The Spacelabs Healthcare Telemetry Receiver, Model 96280, is intended to provide the Spacelabs Healthcare monitoring system with adult, pediatric and neonatal patient data of patients connected to Spacelabs Healthcare telemetry transmitters. Data includes physiological waveforms and calculations, cardiac arrhythmia and ST data, and patient demographic information to monitor adequacy of treatment or to exclude causes of symptoms.
Device Story
The Spacelabs Healthcare Telemetry Receiver (Model 96280) receives physiological data (ECG, SpO2, NIBP) from patient-worn telemetry transmitters. It processes this data to provide waveforms, calculations, and cardiac arrhythmia/ST-segment analysis. The device functions as part of a networked monitoring system, communicating via TCP/IP to the Xhibit Central Station, which serves as the primary alarming interface. Used in clinical settings, the system allows healthcare providers to view patient data on central displays to monitor treatment adequacy and identify symptoms. The device increases capacity over the predicate by supporting up to 16 patients per housing and connecting up to three displays to the central station.
Clinical Evidence
Bench testing only. Performance verified against ANSI/AAMI EC-57:2012 for cardiac rhythm and ST-segment measurement algorithms, and IEC standards for electrical safety (60601-1), electromagnetic compatibility (60601-1-2), alarm systems (60601-1-8), ECG monitoring (60601-2-27), and pulse oximetry (80601-2-61).
Technological Characteristics
Telemetry receiver for physiological monitoring (ECG, SpO2, NIBP). Connectivity via TCP/IP network protocol. Supports up to 16 patients per housing. Software-based arrhythmia and ST-segment analysis. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-61, and ANSI/AAMI EC-57.
Indications for Use
Indicated for adult, pediatric, and neonatal patients connected to Spacelabs Healthcare telemetry transmitters for physiological monitoring, including cardiac arrhythmia and ST-segment analysis.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Spacelabs Medical, Inc. Model 90478 and 90479 Telemetry Receiver Module and Housing (K925510)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002
September 9, 2014
Spacelabs Healthcare % Thomas Kroenke Speed To Market, Inc. P.O. Box 3018 Nederland, CO 80466 US
Re: K141156
> Trade/Device Name: Spacelabs Healthcare Telemetry Receiver Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient (with Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: MHX, DSI, MLD, MSX Dated: June 31, 2014 Received: August 4, 2014
Dear Thomas Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
| 510(k) Number (if known): | K141156 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Spacelabs Healthcare Telemetry Receiver, Model 96280 |
| Indications for Use: | The Spacelabs Healthcare Telemetry Receiver, Model 96280, is<br>intended to provide the Spacelabs Healthcare monitoring system<br>with adult, pediatric and neonatal patient data of patients<br>connected to Spacelabs Healthcare telemetry transmitters. Data<br>includes physiological waveforms and calculations, cardiac<br>arrhythmia and ST data, and patient demographic information to<br>monitor adequacy of treatment or to exclude causes of<br>symptoms. |
X ____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
| Submission Date: | 02 May 2014 |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare<br>35301 SE Center St<br>Snoqualmie, WA 98065 |
| Submitter Contact: | Mr. Al Van Houdt<br>Spacelabs Healthcare<br>Phone: +1 (425) 363- 5970<br>Fax: +1 (425) 363-5762<br>Email: al.vanhoudt@spacelabs.com |
| Application<br>Correspondent: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 |
| Manufacturing Site: | Spacelabs Healthcare<br>35301 SE Center St<br>Snoqualmie, WA 98065 |
| Trade Name: | Spacelabs Healthcare Telemetry Receiver, Model 96280 |
| Common Name: | Telemetry Receiver |
| Classification Name: | Monitor, Physiological, Patient (With Arrhythmia Detection or<br>Alarms); |
| Primary<br>Classification<br>Regulation: | 21 CFR §870.1025 |
| Primary Product<br>Code: | MHX |
| Secondary<br>Classification<br>Regulation: | 21 CFR §870.1025; 21 CFR §870.2300 |
| Secondary Product<br>Code: | DSI; MLD; MSX |
{4}------------------------------------------------
| Substantially<br>Equivalent Devices: | New Spacelabs Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|----------------------------------------------------------------------------------------------|
| | Spacelabs Healthcare<br>Telemetry Receiver,<br>Model 96280 | K925510 | Spacelabs Medical, Inc.<br>Model 90478 and 90479<br>Telemetry Receiver<br>Module and Housing |
