K141070 · Togo Medikit Co., Ltd. · DRE · Jan 15, 2015 · Cardiovascular
Device Facts
Record ID
K141070
Device Name
SUPER SHEATH
Applicant
Togo Medikit Co., Ltd.
Product Code
DRE · Cardiovascular
Decision Date
Jan 15, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1310
Device Class
Class 2
Indications for Use
The device is indicated for use in the introduction of diagnostic and interventional devices inserted into the peripheral and coronary vasculature.
Device Story
Super Sheath is a catheter introducer system consisting of a sheath, dilator, and guidewire. The sheath features a hemostatic valve to minimize blood leakage and a side tube with a three-way stopcock for fluid/medication administration or blood sampling. The dilator facilitates sheath entry into the vasculature and is removed once the sheath is positioned. The device is used in clinical settings by healthcare professionals to provide a conduit for diagnostic or interventional devices. The sheath hub and dilator hub lock via a rotating motion. The device is provided sterile for single-use only. It benefits patients by enabling minimally invasive access to the peripheral and coronary vasculature for diagnostic or therapeutic procedures.
Clinical Evidence
Bench testing only. Testing included sheath shaft tensile, kink, hub-to-shaft tensile, valve integrity, lubricity, radiopacity, cover tube durability, pressure, corrosion resistance, and guidewire performance. Biocompatibility testing included cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, hemocompatibility, pyrogen, and LAL tests. No clinical data provided.
Technological Characteristics
Sheath shaft/hub: ETFE and polyamide. Dilator: Polypropylene (4F-8F) or FEP (9F). Guidewire: Stainless steel coil with inner mandrel. Radiopaque marker: Tantalum. Sterilization: Ethylene oxide. Connectivity: None. Dimensions: 4F-9F diameters, 7-25 cm lengths. Single-use.
Indications for Use
Indicated for introduction of diagnostic and interventional devices into peripheral and coronary vasculature. No specific patient population age or gender restrictions stated.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
Predicate Devices
Super Sheath Introducer Sheath and Introducer Sheath Sets, 4F-9F (k052557)
Super Sheath Introducer Sheath and Introducer Sheath Sets, 3.3F (k121504)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2015
Togo Medikit Co., Ltd. Izumi Maruo Senior Consultant 4-1-17 Hongo Bunkyo-ku, 113-0033 JA
Re: K141070 Trade/Device Name: Super Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DRE Dated: December 8, 2014 Received: December 12, 2014
Dear Izumi Maruo,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141070
Device Name SUPER SHEATH
Indications for Use (Describe)
The device is indicated for use in the introduction of diagnostic and interventional devices inserted into the peripheral and coronary vasculature.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
TOGO MEDIKIT CO., LTD.
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008
#### 510(k) Summary
- a. Owner/Company name, address TOGO MEDIKIT CO., LTD. 17148-6 Aza Kamekawa, Oaza Hichiya Hyuga City, Miyazaki Prefecture, 883-0062, Japan
Daisuke Nagamizu Manager, Regulatory Affairs Section, Quality Assurance Department Phone: 011-81-982-53-8000 011-81-982-53-8008 Fax: Email: qc@togomedikit.co.jp
# b. Contact/Application Correspondent
Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
Phone: 011-81-3-3818-8577 011-81-3-3818-8573 Fax: Email: maruo@mici.co.jp
# c. Date prepared
April 18, 2014
#### d. Name of device
| Trade Name: | SUPER SHEATH |
|----------------------------|----------------------|
| Common Name: | Introducer, Catheter |
| Classification Name: | Catheter Introducer |
| Classification Regulation: | 21 CFR 870.1340 |
| Product Code: | DYB |
| Classification Panel: | Cardiovascular |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with a globe-like structure in the background. A stylized human figure is superimposed on the globe. The figure is white, while the globe is gray. The logo appears to represent a global organization or initiative focused on people.
