The RemotEye Viewer software product is intended to be used as a fully functional, web-based medical image viewer to download, review, interpret, manipulate, visualize and print medical multi-modality image data in DICOM format, also stored in remote locations with respect to the viewing site. When interpreted by a trained physician, the medical images displayed by RemotEye Viewer can be used as an element for diagnosis. Typical users of RemotEye Viewer are trained professionals, including but not limited to radiologists, physicians, nurses and technicians.
Device Story
RemotEye Viewer is a web-based software application for viewing, manipulating, and interpreting DICOM medical images; inputs include multi-modality DICOM image data retrieved from remote network locations; software enables image visualization, window/level adjustment, zooming, panning, rotation, flipping, cine-playback, and multi-planar reconstruction (MPR); output is displayed on a client machine for interpretation by trained professionals (radiologists, physicians, nurses, technicians); device facilitates clinical diagnosis by providing diagnostic-level image review; software operates on Windows, Mac OS X, and Linux platforms; no DICOM files are permanently stored on client PCs; integrates with external DICOM archives.
Clinical Evidence
Bench testing only. Verification and validation were performed according to internal software development processes and documented specifications to ensure functionality and performance equivalence to the predicate.
Technological Characteristics
Web-based software architecture; multi-platform (Windows, Mac OS X, Linux); DICOM-compliant; supports multi-modality image display, MPR, and standard image manipulation tools (zoom, pan, window/level, measurements, annotations).
Indications for Use
Indicated for trained professionals (radiologists, physicians, nurses, technicians) to view, interpret, and manipulate DICOM-formatted medical multi-modality images for diagnostic purposes.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2014
NEOLOGICA S.R.L. MARCO SAMBIN STRADA VILLE, 58 17014 CAIRO MONTENOTTE (SV) ITALY
Re: K141061 Trade/Device Name: RemotEve Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 16, 2014 Received: July 21, 2014
Dear Mr. Sambin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh.f)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141061
Device Name RemotEye Viewer
Indications for Use (Describe)
The RemotEye Viewer software product is intended to be used as a fully functional, web-based medical image viewer to download, review, interpret, manipulate, visualize and print medical multi-modality image data in DICOM format, also stored in remote locations with respect to the viewing site. When interpreted by a trained physician, the medical images displayed by RemotEye Viewer can be used as an element for diagnosis.
Typical users of RemotEye Viewer are trained professionals, including but not limited to radiologists, physicians, nurses and technicians.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sm.h.f)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the text "005 Al 510(k) Summary of Safety and Effectiveness". The text is in bold font and is centered on the page. The text is likely the title of a document or report. The number 005 is likely a document number or identifier.
Date Prepared: April 15, 2014 Date of Additional Information: July 16, 2014
Submitter's Name: NeoLogica s.r.l. Submitter's Address: Strada Ville, 58 - 17014 - Cairo Montenotte (SV) - Italy Submitter's Phone Number: +39 019 505314 Submitter's E-mail: info@neologica.it Submitter's Website: www.neologica.it
Establishment Registration Name: NEOLOGICA S.R.L. Establishment Registration Number: 3005202320
Contact person's Name: Marco Sambin Contact person's Title: Director Contact person's Phone Number: +39 019 505314 Contact person's E-mail: marco.sambin@neologica.it
Device Common Name: DICOM-compliant medical image viewer Device Trade Name: RemotEye Viewer Device Classification Name: System, Image Processing, Radiological Device Regulation Description: Picture archiving and communications system Device Regulation Medical Specialty: Radiology Device Review Panel: Radiology Product Code: LLZ Device Regulation Number: 892.2050 Device Class: II Device 510(k) Number: K141061
Predicate Device: eFilm Workstation Predicate Device Original Applicant: eFilm Medical, Inc. - 500 University Ave., Suite 300, Toronto, Ontario Canada M5G 1V7 Predicate Device 510(k) Number: K012211 Predicate Device Classification Name: System, Image Processing, Radiological Predicate Device Regulation Description: Picture archiving and communications system Predicate Device Classification Advisory Committee: Radiology
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the word "neologica" in a stylized font. Below the word, there are two smaller words, "Research" and "Development", separated by a curved line. The logo has a modern and clean design, with a focus on the company's name and its areas of expertise.