| Device Description: | The Spacelabs Healthcare (Spacelabs) Telemetry Receiver, Model<br>96280, (ETR) is a new version of a currently marketed Spacelabs<br>product. The Spacelabs ETR offers receipt and analysis of patient data<br>for those patients connected to a Spacelabs Healthcare telemetry<br>transmitter. | | |
| | The Spacelabs ETR provides for data communication using the TCP/IP<br>network protocol employed in the Xhibit Central Station, Model<br>96102, (Xhibit) (K122146) network of hardwired and/or ETR<br>monitored patients. | | |
| | Xhibit is the primary alarming device for the ETR telemetry receiver<br>system. | | |
| Intended Use: | The Spacelabs Healthcare Telemetry Receiver, Model 96280, is<br>intended to provide the Spacelabs Healthcare monitoring system with<br>adult, pediatric and neonatal patient data of patients connected to<br>Spacelabs Healthcare telemetry transmitters. Data includes<br>physiological waveforms and calculations, cardiac arrhythmia and ST<br>data, and patient demographic information to monitor adequacy of<br>treatment or to exclude causes of symptoms. | | |
| Technology<br>Comparison: | The Spacelabs ETR employs the same technological characteristics as<br>the predicate device. | | |
| | Characteristic | Predicate Device | Proposed Device |
| | Number of Patients per<br>Receiver Channel | 1 | Same |
| | Patients per Receiver<br>Housing | Up to 8 | Up to 16 |
| | Receivers Connected to<br>Central Station | Two (2)<br>(Ultraview can display 16<br>patients) | Three (3)<br>(Xhibit can display 48 patients) |
| | Number of Displays /<br>Patients per Display | One (1) display / 16 patients<br>per display | Three (3) displays / 16 patients<br>per display |
| | Receiver Capability | ECG, SpO2, NIBP | Same |
| | ECG Analysis in Receiver | Yes | Same |
Same
Yes
Prioritized Alarms
{5}------------------------------------------------
#### Summary of Performance Testing:
| Software | The Spacelabs ETR contains MAJOR level of concern software.<br>Software was designed and developed according to a robust software<br>development process, and was rigorously verified and validated.<br>Software information is provided in accordance with internal<br>requirements and the following standards and guidance documents: FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05; FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99; FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; FDA guidance: Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm, 28 October 2003; FDA guidance: Cybersecurity for networked medical devices containing off-the-shelf (OTS) software, 14 January 2005; and IEC 62304: 2006, Medical device software – Software life cycle processes. Test results indicate that the Spacelabs ETR complies with its predetermined specifications and the applicable standards and guidance documents. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | The Spacelabs ETR was tested for performance in accordance with the following standard: IEC 60601-1: 2005, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. Test results indicate that the Spacelabs ETR complies with the applicable standards. |
| Electromagnetic<br>Compatibility | The Spacelabs ETR was tested for performance in accordance with the following standard: IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests. Test results indicate that the Spacelabs ETR complies with the applicable standards. |
| Performance Testing<br>– Bench | The Spacelabs ETR was tested for performance in accordance with<br>internal requirements and the following standards:<br>ANSI/AAMI EC-57: 2012, Testing and reporting performance<br>results of cardiac rhythm and ST-segment measurement algorithms. IEC 60601-1-8 2006, Am1: 2012, Medical electrical equipment –<br>General requirements, tests and guidance for alarm systems in<br>medical electrical equipment and medical electrical systems. IEC 60601-2-27: 2011, Medical electrical equipment – Part 2-27:<br>Particular requirements for the basic safety and essential<br>performance of electrocardiographic monitoring equipment. IEC 80601-2-61: 2011, Medical electrical equipment – Part 2-61:<br>Particular requirements for basic safety and essential performance<br>of pulse oximeter equipment. IEC 62366: 2007, Medical devices – Application of usability<br>engineering to medical devices. Test results indicate that the Spacelabs ETR complies with its<br>predetermined specifications and the applicable standards. |
| Conclusion | Verification and validation activities were conducted to establish the<br>performance and safety characteristics of the Spacelabs ETR. The<br>results of these activities demonstrate that the Spacelabs ETR is safe<br>and effective when used in accordance with its intended use and<br>labeling. |
{6}------------------------------------------------
Therefore, the Spacelabs ETR is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.