← Product Code [MQE](/productcode/MQE) · K141365

# OTRIEVA TAPERED OVUM ASPIRATION NEEDLE SINGLE LUMEN (K141365)

_William A. Cook Australia Pty. , Ltd. · MQE · Aug 7, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141365

## Device Facts

- **Applicant:** William A. Cook Australia Pty. , Ltd.
- **Product Code:** [MQE](/productcode/MQE.md)
- **Decision Date:** Aug 7, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.6100
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Cook Otrieva™ Tapered Ovum Aspiration Needles are used for laproscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles.

## Device Story

Otrieva™ Tapered Ovum Aspiration Needle is a sterile, single-use device for oocyte retrieval. It consists of a tapered cannula (17G proximal to 20G distal) with a constant 0.60 mm internal diameter, aspiration tubing, and a vacuum line. The needle is inserted transvaginally under ultrasound guidance or via cannula during laparoscopy to access ovarian follicles. The device connects to an external vacuum source to aspirate and flush oocytes. Used by clinicians in clinical settings for in vitro fertilization procedures. The tapered design provides the rigidity of a larger needle with the reduced puncture trauma of a smaller tip. The device facilitates oocyte collection, which is essential for subsequent laboratory fertilization and embryo development.

## Clinical Evidence

Bench testing only. Testing included resistance to breakage, air tightness, tensile strength, and leakage testing. Performance specifications were verified, including a Mouse Embryo Assay (MEA) change from 1-cell (75% blastocyst rate at 96h) to 2-cell (80% blastocyst rate at 72h), confirming no adverse impact on embryo toxicity.

## Technological Characteristics

Materials: Stainless steel #304, polycarbonate, PTFE, silicone, FEP, copolyester. Design: Tapered cannula (17G to 20G), 0.60 mm ID. Sterilization: Ethylene Oxide (SAL 10-6). Single-use. Non-powered.

## Regulatory Identification

Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).

## Predicate Devices

- Ovum Pick-Up Aspiration Needle ([K983593](/device/K983593.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 7, 2014

William A. Cook Australia Pty, Ltd. Katie-Lou Buscher Project Manager, Regulatory Affairs 95 Brandl Street Brisbane Technology Park, Eight Mile Plains Brisbane, QLD 4113 AUSTRALIA

Re: K141365 Trade/Device Name: Otrieva™ Tapered Ovum Aspiration Needle Regulation Number: 21 CFR& 884.6100 Regulation Name: Assisted Reproduction Needles Regulatory Class: II Product Code: MQE Dated: July 14, 2014 Received: July 15, 2014

Dear Katie-Lou Buscher,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K141365

Device Name

Otrieva™ Tapered Ovum Aspiration Needle

Indications for Use (Describe)

The Cook Otrieva™ Tapered Ovum Aspiration Needles are used for laproscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters, and the word "MEDICAL" is written in smaller, white, sans-serif letters below the word "COOK".

## 510(k) Summary - K141365 Otrieva™ Tapered Ovum Aspiration Needle - 14 July 2014

| Submitted by:                            | William A. Cook Australia<br>95 Brandl Street<br>Eight Mile Plains<br>Brisbane<br>Queensland<br>4113<br>Australia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:<br>Email:<br>Telephone:<br>Fax: | Katie-Lou Buscher<br>Katie-Lou.Buscher@CookMedical.com<br>+61 7 3841 1188<br>+61 7 3841 1288                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade name                               | Otrieva™ Tapered Ovum Aspiration Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 510(k) Number                            | K141365                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification name                      | Assisted Reproduction Needles (21 CFR 884.6100, Product Code<br>MQE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Device                         | Ovum Pick-Up Aspiration Needle, cleared under K983593                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Description                       | This device is intended for laparoscopic or ultrasound guided<br>transvaginal aspiration and flushing of oocytes from ovarian follicles<br><br>The needle is passed through a transvaginal ultrasound transducer or<br>placed through a cannula for a laparoscopic procedure to advance into<br>the ovarian follicle.<br><br>The main body of the cannula is 17 gage, which tapers to 20 gage at<br>the distal tip. Although the outer diameter of the needle tapers, a<br>constant 0.60 mm internal diameter is maintained.<br><br>The Otrieva™ needles are available in two lengths: 30cm and 35cm.<br>The needles are provided with an aspiration tube of 90 cm length and a<br>vacuum tube with a length of 50 cm. The tubing is connected to a<br>silicone bung.<br><br>The materials used in the device include: stainless steel, polycarbonate,<br>PTFE, silicone, FEP and copolyester. |
| Indication for Use                       | The Cook Otrieva™ Tapered Ovum Aspiration Needles are used for<br>laparoscopic or ultrasound guided transvaginal aspiration and flushing<br>of oocytes from ovarian follicles.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Comparison to                            | The Otrieva™ Tapered Ovum Aspiration Needle is a modification of the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Device                         | Ovum Pick-up Aspiration Needle (K983593). The modifications are:<br>• Incorporation of a tapered cannula instead of fixed diameter cannula - the tapered cannula is designed to allow the clinician the rigidity of a larger diameter cannula while providing a smaller diameter tip for puncture. As the inner and outer diameter of the cannula at both ends falls within the range of the predicate device, and the internal lumen of the cannula is the same as the predicate device, this change does not represent a new technology that could raise new types of safety or effectiveness questions.                                                                                                                                                                                                                                                                                     |
|                                          | • Addition of a 50 cm vacuum line - the addition of the vacuum line simply provides an extension of the connection to the vacuum source and does not change the use of the device. Attachment of an Ovum Aspiration Needle to the vacuum source using tubing is not a new technology and therefore there are no new safety or effectiveness questions raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                          | • Incorporation of a smaller (1.32 mm) diameter aspiration tube – The smaller diameter aspiration tube is applied to better match the inner lumen of the needle cannula. This modification is not a new technology and does not raise new types of safety or effectiveness questions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                          | In addition to the modifications described above, the Indication for Use statement has been modified from the predicate device to include transvaginal ultrasound and laparoscopic approaches, both of which were specifically stated within the instructions for the predicate device, just not stated directly within the Indications for Use statement. This modification to the Indication for Use statement does not impact safety and efficacy of the device.                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                          | The Mouse Embryo Assay (MEA) specification has changed from one-cell (75% or greater blastocyst rate at 96 h) to 2-cell (80% or greater blastocyst rate at 72 h). This is a change in performance specification, and does not represent a new technology, therefore the safety and effectiveness is not altered.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | A full comparison is provided in Table 1 over the page.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{4}------------------------------------------------

