SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE(HP) SUSCEPTIBILITY PLATES

K140985 · Thermo Fisher Scientific · JWY · Jun 10, 2014 · Microbiology

Device Facts

Record IDK140985
Device NameSENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE(HP) SUSCEPTIBILITY PLATES
ApplicantThermo Fisher Scientific
Product CodeJWY · Microbiology
Decision DateJun 10, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae,Streptococcus pneumoniae and Streptococcus species. This 510(k) is for the addition of Clindamycin in the dilution range of 0.015-1ug/ml to the Sensititre HP MIC susceptibility plate for testing Streptococcus species. The approved primary "Indications for Use" and clinical significance of Clindamycin is for: Faculative Gram-Positive Fastidious Microorganisms: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus

Device Story

Sensititre HP MIC Susceptibility Plate is an in vitro diagnostic tool used in clinical microbiology laboratories. It utilizes microbroth dilution to determine the minimum inhibitory concentration (MIC) of antimicrobial agents against specific fastidious bacteria. The device consists of plates containing pre-dispensed, dried antimicrobial agents (specifically Clindamycin) at various concentrations. Laboratory technicians inoculate the plates with patient bacterial isolates; plates are incubated; bacterial growth is observed to determine the MIC. The resulting MIC values assist clinicians in selecting appropriate antibiotic therapy for patients with infections caused by susceptible organisms. The device provides standardized, quantitative susceptibility results to guide clinical decision-making and improve patient outcomes.

Clinical Evidence

Performance evaluated using 225 clinical and 60 challenge isolates of S. agalactiae, S. pyogenes, and S. anginosus. Testing compared the Sensititre system (manual and AutoRead) against CLSI M07-A9 frozen broth microdilution reference panels. Combined results showed 99.6% Essential Agreement (EA) and 100% Category Agreement (CA) for both reading methods. Reproducibility studies demonstrated 100% performance. Quality control using S. pneumoniae ATCC 49619 met expected ranges.

Technological Characteristics

Microbroth dilution susceptibility plate; contains dried antimicrobial agents (Clindamycin) in specific dilution ranges (0.015-1 ug/ml). Designed for in vitro diagnostic use in clinical laboratories. Standardized plate format for testing fastidious Gram-positive organisms.

