The Single-Site™ Wristed Needle Driver instrument is intended to be used with the da Vinci® Si™ Surgical System (IS3000) to perform needle driving and suture tying tasks.
Device Story
Resposable endoscopic instrument for use with da Vinci Si Surgical System (IS3000); enables needle driving and suture tying. Features straight grip and distal wrist articulation for increased range of motion. Used by trained physicians in operating rooms for single-incision laparoscopic procedures (cholecystectomy, hysterectomy, salpingo-oophorectomy). Surgeon operates via system console; system translates hand movements into instrument articulation. Provides enhanced dexterity for suturing compared to non-wristed instruments. Benefits include improved access to target anatomy and precision in tissue approximation during minimally invasive surgery.
Clinical Evidence
No human clinical trials provided. Evidence includes bench verification (physical, mechanical, electrical, and life testing for 5 uses), cadaver studies (n=5 surgeons, vaginal cuff closure), and live porcine models (n=5 surgeons, uterine horn hysterotomy). Equivalence testing compared subject device to predicate across 52 suturing tasks, evaluating suture line quality, tissue bite, stitch distance, and closure time. Results confirmed performance, safety, and functionality equivalent to predicate.
Technological Characteristics
Resposable endoscopic instrument; semi-rigid shaft; distal wrist articulation; straight grip. Compatible with da Vinci Si Surgical System (IS3000). Sterilization required for reuse (up to 5 cycles). Mechanical actuation via system interface.
Indications for Use
Indicated for use by trained physicians in an operating room for endoscopic tissue manipulation, grasping, cutting, dissection, approximation, clip-ligation, electrocautery, and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy, and salpingo-oophorectomy with da Vinci Single-Site Instruments and Accessories.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing right, with flowing lines that resemble hair or ribbons.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26, 2014
Intuitive Surgical Incorporated Ms. Einav Yemini Associate Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086
Re: K141075
Trade/Device Name: Single-Site™ Wristed Needle Driver Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: August 22, 2014 Received: August 27, 2014
Dear Ms. Yemini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Precautions/Warnings/Contraindications section of the device's labeling:
{1}------------------------------------------------
The safety and effectiveness of this device for use in the performance of general laparoscopic abdominal surgery or general gynecological surgery procedures have not been established. This device is only intended to be used for single incision laparoscopic cholecystectomy, benign hysterectomy, and salpingo-oophorectomy with the da Vinci Single Site Instruments and the da Vinci Si Surgical System (IS3000).
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [OR, for IVDs only] Parts 801 and 809 ] ); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{2}------------------------------------------------
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# William H. Maisel -S
William H. Maisel, MD, MPH Director, Office of Device Evaluation (Acting) Deputy Center Director for Science Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) Not known. To be assigned.
Device Name Single-Site® Wristed Needle Driver
#### Indications for Use (Describe)
The Intuitive Surgical® da Vinci® Single-Site™ Instruments and Accessories used with the da Vinci® Si Surgical System (133000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery and suturing single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, suction irrigators, monopolar cautery, bipolar cautery, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, flexible blunt obturators, and the Single-Site Port.
Type of Use (Select one or both, as applicable)
∑ Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
{5}------------------------------------------------
## 510(k) Summary [As Required by 21 CFR 807.92(c)]
| 510(k) Owner: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Einav Yemini<br>Associate Regulatory Engineer<br>408-523-2656 (phone)<br>408-523-8907 (fax)<br>einav.yemini@intusurg.com |
| Date Summary Prepared: | August 22, 2014 |
| Trade Name: | Single-Site™ Wristed Needle Driver |
| Common Name: | Endoscope and accessories |
| Product Code: | NAY |
| Classification: | Class II<br>Endoscope and Accessories, 21 CFR 876.1500 |
| Classification Advisory<br>Committee: | General and Plastic Surgery |
| Predicate Devices: | Single-Site™ Curved Needle Driver (K122532) |
## Device Description:
The da Vinci® Single-Site™ Instruments and Accessories consist of semi-rigid shaft instruments, two fixed-shape curved cannulae sets (250 mm and 300 mm lengths), an accessory cannula for insertion of manual laparoscopic instruments, semi-rigid blunt obturators (250 mm and 300 mm lengths), a rigid 10 mm Blunt Obturator, and a Single-Site™ Port (with insufflation adapter and stopcock) for the placement and insertion of multiple cannulae/instruments through a single incision.
{6}------------------------------------------------
The da Vincio Single-Site™ Instruments and Accessories include instruments to perform manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clipligation, electrocautery, and suturing. The da Vincio Single-Site™ Instruments and Accessories are intended to be used with the existing da Vinci® Si™ Surgical System (IS3000).
