ACCUMETRICS,INC. VERIFYNOW PRUTEST, ACCUMETRICS, INC. VERIFYNOW PRUTEST
Applicant
Accumetrics, Inc.
Product Code
JOZ · Hematology
Decision Date
Jul 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5700
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K141427 · Jul 24, 2015
ACCUMETRICS,INC. VERIFYNOW PRUTEST, ACCUMETRICS, INC. VERIFYNOW PRUTEST
Accumetrics, Inc.
Whole blood samples from healthy donors; Whole blood samples from patients with acute coronary syndrome (ACS)
Retrospective/routine clinical blood samples were used to establish reference ranges, expected PRU values for treatment-naïve and clopidogrel-treated patients, and to assess intermediate precision and stability.
Healthy Donor PRU Reference Range; Observational study of healthy donors
Healthy donors; Sample Size: 152
Not applicable for this study
Platelet function (PRU)
Treatment-naïve patients in the intended use population; Observational study of ACS patients
Patients with acute coronary syndrome (ACS) not receiving P2Y12 inhibitors; Sample Size: 84
Not applicable for this study
PRU values
ACS patients receiving treatment with clopidogrel; Observational study of ACS patients
Patients with ACS receiving dual treatment with aspirin and clopidogrel; Sample Size: 71
Not applicable for this study
PRU values
Indications for Use
The VerifyNow™ PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.
Device Story
VerifyNow PRUTest measures platelet P2Y12 receptor blockade in citrated whole blood; used in laboratory or point-of-care settings by professionals. Device contains three lyophilized reagent pellets: ADP (for P2Y12-mediated aggregation), TRAP (internal control), and No-Agonist Pellet. Instrument automatically draws blood from closed vacuum collection tube; uses fibrinogen-coated microparticles to bind activated platelet GP IIb/IIIa receptors. Turbidimetric optical detection measures aggregation; output reported in P2Y12 Reaction Units (PRU). Results assist clinicians in assessing platelet inhibition levels. System includes internal controls for sample filling, fluid transfer, and reaction monitoring; requires factory calibration via barcode scanning. Electronic Quality Control (EQC) ensures instrument optics, pneumatics, and calculation accuracy.
Clinical Evidence
Bench testing only. Precision and reproducibility studies (n=9 donors) showed between-instrument variability ≤10%. Method comparison against predicate (n=119 patients) yielded a slope of 1.01 (95% CI: 0.97–1.05) and correlation coefficient R=0.98. Stability studies support 9-month reagent shelf life. Interference testing confirmed no significant impact from common medications/substances. Reference range established using 152 healthy individuals (182-335 PRU).
Technological Characteristics
Turbidimetric optical detection system. Disposable test device contains lyophilized reagent pellets (ADP, TRAP, No-Agonist) with fibrinogen/BSA-coated beads. Instrument-based electronic quality control (EOC) and wet quality controls (WQC). Factory calibration. Results reported in PRU. Standalone instrument. No humidity sensor in test device.
Indications for Use
Indicated for professional use in laboratory or point-of-care settings to measure platelet P2Y12 receptor blockade in whole blood samples from patients, including those with acute coronary syndrome (ACS).
Regulatory Classification
Identification
An automated platelet aggregation system is a device used to determine changes in platelet shape and platelet aggregation following the addition of an aggregating reagent to a platelet-rich plasma.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K141427
B. Purpose for Submission:
Modification of previously cleared instrument, assay, and control material.
Changes to assay reagents include the ADP pellet formulation, TRAP pellet channel location and introduction of a no-agonist pellet. The humidity sensor was removed from the instrument, while a port cover was added. The formulation of the control material was also changed.
C. Measurand:
Platelet aggregation
D. Type of Test:
Automated whole blood platelet aggregation
E. Applicant:
Accumetrics, Inc.
