K141034 · Church & Dwight Co., Inc. · NUC · Jul 21, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141034
Device Name
AZUL PERSONAL LUBRICANT
Applicant
Church & Dwight Co., Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Jul 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The intended use is as a personal lubricant compatible with natural rubber latex and polyisoprene condoms. Not compatible with polyurethane condoms.
Device Story
Azul Personal Lubricant is a hydrous, water-based personal lubricant containing aloe and vitamin E. Formula thickened with sodium hyaluronate and hydroxyethylcellulose; pH-buffered (5.9–6.9); preserved with methylparaben. Packaged in PET bottle with PP flip-top closure and induction seal. Intended for OTC consumer use to enhance comfort during intimate sexual activity. Compatible with natural rubber latex and polyisoprene condoms; incompatible with polyurethane condoms. Device provides lubrication and moisture supplementation.
Clinical Evidence
Bench testing only. Biocompatibility: in vitro cytotoxicity, rabbit vaginal/penile irritation, acute systemic toxicity, guinea pig sensitization, primary rabbit skin irritation. Stability: 2-year shelf life validated via accelerated aging. Condom compatibility: tested per modified ASTM D7661; confirmed compatible with natural rubber latex and polyisoprene, incompatible with polyurethane.
Technological Characteristics
Water-based lubricant; ingredients include aloe, vitamin E, sodium hyaluronate, hydroxyethylcellulose, methylparaben. pH 5.9-6.9. Packaged in PET bottle with PP closure. Compatibility tested per modified ASTM D7661. Non-sterile.
Indications for Use
Indicated for penile and/or vaginal application to lubricate and moisturize, enhance ease and comfort of intimate sexual activity, and supplement natural lubrication. Compatible with natural rubber latex and polyisoprene condoms; not compatible with polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
# JUL 2-1-2014
Pg. I of 3 K 141034 510(k) Premarket Notification Church & Dwight Co., Inc. Azul Personal Lubricant
Confidential July 21, 2014 - Revised Page 2 of 26
## II. 510(k) Summary Azul Personal Lubricant
| Submitter Name: | Church & Dwight Co., Inc. |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 500 Charles Ewing Boulevard<br>Ewing, NJ 08628 |
| Contact Person: | Joseph Ciccone<br>Senior Manager, Regulatory Affairs<br>Church & Dwight Co., Inc.<br>500 Charles Ewing Boulevard<br>Ewing, NJ 08628<br>Tel: (609) 806.1994<br>Fax: (609) 403.7411 |
| Date Prepared: | July 21, 2014 |
| Device Trade Name: | Azul Personal Lubricant |
| Device Common Name: | Personal Lubricant |
| Product Code: | NUC - Condom (21 CFR § 884.5300) |
| Classification: | Class II |
| Predicate Devices: | Lifestyles Natural (K122054)<br>Aloe Cadabra Lubricant (K124044) |
| Device Description: | The Azul Personal Lubricant is a hydrous, clear water-based personal<br>lubricant with aloe and vitamin E that is compatible with natural rubber<br>latex and polyisoprene condoms. This product is not compatible with<br>polyurethane condoms. The lubricant formula is thickened using sodium<br>hyaluronate and hydroxyethylcellulose. It contains a pH buffer system to<br>maintain a slightly acidic pH of 5.9 to 6.9, and is preserved using<br>methylparaben. This medical device is designed to meet product<br>specifications which include: |
| | Color/appearance |
| | Odor<br>� |
| | pl-l |
| | Viscosity |
| | Methylparaben assay<br>� |
| | Osmolality |
| | Total aerobic microbial count<br>● |
| | Total yeast and mold count<br>● |
| | Absence of pathogenic organisms |
| | Antimicrobial effectiveness |
| | The Azul Personal Lubricant is packaged in a polyethylene terephthalate<br>(PET) bottle with a screw on, flip top polypropylene (PP) closure. An<br>induction seal will be placed over the bottle for tamper resistance and<br>preservation. |
| Intended Use: | The intended use is as a personal lubricant compatible with natural<br>rubber latex and polyisoprene condoms. Not compatible with<br>polyurethane condoms. |
| Indications for Use: | Azul Personal Lubricant is a personal lubricant for penile and/or vaginal<br>application, intended to lubricate and moisturize, to enhance the ease and<br>comfort of intimate sexual activity and supplement the body's natural<br>lubrication. This product is compatible with natural rubber latex and<br>polyisoprene condoms. This product is not compatible with<br>polyurethane condoms. |
| Technological Characteristics: | Azul Personal Lubricant contains similar ingredients as legally marketed<br>water-based personal lubricants that are compatible with natural rubber<br>latex and polvisoprene condoms. This product is not compatible with<br>polyurethane condoms. |
| Summary of Studies: | |
| | Biocompatibility studies - Performed on the final 510(k)-subject<br>device. These studies include in vitro cytotoxicity study; rabbit<br>vaginal irritation study; rabbit penile irritation study; acute systemic<br>toxicity study; guinea pig maximization sensitization study; primary<br>rabbit skin irritation study. Based on the results of these studies. Azul<br>Personal Lubricant is considered safe for consumer use. |
(continued on next page)
Image /page/0/Picture/6 description: The image shows the logo for Trojan brand condoms. The logo features the word "TROJAN" in a bold, sans-serif font, with the tagline "BRAND CONDOMS" underneath in a smaller font. To the right of the wordmark is an image of a Trojan warrior's head, complete with a helmet. The logo is simple, recognizable, and conveys a sense of strength and protection.
