K141294 · Kimberly-Clark Corp. · HEB · Aug 25, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141294
Device Name
U BY KOTEX SECURITY UNSCENTED MENSTRUAL TAMPONS
Applicant
Kimberly-Clark Corp.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Aug 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
Kimberly-Clark* U by Kotex® Security® is an unscented menstrual tampon intended to be inserted into the vagina to absorb menstrual fluid.
Device Story
Device is a conventional unscented menstrual tampon; consists of absorbent pledget (rayon fibers and cotton, overwrapped with non-woven fabric), withdrawal cord, and two-piece plastic applicator (inner plunger, outer barrel with flexible petals). Used by patients for menstrual fluid absorption. Device is inserted into the vagina; functions via physical absorption. No electronic or software components. Benefits include menstrual hygiene management.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including syngyna absorbency testing per 21 CFR 801.430, in vitro microbiological testing, and biocompatibility testing (genotoxicity, cytotoxicity, mucosal irritation/sensitization, acute systemic toxicity, and colorant extraction) conducted per ISO 10993 and USP 661 standards.
Technological Characteristics
Materials: rayon fibers, cotton, non-woven fabric, plastic applicator. Design: compressed pledget with withdrawal cord. Energy: none (mechanical). Connectivity: none. Sterilization: not specified. Standards: ISO 10993 (biocompatibility), USP 661 (colorant extraction), 21 CFR 801.430 (syngyna absorbency).
Indications for Use
Indicated for use by individuals requiring vaginal absorption of menstrual fluid.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Kimberly-Clark* U by Kotex® Security® Unscented Menstrual Tampons (K896994)
Reference Devices
Kimberly-Clark* U by Kotex® Sleek® Unscented Menstrual Tampons (K112635)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle are three stylized human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, flowing line style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2014
Kimberly-Clark Corp. Charnelle Thomas Global Regulatory Affairs Specialist 1400 Holcomb Bridge Road Rowell, GA 30076
Re: K141294
> Trade/Device Name: U by Kotex Security Unscented Menstrual Tampons Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: June 17, 2014 Received: June 19, 2014
Dear Charnelle Thomas,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Kimberly-Clark Corporation. The logo consists of a circular symbol with a stylized "K" inside, followed by the company name in bold, sans-serif font. The words "Kimberly-Clark" are in a larger, bolder font than the word "Corporation."
# INDICATIONS FOR USE
| Applicant: | Kimberly-Clark Corporation |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K141294 |
| Device Name: | Kimberly-Clark* U by Kotex® Security® Unscented<br>Menstrual Tampons |
| Indications for Use: | Kimberly-Clark* U by Kotex® Security® is an unscented<br>menstrual tampon inserted into the vagina to absorb<br>menstrual fluid. |
| Prescription Use<br>Per 21CFR 801.109 Subpart D | OR Over-The-Counter <span style="text-decoration: underline;">X</span><br>Per 21CFR 801.109 Subpart C |
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Page 1 of 1 - Section 4. Indications for Use
{3}------------------------------------------------
# ්යු Kimberly-Clark Corporation
# Traditional 510(k) for Kimberly-Clark* U by Kotex® Security® Unscented Menstrual Tampons
# Section 5, 510(k) SUMMARY
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
| Submitter's Name: | Kimberly-Clark Corporation |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 2100 Winchester Road<br>Neenah, WI 54956 |
| | Mailing address for regulatory correspondence:<br>1400 Holcomb Bridge Road<br>Roswell, GA 30076-2199 |
| Submitter's Phone No: | 678-352-6031 |
| Submitter's Fax No. | 920-225-4979 |
| Date of Preparation: | August 25, 2014 |
| Name of Device<br>Trade Name: | U by Kotex® Security® Unscented Menstrual Tampons; Regular,<br>Super and Super Plus absorbencies |
| Common Name: | Menstrual Tampon, Unscented |
| Classification Name: | Tampon, Menstrual, Unscented |
| Product Code: | HEB |
| Classification: | 21 CFR 884.5470 |
| Predicate Device: | Kimberly-Clark* U by Kotex® Security® Unscented Menstrual<br>Tampons K896994 |
| Reference Device: | Kimberly-Clark* U by Kotex® Sleek® Unscented Menstrual<br>Tampons K112635 |
| Description of the device: | This device is a conventional unscented menstrual tampon consisting<br>of an absorbent pledget, a withdrawal cord and an applicator. The<br>absorbent pledget consists of a ribbon of rayon fibers and cotton,<br>overwrapped with a non-woven fabric. A withdrawal cord is placed<br>through the pledget and knotted. The tampon pledget is compressed.<br>The formed pledget is inserted into a two-piece plastic applicator<br>consisting of an inner tube (plunger), and an outer insertion tube<br>(barrel) with flexible petals that form a closed, rounded tip. Each<br>tampon is individually wrapped in a plastic film wrapper and multiple<br>tampons are packaged in sealed cartons for commercial sale.<br>Tampons are available in Regular, Super and Super Plus<br>absorbencies in various counts. |
®Registered Trademark and *Trademark of Kimberly-Clark Worldwide, Inc.
