IN ANATOMIC: The Aequalis Fx2 humeral stem combined with the Flex Shoulder System humeral head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement. The Aequalis Fx2 is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and stability and to relieve pain. The Aequalis Fx2 is indicated for use as a replacement of shoulder joints disabled by: Rheumatoid arthritis with pain Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) Correction of functional deformity Fractures of the proximal humerus Traumatic arthritis Revision of other devices if sufficient bone stock remains IN REVERSE: The Aequalis Fx2 is indicated for use as a replacement of shoulder joints with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: Rheumatoid arthritis Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) Correction of functional deformity Fractures of the proximal humerus Traumatic arthritis Revision of the devices if sufficient bone stock remains. The reversed insert is permitted to be used in the transformation to reverse Aequalis Fx2 without the removal of the humeral stem, and if it is well fixed, during a revision surgery, for patient with a functional deltoid muscle.
Device Story
Aequalis Fx2 is a non-constrained shoulder prosthesis for total or partial glenohumeral replacement. System includes humeral stem (titanium alloy, HAP-coated) and either Flex Shoulder System humeral heads (anatomic) or reversed inserts (UHMWPE/titanium) for reverse configuration. Used in orthopedic surgery for fracture management, primary replacement, or revision. Allows conversion from anatomic to reverse configuration without stem removal if well-fixed. Implanted by orthopedic surgeons in clinical settings. Provides joint mobility, stability, and pain relief. Humeral stem is for cemented use; glenoid sphere uses screw fixation; all components are single-use.
Clinical Evidence
Bench testing only. Testing included pull-out, torque, and fatigue testing to validate locking mechanisms and fatigue strength in both anatomic and reversed configurations.
Technological Characteristics
Humeral stem: Titanium alloy (ISO 5832-3) with HAP coating (ASTM F1185). Reversed insert: UHMWPE (ISO 5834-2) and titanium alloy. Stem fixation: Cemented (distal) and uncemented (proximal). Sterilization: Terminal sterilization. Dimensions: Stem lengths 90-210mm; insert diameters 36/42mm.
Indications for Use
Indicated for patients requiring shoulder joint replacement due to rheumatoid arthritis, non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), functional deformity, proximal humerus fractures, traumatic arthritis, or revision surgery with sufficient bone stock. Anatomic configuration requires intact/reconstructable rotator cuff. Reverse configuration requires functional deltoid muscle and massive, non-repairable rotator cuff tear.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract image of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2014
TORNIER S.A.S. Maud Andriollo Regulatory Affairs Specialist 161 rue Lavoisier 38330 Montbonnot Saint Martin FRANCE
Re: K141345
Trade/Device Name: Aequalis Fx2 Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, HSD Dated: October 28, 2014 Received: October 30, 2014
Dear Maud Andriollo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
## Page 2 – Maud Andriollo
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
#### 510(k) Number (if known)
#### K141345
Device Name Aequalis Fx2
#### Indications for Use (Describe)
#### IN ANATOMIC:
The Aequalis Fx2 humeral stem combined with the Flex Shoulder System humeral head may be used by themselves, as a hemiarthroplasty. if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement. The Aequalis Fx2 is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and stability and to relieve pain. The Aequalis Fx2 is indicated for use as a replacement of shoulder joints disabled by:
- Rheumatoid arthritis with pain
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the proximal humerus
- Traumatic arthritis
- Revision of other devices if sufficient bone stock remains
#### IN REVERSE:
The Aequalis Fx2 is indicated for use as a replacement of shoulder joints with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- Rheumatoid arthritis
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the proximal humerus
- Traumatic arthritis
- Revision of the devices if sufficient bone stock remains.
The reversed insert is permitted to be used in the transformation to reverse Aequalis Fx2 without the removal of the humeral stem, and if it is well fixed, during a revision surgery, for patient with a functional deltoid muscle.
#### Notes:
- all components are single use.
- the humeral stem is for cemented use only.
- the all-poly glenoid components are intended for cemented use only.
- the glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "TORNIER" in large, bold, blue font. Below that, in a smaller, bold, blue font, are the words "Implants Chirurgicaux". The words are stacked on top of each other and centered.
# 510 (k) Summary of Safety and Effectiveness information Traditional 510(k) Premarket – Aequalis Fx2
Date: November, 14th 2014
| 1) Device name | |
|----------------------|---------------------|
| Trade name: | Aequalis Fx2 |
| Common name: | Shoulder Prosthesis |
| Classification name: | |
- · Shoulder joint metal/polymer semi-constrained cemented prosthesis are class II devices under 21 CFR 888.3660 (product code KWS) and are classified by the Orthopedic Devices Panel
- · Shoulder joint metal/polymer semi-constrained cemented prosthesis are class II devices under 21 CFR 888.3660 (product code PHX)
- Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis are class II devices under . CFR 888.3690 (product code HSD) and are classified by the Orthopedic Devices Panel
### 2) Submitter :
TORNIER SAS 161 rue Lavoisier 38330 Montbonnot Saint Martin- France Registration Number: 3000931034
#### 3) Company contact :
Tornier Mrs Maud Andriollo Regulatory Affairs Specialist 161 rue Lavoisier 38334 Montbonnot Tel: 00 33 4 76 61 35 00 Fax: 00 33 4 76 61 35 59 e-mail:maud.andriollo@tornier.com
#### 4) Classification
| Device class: | Class II |
|-----------------------|---------------|
| Classification panel: | Orthopedic |
| Product code: | KWS, PHX, HSD |
# 5) Equivalent / Predicate device : Aequalis Ascend Flex Shoulder System, TORNIER SAS, K122698
Image /page/3/Picture/14 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is formed by geometric shapes, creating a modern and abstract design. The hexagon and the 'T' are both in a light blue or lavender color, contrasting with the white background. The logo appears clean and simple, suitable for a tech company or a brand with a focus on innovation.
