorax drainage) to return the autologous blood to the patient which was removed from the thorax for the volume exchange.
#### Comparison of Technological Characteristics to Predicate Device
The modified Venous Hardshell Cardiotomy Reservoirs have the following similarities to the Venous Hardshell Cardiotomy Reservoirs, which previously received 510(k) clearance:
- have the same intended use,
- use the same operating principle,
{2}------------------------------------------------
Premarket Notification Special 510(k) - Device Modification: Modified Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating
- . incorporate the same reservoir design,
- have the same shelf life, .
- . are packaged and sterilized using the same processes and
- have the same SOFTLINE Coating
There is no change in the fundamental scientific technology and the product change does not affect the intended use.
In summary, the Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating described in this submission are, in our opinion, substantially equivalent to the predicate devices.
#### Nonclinical Data
The modified devices were subjected to design verification tests which are based on the risk assessment. The tests are designed to show that the Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating with the proposed modifications are as safe and effective as the originally cleared devices.
The following design verification tests / evaluations were performed:
Air handling
-
-
Defoaming
Break Through Volume, cardiotomy part and venous part .
· Dynamic Priming Volume, cardiotomy part -
Functional Test, cardiotomy part -
BioBurden -
LAL
Biocompatibility
The evaluation and test results do not show any kind of risk potential for the health or security of the patient or user. Based on the test results and evaluation, the Neonatal and
06-4
{3}------------------------------------------------
Premarket Notification Special 510(k) - Device Modification; Modified Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating
Pediatric Venous Hardshell Cardiotomy Reservoirs with and without SOFTLINE Coating are safe and effective for their intended use.
#### Clinical Data
Clinical results were not submitted to support substantial equivalence.
#### Conclusions
Based on the risk analysis, MAQUET Cardiopulmonary AG has conducted the appropriate . design verification activities and believes that the modified devices are substantially equivalent to the cleared MAQUET predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around the perimeter. In the center is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and modern design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2014
MAQUET Cardiopulmonary, AG Katrin Schwenkglenks Regulatory Affairs Manager Neue Rottenburger Strasse 37 72379 Hechingen, Germany
Re: K141432
Trade/Device Name: Neonatal and Pediatric Venous Hardshell Cardiotomy Reservoir with and without SOFTLINE Coating Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: DTN, DTP Dated: May 23, 2014 Received: June 9, 2014
Dear Ms. Schwenkglenks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Schwenkglenks
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda,gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
M.A.D.A.
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K141432
Device Name: Neonatal Venous Hardshell Cardiotomy Reservoir and Pediatric Venous Hardshell Cardiotomy Reservoir with and without SOFTLINE Coating
#### Indications for Use:
The venous hardshell cardiotomy reservoir is used to collect, store and filter blood in extracorporeal circulation, in cardiopulmonary bypass operations on pediatric palients for up to 6 hours.
The reservoir can also be employed postoperatively as drainage and autotransfusion reservoir (e.g., for thorax drainage) to return the autologous blood to the patient which was removed from the thorax for the volume exchange.
Over-The-Counter Use Prescription Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Zeller
Page __ of ___
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.