K141206 · Edwards Lifesciences, LLC · GAW · Jan 16, 2015 · General, Plastic Surgery
Device Facts
Record ID
K141206
Device Name
EDWARDS LIFESCIENCES THRUPORT KNOTTING SYSTEM
Applicant
Edwards Lifesciences, LLC
Product Code
GAW · General, Plastic Surgery
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for use in the approximation of soft tissue and prosthetic materials.
Device Story
ThruPort OptiClip Knotting System is a sterile, single-use surgical instrument comprising a knotting instrument and a knot loader. Used by surgeons to approximate soft tissue and prosthetic materials. The system utilizes a nitinol knot to secure 2-0 polyester or 2-0 polypropylene sutures. The operator manually loads the suture into the instrument using the knot loader, then uses the instrument to deploy the nitinol knot at the surgical site. The device provides tactile feedback during deployment. The nitinol knot is MR Conditional (1.5T/3.0T), while the instrument and loader are MR Unsafe. The system is designed for durability up to 25 deployment cycles.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included functional design verification (tensile strength, deployment force, shaft stiffness, dimensional interference, pinch point analysis, durability, and corrosion resistance), biocompatibility per ISO 10993-1:2009, sterility, shelf life, and MR safety evaluation.
Technological Characteristics
Sterile, single-use, metal and polymeric construction. Mechanical actuation via trigger-based deployment of nitinol knot. Dimensions include 15cm, 22cm, and 30cm shaft lengths. MR Conditional (nitinol knot) and MR Unsafe (instrument/loader). Biocompatibility per ISO 10993-1:2009.
Indications for Use
Indicated for use in the approximation of soft tissue and prosthetic materials in conjunction with 2-0 polyester or 2-0 polypropylene suture and a knot loader with nitinol knot.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Suture Placement Devices and Accessories, LSI Solutions, Inc. (K100593)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
Edwards Lifesciences LLC Ms. Nina Brooke Regulatory Affairs Associate III One Edwards Way Irvine, California 92614
Re: K141206
Trade/Device Name: ThruPort Knotting Instrument Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable polypropylene surgical suture Regulatory Class: Class II Product Code: GAW, HCF Dated: January 12, 2015 Received: January 14, 2015
Dear Ms. Brooke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K141206
Device Name ThruPort Knotting Instrument
Indications for Use (Describe)
The knotting instrument used in conjunction with 2-0 polyester or 2-0 polypropylene suture and a knot loader with nitinol knot is indicated for use in the approximation of soft tissue and prosthetic materials.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
| 510(k) Summary | |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Edwards Lifesciences LLC |
| Contact Person: | Nina Brooke, Regulatory Affairs Associate III<br>One Edwards Way<br>Irvine, CA 92614<br>Phone: (949) 244-6444<br>Fax: (949)756-4573 |
| Date Prepared: | January 15, 2015 |
| Trade Name: | Edwards ThruPort OptiClip Knotting System |
| Classification Name: | Suture, Nonabsorbable, Synthetic, Polypropylene (21 CFR<br>§878.5010, Product Code GAW, Class II)<br>Instrument Ligature Passing and Knot Tying, General &<br>Plastic Surgery Panel (21 CFR §878.4800, Product Code<br>HCF, Class I) |
| Predicate Device: | K100593, Suture Placement Devices and Accessories, LSI<br>Solutions, Inc. |
#### Device Description:
Edwards Lifesciences' ThruPort OptiClip Knotting System is a sterile, nonpyrogenic, single-use surgical instrument made of metal and polymeric materials. lt consists of the ThruPort OptiClip Knotting Instrument and the ThruPort OptiClip Knot Loader.
# Intended Use:
Intended for use in the approximation of soft tissue and prosthetic materials.
# Indications for Use:
The ThruPort Knotting Instrument is used in conjunction with 2-0 polyester or 2-0 polypropylene suture and a knot loader with nitinol knot and is indicated for use in the approximation of soft tissue and prosthetic materials.
{4}------------------------------------------------
# Comparative Analysis:
The subject device has the same intended use and technological characteristics (i.e., design, material, chemical composition) as the predicate device. It has been demonstrated that the subject device is comparable to the predicate device in fundamental scientific technology, material types, principles of operation, and functional performance evaluations.
### Functional and Safety Testing:
The functional data indicate that the device performs in a substantially equivalent manner when compared to the predicate device. The following functional tests were completed and all results met acceptance criteria:
- . Functional bench testing (design verification)
- Knot Loader Suture Snare Tensile Strength: Demonstrate that the Knot . Loader suture snare can withstand forces encountered during suture loading within the Knotting Instrument.