TORNIER S.A.S. 161, rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the name "TORNIER" in large, bold, blue font. Below that, in a smaller, bold, blue font, is the phrase "Implants Chirurgicaux". The text appears to be a logo or brand name for a company that produces surgical implants.
Aequalis Shoulder Fracture System, TORNIER SAS, K060209, K131231 Aequalis Reversed Fracture Shoulder Prosthesis, TORNIER SAS, K082120, K131231 Comprehensive Reverse Shoulder System, BIOMET, K113069
### 6) Device description :
The Aequalis Fx2 is a non-constrained prosthesis intended for the total or partial replacement of the glenohumeral articulation.
The Aequalis Fx2 consists of:
in an anatomic configuration, a humeral stem compatible with Flex Shoulder System humeral heads ; । Or
- in a reversed configuration, a humeral stem and a reversed insert, compatible with Aequalis । Reversed/Aequalis Reversed II glenoid implants.
The Aequalis Fx2 is intended for use as:
- । traumatic or pathologic conditions of the shoulder resulting in fracture of the glenohumeral joint, including humeral head fracture and displaced 3-or 4-part proximal humeral fractures,
- in case of bone defect in the proximal part of the humerus, ।
- a replacement of shoulder joints in primary anatomic or in primary reverse, ।
- a replacement of other shoulder joints devices in case of revisions if sufficient bone stock remains. -
The Aequalis Fx2 also allows for conversions from anatomic to reverse shoulder prosthesis in case of revision.
### 7) Materials :
The humeral stem is manufactured from titanium alloy (Ti6Al4V) in accordance with ISO 5832-3 and coated with hydroxylapatite (HAP) in accordance with ASTM standard F-1185 or ISO 13779-2.
The reversed insert is manufactured from UHMWPE according to ISO standard 5834-2 and titanium alloy (Ti6Al4V) in accordance with ISO 5832-3.
# 8) Indications :
#### IN ANATOMIC:
The Aequalis Fx2 humeral stem combined with the Flex Shoulder System humeral head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement.
The Aequalis Fx2 is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain. The Aequalis Fx2 is indicated for use as a replacement of shoulder joints disabled by:
- Rheumatoid arthritis with pain -
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) ।
- Correction of functional deformity ।
- Fractures of the proximal humerus ।
- -Traumatic arthritis
- Revision of other devices if sufficient bone stock remains. -
Image /page/4/Picture/26 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is constructed from geometric shapes, giving it a modern and abstract appearance. The color of the logo is a light shade of blue or lavender, providing a soft and clean aesthetic. The hexagon shape provides a clear border for the logo.
TORNIER S.A.S. 161, rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Tornier Implants Chirurgicaux. The word "TORNIER" is in large, bold, blue font on the top line. The words "Implants Chirurgicaux" are in a smaller, bold, blue font on the second line.
### IN REVERSE:
The Aequalis Fx2 is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- Rheumatoid arthritis ।
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) ।
- । Correction of functional deformity
- Fractures of the proximal humerus -
- -Traumatic arthritis
- Revision of the devices if sufficient bone stock remains. ।
The reversed insert is permitted to be used in the transformation from anatomic to reverse Aequalis Fx2 without the removal of the humeral stem, and if it is well fixed, during a revision surgery, for patient with a functional deltoid muscle.
#### Notes:
- all components are single use. ।
- the humeral stem is for cemented use only. -
- the all-poly glenoid components are intended for cemented use only. ।
- । the glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
- the humeral head diameters 37 and 39mm must be utilized only with the stem with a small metaphysis. -The humeral head must cover completely the stem metaphysis.
Image /page/5/Picture/16 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is constructed from geometric shapes, giving it a modern and abstract appearance. The hexagon and the 'T' are both colored in a light blue or lavender hue. The logo is simple and clean, suggesting a focus on design and possibly technology or innovation.
TORNIER S.A.S. 161, rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image displays the word "TORNIER" in large, bold, blue font. Below this, the words "Implants Chirurgicaux" are written in a smaller, blue font. The text appears to be a logo or heading, possibly for a medical or surgical company specializing in implants.