#### e. Predicate devices
The SUPER SHEATH is substantially equivalent to the following legally marketed devices:
| 510(k): | k052557 |
|--------------------|------------------------------------------------------------------|
| Trade name: | Super Sheath Introducer Sheath and Introducer Sheath Sets, 4F-9F |
| Regulation Number: | 21 CFR 870.1310 |
| Product code: | DRE |
| 510(k): | k121504 |
| Trade name: | Super Sheath Introducer Sheath and Introducer Sheath Sets, 3.3F |
| Regulation Number: | 21 CFR 870.1340 |
| Product code: | DYB |
Since the same name as predicate device is used, the new device under application is hereinafter called "the SUPER SHEATH (PROPOSED)" and the predicate devices are called "the SUPER SHEATH (k052557)" and "the 3.3F SUPER SHEATH (k121504)", respectively, for convenience of discussion in this application.
We are claiming substantial equivalence to product codes DRE and DYB.
#### f. Description of the device
Major components of the SUPER SHEATH (PROPOSED) are a sheath and a dilator. In addition, a guidewire is covered under this premarket notification. The guidewire is placed in the gudiewire case including inserter.
The sheath for the SUPER SHEATH (PROPOSED) is named as the SUPER SHEATH (PROPOSED) Introducer Sheath and is available in 4F- 9F diameters and in lengths ranging from 7 to 25 cm. Some types of sheaths have radiopaque markers. The SUPER SHEATH (PROPOSED) are provided sterile and are intended for one procedure use only.
The one-piece construction of the sheath shaft and hub allows smooth passage of concomitantly interventional medical devices. The sheath shaft and hub are made of Ethylene Tetrafluoro ethylene (ETFE) and polyamide.
The hubs, color-coded by French size, contain a hemostatic valve to minimize blood leakage during a procedure. A side tube equipped with a three-way stopcock is attached to the sheath hub. The side tube extension may be used for fluid and medication administration, as well as blood sampling. Some types of sheaths contain a marker band that is embedded in the shaft approximately 2.0 mm from the distal tip.
The dilator is an open tapered plastic tube with an integral luer hub for guidewire insertion. The dilator is inserted into the introducer sheath. The dilator is longer than the sheath with a rounded tapered distal tip. The dilator facilitates and supports the entry of the sheath into the patient's vasculature. Once the sheath is in place the dilator is removed. There are two different dilator distal tip sizes, 0.038" and 0.035" guidewire compatible. The 4F dilators are only available with a 0.035" guidewire lumen dilator. The 5-9F dilators are available with a
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with a globe-like structure made of intersecting lines, resembling a wireframe. Inside the globe, there is a stylized human figure with its arms outstretched, as if embracing the world. The figure is positioned in the center of the globe, creating a sense of global unity or connection.
0.035" or 0.038" guidewire lumen. The 4bes are made from Polypropylene (PP). The 9F dilator tubes are made of Fluorinated Ethylene Propylene (FEP). The dilator tubes are attached to the inner hub with a metal bush. The sheath hub and the dilator hub lock using a rotating motion.
The SUPER SHEATH (PROPOSED) has a polypropylene suture wing which is color coded by French size. The suture wing is a small projection near the hub with a hole in it for correct placement of the sheath using sutures.
The uncoated guidewire is made of a stainless steel coil wrapped tightly around an inner mandrel that tapers at the distal tip. The flexible tip is J-shaped available in diameters of 0.035" and 0.038". The guidewire, with inserter, is available in 45 cm and 80 cm lengths. The inserter is used strictly for guiding the guidewire into a cannula or introducer.
## g. Indications for Use
## Indication for Use
The device is indicated for use in the introduction of diagnostic and interventional devices inserted into the peripheral and coronary vasculature.
## h. Statement of substantial equivalence
The main body of the SUPER SHEATH (PROPOSED) was essentially modified from the SUPER SHEATH (k052557). Thus, most of the characteristics of the SUPER SHEATH (PROPOSED) are similar to those of the SUPER SHEATH (k052557). Regarding packaging, the packaging materials and configuration are the same as the 3.3F SUPER SHEATH (k121504).
