ASTROGLIDE TTC FERTILITY FRIENDLY PERSONAL LUBRICANT
K141132 · Biofilm, Inc. · PEB · Jan 9, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K141132
Device Name
ASTROGLIDE TTC FERTILITY FRIENDLY PERSONAL LUBRICANT
Applicant
Biofilm, Inc.
Product Code
PEB · Obstetrics/Gynecology
Decision Date
Jan 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Astroglide® TTC is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. Astroglide TTC is compatible with sperm, oocytes, and embryos and can be used by couples trying to conceive. Astroglide® TTC is compatible with natural rubber latex, polyisoprene and polyurethane condoms.
Device Story
Astroglide TTC is a non-sterile, water-based personal lubricant; applied by hand or via pre-filled tube applicator to penile or vaginal tissues. Formulated with water, propylene glycol, hydroxyethylcellulose, fructose, methylparaben, sodium phosphate, potassium phosphate, propyl paraben, galactose, and sodium hydroxide. Designed with pH and osmolarity compatible with sperm survival and migration. Used by couples trying to conceive in home settings. Lubricant properties do not impede sperm penetration; does not harm sperm motility, DNA integrity, or fertilization capability. Enhances comfort during sexual activity while maintaining compatibility with condoms and reproductive cells.
Clinical Evidence
Bench testing only. Performance data includes: Mouse Embryo Assay (MEA) showing >80% expanded blastocysts; Human Sperm Survival Assay (HSSA) showing ≥80% survival relative to control; Computer Assisted Sperm Analysis (CASA) confirming progressive motility; DNA integrity testing; hamster zona-free ovum penetration assays; and lubricant barrier assays. Biocompatibility testing per ISO 10993-5, -10, and -11 (cytotoxicity, sensitization, vaginal irritation, systemic toxicity) confirmed non-irritating and non-cytotoxic results. Antimicrobial effectiveness and 8-month stability data provided.
Technological Characteristics
Water-based lubricant; ingredients include propylene glycol, hydroxyethylcellulose, fructose, methylparaben, sodium phosphate, potassium phosphate, propyl paraben, galactose, sodium hydroxide. pH 7.2-7.6; osmolarity 220-400 mOsm/kg. Packaged in pre-filled tube applicators. Biocompatibility testing per ISO 10993 standards. Non-sterile.
Indications for Use
Indicated for couples trying to conceive as a personal lubricant for penile and/or vaginal application to moisturize, lubricate, and enhance comfort during intimate sexual activity; compatible with sperm, oocytes, embryos, and natural rubber latex, polyisoprene, and polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2015
Biofilm, Inc. Sherry Castello Regulatory Affairs and Quality Assurance Associate 3225 Executive Ridge Vista, CA 92081
K141132 Trade/Device Name: Astroglide TTC Fertility Friendly Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: PEB Dated: November 24, 2014 Received: November 25, 2014
Dear Sherry Castello,
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141132
Device Name
Astroglide TTC Fertility Friendly Personal Lubricant
Indications for Use (Describe)
Astroglide® TTC is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. Astroglide TTC is compatible with sperm, oocytes, and embryos and can be used by couples trying to conceive. Astroglide® TTC is compatible with natural rubber latex, polyisoprene and polyurethane condoms.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Biofilm, Inc. The text "BIOFILM, INC." is in a bold, sans-serif font and is purple. Underneath the text is a curved line, also in purple, that resembles a wave.
## 510(k) Summary Astroglide TTC Fertility Friendly Personal Lubricant
- General Information on Submitter i.
| Applicant: | BioFilm, Inc. |
|-----------------------------|---------------------------------------------|
| Address: | 3225 Executive Ridge<br>Vista, CA 92081 USA |
| Telephone: | 760-727-9030 |
| Fax: | 760-727-8080 |
| Contact Person: | Sherry Castello |
| Email: | sherry@biofilm.com |
| Date Prepared: | November 24, 2014 |
| Establishment Registration: | 2025771 |
| ii. | General Information on Device | | |
|-----|-------------------------------|----------------------------------------------------------------------------------------------------------------------|--|
| | Proprietary Name: | Astroglide® TTC Fertility Friendly Personal Lubricant | |
| | 510(k) Number: | K141132 | |
| | Common Name: | Personal Lubricant | |
| | Classification Name: | Lubricant, Personal, Gamete, Fertilization, and Embryo<br>Compatible (21 CFR 884.5300, Condom), product code:<br>PEB | |
