K141187 · Medicrea International SA · OVD · Jan 15, 2015 · Orthopedic
Device Facts
Record ID
K141187
Device Name
IMPIX ALIF S/A
Applicant
Medicrea International SA
Product Code
OVD · Orthopedic
Decision Date
Jan 15, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The stand-alone intersomatic IMPIX ALIF S/A implants are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft. Patients should have a satisfactory course of conservative treatment (usually at least six (6) months of non-operative treatment) prior to treatment with an IMPIX-ALIF-S/A stand-alone cage. The IMPIX ALIF S/A devices should be used with the three (3) bone screws (Small Footprint) or four (4) bone screws (Medium, Large Footprints) accompanying the device. Otherwise, an additional fixation system (e.g., pedicle screw system) should be implanted.
Device Story
Intervertebral body fusion device; used for lumbar spinal fusion; consists of PEEK cage with tantalum markers and titanium alloy (Ti-6Al-4V) bone screws. Implanted via anterior approach; provides stand-alone fixation; requires autogenous bone graft. Used by surgeons in clinical settings. Device stabilizes spinal segment; promotes fusion; reduces discogenic back pain. Mechanical performance verified via bench testing; no clinical or animal studies conducted.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), static/dynamic axial compression, compression shear, subsidence, expulsion, and screw clip-out evaluation.
Technological Characteristics
PEEK OPTIMA LT1 (ASTM F2026) cage; Ti-6Al-4V (ASTM F136, ISO 5832-3) screws; Tantalum (ASTM F560, ISO 13782) markers. Stand-alone intersomatic fusion device with integrated screw fixation. Mechanical testing per ASTM F2077-11 and F2267-04.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1), including those with up to Grade I spondylolisthesis or retrolisthesis. Requires failure of at least 6 months of conservative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure with outstretched arms, which is meant to represent the department's mission of protecting the health of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2015
Medicrea International S.A. Mr. David Ryan Marketing & Product Development Director 14 Porte Du Grand Lyon 01700 Neyron France
Re: K141187
Trade/Device Name: IMPIX ALIF S/A Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MAX Dated: December 17, 2014 Received: December 19, 2014
Dear Mr. Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
# Page 2 – Mr. David Ryan
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141187
Device Name IMPIX ALIF S/A
Indications for Use (Describe)
The stand-alone intersomatic IMPIX ALIF S/A implants are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).
This device is to be used with autogenous bone graft.
Patients should have a satisfactory course of conservative treatment (usually at least six (6) months of non-operative treatment) prior to treatment with an IMPIX-ALIF-S/A stand-alone cage.
The IMPIX ALIF S/A devices should be used with the three (3) bone screws (Small Footprint) or four (4) bone screws (Medium. Large Footprints) accompanying the device. Otherwise, an additional fixation system (e.g., pedicle screw system) should be implanted.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 1. DEVICE SUBMITTER
MEDICREA® INTERNATIONAL S.A. 14 Porte Du Grand Lyon 01700 NEYRON- France
Phone: +33 4 72 01 87 87 Fax: +33 4 72 01 87 88
Contact Person: David Ryan VP Product Development and Marketing dryan@medicrea.com
Date Prepared: 12/01/2015
2. DEVICE
Name of Device: IMPIX ALIF S/A
Common or Usual Name: Intervertebral Fusion Device With Integrated Fixation, Lumbar Intervertebral Fusion Device With Bone Graft, Lumbar
Classification Name: Intervertebral Body Fusion Device
Regulatory Class: 21 CFR 888.3080- Intervertebral Body Fusion Device Product Code: OVD; MAX
# 3. PREDICATE DEVICE
The primary predicate used in this 510(k) notice is:
- The STALIF MIDLINE (Centinel Spine, K101301)
The additional predicates are:
- The IMPIX ALIF Lumbar IBF (MEDICREA TECHNOLOGIE, K083798) ●
- . The IMPIX L (MEDICREA TECHNOLOGIE, K072226)
- The SYNFIX LR (Synthes Spine, K062083)
- . The STALIF TT (Surgigraft Limited,K073109)
These predicates have not been subject to a design-related recall.
{4}------------------------------------------------
## 4. DEVICE DESCRIPTION
The MEDICREA® IMPIX ALIF S/A is an intervertebral fusion device with integrated fixation use as an adjunct to fusion. The implant is composed of:
- one PEEK cage with tantalum markers
- either 3 or 4 IMPIX ALIF S/A screws , depending on the footprint of cage considered.
The IMPIX ALIF S/A screw is manufactured from titanium alloy (Ti-6Al-4V) meeting ASTM F136 and ISO 5832-3 standards, and the PEEK cage is manufactured from PEEK OPTIMA® LT1 meeting the ASTM F 2026.The tantalum markers are manufactured from tantalum meeting the ASTMF560 and ISO 13782 standards.