Comparison table of technological characteristics is as follows;
| Feature | SUPER<br>(PROPOSED) | SHEATH | SUPER<br>(k052557) | SHEATH | 3.3F SUPER SHEATH<br>(k121504) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The device is indicated for use in the introduction of diagnostic and interventional devices inserted into the peripheral and coronary vasculature. | | Introducer Sheaths and Introducer Sheath Sets are intended for use in the introduction of diagnostic and interventional devices into the human vasculature. | | The Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Sets are indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature of adult and pediatric patients of all ages. |
| Regulation | 870.1340 | | 870.1310 | | 870.1340 |
| | | | Comparison table of technological characteristics |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Number | | | |
| FDA Product Code | DYB | DRE | DYB |
| Prescription/OTC Use | Prescription | Prescription | Prescription |
| Single-use/Reusab le | Single-use | Single-use | Single-use |
| Sheath | | | |
| French Sizes Available | 4F- 9F | 4F- 9F | 3.3F |
| Effective Length | 7 cm - 25 cm | 7 cm - 25 cm | 5 cm - 7 cm |
| Sheath Shaft Inner Diameter (ID) | 1.58 mm - 3.23 mm | 1.58 mm - 3.23 mm | 1.30 mm |
| Sheath Shaft Outer Diameter (OD) | 1.98 mm - 3.69 mm | 1.98 mm - 3.69 mm | 1.70 mm |
| Sheath Shaft Materials | Ethylene tetrafluoro ethylene (Neoflon® ETFE EP-526):90%<br>Barium Sulfate: 10% | Ethylene tetrafluoro ethylene (Neoflon® ETFE EP-522):85%<br>Barium Sulfate: 15% | Ethylene tetrafluoro ethylene (Neoflon® ETFE EP-522):85%<br>Barium Sulfate: 15% |
| Cover Tube Material | Ethylene tetrafluoro ethylene (Neoflon® ETFE EP-526) | Ethylene tetrafluoro ethylene (Neoflon® ETFE EP-522) | Not Applicable |
| Radiopaque Marker | Tantalum | Tantalum | Not Applicable |
| Dilator | | | |
| Dilator Tube Materials | 4F -8F:<br>Polypropylene<br>9F: Fluorinated ethylene propylene | 4F -8F:<br>Polypropylene<br>9F: Fluorinated ethylene propylene | Polypropylene |
| Dilator Tube Inner Diameter (ID) | 4F-8F 0.92 mm – 1.00 mm (0.035" *)<br>1.02 mm– 1.10 mm (0.038" *)<br>9F 0.95 mm – 1.05 mm (0.035" *)<br>1.05 mm– 1.15 mm (0.038" *) | 4F-8F 0.92 mm – 1.00 mm (0.035" *)<br>1.02 mm– 1.10 mm (0.038" *)<br>9F 0.95 mm – 1.05 mm (0.035" *)<br>1.05 mm– 1.15 mm (0.038" *) | 0.65 mm – 0.75 mm |
| | | | |
| Dilator Tube<br>Outer Diameter<br>(OD) | 1.38 mm – 3.09 mm | 1.38 mm – 3.09 mm | 1.23 mm |
| Dilator Tube<br>Effective Length | 13 cm – 31 cm<br>(130 mm – 310 mm) | 13 cm – 31 cm<br>(130 mm – 310 mm) | 10.5 cm to 12.5 cm<br>(105 mm – 125 mm) |
| Dilator Distal Tip | 0.035”*<br>0.038” * | 0.035” *<br>0.038” * | 0.018”<br>0.021”<br>0.025” |
| Guidewire and Inserter | | | |
| Guidewire<br>Tip<br>Shape | J-tip | J-tip | Straight type |
| Outer Diameter | 0.035”, 0.038” | 0.035”, 0.038” | 0.025” |
| Inserter | Available | Available | Available |
| Sterilization and Shelf-life | | | |
| Sterilization<br>method | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Sterile Package | Pouch | Blister package | Pouch |
| Shelf-Life | Two (2) years | Three (3) years | Three (3) years |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a globe and a stylized figure in the center. The figure appears to be reaching upwards, and the globe is depicted with lines of latitude and longitude. The logo is rendered in a grayscale color scheme, giving it a simple and professional appearance.
#### OGO MEDIKIT CO., LT 7148-6. Aza Kamekawa. Oaza Hichiya. Hyuga City
Miyazaki Prefecture 883-0062, Japan TELEPHONE : +81-982-53-8027 FACSIMILE : +81-982-53-8008
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with a stylized human figure standing in front of a globe. The globe is represented by a grid of lines, suggesting a network or global reach. The human figure is simplified, with arms raised, creating a sense of connection or outreach. The logo appears to be designed to convey a message of global unity, human connection, or international cooperation.