- Knotting System Suture Loading Force: To establish the forces . encountered during suture loading within the Knotting Instrument
- Knotting Instrument Shaft Stiffness: Demonstrate that the shaft of the Knotting Instrument has adequate stiffness to perform effectively within the intended use.
- . Knotting Instrument Weight: Demonstrate that the weight of the assembled Knotting System allows for adequate tactile feedback.
- Dimensional Interference: Demonstrate that the components of the Knotting ● System (including the un-deployed nitinol knot) do not exhibit dimensional interference relative to the suture prior to deployment.
- . Pinch Point Area: Demonstrate that the area of the exposed Knotting System pinch point is less than that of the predicate device.
- . Knotting Instrument Effective Length: Demonstrate that the effective length of the 15cm, 22cm, and 30cm Knotting Instrument are as labeled
- . Shaft Outer Diameter: Demonstrate that the outer diameter of the Knotting Instrument is less than or equal to the predicate device.
- . Overall Nitinol Knot Height: Demonstrate that the deployed nitinol knot height is less than or equal to the predicate device
- Knot Loader Outer Diameter: Demonstrate that the outer diameter of the . Knot Loader is less than or equal to the predicate device outer diameter.
{5}------------------------------------------------
- Knotting System Deployment Force: Demonstrate that the force required ● to fully depress the trigger and deploy a nitinol knot exceeds the weight of the Knotting System and is low enough to ensure ease of use
- . Suture Tension: Demonstrate that the tension required to cut suture is less than or equal to the predicate device.
- Knot Loader Disposable Tip Push Off Force: Demonstrate that the Knot Loader locking tip will not be pushed off during the deployment of the nitinol knot.
- . Nitinol Knot Push Off Force: Demonstrate that the Knot Loader locking tip will not be pushed off during the deployment of the nitinol knot.
- Nitinol Knot Suture Retention Strength: Demonstrate that the nitinol knot . can effectively secure the suture upon deployment when used for the intended use.
- Knotting Instrument Handle Strength: Demonstrate that the handle of the . Knotting Instrument can withstand the force of applying suture tension and pressure onto tissue or prosthetic material.
- Knotting Instrument Corrosion: Demonstrate that the Knotting Instrument . can function after exposure to a simulated use environment.
- Nitinol Knot Corrosion: Demonstrate that the nitinol knot does not corrode. ●
- Knotting Instrument Durability: Demonstrate that the Knotting Instrument . can function for a maximum of 25 deployment cycles.
- . Knotting System Packaging Integrity: Demonstrate that the packaging configuration of the Knotting System is able to maintain a sterile barrier after worst case conditioning.
- Nitinol Knot Magnetic Resonance Conditional Testing: To substantiate . the claim of magnetic resonance (MR) conditional.
- · Biocompatibility testing was performed per ISO 10993-1:2009
- · Sterility and sterile barrier testing
- · Shelf life testing
- · Packaging testing
- · Design validation
#### MR evaluation testing:
Per testing conducted the following MR safety statement will be added to our labeling
# MRI SAFETY INFORMATION
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows a black triangle with the letters "MR" in bold black font in the center. The triangle is equilateral, and the letters are horizontally centered within it. The background is plain white.
Non-clinical testing has demonstrated that the nitinol knot is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:
- Static magnetic field of 1.5 Tesla and 3.0 Tesla only .
- Maximum spatial gradient magnetic field of 3,000 Gauss/cm (30 T/m) or less .
- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2 W/kg (Normal Operating Mode)
Under the scan conditions defined above, the nitinol knot is expected to produce a maximum temperature rise of less than 1.5 ℃ after 15 minutes of continuous scanning.
In non-clinical testing, the image artifact caused by the nitinol knot extends approximately 1 mm from the nitinol knot when imaged with a spin echo pulse sequence in a 3.0 T MRI system and 5 mm from the nitinol knot when imaged with a gradient echo pulse sequence in a 3.0 T MRI system.
# The knotting instrument and knot loader are MR Unsafe.
A patient identification card is included with the knot loader, which contains the MRI safety information for the nitinol knot. Fill out the patient specific information including patient name, date of implant, doctor name, and doctor phone number and then distribute to the patient at discharge.
#### Conclusion:
The Edwards Thruport OptiClip Knotting System is substantially equivalent to the cited predicate device. Additionally, the device testing confirms the safety and effectiveness of the device when compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.