← Product Code [EHD](/productcode/EHD) · K141293

# PHOT-X IIS MODEL 505 (K141293)

_Takara Belmont Corporation · EHD · Sep 3, 2014 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141293

## Device Facts

- **Applicant:** Takara Belmont Corporation
- **Product Code:** [EHD](/productcode/EHD.md)
- **Decision Date:** Sep 3, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The PHOT-XIIs MODEL 505 is an extraoral source dental radiographic x-ray unit. This unit works as a diagnostic purpose x-ray source for human teeth with the resultant image recorded on intraoral dental x-ray film or image receptor.

## Device Story

The PHOT-X IIs Model 505 is an extraoral dental radiographic x-ray unit consisting of an x-ray generating tubehead, control box, and supporting hardware. It functions as a diagnostic x-ray source for human teeth, producing images recorded on intraoral film or digital image receptors. Operated by dental professionals in clinical settings, the device uses a DC constant potential x-ray generator with a pre-heating filament system. The operator selects exposure parameters via the control box; the device emits x-rays to capture dental anatomy. The output is a radiographic image used by clinicians to diagnose dental conditions. The device benefits patients by providing diagnostic information for treatment planning.

## Clinical Evidence

Bench testing only. The device's performance was verified through compliance testing with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-65, and ISO14971. No clinical data was required or provided.

## Technological Characteristics

Extraoral dental x-ray system; DC constant potential generator; 0.4 mm focal point; 60/70 kV tube potential; 3/6 mA tube current; 1.7 mmAl inherent filtration + 0.3 mmAl added filtration; 120 VAC power; 1:30 duty cycle. Standards: AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-65, ISO14971.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- Belmont, PHOT-X II Model 303 ([K042260](/device/K042260.md))
- Progeny, PREVA ([K043092](/device/K043092.md))
- Sirona, HELIODENT PLUS ([K083344](/device/K083344.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. The logo is simple and professional, reflecting the department's role in public health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 3, 2014

Takara Belmont Corporation % Robert Schiff, Ph. D. President Schiff & Company, Inc. 1120 Bloomfiled Ave. WEST CALDWELL, NJ 07006

Re: K141293

> Trade/Device Name: PHOT-X IIs Model 505 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: II Product Code: EHD Dated: August 1. 2014 Received: August 4, 2014

Dear Dr. Schiff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Smh.f)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K141293

Device Name

PHOT-X IIs Model 505

Indications for Use (Describe)

PHOT-X IIs Model 505 is a extraoral source dental radiographic x-ray unit. This unit works as a diagnostic purpose x-ray source for human teeth with the resultant image recorded on intraoral dental x-ray film or image receptor.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary (as required by 807.92)

(1) SUBMITTER : Takara Belmont Corporation

Address : 1-1-2 Chome, Higashi-Shinsaibashi, Chuo-ku Telephone : 81-6-6213-5945 Contact person : Toshinori Kiyomatsu Date prepared : April 18, 2014

(2) DEVICE NAME : PHOT-X IIs Model 505 Trade Name : PHOT-X IIs Model 505 Common Name : Dental Periapical X-ray Classification Name : UNIT, X-RAY, EXTRAORAL WITH TIMER (per 21CFR section 872.1800, Product Code EHD)

- (3) PREDICATE DEVICE : Substantial equivalence is based on following legally marketed devices. Belmont, PHOT-X II Model 303 (K042260) Progeny, PREVA (K043092) Sirona, HELIODENT PLUS (K083344)
- (4) DESCRIPTION OF THE DEVICE : PHOT-X IIs Model 505 dental x-ray contains; x-ray generating tubehead, control box and supporting device. Package includes Operator's Manual, Installation Manual and Warranty.
- (5) INTENDED USE : The PHOT-XIIs MODEL 505 is an extraoral source dental radiographic x-ray unit. This unit works as a diagnostic purpose x-ray source for human teeth with the resultant image recorded on intraoral dental x-ray film or image receptor.
- (6) COMPARISON WITH PREDICATE DEVICES : Following table is a comparison of our new PHOT-X Ils Model 505 x-ray and predicate devices.

