The Vitrea® CT Multi-Chamber CFA option is intended to be used with CT studies of the heart to assist cardiologists and radiologists in assessing function when producing a cardiac evaluation. The CT Multi-Chamber CFA option includes semi-automatic heart segmentation including three chambers (left ventricle, right ventricle, and left atrium) segmentation, including identification of long axis and mitral valve boundaries across multiple phases; calculation of global metrics, including end diastolic volume, end systolic volume, stroke volume, ejection fraction, cardiac output, cardiac index, stroke index, and myocardial mass; and calculation of regional metrics; including wall motion, percentage of wall thickening, regional ejection fraction, and polar plots.
Device Story
The Vitrea CT Multi-Chamber Cardiac Functional Analysis (CFA) is a post-processing software application for the Vitrea platform. It processes DICOM 3.0 CT angiographic images to assist cardiologists and radiologists in cardiac evaluation. The software performs semi-automatic segmentation of the left ventricle, right ventricle, and left atrium, identifying long axis and mitral valve boundaries across multiple phases. It calculates global metrics (e.g., ejection fraction, cardiac output, myocardial mass) and regional metrics (e.g., wall motion, polar plots). Used in medical facilities on radiology workstations, the device provides clinicians with quantitative data and visual reports to aid in diagnosing and managing cardiovascular disease. It benefits patients by providing non-invasive, automated functional assessments of multiple heart chambers.
Clinical Evidence
No clinical studies were required. Evidence consists of bench testing, including software verification (performance and safety testing) and validation (simulated usability testing by experienced medical professionals).
Technological Characteristics
Software-based image processing application for radiology workstations. Operates on DICOM 3.0 compliant CT datasets. Features semi-automatic segmentation, multi-planar reformatting (MPR), and automated cardiac parameter calculation. Complies with NEMA PS 3.1-3.20 (DICOM), ISO 14971:2007 (Risk Management), and IEC 62304:2006 (Software Life Cycle).
Indications for Use
Indicated for cardiologists and radiologists to assess cardiac function in patients with suspected or known cardiac diseases using CT studies of the heart.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Vitrea 2, version 3.7 Medical Image Processing Software - Cardiac Functional Analysis (K043333)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that overlap and create a sense of movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1. 2014
VITAL IMAGES, INC. PARTHIV SHAH SR. REGULATORY AFFAIRS SPECIALIST 5850 OPUS PARKWAY SUITE 300 MINNETONKA MN 55343
Re: K141302
Trade/Device Name: Vitrea CT Multi-Chamber Cardiac Functional Analysis (CFA) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 2, 2014 Received: July 7, 2014
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh.P)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K141302
# Device Name
Vitrea® CT Multi-Chamber Cardiac Functional Analysis (CFA)
# Indications for Use (Describe)
The Vitrea® CT Multi-Chamber CFA option is intended to be used with CT studies of the heart to assist cardiologists and radiologists in assessing function when producing a cardiac evaluation. The CT Multi-Chamber CFA option includes semi-automatic heart segmentation including three chambers (left ventricle, and left atrium) segmentation, including identification of long axis and mitral valve boundaries across multiple phases; calculation of global metrics, including end diastolic volume, end systolic volume, ejection fraction, cardiac output, cardiac index, stroke index, and myocardial mass; and calculation of regional metrics; including wall motion, percentage of wall thickening, regional ejection fraction, and polar plots.
# Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/15 description: The image shows a signature over the letters FDA. The signature is illegible, but it appears to be written in cursive. The letters FDA are in a light gray color and are in the background. The signature is in black and is in the foreground.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officen Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in smaller, red letters. The word "VITAL" is the main focus of the image, with the company affiliation provided as additional context.