F. Proprietary and Established Names:
VerifyNow™ PRUTest
G. Regulatory Information:
1. Regulation section:
21 CFR 864.5700 – Automated platelet aggregation system
2. Classification:
Class II
3. Product code:
JOZ – System, Automated Platelet Aggregation
{1}
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
The VerifyNow™ PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
Not applicable
4. Special instrument requirements:
The VerifyNow™ PRUTest is designed to be run on the VerifyNow instrument
I. Device Description:
The VerifyNow™ PRUTest device contains three lyophilized reagent pellets in separate reaction chambers within the test device:
1) ADP pellet consisting of a preparation of Fibrinogen and BSA coated beads, adenosine-5-diphosphate (ADP), prostaglandin E1 (PGE1), dye, buffer, and a preservative
2) TRAP pellet (internal control) consisting of a preparation of iso-TRAP (Thrombin Receptor Activating Peptide), Fibrinogen and BSA coated beads, buffer, dye, and a preservative
3) No-Agonist Pellet (NAP) consisting of a preparation of BSA coated beads, dye, buffer, and a preservative
J. Substantial Equivalence Information:
1. Predicate device name(s):
Accumetrics, Inc. VerifyNow™ P2Y12 Assay
2. Predicate 510(k) number(s):
K051231
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | A whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only. | Same |
| Principle of Operation | Fibrinogen-coated microparticles, ADP/PGE1 platelet activation with turbidimetric optical detection of P2Y12-mediated platelet aggregation | Same |
| Specimen Type | Citrated whole blood | Same |
| Testing site | Point of care or laboratory | Same |
| Device Output | P2Y12 Reaction Units (PRU) | Same |
| Controls | 1. Bi-level external/ wet quality control (WQC) 2. Internal control | Same |
| Calibration | Factory | Same |
| Differences | | |
| Item | Device | Predicate |
| ADP pellet formulation | 80% Fibrinogen/20% BSA | 100% Fibrinogen |
| TRAP control formulation | 20 μM TRAP & 800 μM Par-4 | 3.4 μM TRAP |
| NAP pellet | No-agonist pellet (NAP) introduced | No-agonist pellet absent |
| Control Level 1 | Contains BSA | Contains fish skin gelatin |
| Control Level 2 | GRP conjugate truncated | GRP conjugate original length |
| Instrument humidity sensor | Removed | Present |
K. Standard/Guidance Document Referenced (if applicable):
CLSI C28-A3c Defining, Establishing, Verifying Reference Intervals in the Clinical Lab, Approved Guideline-3rd Edition
CLSI EP05-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-2nd Edition
{3}
CLSI EP09-A3 Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline
CLSI EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
IEC 60601-1-2 Medical Electrical Equipment- Part1-2: General Requirements for Safety-Collateral Standard: Electromagnetic Compatibility-Requirements and Tests
# L. Test Principle:
The patient sample is citrate-anticoagulated whole blood, which is automatically dispensed from the blood collection tube into the test device by the instrument, with no blood handling required by the user. Fibrinogen-coated microparticles are used in the VerifyNow PRUTest device to bind activated platelet GP IIb/IIIa receptors. ADP is incorporated into the assay to activate platelets, and the reagent is formulated to specifically measure P2Y12-mediated platelet aggregation. When the activated platelets are exposed to the fibrinogen-coated microparticles, aggregation occurs in proportion to the number of activated platelet receptors. The VerifyNow PRUTest reports results in P2Y12 Reaction Units (PRU).
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
# Reagent precision- whole blood:
Between-lot precision and between-instrument precision were determined in two separate experiments. Duplicate whole blood samples from each of 9 donors covering the device's measuring range were tested on a minimum of five (5) non-consecutive days on three VerifyNow™ Instruments. A different device lot was tested on each instrument, and lot/instrument combinations were reassigned each day for randomization purposes. Precision estimates in this experiment include the effects of both instrument and lot variability. Within-run, between-run, between-day, and between-lot variability was determined for each donor. With the exception of donor 7, the data met the predetermined acceptance criteria of $\leq 10\%$ CV.