|                                                                                         | New Device                                                                                                                                                                                                                                                                                                        | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                   |                                                                                                                                                                                                                |
|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                         | BELMONT<br>PHOT-XIIs Model 505                                                                                                                                                                                                                                                                                    | BELMONT<br>PHOT-XII Model 303<br>(K042260)                                                                                                                                                                                                                                                                                                                                                                                                                                       | Progeny<br>Preva<br>(K043092)                                                                                                                     | Sirona<br>HELIODENT PLUS<br>(K083344)                                                                                                                                                                          |
| Indications for Use                                                                     | The PHOT-XIIs MODEL<br>505 is an extraoral<br>source dental<br>radiographic x-ray<br>unit. This unit<br>works as a<br>diagnostic purpose<br>x-ray source for<br>human teeth with<br>the resultant image<br>recorded on<br>intraoral dental<br>x-ray film or image<br>receptor.                                    | PHOT-X II MODEL 303<br>is a extraoral<br>source dental<br>radiographic x-ray<br>unit. This unit<br>works<br>as a diagnostic<br>purpose x-ray<br>source for human<br>teeth with the<br>resultant image<br>recorded on<br>intraoral dental<br>x-ray film or image<br>receptor. The<br>design, function<br>and positioning of<br>the x-ray<br>unit is similar to<br>most all other x-ray<br>machines<br>manufactured for<br>this specific<br>purpose over the<br>past thirty years. | The intended use of<br>the Progeny PREVA<br>Extra-Oral X-Ray<br>system is to act as<br>a diagnostic source<br>for radiographic<br>dental imaging: | The HELIODENT Plus<br>is an extraoral<br>X-Ray source System<br>intended to be used<br>for dental<br>radiographic<br>examination and<br>diagnosis of<br>diseases of the<br>teeth, jaw, and<br>oral structures. |
| A. X-ray Wave Form                                                                      | DC Constant<br>potential                                                                                                                                                                                                                                                                                          | DC Constant<br>potential                                                                                                                                                                                                                                                                                                                                                                                                                                                         | DC Constant<br>potential                                                                                                                          | DC Constant<br>potential                                                                                                                                                                                       |
|                                                                                         | New Device                                                                                                                                                                                                                                                                                                        | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                   |                                                                                                                                                                                                                |
|                                                                                         | BELMONT<br>PHOT-XIIs Model 505                                                                                                                                                                                                                                                                                    | BELMONT<br>PHOT-XII Model 303<br>(K042260)                                                                                                                                                                                                                                                                                                                                                                                                                                       | Progeny<br>Preva<br>(K043092)                                                                                                                     | Sirona<br>HELIODENT PLUS<br>(K083344)                                                                                                                                                                          |
| B. Filament                                                                             | Pre-heating System                                                                                                                                                                                                                                                                                                | Pre-heating System                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Pre-heating System                                                                                                                                | Pre-heating System                                                                                                                                                                                             |
| C. Focal Point Measurement                                                              | 0.4 mm                                                                                                                                                                                                                                                                                                            | 0.7 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 0.4 mm                                                                                                                                            | 0.4 mm                                                                                                                                                                                                         |
| D. Rated tube potential                                                                 | 60 or 70 kV                                                                                                                                                                                                                                                                                                       | 60 or 70 kV                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 60, 65 or 70 kV                                                                                                                                   | 60 or 70 kV                                                                                                                                                                                                    |
| E. Rated tube current                                                                   | 3 or 6 mA                                                                                                                                                                                                                                                                                                         | 4 or 7 mA                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 4, 5, 6 or 7 mA                                                                                                                                   | 7 mA                                                                                                                                                                                                           |
| F. Maximum rated tube potential                                                         | 70 kV                                                                                                                                                                                                                                                                                                             | 70 kV                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 70 kV                                                                                                                                             | 70 kV                                                                                                                                                                                                          |
| G. Rated line voltage                                                                   | 120 VAC                                                                                                                                                                                                                                                                                                           | 120 VAC                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 120 VAC                                                                                                                                           | 120 VAC                                                                                                                                                                                                        |
| H. Line voltage range                                                                   | 108 VAC-132 VAC                                                                                                                                                                                                                                                                                                   | 108 VAC-132 VAC                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 108 VAC-132 VAC                                                                                                                                   | 108 VAC-132 VAC                                                                                                                                                                                                |