This 510(k) summary is submitted in accordance with the requirements of 21 C.F.R. Part 807.92.(c)
| Purpose of<br>Submission: | Vital Images, Inc. hereby submits this traditional 510(k) to provide a<br>notification submission of a new software application preset called "CT<br>Multi-Chamber Cardiac Functional Analysis (CFA)". This new application<br>has been added into the Cardiac Functional CT protocol available on the<br>Vitrea platforms to provide cardiac functions of Right Ventricle (RV) and<br>Left Atrium (LA) chambers of heart in addition to the previously cleared<br>cardiac evaluation of Left Ventricle (LV) chamber of heart. |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Vital Images, Inc.<br>5850 Opus Parkway<br>Suite 300<br>Minnetonka, MN 55343-4414 |
| Establishment<br>Registration: | 2134213 |
| Contact Person: | Parthiv Shah<br>Sr. Regulatory Affairs Specialist<br>Phone : 952-487-9574<br>Fax: 952-487-9510<br>E-mail: pshah@vitalimages.com |
| 510(k) Type: | Traditional |
| Summary Date: | July 2, 2014 |
| Device Name<br>Trade Name: | Vitrea® CT Multi-Chamber Cardiac Functional Analysis (CFA) |
| Common Name: | System, Image Processing, Radiological |
| Classification Name: | System, Image Processing, Radiological (21 C.F.R. 892.2050, LLZ) |
| Regulatory Description: | Picture Archiving and Communications System |
# Predicate Devices:
| Predicate Device | Manufacturer | FDA 510(k) number |
|--------------------------------------------------------------------------------------------------------------------------|------------------------|-------------------|
| Vitrea 2, version 3.7 Medical Image<br>Processing Software - Cardiac<br>Functional Analysis<br>(Legally Marketed Device) | Vital Images, Inc. | K043333 |
| CardIQ Function Xpress<br>(Legally Marketed Device) | GE Medical Systems SCS | K073153 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The letters in "VITAL" are all capitalized except for the "i", which is lowercase and has a dot above it.
# Device Description:
The Vitrea CT Multi-Chamber Cardiac Functional Analysis (CFA) software application is a new post processing licensable software preset available on the Cardiac Functional CT protocol for the Vitrea Platforms. This product can be used in the analysis of CT angiographic images to calculate and display several functional cardiac parameters. The software has the ability to select the chambers of the heart and diastolic and/or systolic phases to evaluate the heart's function. The Vitrea CT Multi-Chamber CFA software application contains both graphic and text report capabilities with predefined templates for ease of use.
# Intended Use / Indications for Use:
The Vitrea® CT Multi-Chamber CFA option is intended to be used with CT studies of the heart to assist cardiologists and radiologists in assessing function when producing a cardiac evaluation. The CT Multi-Chamber CFA option includes semi-automatic heart segmentation including three chambers (left ventricle, right ventricle, and left atrium) segmentation, including identification of long axis and mitral valve boundaries across multiple phases; calculation of global metrics, including end diastolic volume, end systolic volume, stroke volume, ejection fraction, cardiac index, stroke index, and myocardial mass; and calculation of regional metrics; including wall motion, percentage of wall thickening, regional ejection fraction, and polar plots.
# Intended for Disease / Condition / Patient Population:
The software assists cardiologists and radiologists during a cardiac evaluation to assess various cardiac functions of patients with suspected or known cardiac diseases.
# Substantial Equivalence:
The Vitrea CT Multi-Chamber Cardiac Functional Analysis (CFA) software application is substantial equivalent to a combination of the predicate Vital Images, Inc. Cardiac Functional Analysis (K043333) and GE Medical Systems SCS CardIQ Function Xpress (K073153) devices and is safe and effective for use.