Materials: Titanium alloy (Ti-6Al-4V) and PEEK OPTIMA® LT1 and tantalum
## 5. INDICATIONS FOR USE
The stand-alone intersomatic IMPIX ALIF S/A implants are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine from L2-S1. DDD is defined as discogenic back pain with deqeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).
This device is to be used with autogenous bone graft.
Patients should have a satisfactory course of conservative treatment (usually at least six (6) months of non-operative treatment) prior to treatment with an IMPIX-ALIF-S/A stand-alone cage.
The IMPIX ALIF S/A devices should be used with the three (3) bone screws (Small Footprint) or four (4) bone screws (Medium, Large Footprints) accompanying the device. Otherwise, an additional fixation system (e.g., pedicle screw system) should be implanted.
{5}------------------------------------------------
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The table below compares the features and characteristics of the IMPIX ALIF SA to its predicate device.
| Device | IMPIX<br>ALIF SA -<br>New | IMPIX ALIF<br>Lumbar IBF | IMPIX L | STALIF<br>MIDLINE | STALIF TT | SYNFIX LR |
|----------------------------|-------------------------------------------------|--------------------------|----------------------|----------------------------------------------------------------|-------------------------------|--------------------------------------------------|
| 510(k)<br>number | K062083 | K083798 | K072226 | K101301 | K073109 | K062083 |
| Lumbar | Yes | Yes | Yes | Yes | Yes | Yes |
| Anterior<br>Approach | Yes | Yes | No | Yes | Yes | Yes |
| Design | | | | | | |
| Anchorage<br>mean | 3 or 4<br>screws | No | No | 3 screws | 3 screws | 4 screws |
| Security<br>features | Screws<br>head<br>clipped<br>within the<br>cage | No | No | Titanium<br>ring on the<br>screws<br>clipped<br>in<br>the cage | No | Screw head<br>threaded in<br>a titanium<br>plate |
| Shape | Anatomical<br>shape | Anatomical<br>shape | Rectangular<br>shape | Semi-<br>cylindrical<br>shape | Semi-<br>cylindrical<br>shape | Rectangular<br>shape |
| Standalone<br>device | Yes | No | No | Yes | Yes | Yes |
| Materials | | | | | | |
| Ti-6Al 4V | Yes | No | No | Yes | Yes | No |
| Titanium<br>alloy<br>(TAN) | No | No | No | No | No | Yes |
| Tatalum | Yes | Yes | Yes | Yes | Yes | Yes |
| PEEK<br>OPTIMA®<br>LT1 | Yes | Yes | Yes | Yes | Yes | Yes |
Material composition is identical to other MEDICREA® INTERNATIONAL products that have been cleared via the 510(k) process.
{6}------------------------------------------------
## 7. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The biocompatibility evaluation for the IMPIX ALIF S/A system was conducted in accordance with the FDA blue book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- く Cytotoxicity
- V Sensitization
- V Irritation
- V Systemic toxicity
- V Pyrogen Testing
According to the standard ISO 10993-1, the IMPIX ALIF S/A System is defined as implantable device
in contact with tissue and bone and as a permanent contact with the patient.
For chemical composition, the material conform to Ti-6Al-4V ELI, following ASTM F136 and ISO 5832-3 standards, PEEK OPTIMA LT1 following ASTM F2026 standard and Tantalum following ASTM F560 and ISO 13782 standards.
#### Mechanical testinq
Testing in compliance with the FDA's June 12, 2007 "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device was performed for the IMPIX ALIF S/A system and demonstrated substantially equivalent performance to the identified predicate device systems.
The following Mechanical tests were performed:
- . FEA testing to determine worst case
- Static and dynamic axial compression (per ASTM F2077-11) ●
- Static and dynamic compression shear (per ASTM F2077-11) ●
- Subsidence (per ASTM F2267-04) ●
- Expulsion (per ASTM Draft Standard F-04.25.02.02) ●
- Screw clip-out evaluation o
#### Clinical study
No clinical studies were performed.
#### Animal study
No animal studies were performed.
#### 8. CONCLUSION
MEDICREA® INTERNATIONAL S.A. IMPIX ALIF S/A Spinal System - additional components are substantially equivalent to the already cleared STALIF MIDLINE(Centinel Spine, K101301); STALIF TT (Surgigraft Limited.K073109); SYNFIX LR (Synthes Spine, K062083); IMPIX ALIF Lumbar IBF (MEDICREA TECHNOLOGIE, K083798); IMPIX L (MEDICREA TECHNOLOGIE, K072226) in terms of intended use, materials used, sterilization, mechanical safety and performances
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.