Predicate Device Review Advisory Committee: Radiology Predicate Device Product Code: LLZ Predicate Device Requlation Number: 892.2050 Predicate Device Class: II Date Received: 07/16/2001 Decision Date: 07/31/2001 Decision: Substantially Equivalent (SE)
## Device Description
The RemotEye Viewer software product is a feature-rich, diagnostic-level, web-based DICOM medical image viewer that allows downloading, reviewing, interpreting, manipulating, visualizing and printing medical multi-modality image data in DICOM format, from a client machine. The DICOM images may be physically remote with respect to the viewing client, but reachable through a network.
Typical users of RemotEye Viewer are trained professionals, including but not limited to radiologists, physicians, nurses and technicians.
## Device Intended Use
The RemotEye Viewer software product is intended to be used as a fully functional, webbased medical image viewer to download, review, interpret, manipulate, visualize and print medical multi-modality image data in DICOM format. also stored in remote locations with respect to the viewing site. When interpreted by a trained physician, the medical images displayed by RemotEye Viewer can be used as an element for diagnosis.
Typical users of RemotEye Viewer are trained professionals, including but not limited to radiologists, physicians, nurses and technicians.
## Technological Characteristics
Both the RemotEye Viewer and the eFilm Workstation are stand-alone medical imaging software packages which can be used on more than one hardware platform. As long as minimum requirements are met, the user is free to choose his/her own hardware platform. Hardware and software requirements for RemotEye Viewer are described both in the Installation Manual and in the User Manual. Please refer to pages 008 Al - 4, 008 Al - 5, 008_Al - 15, 008_Al - 16 of the 008_Al RemotEye Viewer Installation Manual and pages 007_Al - 6, 007_Al - 7, 007_Al - 8 of the 007_Al RemotEye Viewer User Manual.
## Technological Characteristics in common:
Studies and patients search, Display of grayscale and color DICOM images, On-screen presentation of images with several different layouts, Image navigation, Synchronized series navigation for CT and MR, Textual overlays of main image info, Image zooming, Image panning, Window/level contrast manipulation, Image rotation, Image flipping,
{5}------------------------------------------------
Image enhancement, Display of references lines, Cine-playback, Distance measurements, Area measurements, Angle measurements, Graphical annotations, Normal print, DICOM print, Image export, DICOM CD/DVD burning, Report creation and viewing, MPR (Multi-Planar Reconstruction).
Different Technological Characteristics:
- RemotEye Viewer is a software product based on a web-enabled architecture. . eFilm Workstation is a traditional Windows native desktop application.
- . RemotEye Viewer is a multi-platform viewer, it can successfully run on Windows, Mac OS X and Linux.
eFilm Workstation is a Windows-only software solution.
- . RemotEye Viewer is a web-based software designed to integrate with an external DICOM archive. No DICOM files are permanently stored on the client PCs by RemotEye Viewer.
eFilm Workstation includes an embedded DICOM server, and stores DICOM files locally on the host PC.
These differences don't constitute any new intended use and don't raise new questions of safety and effectiveness.
# Testing
RemotEye Viewer is tested according the specifications that are documented in 013-139 and 009_Al Verification and Validation document.
Testing is an integral part of NeoLogica's software development process as described in 013-126.
## Conclusions
eFilm Workstation and RemotEye Viewer have similar intended use, typical users, functionality and technological characteristics.
RemotEye Viewer provides some different technological characteristics compared to eFilm Workstation; however these differences neither constitute any new intended use, nor raise any questions that affect safety and effectiveness.
Hence RemotEye Viewer is substantially equivalent to eFilm Workstation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.