APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH SDS SOFTWARE

K141220 · Thermo Fisher Scientific (Life Technologies Holdin · NSU · May 22, 2014 · Clinical Chemistry

Device Facts

Record IDK141220
Device NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH SDS SOFTWARE
ApplicantThermo Fisher Scientific (Life Technologies Holdin
Product CodeNSU · Clinical Chemistry
Decision DateMay 22, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.2570
Device ClassClass 2

Indications for Use

The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument with the SDS Software is a real-time nucleic acid amplification and five color fluorescence detection system for use with FDA cleared or approved tests on human-derived specimens. The 7500 Fast Dx Real-Time PCR instrument and SDS Software are intended for use in combination with in vitro diagnostic tests labeled for use on this instrument. The 7500 Fast Dx instrument is intended for use by laboratory professionals trained in laboratory techniques, procedures, and on use of the system.

Device Story

Real-time PCR instrument; performs nucleic acid amplification and fluorescence detection. Inputs: human-derived specimens in 96-well Fast PCR plates. Operation: integrated thermal cycler and microvolume fluorimeter; automated control of amplification, detection, and data analysis. Output: comparative quantitative readouts. Used in clinical laboratories by trained professionals. Modification: software update to support Windows 7 OS, updated camera driver, corrected audit/e-signature print preview, and enabled network drive/printer access. Healthcare providers use quantitative readouts to support diagnostic decisions based on FDA-cleared assays.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical verification testing of the updated operating system.

Technological Characteristics

Real-time PCR instrument; integrated thermal cycler and microvolume fluorimeter. Amplification volume: 10-30 uL in 96-well plates. Five-color fluorescence detection. Software: SDS Software v1.4.1 running on Microsoft Windows 7 Professional (32/64 bit). Connectivity: network drive and network printer support.

Indications for Use

Indicated for laboratory professionals to perform real-time nucleic acid amplification and five-color fluorescence detection on human-derived specimens using FDA-cleared or approved in vitro diagnostic tests.

Regulatory Classification

Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Review Memorandum To: THE FILE RE: DOCUMENT NUMBER K141220 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. The Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument with SDS Software 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S): The SDS software will be changed to support an additional OS (Windows 7 in addition to Windows XP). Update camera driver for Win 7/64 bit compatibility. The camera driver used in the current 7500 Fast Dx Real-time PCR Software v1.4 is not compatible with Windows 7 OS. The camera driver will be updated to make the software compatible with Windows 7 OS. An audit and e-Signature report print preview & print defect will be corrected. The current instrument software does not allow the user to preview and print the Audit and e-Signature reports. This anomaly will be fixed to allow the user to preview and print the Audit and e-Signature reports The current version of the SDS Software does not allow the user to open a file from the network drive or print a file to a network printer when it's running on a computer with the Window 7 OS. This function will be implemented to allow the user to open a file from a network drive and print to a network printer once the user logins in with their user name and password credentials. A minor cosmetic change will be implemented by removing the "AB" button and "myScience" option from the software. These functions are no longer supported. 4. The FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. 5. Comparison Information (similarities and differences): | | New Device: Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software (K141220) | Predicate Device: Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software version 1.4 (K082562) | | --- | --- | --- | | Intended Use | Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument with the SDS Software is a real-time nucleic acid amplification and five color fluorescence detection | Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument with the SDS Software version 1.4 is a real-time nucleic acid amplification and detection system | {1} Page 2 of 3 | | New Device: Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software (K141220) | Predicate Device: Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software version 1.4 (K082562) | | --- | --- | --- | | | system for use with FDA cleared or approved tests on human-derived specimens. The 7500 Fast Dx Real-Time PCR instrument and SDS Software are intended for use in combination with in vitro diagnostic tests labeled for use on this instrument. The 7500 Fast Dx instrument is intended for use by laboratory professionals trained in laboratory techniques, procedures, and on use of the system. | that measures nucleic acid signals from reverse transcribed RNA and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes. The 7500 Fast Dx instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer. | | Fundamental Technology/Detection | Real-time PCR | Same | | Instrument Computer Operating System | Microsoft Windows 7 | Microsoft Windows XP | | Degree of Automation | Requires manual transfer of amplification mixture to amplification/detection instrument. Automated control of amplification, detection and data analysis | Same | | Primary Operational Amplification and Detection Components | Integrated thermal cycler and microvolume fluorimeter for walk away PCR amplification and detection | Same | | Amplification Reaction Volume | 10-30 uL in 96-well Fast PCR plates | Same | 6. A Design Control Activities Summary which includes: a) Software architecture document describing the software design structure and programming environment. b) Software lifecycle management plan contains requirements for the software development lifecycle of the product. It defines the approach for software development as a subset of Life Technologies' Product Commercialization Process (PCP). It also defines the approach to software maintenance through product decommissioning and disposal. c) Impact assessment report assessing the impact and user safety risk for 7500 Fast Dx Real-Time PCR Instrument software change to support new computer models with Windows 7 Professional 32/64 bit Operating System Service Pack 1. d) Sustaining project master plan describes the software development plan and test plan. {2} Page 3 of 3 e) Software verification test report describes the verification results of the 7500 Fast Dx SDS Software v1.4.1 in accordance to the Sustaining Project Master Plan. f) Revision level history. g) Software configuration management plan provides the guidance for the control of evolving software systems, including creation of a Software Configuration Management Plan (SCMP). It controls versioning and configuration management of software products. It defines an approach for software configuration management and version control system use as a subset of Software Development Lifecycle Procedure. h) Service and support plan along with a software release for deployment guideline. i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met. j) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 7. A Truthful and Accurate Statement, a 510(k) Summary and the Indications for Use Enclosure. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...