← Product Code [NKB](/productcode/NKB) · K141186

# FIREBIRD SPINAL FIXATION SYSTEM (K141186)

_Orthofix, Inc. · NKB · Jan 2, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K141186

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jan 2, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle, and non pedicle fixation (T1 - S2/Ilium). Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications: Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis: Trauma (i.e., fracture or dislocation); Spinal stenosis: Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); Tumors; Pseudoarthrosis, and Failed previous fusion. When used for fixation to the ilium, the offset connectors of the Firebird Spinal Fixation System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level. The Phoenix MIS Fixation System when used with the Firebird Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery. The Firebird Spinal Fixation System components are used with certain components of the Orthofix Spinal Fixation System, including rods, rod connectors and cross-connectors. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Firebird Spinal Fixation System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Firebird Spinal Fixation System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

Temporary spinal fixation system; titanium alloy or cobalt chrome alloy components; includes rods, multi-axial/mono-axial pedicle screws, set screws, lateral offsets, bone screws, screw bodies, hooks, iliac connectors. Surgeon builds spinal implant construct; attaches to vertebral body/ilium via screw or hook fixation. Modification adds longitudinal line to straight titanium alloy rods for improved identification of induced curvature. Used in posterior spinal surgery; provides stabilization as adjunct to fusion. Benefits patient by correcting spinal deformities, stabilizing fractures, or supporting fusion in degenerative/pathological conditions.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717-14 (static and fatigue for spinal implant constructs) and ASTM F1798-13 (static and fatigue properties of interconnection mechanisms). Results demonstrate substantial equivalence to the predicate device.

## Technological Characteristics

Materials: Titanium alloy (Ti-6AL-4V ELI) or cobalt chrome alloy. Components: Rods, pedicle screws (multi-axial/mono-axial), set screws, lateral offsets, bone screws, screw bodies, hooks, iliac connectors. Design: Posterior non-cervical spinal fixation system. Modification: Longitudinal line added to straight titanium alloy rods. Standards: ASTM F1717-14, ASTM F1798-13.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Orthofix Firebird Spinal Fixation System ([K130932](/device/K130932.md))
- LDR Spine USA SpineTune TL System ([K120760](/device/K120760.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of a human figure with three faces in profile, representing health and well-being. The figure is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circle's perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 2, 2015

Orthofix, Incorporated Ms. Jacki Geren Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056

Re: K141186

Trade/Device Name: Firebird Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP Dated: October 31, 2014 Received: November 6, 2014

Dear Ms. Geren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Jacki Geren

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K141186

Device Name Firebird Spinal Fixation System

#### Indications for Use (Describe)

The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle fixation (T1 - S2/ Ilium). Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications:

a) Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);

b) Spondylolisthesis:

- c) Trauma (i.e., fracture or dislocation);
- d) Spinal stenosis;

e) Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis);

f) Tumors:

g) Pseudoarthrosis, and

h) Failed previous fusion

When used for fixation to the ilium, the offset connectors of the Firebird Spinal Fixation System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level.

The Phoenix MIS Fixation System when used with the Firebird Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

The Firebird Spinal Fixation System components are used with certain components of the Orthofix Spinal Fixation System, including rods, rod connectors and cross-connectors.

When used for posterior non-cervical pedicle screw fixation in pediatic patients, the Firebird Spinal Fixation System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Firebird System for pediatric use is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Orthofix logo. The logo consists of a blue abstract shape above the word "ORTHOFIX" in black, with a registered trademark symbol next to it. The abstract shape is made up of curved lines that form a circular shape with a gap in the middle.