#### 9) Summary of technological characteristics
### Table 1: Main features comparison
| Main features or<br>system characteristics | | Aequalis Fx2 | Aequalis Ascend Flex<br>Shoulder System | Aequalis Shoulder<br>Fracture System | Aequalis Reversed<br>Fracture Shoulder<br>Prosthesis | Comprehensive Reverse<br>Shoulder System |
|--------------------------------------------|--------------------|-----------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------|------------------------------------------|
| Material | Stem | Titanium + HAP | Titanium + Ti Plasma<br>Spray | Titanium + HAP | Titanium + HAP | Titanium + Ti Porous<br>Plasma spray |
| | Reversed<br>Insert | UHMWPE + Titanium | UHMWPE + Titanium | NA | UHMWPE | Titanium + UHMWPE |
| Standard | Stem | ASTM F136 / ISO 5832-3 | ASTM F136 / ISO 5832-3 | ASTM F136 / ISO 5832-3 | ASTM F136 / ISO 5832-3 | unknown |
| | Coating | ASTM F1185 | ASTM F1580 | ASTM F1185 | ASTM F1185 | ASTM F1580 |
| | Reversed<br>Insert | ISO 5834-2 + ISO 5832-3 | ISO 5834-2 + ISO 5832-3 | NA | ISO 5834-2 | unknown |
| Stem fixation | | Cemented on distal part<br>Uncemented on proximal<br>part | Cemented or uncemented | Cemented on distal part<br>Uncemented on proximal<br>part | Cemented use | Cemented or uncemented |
| Stem length | | 90; 130; 170; 210 mm | 66; 70; 74; 78; 82; 86; 88;<br>90; 93; 94; 98; 104; 109;<br>115; 120; 125 mm | 130; 170; 180; 210 mm | 130; 170; 180; 210 mm | 55; 83; 122; 194 mm |
| Diameter of reversed<br>insert | | 36mm, 42mm<br>Thickness 6; 9; 12 mm | 36mm, 42mm<br>Thickness 6; 9 mm | NA | 36mm, 42mm<br>Thickness 6; 9; 12mm | 31mm, 36mm, 41mm |
| Terminal sterilization | | Yes | Yes | Yes | Yes | Yes |
| Manufacturer | | Tornier | Tornier | Tornier | Tornier | Biomet |
| K-number | | Pending | K122698 | K060209, K131231 | K082120, K131231 | K113069 |
Image /page/6/Picture/4 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is rendered in a light purple or lavender color and appears to be constructed from geometric shapes, giving it a modern and somewhat abstract look. The hexagon provides a clean, geometric border around the 'T', further emphasizing the logo's contemporary design. The background is plain white, which makes the logo stand out.
TORNIER S.A.S. Tél. : 33 (0)4 76 61 35 00 161, rue Lavoisier Fax : 33 (0)4 76 61 35 33 38334 MONTBONNOT CEDEX FRANCE
S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the words "TORNIER Implants Chirurgicaux" in a dark blue font. The word "TORNIER" is in a larger font than the words "Implants Chirurgicaux". The words are stacked on top of each other, with "TORNIER" on top and "Implants Chirurgicaux" on the bottom. The background is white.
The design, the indication for use, the manufacturing principle, the method of fixation of the stem, the packaging and the sterilization process of the Aequalis Fx2 are identical or equivalent to the predicate devices.
### 10) Non-clinical testing
Pull out, torque testing, and fatigue testing were performed.
- The aim of pull out and torque testing is to validate the locking mechanism between a metal part simulating an Aequalis Fx2 humeral stem and an UHMWPE simulating reversed insert.
- -The aim of fatigue testing in anatomic configuration is to evaluate the fatigue strength of the Aequalis Fx2 implant in an anatomic configuration.
- -The aim of fatigue testing in reversed configuration is to evaluate the fatigue strength of the Aequalis Fx2 implant in a reversed configuration after anatomic configuration.
The results of these tests demonstrate the equivalence between the Aequalis Fx2 and the predicate devices.
### 11) Substantial equivalence conclusion
Based upon this comparative study, substantial equivalence of the Aequalis Fx2 to the predicates can be demonstrated on the following grounds, according to the FDA's Guidelines for Substantial Equivalence Decision Making Process:
- -The Aequalis Fx2 is compared to the predicate devices.
- The Aequalis Fx2 has the same intended use as the cleared predicates. -
- -Major technological characteristics are equivalent between the Aequalis Fx2 and the predicate devices:
- Equivalence of general features -
- -Equivalent means of fixation
- -Equivalent material
- -Equivalent prosthetic dimensions
Therefore, in the light of the above information, the Aequalis Fx2 is found to be equivalent to the predicate devices.
Image /page/7/Picture/18 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is rendered in a light blue or lavender color and appears to be constructed from geometric shapes, giving it a modern and somewhat abstract look. The hexagon provides a clear, defined border for the logo, making it stand out.
TORNIER S.A.S. 161, rue Lavoisier 38334 MONTBONNOT CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 35 043 008 € SIRET : 070 501 275 000 21 R.C.S. : 070 501 275 CODE APE : 3250 A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.