TEVADAPTOR; CONNECTING SET W/ ULTRASITE INJECTION SITE, SPIKE PORT ADAPTOR SET W/ ULTRASITE INJECTION SITE
K141306 · Teva Medical Ltd., Migada Plant · FPA · Jun 12, 2014 · General Hospital
Device Facts
Record ID
K141306
Device Name
TEVADAPTOR; CONNECTING SET W/ ULTRASITE INJECTION SITE, SPIKE PORT ADAPTOR SET W/ ULTRASITE INJECTION SITE
Applicant
Teva Medical Ltd., Migada Plant
Product Code
FPA · General Hospital
Decision Date
Jun 12, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The I.V. Administration Set is a single use, sterile I.V. set for administration of drugs and/or fluids from a container to a patient vascular system.
Device Story
The Teva Medical I.V. Administration Set is a single-use, sterile, non-pyrogenic device designed to deliver intravenous fluids or medications from a container (e.g., I.V. bag) to a patient's vascular system. The device consists of generic components including a hollow spike for bag penetration, tubing, Y-sites, clamps, Luer connectors, and needleless injection sites (specifically ULTRASITE). The subject submission introduces two new configurations: a connecting set and a spike port adaptor, both incorporating the ULTRASITE injection site. These devices are used in clinical settings to facilitate fluid delivery. The healthcare provider connects the set to the fluid source and the patient's vascular access device. The device functions as a passive conduit for fluid administration; it does not contain active electronic components, sensors, or software. The design changes relative to the predicate include shorter tubing lengths, the removal of drip chambers/flow control/air venting in the new configurations, and the addition of a 'twist-off' component.
Clinical Evidence
No clinical data. Bench testing only. The manufacturer performed verification and validation of design changes under their quality management system in compliance with 21 CFR 820.
Technological Characteristics
Single-use, sterile, non-pyrogenic I.V. administration set. Components include spike, Y-site, tubing, clamp, needleless injection site (ULTRASITE), twist-off, and Luer connections. Materials are identical to those previously cleared in K121269 and K071741. No electronic components, software, or energy sources.
Indications for Use
Indicated for the administration of drugs and/or fluids from a container to a patient's vascular system via a needle or catheter.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains two logos. The first logo on the left is a simple, abstract design featuring the letters "VN" in a bold, sans-serif font. The second logo on the right is more complex, with the word "Medical" in a stylized font at the top. Below "Medical" is smaller text that reads "1993 Medical Ltd."
# Appendix E
# JUN 1 2 2014
ﺮ
510(k) Summary in accordance with 21 CFR 807.92(c)
| Device Name: | Teva Medical I.V. Administration Set |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Special |
| Date of Submission: | 16 May 2014 |
| Manufacturer: | Teva Medical Ltd., MIGADA Plant<br>North Industrial Zone<br>Kiryat Shmona 10258<br>ISRAEL |
| FDA Registration Number: | 9611423 |
| Owner/Operator Number: | 9001925 |
| 510(k) Owner: | Teva Medical Ltd., MIGADA Plant<br>North Industrial Zone<br>Kiryat Shmona 10258<br>ISRAEL |
| Phone:<br>Fax: | +972 (0)46 908822<br>+972 (0)39 267865 |
| 510(k) Submitter and Contact: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy<br>Phone: +39 06 578 2665<br>Fax: +30 06 574 3786<br>Email: rgray@donawa.com |
| FDA Product Code: | FPA |
| FDA Regulation Number: | 880.5440 |
| FDA Classification Name: | Set, Administration, Intravascular |
| Classification Panel: | General Hospital |
| Common Name: | I.V. Administration Set |
| FDA Classification: | Class II |
| FDA Identification: | An intravascular administration set is a device used to<br>administer fluids from a container to a patient's vascular system<br>through a needle or catheter inserted into a vein. The device. |
Special 510(k) for IV Teva Medical Administration Set
(
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the text "Medical" in a stylized font. The word "Medical" is in white text on a black background. Below "Medical" is the text "love medical" in a smaller font. To the left of the text is a symbol that appears to be "VN".
may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
The I.V. Administration Set is a single use, sterile I.V. set for administration of Indications for Use: drugs and/ or fluids from a container to a patient vascular system.