| Days | Runs/day | Reps | Mean PRU | Within-run | | Between-run | | Between-day | | Between-lot | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 5 | 1 | 30 | 64.1 | 4.7 | 7.4 | ND | ND | 9.6 | 15.0 | 0.0 | 0.0 |
| 8 | 2 | 96 | 123.4 | 9.6 | 7.8 | 18.7 | 15.1 | 57.6 | 46.7 | 0.0 | 0.0 |
| 5 | 2 | 60 | 161.9 | 6.3 | 3.9 | 5.1 | 3.1 | 17.7 | 10.9 | 0.0 | 0.0 |
| 5 | 1 | 30 | 162.5 | 10.9 | 6.7 | ND | ND | 13.3 | 8.2 | 0.0 | 0.0 |
| 5 | 1 | 30 | 190.4 | 12.0 | 6.3 | ND | ND | 15.1 | 7.9 | 0.0 | 0.0 |
| 10 | 2 | 114 | 215.6 | 13.0 | 6.0 | 7.3 | 3.4 | 16.5 | 7.6 | 0.0 | 0.0 |
| 5 | 1 | 30 | 220.6 | 8.9 | 4.0 | ND | ND | 17.0 | 7.7 | 0.0 | 0.0 |
| 10 | 2 | 118 | 243.8 | 10.4 | 4.2 | 0.0 | 0.0 | 21.9 | 9.0 | 0.0 | 0.0 |
{4}
5
| Days | Runs/day | Reps | Mean PRU | Within-run | | Between-run | | Between-day | | Between-lot | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 10 | 2 | 120 | 289.0 | 11.2 | 3.9 | 7.7 | 2.7 | 27.0 | 9.3 | 0.0 | 0.0 |
No data (ND) for between-run precision for four subjects that had only one run per day. One of these subjects had insufficient blood volume to complete the between instrument experiment.
For evaluation of between-instrument variability, nine donors were tested with one reagent lot in duplicate across all three instruments for a minimum of five days, with two runs per day using a different operator for each run. The between-instrument variability was ≤ 10%.
| Days | Runs/day | Reps | Mean PRU | Within-run | | Between-run | | Between-day | | Between-instrument | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 5 | 1 | 30 | 67.1 | 7.3 | 10.8 | ND | ND | 9.3 | 13.8 | 0.0 | 0.0 |
| 5 | 2 | 60 | 115.4 | 8.8 | 7.6 | 20.4 | 17.7 | 65.9 | 57.1 | 0.0 | 0.0 |
| 5 | 2 | 60 | 159.9 | 9.3 | 5.8 | 4.1 | 2.5 | 18.1 | 11.3 | 3.6 | 2.3 |
| 5 | 1 | 30 | 160.2 | 11.2 | 7.0 | ND | ND | 11.6 | 7.3 | 0.0 | 0.0 |
| 5 | 1 | 30 | 228.7 | 7.9 | 3.5 | ND | ND | 14.4 | 6.3 | 8.1 | 3.5 |
| 5 | 2 | 54 | 230.7 | 14.7 | 6.4 | 12.1 | 5.3 | 9.2 | 4.0 | 7.9 | 3.4 |
| 8 | 2 | 89 | 252.8 | 13.6 | 5.4 | 0.0 | 0.0 | 10.4 | 8.1 | 9.3 | 3.7 |
| 9 | 2 | 101 | 289.6 | 10.7 | 3.7 | 5.4 | 1.9 | 21.4 | 7.4 | 11.4 | 3.9 |
No data (ND) for between-run precision for four subjects that had only one run per day. One of these subjects had insufficient blood volume to complete the between instrument experiment.