#### iii. Predicate Device
Pre-Va Vaginal Lubricant, K072741, manufactured by INGFertility, LLC
#### Description of Device iv.
Astroglide® TTC Fertility Friendly Personal Lubricant is a non-sterile, clear, and water-based personal lubricant for penile and/or vaginal application. Astroglide® TTC has a pH and osmolarity that is compatible with sperm survival and migration. The device is packaged in pre-filled tube applicators which may be used for single use intra-vaginal application or applied directly to penis or vagina.
Astroglide® TTC is formulated using water, propylene glycol, hydroxyethylcellulose, fructose, methylparaben, sodium phosphate, potassium phosphate, propyl paraben, galactose, and sodium hydroxide.
This device is batch lot tested for appearance, color, viscosity, pH, microbial count, osmolality, endotoxin, mouse embryo assay and human sperm survival assay.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Biofilm, Inc. The logo is purple and features the company name in a bold, sans-serif font. Underneath the name is a curved line, also in purple, that adds a visual element to the logo. The registered trademark symbol is present next to the "C" in "INC."
#### Indications for Use V.
Astroglide® TTC is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. Astroglide TTC is compatible with sperm, oocytes, and embryos and can be used by couples trying to conceive. Astroglide® TTC is compatible with natural rubber latex, polyisoprene and polyurethane condoms.
#### vi. Technological Characteristics of Astroglide® TTC Compared to Predicate
Technological Characteristics of Astroglide TTC and the predicate are very similar. Any minor technological differences noted do not affect the safety or effectiveness of the device.
| Attribute | TTC, Fertility Friendly<br>Personal Lubricant | Predicate: Pre-Va Vaginal<br>Lubricant |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K141132 | K072741 |
| Indications for use | Astroglide® TTC is a<br>personal lubricant for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate<br>sexual activity and<br>supplement the body's<br>natural lubrication.<br>Astroglide TTC is<br>compatible with sperm,<br>oocytes, and embryos and<br>can be used by couples trying<br>to conceive. Astroglide®<br>TTC is compatible with<br>natural rubber latex,<br>polyisoprene and<br>polyurethane condoms. | ● To lubricate vaginal<br>tissues to facilitate entry of<br>diagnostic or therapeutic<br>devices including those<br>used in fertility<br>interventions. Pre-Va may<br>be applied directly to the<br>device or may be<br>deposited intravaginally<br>using the applicator prior<br>to the insertion of<br>diagnostic or therapeutic<br>devices used in fertility<br>interventions.<br>● As a personal lubricant<br>Pre-Va supplements the<br>body's own natural<br>lubrication fluids, to<br>moisturize, relieve friction<br>and to enhance the ease<br>and comfort of intimate<br>sexual activity. Pre-Va is<br>safe for use by couples<br>who are trying to conceive<br>and may be applied to<br>vaginal or penile tissues<br>for lubrication and |
{5}------------------------------------------------
| Attribute | TTC, Fertility Friendly<br>Personal Lubricant | Predicate: Pre-Va Vaginal<br>Lubricant |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ingredients | Water, Propylene Glycol,<br>Hydroxyethylcellulose, Fructose,<br>Methylparaben, Sodium<br>Phosphate, Potassium Phosphate,<br>Propyl Paraben, Galactose,<br>Sodium Hydroxide | Water, Hydroxyethylcellulose,<br>Pluronic 127, Sodium Chloride,<br>Arabinogalactan, Sodium<br>Phosphate, Carbopol 934P,<br>Methyl Paraben, Sodium<br>Hydroxide, Potassium Phosphate |
| Method of<br>Application | Directly by hand or intra-<br>vaginally with pre-filled tube<br>applicator | Directly from tube or with<br>separate disposable applicator |
| Storage Instruction | 36°F to 86°F | 36°F to 86°F |
| pH | 7.2-7.6 | 7.20-7.45 |
| Osmolarity | 220-400 mOsm/kg | 260 to 370 mOsm/kg |
| Endotoxin by LAL<br>methodology | $≤$ 0.7 EU/mL | $≤$ 0.7 EU/mL |
| Mouse Embryo<br>Assay (MEA) | Using 1-cell MEA exposed to<br>10% solution for 30 minutes<br>$>$ 80% expanded blastocysts at 96<br>hrs. | Using 1-cell MEA exposed to 5%<br>solution for 30 minutes, $>$ 80%<br>expanded blastocysts at 96 hrs. |
| Human Sperm<br>Survival Assay<br>(HSSA) | After exposure to 10% TTC for 2<br>hours, $≥$ 80% of the control. | After exposure to 10% Pre-Va for<br>30 min, $≥$ 80% of the control. |
| Condom<br>Compatibility | Compatible with latex,<br>polyurethane, and polyisoprene<br>condoms | Compatible with latex and<br>polyurethane condoms. |