The subject device, Single-Site™ Wristed Needle Driver (P/N 428115), is a new addition to the Single-Site" Instruments and Accessories family. It is a resposable endoscopic instrument used in conjunction with the da Vinci® Si " Surgical System (IS3000) to handle needles and suture during endoscopic surgical procedures. The distal end configuration of the instrument includes a straight grip and a wrist to enable articulation of the grip.
## Intended Use:
The Single-Site™ Wristed Needle Driver instrument is intended to be used with the da Vinci@ Si" Surgical System (IS3000) to perform needle driving and suture tying tasks.
## Indications for Use
The Intuitive Surgical® da Vincio Single-Site™ Instruments and Accessories used with the da Vinci@ Si™ Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery, and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingooophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators, monopolar cautery, bipolar cautery, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, flexible blunt obturators, and the Single-Site™ Port.
The indications for use are unchanged from the predicate (K122532).
## Technological Characteristics
The Single-Site™ Wristed Needle Driver instrument is substantially equivalent to the cleared Single-Site™ Curved Needle Driver (K122532). The design of the Single-Site Wristed Needle Driver is a modification of the da Vinci@ Single-Site" Curved Needle Driver (K122532) to incorporate a wrist at the distal end of the instrument. The wrist provides additional range of motion and facilitates suturing tasks. Additionally, the grip configuration was revised from a curved to a straight grip. The Single-Site" Wristed Needle Driver is equivalent to the predicate Single-Site" Curved Needle Driver in terms of its technological characteristics, and the intended use is identical.
{7}------------------------------------------------
#### Performance Data
Performance data (bench, animal and cadaver testing) demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. The tests included dimensional measurements, mechanical and functional verification, and simulated use in animal and cadaver models as follow.
#### Bench Verification
The Single-Site™ Wristed Needle Driver instrument was subjected to design verification and reliability (life) testing to evaluate performance and demonstrate that the design output meets design input requirements. The tables below provide a summary of the bench verification testing.
| Category | Description |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose / Scope | Verification that the design of the instrument meets its physical,<br>mechanical, electrical, user interface, and equipment interface<br>requirements. |
| # of Test Samples | Five (5) |
| Models | Bench Testing |
| Test Methodology | Physical measurements, verification of mechanical requirements,<br>functional performance requirements, electrical requirements, user<br>interface requirements, and equipment interface requirements. |
#### Design Verification Testing
#### Reliability (life) Testing
| Category | Description |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose / Scope | Testing to demonstrate that the instrument will operate properly<br>throughout its intended life of five uses. |
| # of Test Samples | Eight |
| Models | Bench testing |
| Test Methodology | Instruments are subjected to repeated cycles of simulated life, including<br>cleaning, sterilization, and simulated surgical tasks that mimic actual<br>maneuvers performed during minimally invasive surgical operations.<br><br>Performance requirements are evaluated during each cycle and more<br>extensive mechanical measurements are done at cycles #1, #5, and #7, to<br>confirm performance remains within specification. |
{8}------------------------------------------------
#### Cadaver and Animal Validations
A cadaver model was used to demonstrate clinical performance for anatomical access and reach. A live animal model was used to assess safety and performance in cases where a live tissue model was appropriate. The live animal model replicates factors experienced during clinical use, including working with perfused organs, bleeding, normal tissue handling, and ensure that appropriate hemostasis is achieved and maintained.
For design validation, the cadaver model allowed evaluation of vaginal cuff closure, and a live porcine model allowed evaluation of suturing live tissue that is representative of dense muscular tissue, similar to the vaginal cuff. The table below provides a summary of the design validation testing.
| Category | Description |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Purpose / Scope</b> | Confirm that instrument meets product requirements, specifications, and user needs |
| <b># of Test Samples</b> | Two (2) |
| <b>Models Used</b> | Cadaver (3) and animal (2) |
| <b>Test Methodology</b> | Utilize instrument in representative uses to confirm that device meets general clinical use requirements, performance, functionality, compatibility, and user-related requirements |
#### Design Validation Testing
For equivalency testing, five (5) independent practicing surgeons participated in a study completing a comprehensive set of suturing tasks using both the Single-Site™ Wristed Needle Driver (subject) and the Single-Site™ Curved Needle Driver (predicate). With each instrument, each surgeon completed at least two gynecologic suturing tasks in cadaver models and three gynecologic suturing tasks in porcine models, for a total of 52 completed suturing tasks with both instruments. Surgeons completed questionnaires that evaluated the suturing performance, suture line quality, and safety with the each needle driver; study observers recorded objective assessments of the suture lines quality. The tables below provide a summary of the equivalence testing.