## Repeatability and Reproducibility- Wet Quality Controls:
Repeatability: Due to changes to the quality control material, repeatability for one lot of Wet Quality Control (WQC) Level 1 and 2 was tested in duplicate with three instruments over 20 non-consecutive days. Within-run, between-run, between-day, and total variability were assessed. Level 2 WQC meet the acceptance criteria of %CV ≤10%.
| | | Within-run | | Between-run | | Between-day | | Between-lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| n | Mean PRU | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 240 | 300.5 | 13.6 | 4.5 | 7.7 | 2.5 | 8.3 | 2.8 | 13.5 | 4.5 | 22.2 | 7.4 |
Level 1 WQC simulates a patient samples with highly inhibited platelets and does not produce a significant aggregation profile. The CVs for Level 1 WQC were not calculated because the nominal PRU value is zero. Values generated for Level 1 WQC did not exceed 30 PRU.
{5}
Reproducibility: Reproducibility was assessed for one lot of WQC Level 1 and 2 run on three instruments, tested twice a day, in duplicate over six non-consecutive days. The acceptance criterion for WQC Level 2 precision of $\% \mathrm{CV} \leq 10\%$ was met. Values generated for Level 1 WQC did not exceed 30 PRU.
| | | Within-run | | Between-run | | Between-day | | Between-instrument | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| n | Mean PRU | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 72 | 302.7 | 22.6 | 7.5 | 0.0 | 0.0 | 8.5 | 2.8 | 0.0 | 0.0 | 24.2 | 8.0 |
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Real time stability: Real time stability studies with control lot 3 are currently ongoing. Three lots of WQC Level 2 controls and three lots of reagent are stored at $20 \pm 5^{\circ} \mathrm{C}$ . Each of the three lots of WQC Levels 1 and 2 were tested at time 0, 3, 6, and 9 months with each of three reagent lots. The results did not exceed the pre-specified $10\%$ deviation from the initial measurement, supporting a stability claim of 9 months at $20 \pm 5^{\circ} \mathrm{C}$ for both WQC Level 2 and reagent.
Level 1 WQC simulates a patient samples with highly inhibited platelets and does not produce a significant aggregation profile. The CVs for Level 1 WQC were not calculated because the nominal PRU value is zero. The PRU range of the Level 1 WQC does not exceed 30 PRU in the 9-month testing timeframe.
Isochronous stability reagent stability study with patient samples: In addition to real time stability with the WQC material, blood from two donors and the two levels of WQC material were tested at 3, 6, and 9 months with three (3) lots of reagent stored at both $-70^{\circ}\mathrm{C}$ and $20\pm 5^{\circ}\mathrm{C}$ . The difference in means between whole blood tested with the $-70^{\circ}\mathrm{C}$ and $20\pm 5^{\circ}\mathrm{C}$ conditions was within $\leq 10\%$ further supporting the 9-month stability claim for VerifyNow™ PRUTest reagent storage at $20\pm 5^{\circ}\mathrm{C}$ .
Sample stability: A sample stability study was provided with blood from 11 donors spanning the analytical measuring range (AMR) of the device. Samples were run in duplicate after wait times of $\leq 1, 5, 10, 20, 30, 60, 120, 240,$ and 300 minutes of storage. The acceptance criterion was defined as PRUTest values that deviated $< 10\%$ from the one-sided $95\%$ lower confidence boundary of the PRUTest value predicted from the regression line. All but one sample met the VerifyNow™ PRUTest sample stability acceptance criterion for the 10-240 minute timeframe. The failed sample had extremely low PRU values ( $< 20$ PRU). The VerifyNow™ package insert instructs the user to wait a minimum of 10 minutes after blood draw before running the assay.
{6}
Shipping stability: Device kits were packaged by facilities personnel responsible for shipping customer material. Effects of initial manual handling, vehicle stacking, loose load vibration, low pressure (high altitude), vehicle vibration, impacts and final manual handling were examined. Two test lots of reagent were subjected to distribution testing. For each donor sample, the mean PRU result must be within $\pm 10\%$ of the mean of the control (un-shipped) treatment. The mean PRU result for the shipped treatment was within $\pm 10\%$ of the mean of the control treatment, meeting the pre-determined acceptance criteria.