# Intended Use Comparison:
# For Cardiac Evaluation of Left Ventricle (LV) Chamber:
| Vitrea CT Multi-Chamber Cardiac<br>Functional Analysis (CFA)<br>(K141302)<br>(Submission Subject) | Vitrea 2, version 3.7 Medical<br>Image Processing Software –<br>Cardiac Functional Analysis<br>(K043333)<br>(Predicate Device-1) | Differences | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------|
| Intended Use / Indications for Use: | | | |
| Vitrea CT Multi-Chamber Cardiac<br>Functional Analysis (CFA)<br>(K141302)<br>(Submission Subject) | Vitrea 2, version 3.7 Medical<br>Image Processing Software –<br>Cardiac Functional Analysis<br>(K043333)<br>(Predicate Device-1) | Differences | |
| The Vitrea® CT Multi-Chamber CFA<br>option is intended to be used with CT<br>studies of the heart to assist<br>cardiologists and radiologists in<br>assessing function when producing a<br>cardiac evaluation.<br><br>The CT Multi-Chamber CFA option<br>includes semi-automatic heart<br>segmentation including three<br>chambers (left ventricle, right<br>ventricle, and left atrium)<br>segmentation, including identification<br>of long axis and mitral valve<br>boundaries across multiple phases;<br>calculation of global metrics, including<br>end diastolic volume, end systolic<br>volume, stroke volume, ejection<br>fraction, cardiac output, cardiac index,<br>stroke index, and myocardial mass;<br>and calculation of regional metrics;<br>including wall motion, percentage of<br>wall thickening, regional ejection<br>fraction, and polar plots. | The CT CFA option is intended to<br>be used with CT studies of the heart<br>to assist cardiologists and<br>radiologists in assessing function<br>when producing a cardiac<br>evaluation.<br><br>The CFA option includes semi-<br>automatic heart and left ventricle<br>segmentation, including<br>identification of long axis and mitral<br>valve boundaries across multiple<br>phases; calculation of global<br>metrics, including end diastolic<br>volume, end systolic volume, stroke<br>volume, ejection fraction, cardiac<br>output, cardiac index, stroke index,<br>and myocardial mass; and<br>calculation of regional metrics;<br>including wall motion, percentage of<br>wall thickening, regional ejection<br>fraction, and polar plots. | The predicate device-1<br>does not provide<br>cardiac evaluation of<br>right ventricle and left<br>atrium compared to the<br>subject device. | |
| Intended Users: | | | |
| Cardiologists and radiologists. | Cardiologists and radiologists. | None | |
| Modality Support: | | | |
| CT | CT | None | |
| For Cardiac Evaluation of Right Ventricle (RV) and Left Atrium (LA) Chambers: | | | |
| Vitrea® CT Multi-Chamber Cardiac<br>Functional Analysis (CFA)<br>(K141302)<br>(Submission Subject) | CardIQ Function Xpress<br>(K073153)<br>(Predicate Device-2) | Differences | |
| Intended Use / Indications for Use: | | | |
| The Vitrea® CT Multi-Chamber CFA<br>option is intended to be used with CT<br>studies of the heart to assist<br>cardiologists and radiologists in<br>assessing function when producing a<br>cardiac evaluation.<br>The CT Multi-Chamber CFA option<br>includes semi-automatic heart<br>segmentation including three<br>chambers (left ventricle, <b>right<br/>ventricle, and left atrium</b> )<br>segmentation, including identification<br>of long axis and mitral valve<br>boundaries across multiple phases;<br>calculation of global metrics, including<br>end diastolic volume, end systolic<br>volume, stroke volume, ejection<br>fraction, cardiac output, cardiac index,<br>stroke index, and myocardial mass;<br>and calculation of regional metrics;<br>including wall motion, percentage of<br>wall thickening, regional ejection<br>fraction, and polar plots. | CardIQ Function Xpress is intended<br>to provide an optimized noninvasive<br>application to analyze<br><b>cardiovascular anatomy and<br/>pathology</b> and aid in determining<br>treatment paths from a set of<br>Computed Tomography (CT)<br>Angiographic images.