{4}------------------------------------------------
# Traditional 510(k) for Kimberly-Clark* U by Kotex® Security® Unscented Menstrual Tampons
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION (Continued)
| Indications for<br>Use: | Kimberly-Clark* U by Kotex® Security® is an unscented menstrual<br>tampon intended to be inserted into the vagina to absorb<br>menstrual fluid. | | |
|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-------------|
| Summary of<br>technological<br>characteristics<br>compared to the<br>predicate device: | The device is the same as the predicate device with regards to the<br>tampon component materials and overall design. The difference<br>between the subject and predicate device applicator is in the addition<br>of two new applicator colors, emerald green and royal blue. The<br>subject device applicator is the same as the predicate device applicator<br>in all other respects. There will also be minor changes to the<br>brandname, packaging and labeling. The brandname will change from<br>Kotex® Security® Natural Balance to U by Kotex® Security®. | | |
| Brief description<br>of preclinical<br>toxicology:<br>(biocompatibility)<br>tests | Testing from the reference device, U by Kotex® Sleek® K112635,<br>was used to support the safety of new applicator colors.<br><br>Preclinical Tests<br>Genotoxicity Test (MLA)<br>Genotoxicity Test (AMES)<br>Genotoxicity Test<br>(Mouse Micronucleus)<br>Cytotoxicity Test<br>Mucosal Irritation Test<br>Mucosal Sensitization Test<br>Acute Systemic Toxicity Test<br>Colorant Extraction Test | Standard | Performance |
| | | ISO 10993, Part 3 | Meets |
| | | ISO 10993, Part 3 | Meets |
| | | ISO 10993, Part 3 | Meets |
| | | ISO 10993, Part 5 | Meets |
| | | ISO 10993, Part 10 | Meets |
| | | ISO 10993, Part 10 | Meets |
| | | ISO 10993, Part11 | Meets |
| | | USP 661 | Meets |
| Safety Assessment: | The subject 510(k) device has undergone an extensive series of<br>safety tests, including an assessment of performance<br>characteristics, preclinical microbiological testing and<br>biocompatibility testing. Testing from the reference device, U by<br>Kotex® Sleek® K112635, was used to support the safety of new<br>applicator colors. The results of these studies support the<br>conclusion that the subject 510(k) device is equivalent and as safe<br>as the predicate device, the Kimberly- Clark* Kotex® Security®<br>Tampon. | | |
®Registered Trademark and *Trademark of Kimberly-Clark Worldwide, Inc.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Kimberly-Clark Corporation. The logo consists of a stylized "K" symbol to the left of the company name. The company name is written in a combination of bold and regular font weights, with "Kimberly-Clark" in bold and "Corporation" in a lighter weight.
# Traditional 510(k) for Kimberly-Clark* U by Kotex® Security® Unscented Menstrual Tampons
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION (Continued)
| Effectiveness: | Kimberly-Clark* U by Kotex® Security® Unscented Menstrual<br>Tampons comply with the syngyna absorbency requirements of 21<br>CFR §801.430 as does the predicate device, the Kimberly-Clark*<br>Kotex® Security® Tampon. |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions: | The results of the extensive performance assessments, in vitro<br>microbiological and biocompatibility testing of the subject device,<br>Kimberly-Clark*U by Kotex® Security® Unscented Menstrual<br>Tampons, support the conclusion that it is safe for its intended use<br>and that it is substantially equivalent to the predicate device, the<br>Kimberly- Clark* Kotex® Security® Tampon. |
® Registered Trademark and *Trademark of Kimberly-Clark Worldwide, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.