Expected values: 84 patients with acute coronary syndrome, who are candidates for, but are not receiving a P2Y12 receptor inhibitor and 71 patients receiving aspirin and clopidogrel dual therapy were tested in order to determine expected value ranges. The resulting data and statistical analysis are provided below:

| | No P2Y12 Inhibitor | Clopidogrel |
| --- | --- | --- |
| N | 84 | 71 |
| Mean | 274 | 156 |
| SD | 48 | 73 |
| Lower (95% CI) | 180 (160-200) | 6 (0-34) |
| Upper (95% CI) | 376 (358-395) | 300 (269-329) |
| Mean (SD) clopidogrel dose | | 79 (28) |
| Mean (SD) ASA dose | 137 (103) | 167(117) |
d. Detection limit:
Not applicable
{7}
# e. Analytical specificity:
Interference testing was performed in-house to evaluate the effect of common substances in patient samples on the VerifyNow™ PRUTest Assay. The effect on medications and dietary substances commonly present in the intended use population were examined in this study. Blood samples devoid of platelet inhibitor and containing platelet inhibitor were tested pre and post spiking with the test compound. Platelet inhibitor containing samples were spiked with $14\mathrm{mM}$ MeSAMP to produce PRU values of about 130 PRU. One compound (Prilosec) was tested two hours following ingestion by healthy volunteers. The effect of the potential interferent (spiked or ingested) on the assay result was expressed as percent recovery. Samples were tested in randomized fashion with seven (7) instruments. The following substances were evaluated at the indicated concentrations and met the acceptance criteria of $<90\%$ to $110\%$ of baseline PRUTest results:
| Substance | Concentration | PASS/FAIL |
| --- | --- | --- |
| A3P5PS | 100 μM | PASS |
| Acetaminophen | 0.2 and 1.32 mM | PASS |
| Betamethazone | 64 μM | PASS |
| Caffeine | 306 μM | PASS |
| Captopril | 23 μM | PASS |
| Catechin | 86 μM | PASS |
| Cimetidine | 79 μM | PASS |
| Dipyridamole | 20 μM | PASS |
| Diltiazem | 15 μM | PASS |
| Ethanol | 87 mM | PASS |
| Fish Oil | 32 mg/mL | PASS |
| Pietal (Cilostazol) | 14.9 and 60 μM | PASS |
| Glycosamine HCl | 9.40 μM | PASS |
| Hydrochlorothiazide | 20 μM | PASS |
| Ibuprofen | 2.4 mM | PASS |
| Insulin | 3 ng/mL | PASS |
| Lidocaine | 51 μM | PASS |
| LMW Heparin | 1833 U/L | PASS |
| L-Thyroxine | 32 nM | PASS |
| Nitroglycerin | 0.1 μg/mL | PASS |
| Norfloxetine | 7.27 μM | PASS |
| Norverapamil | 4.5 μM | PASS |
| Oxypurinol | 99 μM | PASS |
| Pravastatin | 56 μM | PASS |
| Propranolol | 7.7 μM | PASS |
| Salicyclic Acid | 4.3 mM | PASS |
| Streptokinase | 400 U/mL | PASS |
| Theophylline | 220 μM | PASS |
| Triglycerides | 37 mM | PASS |
{8}
9
| Warfarin | 32 μM | PASS |
| --- | --- | --- |
| A- Tocopherol | 58 μM | PASS |
| DMSO | 0.11% | PASS |
| Celecoxib | 8.5 μg/mL | PASS |
| Prilosec | One 20 mg tablet | PASS |
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study, comparing performance of the Accumetrics, Inc. VerifyNow™ PRUTest and the predicate device was executed at three investigational sites with a total of six (6) operators. Citrated whole blood was collected from 119 patients receiving platelet inhibitors. Samples in the low, medium and high ranges of PRU values and covering the AMR of the device were tested.
