← Product Code [NKB](/productcode/NKB) · K141282

# PGS 5.5 MM CANNULATED PEDICLE SCREW (K141282)

_Pinnacle Spine Group · NKB · Dec 2, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141282

## Device Facts

- **Applicant:** Pinnacle Spine Group
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Dec 2, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PSG 5.5mm Cannulated Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformity, or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudoarthrosis, and previous failed fusion. The PSG 5.5mm Cannulated Pedicle Screw System is a non-cervical spinal fixation system. Pedicle screw fixation is limited to skeletally mature patients.

## Device Story

Posterior spinal fixation system; consists of cannulated pedicle screws, rods, cross-connectors, and locking cap set screws. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Components available in various sizes to accommodate patient anatomy. Mechanical stabilization achieved through rigid fixation of spinal segments. Benefits include structural support for fusion in patients with acute/chronic spinal instabilities or deformities.

## Clinical Evidence

Bench testing only. Performed static axial compression bending, static torsion, and dynamic axial compression bending fatigue testing per ASTM F1717-13. Results confirm mechanical strength is sufficient for intended use and equivalent to predicate devices.

## Technological Characteristics

Posterior spinal fixation system; components include cannulated pedicle screws, rods, cross-connectors, and locking cap set screws. Materials: Titanium alloy (ASTM F136). Mechanical stabilization principle. Non-cervical application.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Synergy VLS Open ([K000236](/device/K000236.md))
- Moss Miami Titanium ([K103490](/device/K103490.md), [K033901](/device/K033901.md), [K955348](/device/K955348.md))
- TSRH ([K081080](/device/K081080.md))
- OPTIMATM ([K020279](/device/K020279.md), [K051971](/device/K051971.md), [K024096](/device/K024096.md))
- Spine Proliant Screw System ([K102870](/device/K102870.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a central emblem. The emblem consists of a stylized design of three human profiles facing right, with flowing lines beneath them, resembling a bird or abstract form.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 2, 2014

Pinnacle Spine Group % Meredith Mav. MS. RAC Empirical Consulting, LLC 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K141282

Trade/Device Name: PSG 5.5mm Cannulated Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: October 28, 2014 Received: October 31, 2014

Dear Ms. May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Meredith May, MS, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120  |
|-----------------------------------------|-----------------------------------|
| Food and Drug Administration            | Expiration Date: January 31, 2017 |
| Indications for Use                     | See PRA Statement on last page.   |

| 510(k) Number (if known)       | K141282                                   |
|--------------------------------|-------------------------------------------|
| Device Name                    | PSG 5.5mm Cannulated Pedicle Screw System |
| Indications for Use (Describe) |                                           |

The PSG 5.5mm Cannulated Pedicle Screw System is intended to provide immobilization and
stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and
chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease
(defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic
studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformity, or curvature (i.e., scoliosis,
kyphosis, and lordosis), tumor, stenosis, pseudoarthrosis, and previous failed fusion.

The PSG 5.5mm Cannulated Pedicle Screw System is a non-cervical spinal fixation system. Pedicle
screw fixation is limited to skeletally mature patients.

| Type of Use (Select one or both, as applicable)                                                                                                                                            |  |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |  |

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

| FORM FDA 3881 (9/13) | Page 1 of 2 |
|----------------------|-------------|
|----------------------|-------------|

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 5.510(K) SUMMARY

| Submitter's Name:          | Pinnacle Spine Group                                                                                                                                    |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address:       | 1601 Elm Street, Suite 1930<br>Dallas, TX 75201                                                                                                         |
| Submitter's Telephone:     | 214.466.1428                                                                                                                                            |
| Contact Person:            | Meredith L. May, MS<br>Empirical Testing Corp.<br>719.337.7579                                                                                          |
| Date Summary was Prepared: | 9 May 2014                                                                                                                                              |
| Trade or Proprietary Name: | PSG 5.5mm Cannulated Pedicle Screw System                                                                                                               |
| Common or Usual Name:      | Orthosis, Spinal Pedicle Fixation<br>Orthosis, Spondylolisthesis Spinal Fixation<br>Orthosis, Spinal Pedicle Fixation, For Degenerative Disc<br>Disease |
| Classification:            | Class III per 21 CFR §888.3070                                                                                                                          |
| Product Code:              | PMD.'MNI, MNH                                                                                                                                           |
| Classification Panel:      | 87 Orthopedic                                                                                                                                           |

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The PSG 5.5mm Cannulated Pedicle Screw System is a multiple component, posterior spinal fixation system which consists of pedicle screws, rods, cross-connectors, and locking cap set screws. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are made from titanium alloy described by such standards as ASTM F136.

### INDICATIONS FOR USE

The PSG 5.5 mm Cannulated Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformity, or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudoarthrosis, and previous failed fusion.

The PSG 5.5mm Cannulated Pedicle Screw System is a non-cervical spinal fixation system. Pedicle screw fixation is limited to skeletally mature patients.

The indications for use for the PSG 5.5mm Cannulated Pedicle Screw System is similar to that of the predicate devices listed in Table 5-1.

{5}------------------------------------------------

#### VGEJ PQNQI KECNEJ CTCEVGTKVKEU

The intended use and technological features of the PSG 5.5mm Cannulated Pedicle Screw System do not substantially differ from the legally marketed predicate devices. The predicate devices and the PSG 5.5mm Cannulated Pedicle Screw System are designed for posterior stabilization to provide immobilization and stabilization of spinal segments as an adjunct to fusion.

| 510k Number                 | Trade or Proprietary or Model Name | Manufacturer     |
|-----------------------------|------------------------------------|------------------|
| K000236                     | Synergy VLS Open                   | Interpore        |
| K103490, K033901, K955348   | Moss Miami Titanium                | DePuy Spine Inc. |
| K081080 (Primary Predicate) | TSRH                               | Medtronic        |
| K020279, K051971, K024096   | OPTIMATM                           | U&I Corporation  |
| K102870                     | Spine Proliant Screw System        | Exactech         |

#### Table 5-1 Predicate Devices

#### PERFORMANCE DATA

The PSG 5.5mm Cannulated Pedicle Screw System has been tested in the following test modes:

- Static axial compression bending per ASTM F1717-13
- Static torsion per ASTM F1717-13 ●
- Dynamic axial compression bending fatigue per ASTM F1717-13 ●

The results of this non-clinical testing show that the strength of the PSG 5.5mm Cannulated Pedicle Screw System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

#### CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the PSG 5.5mm Cannulated Pedicle Screw System is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K141282](https://fda-staging.innolitics.com/device/K141282)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).