#### Device Description:
The I.V. Administration Set is single use, sterile, non-pyrogenic device used to administer intravenous solutions and/or drugs solutions from a container to a patient's vascular system.
The I.V Administration set is comprised of various generic components which are broadly used through the industry such as: Spike, Y-site, tubing, clamp, needless injection site, 'twist-off' and Luer connection.
The purpose of this Special 510(k) is to add the following two new designs of IV Administration Sets to the one already cleared for sale in the US under K121269:
- TEVADAPTOR Connecting Set with ULTRASITE Injection Site �
- TEVADAPTOR Spike Port Adaptor with ULTRASITE Injection Site .
All components used in the two new Administration Sets were used in the predicate device cleared under K121269, except for two, which are identical to components cleared under K071741.
Because none of the materials are new, and all components have been previously cleared for use in V administration sets, no additional biocompatibility data is included in this submission.
Comparison with predicate device:
The predicate device selected for comparison with the I.V. Administration Set is:
| Predicate Device: | I.V. Administration Set |
|----------------------|------------------------------------|
| 510(k) Sponsor: | Teva Medical |
| 510(k) Number: | K121269 |
| Clearance Date: | 22 January 2013 |
| FDA Product Code: | FPA |
| Classification Name: | Set, Administration, Intravascular |
| Regulation No: | 880.5440 |
The following aspects of the subject and predicate devices are identical:
- . Indications for use
- . Fundamental technology
- Single use only .
- . Sterilization
- Biocompatibility �
- . Interconnecting features
Special 510(k) for IV Teva Medical Administration Set
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains two distinct elements. On the left, there's a stylized graphic that appears to be a logo or symbol. To the right, there's a rectangular graphic with the word "Medical" prominently displayed in a stylized font, with the words "Iowa Medical, Ltd." underneath it.
- Interaction with patient and other devices .
- . Safety features
The following aspects of the devices are different:
- The lengths of the new devices are shorter than the predicate. .
- No drip chamber, flow control or air venting in the new devices. �
- Addition of the twist-off component in one of the new devices.
- Labeling has changed as a result of the above differences. 해
- . Packaging materials and process has changed
The two additional (subject) devices have been reviewed under the control of Teva Medical's quality management system and none of the changes from the predicate device have been identified as having any significant effect on safety and effectiveness compared with the original FDA-cleared device. Where verification/validation of applicable changes was required, these have been carried out under the control of Teva Medical's quality system, which is in compliance with 21 CFR 820.
#### Substantial Equivalence Conclusion:
Based on the information contained within this submission, the additional subject device administration sets do not raise any additional safety and effectiveness issues. It is concluded that the additional devices are substantially equivalent to the identified predicate device cleared under K121269, which is already in interstate commerce within the USA.
・・
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or protecting another figure, represented by three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2014
:
Teva Medical Limited, MIGADA Plant C/O Mr. Roger Gray Donawa Lifescience Consulting Piazza Albania 10 00153 Rome Italy
Re: K141306
Trade/Device Name: Teva Medical I.V. Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Set, Administration, Intravascular Regulatory Class: II Product Code: FPA Dated: May 16, 2014 Received: May 19, 2014
Dear Mr. Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mary S. Runner -S
· Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
.
* *
*
*
.
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K141306
Device Name
Teva Medical I.V. Administration Set
Indications for Use (Describe)
The I.V. Administration Set is a single use, sterile I.V. set for administration of drugs and/or fluids from a container to a patient vascular system.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY |
|------------------------------------------------------------------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) |
| Digitally signed by |
| Richard C. Chapman -S |
| Date: 2014.06.12 |
| 13:24:34 -04'00' |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.