TOGO MEDIKIT CO., LTD.
7148-6, Aza Kamekawa, Oaza Hichiya, Hyuga Ci
Miyazaki Prefecture 883-0062, Japan
TELEPHONE : +81-982-53-8027
Note*: There are two different dilator distal tip sizes, 0.038" and 0.035" guidewire compatible. Note that the 4F dilators are only available with a 0.035"guidewire lumen dilator. The 5-9F dilators are available with a 0.035" or 0.038" guidewire lumen.
SUPER SHEATH (k052557) referenced regulation number 870.1310 and product code DRE. However 3.3F SUPER SHEATH (k121504) for additional sizes utilized regulation number 870.1340 and product code DYB, which are referenced by the current submission. Citation of 870.1340 /DYB is presumably the most current regulatory practice as evidenced by 3.3F SUPER SHEATH (k121504).
Regarding the main body of the SUPER SHEATH (PROPOSED), major components are a sheath and a dilator. In addition, a guidewire is covered under this premarket notification. The guidewire is placed in the Gudiewire case including inserter. The only difference from the SUPER SHEATH (k052557) is the modifications of material change of the sheath shaft.
In order to evaluate any effects of the above changes on safety and effectiveness for the main body, biocompatibility testing and bench testing including shelf-life testing were performed.-Biocompatibility testing demonstrated that the SUPER SHEATH (PROPOSED) does
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with a globe-like structure. The globe is made of vertical and horizontal lines, creating a grid-like pattern. In the center of the globe, there is a stylized human figure with its arms raised. The figure appears to be integrated into the globe's structure, suggesting a connection between humanity and the world.
TOGO MEDIKIT CO., LTD.
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008
not raise any biocompatibility concern. testing including shelf-life testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (k052557).
Regarding packaging of the SUPER SHEATH (PROPOSED), packaging integrity test and sterility test demonstrates the packaging works as a sterile barrier for the SUPER SHEATH (PROPOSED) and is substantial equivalent to the packaging of the 3.3F SUPER SHEATH (k121504).
Based on above, we concluded that the modification did not raise any new safety or effectiveness concerns compared to the SUPER SHEATH (k052557) for main body and the 3.3F SUPER SHEATH (k121504) for packaging.
#### i. Bench Testing
The following bench tests were performed to ensure the safety and effectiveness of the SUPER SHEATH (PROPOSED), verify conformity to the international standards and in-house requirements, and demonstrate substantial equivalence to the SUPER SHEATH (k052557).
- Sheath Shaft Tensile Test
- Sheath Kink Test
- Sheath Hub to Shaft Tensile Test
- Valve Integrity of Sheath
- Sheath Lubricity Test
- Sheath Radiopacity Test
- Cover Tube Durability Test
- Hemostatic Valve Integrity/Sheath Pressure Test
- Sheath/Dilator Corrosion Resistance Test
- Dilator Shaft Tensile Test
- . . . . . . . . . . . . . Dilator Hub to Shaft Tensile Test
- Guidewire Tensile Test
- Guidewire Combined Load Test
- Guidewire Torqueability Test
- Guidewire Radiopacity Test
- Guidewire Corrosion Resistance Test
#### Biocompatibility Testing i.
We performed following biocompatibility tests because of material change of the sheath shaft;
- Cytotoxicity
- Intracutaneous reactivity
- Sensitization
- .... Acute Systemic Toxicity
- Hemocompatibility
- Pyrogen test
- LAL test
SUPER SHEATH APRIL 18, 2014
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with a globe and a stylized human figure inside it. The globe is represented by a grid of lines, and the human figure is white. The figure is standing upright with its arms outstretched, forming a T-shape. The logo is simple and conveys a sense of global reach and human connection. The logo is in grayscale.
TOGO MEDIKIT CO., LTD.
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008
In the biocompatibility testing reports, no biocompatibility concern was raised.
# k. Conclusion
Based on the above discussion and enclosed sections regarding substantial equivalence to predicate devices, TOGO MEDIKIT concludes that the SUPER SHEATH (PROPOSED) is substantially equivalent to the SUPER SHEATH (k052557) and the 3.3F SUPER SHEATH (k121504) , and the SUPER SHEATH (PROPOSED) does not raise any new questions regarding safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.