K141254 · Elcam Medical Acal · FMG · Jul 30, 2014 · General Hospital
Device Facts
Record ID
K141254
Device Name
ELCAM STOPCOCKS AND MANIFOLDS
Applicant
Elcam Medical Acal
Product Code
FMG · General Hospital
Decision Date
Jul 30, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
Elcam Stopcocks and Manifolds with Safe2 Rotator are indicated for fluid flow directional control and for providing access port(s) for administration of solutions. Typical uses include pressure monitoring, intravenous fluid administration and transfusion.
Device Story
Stopcocks and manifolds with integrated male luer lock rotator; used for fluid flow control, pressure monitoring, IV administration, and transfusion. Safe2 Rotator component features internally threaded rotator fixed to stopcock body; allows free rotation on male connector axis. Prevents IV tubing kinks/twists; provides convenient access to side ports during injection/aspiration. Used in clinical settings by healthcare providers. Rotator ensures secure connection to tubing sets, preventing inadvertent disconnection. Benefits include improved ease of use for clinicians and reduced risk of line complications.
Clinical Evidence
Bench testing only. Evaluated design modification via ISO 594-2 functional requirements: initial/secondary twisting force, liquid pressure resistance, and disconnection force. Results confirmed device meets specifications and is as safe and effective as the predicate.
Technological Characteristics
Stopcock and manifold assembly with male luer lock rotator. Materials: standard medical-grade plastics. Sensing/actuation: mechanical fluid flow control. Connectivity: none. Sterilization: not specified. Software: none. Compliance: ISO 594-2 (lock fitting).
Indications for Use
Indicated for patients requiring fluid flow directional control, pressure monitoring, intravenous fluid administration, or transfusion. No specific age or gender restrictions stated.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
JUL 3 0 2014
Page 1 of 3
Image /page/0/Picture/3 description: The image shows the logo for Elcam Medical. The logo consists of a stylized "E" symbol to the left of the text "Elcam MEDICAL". Below the text is the tagline "Where everything connects".
# Special 510(k) summary for Elcam Medical Stopcocks and Manifolds with Safe2 Rotator
In accordance with 21 CFR 807.92 the following summary of information is provided
### 1. Date prepared:
May 07, 2014
### 2. Contact information for the 510(k) submission:
Owner: Elcam Medical ACAL Kibbutz BarAm, Merom HaGalil 1386000, Israel T: (972) - 4698-8120/1 F: (972) - 4698-0777 Email: sales@elcam.co.il
Contact Mrs. Natasha Branzburg person: Q.A. and R.A. Engineer Elcam Medical ACAL T: (972) - 77-2568-625 F: (972) - 4698-0777 Email: natasha.ev@elcam.co.il
Elcam Mr. Amir Halperin Medical's President of Elcam Medical, Inc. U.S. agent: 2 University Plaza, Suite 620, Hackensack, NJ 07601, USA T: (201) 457-1120 F: (201) 457-1125 Email: amir.halperin@elcam-medical.com
#### 3. Device name:
Proprietary/ Trade name: Elcam Stopcocks and Manifolds with Safe2 Rotator
Common/ usual name: Stopcocks and manifolds with male luer lock rotator.
Classification: Elcam Stopcocks and Manifolds with Safe2 Rotator have been classified as Class II devices under the following classification names:
el: 972-4-698-8120/1/2 www.elcam-medical.com Fax: 972-4-698-0777 Baram 13860, Israel | sales@eicam.co.il Elcam Stopcocks El Lucomed Mass Injectech
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Elcam Medical. The logo consists of a stylized letter "E" on the left, followed by the words "Elcam" and "MEDICAL" stacked on top of each other. Below the words is the tagline "Where everything connects."
| Classification name | Product code | 21 CFR ref. | Panel |
|-------------------------------------|--------------|-------------|------------------|
| Intravascular<br>administration set | FMG | 880.5440 | General Hospital |
#### 4. Predicate Device
Elcam Stopcocks and Manifolds with Safe2 Rotator were found to be substantially equivalent to Elcam's stopcocks and manifold.
Elcam's stopcocks and manifolds are legally marketed devices which are cleared under 510(k) number K022895.