The ordinary least squares regression analysis results are summarized below:
| | n | Slope (95% CI) | Slope p-value | Intercept (95% CI) | Correlation Coefficient R | R² | Sy|x |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| P2Y12 vs PRU | 119 | 1.01 (0.97–1.05) | 0.56 | -0.77 (-8.00–6.50) | 0.98 | 0.96 | 19.0 |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
{9}
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
A reference range for the VerifyNow™ PRUTest results was determined by testing 152 healthy individuals. The reference range of PRU results from the healthy donor population was determined by calculating the 95% confidence interval (2.5th percentile- 97.5th percentile) in accordance with CLSI C28-A3c and found to be from 182-335 PRU.
| N | Mean (PRU) | SD | 0.025 quantile (95% CI) | 0.975 quantile (95% CI) |
| --- | --- | --- | --- | --- |
| 152 | 266 | 42 | 182(116-197) | 335 (324-354) |
N. Instrument Name:
Accumetrics, Inc., VerifyNow™ System
O. System Descriptions:
1. Modes of Operation:
The operator inserts the sample tube into the test device. This is an individually run, closed-tube assay.
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ x ☑ or No ☐
3. Specimen Identification:
The VerifyNow™ System's default setting does not require the user to enter a Patient ID, however the system can be programmed to require a Patient ID before an assay can be performed.
4. Specimen Sampling and Handling:
{10}
Fresh citrated whole blood samples are required for use with the VerifyNow PRUTest. The sample well contains a needle for sampling directly from a closed vacuum collection tube.
## 5. Calibration:
The VerifyNow™ PRUTest devices are calibrated at the factory and require no calibration by the user. The calibration information is contained in the barcode on the pouch of each assay device. The barcode must be scanned whenever a new lot of assay devices are to be tested.
## 6. Quality Control:
Each assay device includes internal controls to determine whether the VerifyNow™ PRUTest result is valid and causes the instrument to report an error message for an invalid test. The negative internal control test is performed during the first five seconds of the assay, before platelet activation and binding to the fibrinogen-coated beads occurs. The positive internal control consists of a reagent channel that is similar to the P2Y12 assay channel, except it uses Thrombin Receptor Activating Peptide (TRAP) instead of ADP as the agonist. The positive internal control test is performed during the active phase of the assay.
The VerifyNow™ PRUTest utilizes VerifyNow™ Wet Quality Controls (WQC) Levels 1 and 2 for verifying the integrity of the VerifyNow™ PRUTest reagents. Controls have been formulated at clinically relevant levels with WQC Level 1 providing low level PRU results. WQC Level 1 is a carbon sol suspension, used to mimic the optical properties of whole blood. WQC Level 2 is designed to provide high level PRU results, comparable to results obtained from patient samples prior to ingestion of clopidogrel. WQC Level 2 is a lyophilized pellet, which is added to the diluent vial and reconstituted before use. WQC Level 2 uses a peptide that binds to fibrinogen to induce aggregation, which simulates the binding of fibrinogen-coated microparticles to platelets in the VerifyNow™ PRUTest. Assigned values for WQC Level 1 (≤ 30 PRU) and Level 2 (228 – 372 PRU) are included in the package insert.
The VerifyNow™ System contains the following internal controls:
- The instrument automatically verifies sample filling, correct fluid transfer and mixing. It also monitors the electronic and mechanical components.
- Each assay device incorporates two levels of quality control to identify invalid test runs caused by random error, reagent degradation or inappropriate blood samples. Before platelet activation and fibrinogen binding begin, the negative internal control performs a test for non-specific aggregation. During the active phase of the assay, the positive internal control channel monitors the reaction and calculates Control Units, which must fall within specified limits.
Electronic Quality Control (EQC) is one quality control mechanism for the VerifyNow™ instrument. It is a reusable device that is inserted by the operator into the assay port and is used to perform a comprehensive testing routine that confirms performance of instrument optics, the pneumatics system, reagent mixing parameters, calibration parameters,
11
{11}
simulates assay testing at two levels to check correct data acquisition and calculation and the required EQC testing frequency. Accumetrics recommends that EQC be run at least on a daily basis, although a preferred testing frequency may be selected by the institution. When the established test interval has elapsed, the user is locked out from running a patient test until the EQC test has been successfully completed.
P. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
Q. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.