<br>CardIQ Function Xpress in<br>conjunction with CT cardiac images<br>to automatically calculate and<br>display various left ventricular and<br><b>right ventricular functional<br/>parameters</b> as ejection fraction,<br>end systolic and end diastolic<br>volumes, stroke volumes, wall<br>motion, wall thickening, cardiac<br>output, myocardial mass, systemic<br>and pulmonary vascular resistance.<br><b>Volume measurement of each<br/>chamber of the heart is also<br/>available.</b> With CardIQ Function<br>Xpress atrium volumes may be<br>used to determine volume<br>assessment of atrial disease to<br>include but not limited to atrial<br>fibrillation. CardIQ Function Xpress<br>is a CT, non-invasive image<br>analysis software package, which<br>aids in the assessment of cardiac<br>function and in determination of<br>cardiovascular disease diagnosis<br>and management.<br>CardIQ Function Xpress is for use<br>on the Advantage Workstation (AW)<br>platform, CT Scanner, PAC or<br>Centricity stations, which can be<br>used in the analysis of 2D or 3D CT<br>angiography images/data derived<br>from DICOM 3.0 CT scans. | None | |
| Vitrea CT Multi-Chamber Cardiac<br>Functional Analysis (CFA)<br>(K141302)<br>(Submission Subject) | CardIQ Function Xpress<br>(K073153)<br>(Predicate Device-2) | Differences | |
| Intended Users: | Cardiologists and radiologists. | Cardiologists and radiologists. | None |
| Modality Support: | CT | CT | None |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word, in smaller, red letters, is the phrase "A Toshiba Medical Systems Group Company." The text is centered in the image.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The text is centered and the background is white.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, lighter font. The word "VITAL" is the main focus of the image, with the company affiliation provided as additional context.
# Technological Characteristics Comparison:
| | (Submission<br>Subject) | (Predicate<br>Device-1) | (Predicate<br>Device-2) | |
|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-------------|
| Characteristic | Vitrea CT<br>Multi-Chamber<br>Cardiac<br>Functional<br>Analysis<br>(CFA) | Vitrea 2, version<br>3.7 Medical<br>Image<br>Processing<br>Software –<br>Cardiac<br>Functional<br>Analysis | CardIQ<br>Function<br>Xpress | Differences |
| | | For Cardiac<br>Evaluation of<br>Left Ventricle<br>(LV) Chamber | For Cardiac<br>Evaluation of<br>Right<br>Ventricle (RV)<br>and Left<br>Atrium (LA)<br>Chambers | |
| | (K141302) | (K043333) | (K073153) | |
| Device Description: | | | | |
| Input Data: | | | | |
| Characteristic | (Submission<br>Subject) | (Predicate<br>Device-1) | (Predicate<br>Device-2) | Differences |
| Characteristic | Vitrea CT<br>Multi-Chamber<br>Cardiac<br>Functional<br>Analysis<br>(CFA) | Vitrea 2, version<br>3.7 Medical<br>Image<br>Processing<br>Software –<br>Cardiac<br>Functional<br>Analysis | CardIQ<br>Function<br>Xpress | Differences |
| | | For Cardiac<br>Evaluation of<br>Left Ventricle<br>(LV) Chamber | For Cardiac<br>Evaluation of<br>Right<br>Ventricle (RV)<br>and Left<br>Atrium (LA)<br>Chambers | |
| | (K141302) | (K043333) | (K073153) | |
| Use in the analysis of CT<br>angiographic images to calculate and<br>display analysis of several functional<br>cardiac parameters | Yes | Yes | Yes | None |
| DICOM 3.0 compliant image data | Yes | Yes | Yes | None |
| Where Used: | | | | |
| Medical facility | Yes | Yes | Yes | None |
| User Interface: | | | | |
| Designed for use on a radiology<br>workstation | Yes | Yes | Yes | None |
| Loading DICOM Datasets: | | | | |
| Processes multi-phase, multi-slice<br>cardiac CT images | Yes | Yes | Yes | None |