{5}------------------------------------------------

## Table 1 - Comparison of Otrieva and predicate device

| Device Name                            | Otrieva™ Tapered<br>Ovum Aspiration<br>Needle                                                                                                                                                                                                    | Ovum Pick-Up<br>Aspiration Needle                                                                                                                                             | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use                     | The Cook Otrieva™<br>Tapered Ovum<br>Aspiration Needles are<br>used for laparoscopic<br>or ultrasound guided<br>transvaginal aspiration<br>and flushing of oocytes<br>from ovarian follicles.                                                    | The Ovum Pick-Up<br>Aspiration Needles are<br>used for aspiration and<br>flushing of oocytes from<br>ovarian follicles.                                                       | The Otrieva Indication for<br>Use has the addition of<br>ultrasound and<br>laparoscopic approaches<br>explicitly stated. The<br>predicate device is<br>intended to be used<br>laparoscopically and with<br>transvaginal ultrasound,<br>as described in the<br>predicate 510(k)<br>summary, and these<br>approaches were also<br>included in the cleared<br>labelling for the predicate<br>device.<br>This is not a new<br>intended use, as the<br>clinical practise has<br>always included<br>transvaginal ultrasound<br>and laparoscopy. |
| Contraindications                      | This device should not<br>be used on a patient<br>with an active vaginal<br>or intrauterine infection,<br>a sexually transmitted<br>disease, a recent<br>uterine perforation, a<br>recent caesarean<br>section, or who is<br>currently pregnant. | This device should not<br>be used on a patient<br>with an active vaginal or<br>intrauterine infection, a<br>sexually transmitted<br>disease, or who is<br>currently pregnant. | These patient populations<br>were added to ensure<br>consistency with clinical<br>practices for in vitro<br>fertilization.                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Description                    | Needle passed through<br>transvaginal ultrasound<br>transducer or placed<br>through a cannula for a<br>laparoscopic procedure<br>to advance into the<br>ovarian follicle.                                                                        | Needle passed through<br>transvaginal ultrasound<br>transducer or placed<br>through a cannula for a<br>laparoscopic procedure<br>to advance into the<br>ovarian follicle.     | The description of the<br>Otrieva™ is substantially<br>equivalent to the<br>predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Dimensions                             |                                                                                                                                                                                                                                                  |                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Gage                                   | 17 G tapering to 20 G                                                                                                                                                                                                                            | 12 G - 20 G                                                                                                                                                                   | The Otrieva™ Tapered<br>Ovum Aspiration Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Cannula Inner<br>Diameter              | 0.60 mm                                                                                                                                                                                                                                          | 2.10 mm - 0.60 mm                                                                                                                                                             | has an increased wall<br>thickness at the 17 G<br>proximal portion of the                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Needle Length                          | 30 & 35 cm                                                                                                                                                                                                                                       | 20 cm - 40 cm                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Aspiration Tubing<br>Length            | 90 cm                                                                                                                                                                                                                                            | 25cm - 90 cm                                                                                                                                                                  | needle cannula which is<br>necessary to ensure an<br>appropriate wall<br>thickness at the distal,<br>tapered end of the needle<br>while maintaining a 0.60 mm<br>inner diameter.                                                                                                                                                                                                                                                                                                                                                          |
| Device Name                            | Otrieva™ Tapered<br>Ovum Aspiration<br>Needle                                                                                                                                                                                                    | Ovum Pick-Up<br>Aspiration Needle                                                                                                                                             | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Aspiration Tubing<br>External Diameter | 1.32 mm                                                                                                                                                                                                                                          | 1.67 mm - 2.7 mm<br>(5 Fr - 8 Fr)                                                                                                                                             | The diameter of the<br>Otrieva aspiration tubing<br>is smaller than the<br>predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Vacuum Tubing<br>Length                | 50 cm                                                                                                                                                                                                                                            | N/A                                                                                                                                                                           | Vacuum tubing has been<br>added to the design of the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Vacuum Tubing<br>External Diameter     | 2.06 mm                                                                                                                                                                                                                                          | N/A                                                                                                                                                                           | Otrieva™ Tapered Ovum<br>Aspiration Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Materials                              |                                                                                                                                                                                                                                                  |                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Needle Cannula                         | Stainless Steel #304                                                                                                                                                                                                                             | Stainless Steel #304                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Needle Hub                             | Polycarbonate                                                                                                                                                                                                                                    | Polycarbonate                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Aspiration Tubing                      | PTFE                                                                                                                                                                                                                                             | PTFE                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Vacuum Tubing                          | FEP                                                                                                                                                                                                                                              | N/A                                                                                                                                                                           | Vacuum tubing has been<br>added to the design of the<br>Otrieva™ Tapered Ovum<br>Aspiration Needle                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Test Tube Stopper                      | Silicone                                                                                                                                                                                                                                         | Silicone                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Vacuum Line Female<br>Conical Fitting  | Copolyester or<br>Polycarbonate                                                                                                                                                                                                                  | N/A                                                                                                                                                                           | Vacuum tubing has been<br>added to the design of the<br>Otrieva™ Tapered Ovum<br>Aspiration Needle                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Vacuum Line Cannula                    | Stainless steel #304                                                                                                                                                                                                                             | Stainless steel #304                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Other characteristics                  |                                                                                                                                                                                                                                                  |                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Lumens                                 | Single                                                                                                                                                                                                                                           | Single or Double                                                                                                                                                              | Otrieva™ Tapered Ovum<br>Aspiration Needle lumen<br>type falls within the range<br>of the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Echo-tip                               | Echo-Tip only                                                                                                                                                                                                                                    | With or without Echo-Tip                                                                                                                                                      | Otrieva™ Tapered Ovum<br>Aspiration Needle lumen<br>type falls within the range<br>of the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Shelf Life                             | 3 years                                                                                                                                                                                                                                          | 3 years                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Single Use/Reusable                    | Single Use                                                                                                                                                                                                                                       | Single Use                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterility                              | Sterile (SAL 10-6)                                                                                                                                                                                                                               | Sterile (SAL 10-6)                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Endotoxin<br>Specification             | USP endotoxin (LAL)<br>tested and passed with<br>20EU or less per<br>device                                                                                                                                                                      | USP Endotoxin (LAL)<br>tested and passed with<br>20EU or less per device                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| MEA specification                      | Two-cell MEA tested<br>and passed with 80%<br>or greater Blastocyst<br>rate at 72h.                                                                                                                                                              | Once cell MEA tested<br>and passed with 75% or<br>greater Blastocyst at 96<br>h.                                                                                              | MEA test has changed<br>from 1-cell MEA (75% or<br>greater Blastocyst at 96<br>h) to 2-cell MEA (80% or<br>greater Blastocyst at 72<br>h). Both techniques are<br>acceptable and provide<br>indication of embryo<br>toxicity.                                                                                                                                                                                                                                                                                                             |
| Sterilization Method                   | Ethylene Oxide                                                                                                                                                                                                                                   | Ethylene Oxide                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Packaging                              | Polyester/low density<br>polyethylene film with<br>Tyvek-Poly pouch                                                                                                                                                                              | Tyvek-Poly pouch                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{6}------------------------------------------------

{7}------------------------------------------------

| Non-clinical testing<br>summary | The Otrieva™ Tapered Ovum Aspiration Needle has undergone non-clinical testing to verify the design modifications did not affect the performance of the needle. The testing included  Resistance to breakage testing Air tightness testing Tensile strength testing between the tubing and handle junction Leakage testing |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                 | The design modifications were shown to have no adverse effect on the needle performance in any of these tests.                                                                                                                                                                                                             |
| Safety and Efficacy             | The results of the testing provide reasonable assurance that the Otrieva™ Tapered Ovum Aspiration Needle is as safe and effective as the predicate device and support a determination of substantial equivalence.                                                                                                          |

---

**Source:** [https://fda-staging.innolitics.com/device/K141365](https://fda-staging.innolitics.com/device/K141365)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).