#### Summary of Performance Data vii.
Lubricant barrier assays performed on Astroglide® TTC demonstrated that the properties of the device do not impede sperm penetration into the lubricant. Mouse In vitro fertilization assays in combination with Mouse embryo assays (IVF-MEA) demonstrated normal fertilization and embryo development with no suggestion of toxicity. Mucosal penetration studies using a hyaluronic acid (that mimics cervical mucus) indicate that the "swim-up" ability of sperm exposed to Astroglide TTC is not affected. Computer Assisted Sperm Analysis demonstrates the progressive motility of sperm is not harmed by Astroglide TTC. DNA integrity testing after exposure to Astroglide TTC demonstrated the device does not harm human sperm chromatin (DNA). Sperm penetration assays (hamster zona-free ovum tests) demonstrated that the presence of Astroglide TTC did not affect the penetration of hamster ova by human sperm.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Biofilm, Inc. The logo is purple and features the company name in bold, sans-serif font. A curved line is below the company name. The registered trademark symbol is to the right of the company name.
Condom compatibility studies indicate that TTC is safe to use with natural rubber latex, polyisoprene, and polyurethane condoms.
Biocompatibility tests and results are shown in the table below:
| Biocompatibility Testing | Astroglide TTC Result |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity, Direct Contact, ISO 10993-5,<br>formula | The test article was not considered to have a<br>cytotoxic effect. |
| Guinea Pig Maximization, ISO 10993-10,<br>formula | The test article did not elicit sensitization<br>reactions. |
| Vaginal Irritation, ISO 10993-10, formula | The test article was considered non-irritating<br>to the vaginal mucosa in New Zealand White<br>Rabbits. |
| Systemic Toxicity, ISO 10993-11, formula | Test article met the requirements of ISO<br>10993-11. No test animals exhibited any<br>biological reactivity. |
| Cytotoxicity, MEM Elution, ISO 10993-5,<br>applicator tube | The test article was not considered to have a<br>cytotoxic effect. |
| Guinea Pig Maximization, ISO 10993-10,<br>applicator tube | The test article did not elicit sensitization<br>reactions. |
| Vaginal Irritation, ISO 10993-10, applicator<br>tube | In New Zealand White Rabbits, the test<br>article when extracted in saline was<br>considered minimally irritating. The test<br>article when extracted in cottonseed oil was<br>considered non-irritating. |
On-going stability studies show the device has met acceptance criteria for appearance, color, clarity, odor, viscosity, pH, microbial count, osmolality, endotoxin, mouse embryo assay, and human sperm survival assay for an 8 month shelf-life. Antimicrobial effectiveness testing has been performed and the preservative system is effective.
The performance data concludes that Astroglide® TTC is substantially equivalent to the predicate device Pre-Va Vaginal Lubricant. Astroglide® TTC does not harm sperm, oocytes, or embryos and is safe for use by couples trying to conceive.
#### viii. Conclusion
The proposed and predicate indication statements do not represent a new intended use. Astroglide TTC personal lubricant has the same intended use as PreVa for use in a nonclinical environment. The technological characteristics of Astroglide TTC and the predicate are very similar. Any minor technological differences noted do not affect the safety or efficacy of the device. Performance testing demonstrates that Astroglide TTC does not harm sperm, oocytes, or embryos and is safe to use by couples trying to conceive.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.