#### Device Description 5.
Elcam Stopcocks and Manifolds with Safe2 Rotator are an additional variation in the Elcam's Stopcocks and Manifolds product-family.
The Safe2 Rotator component is male luer with a mounted and internally threaded rotator that is fixed on the stopcocks body with a modified male connector. Both components create a uniform straight fluid path.
The device uniqueness is that the Safe2 Rotator can freely rotate on the stopcock male connector axis preventing the formation of kinks and twists in the IV sets during or following manipulations. In addition, the rotation ability of the Safe2 Rotator provides health care providers with a convenient access to a stopcock side (vertical) port during injection and aspiration.
#### 6. Intended use
Elcam Stopcocks and Manifolds with Safe2 Rotator are indicated for fluid flow directional control and for providing access port(s) for administration of solutions. Typical uses include pressure monitoring, intravenous fluid administration and transfusion.
#### 7. Technological characteristics and substantial equivalence
Elcam Stopcocks and manifolds with Safe2 Rotator are substantially equivalent to Elcam stopcocks and manifolds cleared under 510(k) number K022895. This new stopcocks variation has the same indication for use, principle of operation.
overall shape, sterilization method and shelf life as its predicate. Without changing the way of use from the health care provider perspective, Safe2
Rotator component ensures a secured connection of the stopcocks with other tubing set components preventing inadvertent disconnection and allows performing easy injection/ aspiration through the male luer by twisting the rotator towards a health care provider.
There were no new questions regarding Elcam Stopcocks and Manifolds with Safe2 Rotator safety and effectiveness that were raised due to this modification. Therefore, the modified and the original devices were judged as substantially equivalent.
Image /page/1/Picture/17 description: The image shows contact information and brand names. The contact information includes a telephone number (972-4-698-8120/1/2), fax number (972-4-698-0777), website (www.elcam-medical.com), physical address (Baram 13860, Israel), and email address (sales@elcam.co.il). The brand names listed are Elcam Stopcocks, Lucomed, and Injectech.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Elcam Medical. The logo consists of a stylized "E" symbol to the left of the word "Elcam" in a bold, sans-serif font. Below "Elcam" is the word "MEDICAL" in a similar font, and below that is the tagline "Where everything connects."
#### 8. Performance data
Testing related to functionality of Elcam Stopcocks and Manifold with Safe2 Rotator has been conducted in a series of non-clinical tests. It has demonstrated that the modified device is as safe and effective as its predicate.
Following is a list of the bench tests that were conducted following risk management process to evaluate the effect of the design modification:
Compliance with the functional requirements of ISO 594-2 (lock fitting) ı
- Initial twisting force
- Secondary twisting force
- Liquid pressure resistance
- Disconnection force
#### Conclusion 9.
The evaluation of our modified device performances demonstrated that it meets its specifications, labeling claims and that it is as safe and as effective as the predicate devices.
Image /page/2/Picture/12 description: This image contains contact information for Elcam Medical. The image includes the telephone number 972-4-698-8120/1/2 and fax number 972-4-698-0777. The website is www.elcam-medical.com and the email address is sales@elcam.co.il. The address is Baram 13860, Israel.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
July 30, 2014
Elcam Medical ACAL Natasha Branzburg Q.A. and R.A. Engineer Kibbutz Bar-Am MP Merom Hagalil lsrael 13860
Re: K141254
Trade/Device Name: Eleam Stopcocks and Manifolds with Safe2 Rotator Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: FMG Dated: July 6, 2014 Received: July 9, 2014
Dear Mrs. Branzburg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mrs. Branzburg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MarySRunner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K141254
Device Name
Elcam Stopcocks and Manifolds with Safe2 Rotator
Indications for Use (Describe)
Elcam Stopcocks and Manifolds with Safe2 Rotator are indicated for fluid flow directional control and for providing access port(s) for administration of solutions.
Typical uses include pressure monitoring, intravenous fluid administration and transfusion.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
्रिंग पु 11 FOR FDA USE ONLY . 110 134 r - . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ﺔ ﻓ Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/5/Picture/13 description: The image shows a stylized logo of the letters FDA. The letters are formed by multiple parallel lines, giving them a three-dimensional appearance. The letters are bold and stand out against the white background.
Digitally signed by Richard C. Chapman -S Date: 2014.07.29 13:06:05 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.