| Characteristic | (Submission<br>Subject) | (Predicate<br>Device-1) | (Predicate<br>Device-2) | Differences |
| | Vitrea CT<br>Multi-Chamber<br>Cardiac<br>Functional<br>Analysis<br>(CFA) | Vitrea 2, version<br>3.7 Medical<br>Image<br>Processing<br>Software –<br>Cardiac<br>Functional<br>Analysis | CardIQ<br>Function<br>Xpress | |
| | | For Cardiac<br>Evaluation of<br>Left Ventricle<br>(LV) Chamber | For Cardiac<br>Evaluation of<br>Right<br>Ventricle (RV)<br>and Left<br>Atrium (LA)<br>Chambers | |
| | (K141302) | (K043333) | (K073153) | |
| Provides ability to select Diastolic<br>phase | Yes | Yes | Yes | None |
| Provides ability to select Systolic<br>phase | Yes | Yes | Yes | None |
| Automatically identifies the long axis<br>across multiple phases | Yes | Yes | Yes | None |
| Automatically identifies the plane of<br>the Mitral valve | Yes | Yes | Yes | None |
| Segmentation: | | | | |
| Semi-automatic segmentation of Left<br>Ventricle (LV) chamber | Yes | Yes | Yes | None |
| Semi-automatic segmentation of<br>Myocardium of Left Ventricle (LV)<br>chamber | Yes | Yes | Yes | None |
| | (Submission<br>Subject) | (Predicate<br>Device-1) | (Predicate<br>Device-2) | |
| Characteristic | Vitrea CT<br>Multi-Chamber<br>Cardiac<br>Functional<br>Analysis<br>(CFA) | Vitrea 2, version<br>3.7 Medical<br>Image<br>Processing<br>Software –<br>Cardiac<br>Functional<br>Analysis | CardIQ<br>Function<br>Xpress | Differences |
| | | For Cardiac<br>Evaluation of<br>Left Ventricle<br>(LV) Chamber | For Cardiac<br>Evaluation of<br>Right<br>Ventricle (RV)<br>and Left<br>Atrium (LA)<br>Chambers | |
| | (K141302) | (K043333) | (K073153) | |
| Key Functions: | | | | |
| Automatic determination of End<br>Diastolic (ED) and End Systolic (ES)<br>phases by volume | Yes | Yes | Yes | None |
| Volume Rendering | Yes | Yes | Yes | None |
| Phase Navigator | Yes | Yes | Yes | None |
| Semi-automatic determination of<br>Epicardial and Endocardial contours | Yes | Yes | Yes | None |
| Automated Calculation of Cardiac Parameters | | | | |
| End Diastolic Volume (EDV) | Yes | Yes | Yes | None |
| End Systolic Volume (ESV) | Yes | Yes | Yes | None |
| | (Submission<br>Subject) | (Predicate<br>Device-1) | (Predicate<br>Device-2) | |
| Characteristic | Vitrea CT<br>Multi-Chamber<br>Cardiac<br>Functional<br>Analysis<br>(CFA) | Vitrea 2, version<br>3.7 Medical<br>Image<br>Processing<br>Software –<br>Cardiac<br>Functional<br>Analysis | CardIQ<br>Function<br>Xpress | Differences |
| | | For Cardiac<br>Evaluation of<br>Left Ventricle<br>(LV) Chamber | For Cardiac<br>Evaluation of<br>Right<br>Ventricle (RV)<br>and Left<br>Atrium (LA)<br>Chambers | |
| | (K141302) | (K043333) | (K073153) | |
| Stroke Volume (SV) | Yes | Yes | Yes | None |
| Stroke Index (SI) | Yes | Yes | Yes | None |
| Ejection Fraction (EF) | Yes | Yes | Yes | None |
| Cardiac Output (CO) | Yes | Yes | Yes | None |
| Myocardial Mass | Yes | Yes | Yes | None |
| Wall Motion | Yes | Yes | Yes | None |
| Percent Wall Thickening | Yes | Yes | Yes | None |
| Myocardial Volume (MV) | Yes | Yes | Yes | None |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in smaller, red letters. The logo is simple and professional, likely representing a company or organization.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The word "VITAL" is the main focus of the image. The phrase below it provides additional context about the company.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "VITAL" in a bold, sans-serif font. The "i" in "VITAL" is lowercase. Below the word "VITAL" is the text "A Toshiba Medical Systems Group Company" in a smaller, red font. The text is centered below the word "VITAL".
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, li…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.