Church & Dwight Co.. Inc.
{1}------------------------------------------------
Pg. 2 of 3 K 141034 510(k) Premarket Notification Church & Dwight Co., Inc. Azul Personal Lubricant
#### Confidential July 21, 2014 - Revised Page 3 of 26
Image /page/1/Picture/3 description: The image shows the logo for Trojan brand condoms. The logo features the word "TROJAN" in bold, sans-serif font, with the tagline "BRAND CONDOMS" in smaller font below. To the right of the wordmark is an image of a Trojan helmet.
CHURCH & DWIGHT CO., INC.
{2}------------------------------------------------
Pg. 3 of 3 K 141034 510(k) Premarket Notification Church & Dwight Co., Inc. Azul Personal Lubricant
Confidential July 21, 2014 - Revised Page 4 of 26
Shelf life - Stability and shelf life testing was performed on the Azul Personal Lubricant to validate the stability of the device during its intended shelf life of 2 years. Accelerated aging was performed in accordance with accepted standards for medical devices to simulate longterm storage. To evaluate stability, this in vitro study used sealed and unsealed sample bottles to simulate actual device usage.
Condom compatibility - The results for laboratory testing using a modification of the methodology found in ASTM D7661; Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. Laboratory test results demonstrated that the proposed device is compatible with leading commercial brands of natural rubber latex and polyisoprene condoms, but not compatible with polyurethanc condoms.
Substantial Equivalence:
The proposed device is substantially equivalent to the predicate devices. Lifestyles Natural and Aloe Cadabra in technology, intended use, safety.
Image /page/2/Picture/6 description: The image shows the word "TROJAN" in bold, sans-serif font. To the right of the word is a stylized image of a Trojan helmet. Below the word "TROJAN" are the words "BRAND CONDOMS" in a smaller font.
### CHURCH & DWIGHT CO., INC.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2014
Church & Dwight Co., Inc. Joseph Ciccone Senior Manager, Regulatory Affairs 500 Charles Ewing Boulevard Ewing, NJ 08628
- K141034 Re:
Trade/Device Name: Azul Personal Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: April 21, 2014 Received: April 22, 2014
Dear Joseph Ciccone,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
Page 2 - Joseph Ciccone
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/4/Picture/8 description: The image shows the name "Benjamin R Fisher -S" in a bold, sans-serif font. The letters "R" and "F" are stylized with a decorative pattern, possibly indicating a logo or design element. The text is horizontally aligned and appears to be part of a title or heading.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K 141034 510(k) Premarket Notification Church & Dwight Co., Inc. Azul Personal Lubricant
Confidential July 2, 2014 Page 1 of 26
#### I. Indications for Use
510(k) Number (if known): N/A
・
Device Name: Azul Personal Lubricant
INDICATIONS FOR USE:
Azul Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (21 CFR 801 Subpart D) OR
Over-The-Counter (OTC)_X_ (21 CFR 801 Subpart C)
Beniamin R.F. 2014 1-5:06:53 -04'00'
CHURCH & DWIGHT CO., INC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.