| I. Apparent resistance of supply mains                                                  | 0.52 Ω                                                                                                                                                                                                                                                                                                            | 0.48 Ω                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 0.4 Ω                                                                                                                                             | 0.3 Ω                                                                                                                                                                                                          |
| J. Rated line current                                                                   | 10 A at 70kV, 6 mA                                                                                                                                                                                                                                                                                                | 10.8A at 70 kV, 7 mA                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 10 A                                                                                                                                              | 10 A at 70 kV, 7 mA                                                                                                                                                                                            |
| K. Maximum line current                                                                 | 11A at 70kV, 6 mA                                                                                                                                                                                                                                                                                                 | 12A at 70kV, 7 mA                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Unknown                                                                                                                                           | Unknown                                                                                                                                                                                                        |
| L. Exposure time                                                                        | 0.01-2.0 sec.,<br>37 steps                                                                                                                                                                                                                                                                                        | 0.01-3.2 sec.,<br>23 steps                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 0.01-2.0 sec                                                                                                                                      | 0.01-3.2 sec.,<br>23 steps                                                                                                                                                                                     |
| M. Timer accuracy                                                                       | ±5 msec.<br>(below 0.1 sec.<br>setting)<br>±10 msec.<br>(0.1 sec. setting &<br>up)                                                                                                                                                                                                                                | ±5 msec.<br>(below 0.1 sec.<br>setting)<br>±10 msec.<br>(0.1 sec. setting &<br>up)                                                                                                                                                                                                                                                                                                                                                                                               | 5 % ±1 msec                                                                                                                                       | ±10 % + 1 msec                                                                                                                                                                                                 |
| N. Inherent filtration                                                                  | 1.7 mmAl Equivalent                                                                                                                                                                                                                                                                                               | 1.7 mmAl Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Unknown                                                                                                                                           | Unknown                                                                                                                                                                                                        |
| O. Added filtration                                                                     | 0.3 mmAl                                                                                                                                                                                                                                                                                                          | 0.3 mmAl                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Unknown                                                                                                                                           | Unknown                                                                                                                                                                                                        |
| P. Minimum filtration permanently in useful beam.                                       | 2.0 mmAl Equivalent<br>at 70 kV                                                                                                                                                                                                                                                                                   | 2.0 mmAl Equivalent<br>at 70 kV                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 2.0 mmAl Equivalent<br>at 70 kV                                                                                                                   | Unknown                                                                                                                                                                                                        |
| Q. Nominal roentgen output<br>1-Distal end of regular cone<br>2-Distal end of long cone | 1-<br>4.6 mGy/s ± 40 %<br>(60 kV, 3 mA)<br>9.1 mGy/s ± 40 %<br>(60 kV, 6 mA)<br>5.9 mGy/s ± 40 %<br>(70 kV, 3 mA)<br>11.8 mGy/s ± 40 %<br>(70 kV, 6 mA)<br>2-<br>2.0 mGy/s ± 40 %<br>(60 kV, 3 mA)<br>4.1 mGy/s ± 40 %<br>(60 kV, 6 mA)<br>2.6 mGy/s ± 40 %<br>(70 kV, 3 mA)<br>5.2 mGy/s ± 40 %<br>(70 kV, 6 mA) | 1-<br>5.4 mGy/s ± 40 %<br>(60 kV, 4 mA)<br>9.4 mGy/s ± 40 %<br>(60 kV, 7 mA)<br>7.1 mGy/s ± 40 %<br>(70 kV, 4 mA)<br>12.4mGy/s±40%<br>(70kV, 7 mA)<br>2-<br>2.4 mGy/s ± 40 %<br>(60 kV, 4 mA)<br>4.2 mGy/s ± 40 %<br>(60 kV, 7 mA)<br>3.1 mGy/s ± 40 %<br>(70 kV, 4 mA)<br>5.5 mGy/s ± 40 %<br>(70 kV, 7 mA)                                                                                                                                                                     | Unknown                                                                                                                                           | 1-<br>8.5 mGy/s ± 40 %<br>(60 kV, 7 mA)<br>11 mGy/s ± 40 %<br>(70 kV, 7 mA)                                                                                                                                    |
| R. Source to skin distance<br>1-Regular cone<br>2-Long cone                             | 1-203 mm<br>2-305 mm                                                                                                                                                                                                                                                                                              | 1-203 mm<br>2-305 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 1-200 mm<br>2-300 mm                                                                                                                              | 1-200 mm<br>2-300 mm                                                                                                                                                                                           |
| S. Leaking technique factor                                                             | 70 kV, 697 mAs                                                                                                                                                                                                                                                                                                    | 70 kV, 494 mAs                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 70 kV, 1350 mAs                                                                                                                                   | 70 kV, 413 mAs                                                                                                                                                                                                 |
| T. Duty cycle                                                                           | 1 : 30                                                                                                                                                                                                                                                                                                            | 1 : 50                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 1 : 15                                                                                                                                            | 1 : 60                                                                                                                                                                                                         |
| U. Maximum deviation of tube potential and tube current.                                | ±10 kV, ± 2 mA<br>(below 0.1sec.<br>setting)<br>±5 kV, ±1 mA<br>(0.1 sec. setting &<br>up)                                                                                                                                                                                                                        | ±10 kV, ± 2 mA<br>(below 0.1sec.<br>setting)<br>±5 kV, ±1 mA<br>(0.1 sec. setting &<br>up)                                                                                                                                                                                                                                                                                                                                                                                       | ±5 %<br>(tube potential)<br>±1 mA<br>(tube current)                                                                                               | ±5 kV<br>(tube potential)<br>±1.4 mA<br>(tube current)                                                                                                                                                         |

{4}------------------------------------------------

{5}------------------------------------------------

- (7) PERFORMANCE STSTANDARDS APPLIED : AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-65, ISO14971
(Testing for AAMI/ANSI ES60601-1, IEC60601-1-3 and IEC60601-1-6 will be conducted by Intertek, 70 Codman Hill Road, Boxborough, MA 01719 before marketing. Testing for IEC60601-1-2 was conducted by UL Japan Inc. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN and confirmed the compliance with the standard. )

- (8) CONCLUSION : The PHOT-XIIs Model 505 has the same intended use and technology characteristics as the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K141293](https://fda-staging.innolitics.com/device/K141293)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