# 510(k) Summary

| Name:                           | Orthofix Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                        | 3451 Plano Parkway<br>Lewisville, Texas 75056                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Telephone Number:               | 214-937-2000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Fax Number:                     | 214-937-3322                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Email:                          | jackigeren@orthofix.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Registration Number:            | 3008524126                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Contact Person:                 | Jacki Geren<br>Regulatory Affairs Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date Prepared:                  | May 6, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Name of Device                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Trade Name/Proprietary<br>Name: | Firebird Spinal Fixation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Common Name:                    | Spinal Fixation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:                   | NKB; OSH; KWP; MNH; MNI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulatory Classification:      | 21 CFR §888.3070<br>Pedicle screw spinal system (Class III)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Review Panel:                   | Orthopedic Device Panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Devices:              | Orthofix Firebird Spinal Fixation System - K130932 (primary predicate)<br>LDR Spine USA SpineTune TL System - K120760 (additional predicate)<br>No reference devices were used in this submission                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Reason for 510(k) Submission:   | Modification to all straight titanium alloy (Ti-6AL-4V ELI) rods to place<br>a line longitudinally along the length of the rod to provide better<br>identification of induced curvature.                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description:             | The Firebird Spinal Fixation System is a temporary, titanium alloy,<br>multiple component system comprised of a variety of non-sterile, single<br>use components, made of titanium alloy or cobalt chrome alloy, that<br>allow the surgeon to build a spinal implant construct. The system is<br>attached to the vertebral body and ilium by means of screw or hook<br>fixation to the non-cervical spine. The Firebird Spinal Fixation System<br>consists of an assortment of rods, multi-axial and mono-axial pedicle<br>screws, set screws, lateral offsets, bone screws, screw bodies, hooks and<br>iliac connectors. |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape above the word "ORTHOFIX" in black, bold letters. The shape is a curved, three-dimensional design, and the word "ORTHOFIX" is in a sans-serif font, with a registered trademark symbol next to the "X".

#### Intended Use / Indications for Use:

The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle, and non pedicle fixation (T1 - S2/Ilium). Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications:

- Degenerative disc disease (defined as discogenic back pain with degeneration of the disc a) confirmed by history and radiographic studies);
- Spondylolisthesis: b)
- Trauma (i.e., fracture or dislocation); C)
- Spinal stenosis: d)
- Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); e)
- f) Tumors;
- Pseudoarthrosis, and g)
- Failed previous fusion h)

When used for fixation to the ilium, the offset connectors of the Firebird Spinal Fixation System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level.

The Phoenix MIS Fixation System when used with the Firebird Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

The Firebird Spinal Fixation System components are used with certain components of the Orthofix Spinal Fixation System, including rods, rod connectors and cross-connectors,

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Firebird Spinal Fixation System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Firebird Spinal Fixation System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

| Characteristics              | Firebird Spinal Fixation<br>System<br>(Subject Device)                 | Predicate Device:<br>Firebird Spinal Fixation<br>System<br>(K130932)   | Predicate Device:<br>LDR Spine USA<br>SpineTune TL System<br>(K120760) |
|------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|
| Device Name                  | Firebird Spinal Fixation<br>System                                     | Firebird Spinal Fixation<br>System (K130932)                           | SpineTune TL System<br>(K120760)                                       |
| Method of<br>Fixation        | Non-Cervical Fixation                                                  | Non-Cervical Fixation                                                  | Non-Cervical Fixation                                                  |
| Implantation                 | Posterior Approach                                                     | Posterior Approach                                                     | Posterior Approach                                                     |
| Design                       | This system allows a<br>surgeon to build a spinal<br>implant construct | This system allows a<br>surgeon to build a spinal<br>implant construct | This system allows a<br>surgeon to build a spinal<br>implant construct |
| Material                     | Titanium alloy (Ti-6AL-<br>4V ELI)                                     | Titanium Alloy (Ti-6AL-<br>4V ELI)                                     | Titanium alloy                                                         |
| Longitudinal<br>Line on Rods | Yes                                                                    | No                                                                     | Yes                                                                    |

### Summary of Technological Characteristics of the Device Compared to the Selected Predicate Device

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape above the word "ORTHOFIX" in black, with a registered trademark symbol next to the word. The abstract shape is a curved, three-dimensional shape that resembles a stylized "O".

## PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence

| Characteristics                        | Standard / Test / FDA Guidance          |
|----------------------------------------|-----------------------------------------|
| Static Torsion Test                    | N/A No change from the predicate device |
| Static Axial Compression Bending Test  | N/A No change from the predicate device |
| Dynamic Axial Compression Bending Test | ASTM F1717-14                           |
| Axial Rod Gripping Test                | ASTM F1798-13                           |

### Performance Data Summary

Mechanical testing for the subject Firebird Spinal Fixation System lined rods were conducted in accordance with ASTM F1717-14 Standard Test Method for Static and Fatigue for Spinal Implant Constructs in a Vertebrectomy Model and ASTM 1798-13 Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants. Test results demonstrated that the subject lined rods are substantially equivalent to the predicate device Firebird Spinal Fixation System (K130932) which has the same intended use, similar indications, technological characteristics, and principles of operations

#### Basis of Substantial Equivalence

The subject Firebird Spinal Fixation System straight titanium alloy (Ti-6AL-4V ELI) lined rods are substantially equivalent to the predicate devices based on indications for use, intended use, design, materials and function and mechanical performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K141186](https://fda-staging.innolitics.com/device/K141186)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).