{9}------------------------------------------------
| Category | Description |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose / Scope | Demonstrate substantial equivalence of addition of a wrist to the<br>needle driver instrument ( <i>Single-Site Wristed Needle Driver</i> ). |
| Comparison Device(s) | <i>Single-Site Curved Needle Driver</i> |
| Models used /<br>Evaluation areas | The ability to perform a complete procedure (benign hysterectomy<br>with bilateral salpingo-oophorectomy) was evaluated using a<br>cadaver model. Cadavers were used to represent human anatomy<br>and size, and to confirm that the instrument can adequately access<br>the necessary structures and perform the complete procedure.<br>Comparison of the safety and performance of the applicable needle<br>driver to the comparative device was done as follows:<br>• Ability to perform a vaginal cuff closure in a cadaver model. |
| Evaluators / Number of<br>Evaluators | 5 independent surgeons (5 cadavers) |
| Procedures / Tasks<br>evaluated | Full Procedures:<br>Benign hysterectomy with bilateral salpingo- oophorectomy (n=5)<br>Cadaver suturing task:<br>Vaginal cuff closure (n=11) |
| Procedure Completion | Confirmation that surgeon was able to complete full procedure |
| Suture Line Quality and<br>Effectiveness | Likert Scale and yes/no questions used to compare surgeons visual<br>assessment of quality and effectiveness of the suture lines.<br>Intraoperative measurements:<br>• Tissue Bite<br>• Number of stitches<br>• Distance between stitches<br>• Closure completion time |
| Suturing Task Elements | Likert Scale used to compare the ability to perform suturing tasks. |
| Ease of Use | Subjective surgeon assessment of ability to perform task/procedure |
| Safety Assessments | Study observer, surgeon participant, and independent reviewer<br>assessments<br>• Ability to operate safely<br>• Any instances of harm |
#### Equivalence Testing – Cadaver
{10}------------------------------------------------
| Category | Description |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose / Scope | Demonstrate substantial equivalence of addition of a wrist to the needle driver instrument ( <i>Single-Site Wristed Needle Driver</i> ). |
| Comparison Device(s) | <i>Single-Site Curved Needle Driver</i> |
| Models used / Evaluation areas | Comparison of the safety and performance of the applicable needle driver to the comparative device was done as follows:<br>• Ability to perform applicable suturing tasks in a live porcine model.<br>Live animals (porcine) were used to represent the effects of live tissue during clinical use and compare performance and safety for representative suturing tasks. |
| Evaluators / Number of Evaluators | 5 independent surgeons (5 animals) |
| Procedures / Tasks evaluated | Porcine suturing task:<br>Uterine horn hysterotomy closure (n=15, 3/surgeon)<br>Surgeon's assessment (yes/no) and Likert Scale questions used to compare quality and effectiveness of the suture lines. |
| Suture Line Quality and Effectiveness | Intraoperative measurements:<br>• Tissue Bite<br>• Number of stitches<br>• Distance between stitches<br>• Closure completion time |
| Suturing Task Elements | Likert Scale used to compare the ability to perform suturing tasks. |
| Ease of Use | Subjective surgeon assessment of ability to perform task/procedure |
| Safety Assessments | Study observer, surgeon participant, and independent reviewer assessments<br>• Ability to operate safely<br>• Any instances of harm |
#### Equivalence Testing - Animal
#### Clinical Evaluation
No clinical testing was provided with this submission using the subject device. Safety, effectiveness and substantial equivalence were demonstrated through bench testing, cadaver and animal validations, and comparison testing, performed with the subject device. The testing performed demonstrated that:
- The device can perform its intended function ●
- The device can perform the surgical tasks proposed in the Indications for Use . safely and effectively
- The clinical performance and safety of the device is equivalent to the predicate . device
{11}------------------------------------------------
#### Human Factors Evaluation
Risk analysis was performed for Human Factors. No human factors testing was provided with this submission using the subject device. The device does not include any new user interfaces and does not present any new usability risks compared to the predicate.
#### Conclusion
Based on the intended use, indications for use, technological characteristics, and performance data, the Single-Site™ Wristed Needle Driver is substantially equivalent (SE) to the predicate device. This SE determination is based on bench testing including reliability testing, animal/cadaver validation, simulated clinical procedures in live animals, and human factors assessment. The bench/reliability testing verified that the design requirements and specifications for the new and/or changed components of the instrument were met. The animal and cadaver validation demonstrated the users' ability to use the system to accurately control the instrument, to reach the necessary target anatomy, and to perform the surgical task. The simulated clinical procedures in cadavers and live animals provided clinical validation that the instrument can safely and effectively complete surgical tasks in procedure steps encompassed by the Indications for Use statement.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.