Indications for Use

Indicated for in vitro clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus species (specifically S. pyogenes, S. agalactiae, and S. anginosus) to determine minimum inhibitory concentrations (MIC) of Clindamycin.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K140985 B. Purpose for Submission: Addition of *Streptococcus* species to the Intended Use of Clindamycin (0.015 – 1 µg/mL) in the Sensititre *Haemophilus*/*Streptococcus pneumoniae* (HP) MIC susceptibility panel. C. Measurand: Clindamycin concentrations of 0.015 – 1 µg/mL D. Type of Test: Quantitative antimicrobial susceptibility test (AST), growth-based fluorescence E. Applicant: ThermoFisher Scientific F. Proprietary and Established Names: Sensititre *Haemophilus influenza*/*Streptococcus pneumoniae* MIC Susceptibility Plate G. Regulatory Information: 1. Regulation section: 21 CFR 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: JWY – Manual Antimicrobial Test LRG – Instrument for Autoreader and Interpretation of Overnight Susceptibility Systems {1} LTT – Panels, Test, Susceptibility, Antimicrobial LTW – Susceptibility Test Cards, Antimicrobial 4. Panel: 83, Microbiology H. Intended Use: 1. Intended use(s): The Sensititre Haemophilus influenza/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species. 2. Indication(s) for use: The Sensititre Haemophilus influenza/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenza, Streptococcus pneumoniae and Streptococcus species. This 510(k) is for the addition of Clindamycin in the dilution range of 0.015 – 1 µg/mL to the Sensititre HP MIC Susceptibility plate for testing Streptococcus spp. The approved primary “Indications for Use” and clinical significance of Clindamycin is for: Facultative Gram-Positive Fastidious Microorganisms: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus 3. Special conditions for use statement(s): Prescription use only 4. Special instrument requirements: Autoinoculator/AIM only Manual read (Vizion) or AutoRead (ARIS/AutoReader OptiRead) {2} # I. Device Description: The Sensititre 18-24 hours MIC Breakpoint Susceptibility System is a micro-version of the classic broth dilutions method and can provide both qualitative and quantitative susceptibility results in a dried microtitre plate format. Each micro-broth dilution plate is dosed with antimicrobial agents at specific dilutions and then dried. The organism to be tested must be in pure culture and identified as a Streptococcus species. A standardized suspension is prepared from colonies in pure growth and inoculated into the microtitre plate using the Sensititre AutoInoculator/AIM instrument (AutoInoculator). After the indicated hours of incubation, the microtitre plate is examined for growth to determine the MIC either manually (Vizion or manual viewer) or using the ARIS/AutoReader/OptiRead (AutoReader). The FDA recommended interpretive criteria for clindamycin are listed in Table 1. Table 1: FDA interpretive criteria for clindamycin | Organism | Susceptibility Interpretive Criteria (MIC in μg/mL)*† | | | | --- | --- | --- | --- | | | S | I | R | | S. pneumoniae, Streptococcus species (S. anginosus, S. pyogenes, S. agalactiae) | ≤0.25 | 0.5 | ≥1 | * These interpretive standards for S. pneumoniae and other Streptococcus spp. are applicable only to tests performed by broth microdilution using cation-adjusted Mueller Hinton broth with 2 to $5\%$ lysed horse blood inoculated with a direct colony suspension and incubated in ambient air at $35^{\circ}\mathrm{C}$ for 20 to 24 hours. † S, Susceptible; I, Intermediate; R, Resistant # J. Substantial Equivalence Information: 1. Predicate device name(s): MicroScan MICroSTREP plus Panel - Clindamycin 2. Predicate $510(\mathbf{k})$ number(s): K021184 {3} # 3. Comparison with predicate: Table 2: Comparison with the predicate device | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Name | SensititreHaemophilus/Streptococcus pneumoniae (HP) MIC Susceptibility Plates | MicroScan MICroSTREP plus panel, K021184 | | Intended Use | SensititreHaemophilus/Streptococcus pneumoniae (HP) MIC Susceptibility plates is an in vitro diagnostic product for clinical susceptibility testing of Haemophilus influenza, Streptococcus pneumoniae and Streptococcus species. This 510(k) is for the addition of clindamycin in the dilution range of 0.015 – 1 μg/mL to the panel. | To determine antimicrobic susceptibility to Clindamycin with aerobic streptococci other than Streptococcus pneumoniae. | | Inoculum Preparation | Nephelometer | Same | | Specimen | Isolated colonies from cultures | Same | | Test Organism | Streptococcus species other than S. pneumoniae | Same | | Test Medium | Cation adjusted Mueller-Hinton broth with TES buffer and lysed horse blood | Same | | Antibiotic | Clindamycin | Same | | Incubation Time | 20-24 hours | Same | | Incubation Temperature | 34 – 36 °C | Same | | Incubation Atmosphere | Aerobic, non-CO2 | Same | | Reading Method | Manual or Automated | Same | | Results | Results reported as minimum inhibitory concentration (MIC) and categorical interpretation (SIR). | Same | {4} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Technology | Fluorescence-based | Growth-based | | Panel Inoculation | Automated inoculation using the AIM | RENOK Rehydrator/Inoculator | | Instrument | Autoread or manual read | Manual read or MicroScan WalkAway System | ## K. Standard/Guidance Document Referenced (if applicable): M07-A9 Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – Ninth Edition, 2012 M100-S24 Performance Standards for Antimicrobial Susceptibility Testing; Twenty-Fourth Informational Supplement, 2014 Guidance for Industry and FDA: Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009. ## L. Test Principle: Each plate (panel) is dosed with antimicrobial agents at appropriate dilutions. A standardized suspension is prepared from a pure growth of colonies and inoculated into the microtitre plate using the Sensititre AutoInoculator/AIM instrument. After the indicated hours of incubation, the microtitre plate is examined for growth to determine the MIC. Results can be read manually by visual reading of growth or automatically on an AutoRead using fluorescence. The Sensititre AutoRead system utilizes fluorescence technology. The technology involves the detection of bacterial growth by monitoring the activity of specific surface enzymes produced by the test organism. Growth is determined by generating a fluorescent product from a non-fluorescent (fluorogenic) substrate. The non-fluorogenic substrate is prepared by conjugating a fluorescent compound to the specific enzyme substrates with a bond which prevents fluorescence. The fluorophore is then said to be quenched. The plates are prepared with the substrate already added to the plate. Enzymatic action of the bacterial surface enzymes on the specific substrates cleaves this bond releasing the fluorophore, which is now capable of fluorescence. The amount of fluorescence detected is directly related to the activity of the bacterial surface enzymes and therefore, to the bacterial growth. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Reproducibility was conducted at three sites using 17 *Streptococcus* isolates which included *S. anginosus*, *S. pyogenes* and *S. agalactiae*. Two additional isolates were {5} tested that included a species not indicated in the FDA drug label (S. dysgalactiae) and six isolates that had off-scale MICs. The sponsor was asked to remove these isolates from the reproducibility report. Because the reproducibility results were acceptable for all isolates, additional testing was not requested. The panels were inoculated using the Autoinoculator only. Panels were read both manually using the Vizion and using the AutoRead. The mode MIC value was determined and the reproducibility was calculated based on MICs falling within $\pm 1$ dilution of the mode MIC value. The reproducibility studies for both the manual read method and the AutoRead method demonstrated best case acceptable performance of $100\%$ . # b. Linearity/assay reportable range: Not applicable. # c. Traceability, Stability, Expected values (controls, calibrators, or methods): The recommended quality control isolate (S. pneumoniae ATCC 49619) was tested 20 times at each site using the Sensititre panels with inocula prepared using the AutoInoculator. Readings were performed using both the manual read method and the AutoRead method. The QC isolate was tested using the reference method 20 times at each site using manual inoculation and manual read, with all results falling within the acceptable range. The clindamycin test results demonstrate that the system can produce QC results within the expected range. The FDA recommended quality control organism and the acceptable QC range is noted in Table 3. QC results obtained with the Sensititre panel and with the reference method are listed in Table 4. Table 3: FDA recommended QC organism and expected range | QC organism | Expected range in μg/mL | | --- | --- | | S. pneumoniae ATCC 49619 | 0.03 – 0.12 | Table 4: QC results obtained with the Sensititre panel using manual and AutoRead methods | QC organism | Expected MIC range (μg/mL) | Concentration (μg/mL) | | Read method | | | --- | --- | --- | --- | --- | --- | | S. pneumoniae ATCC 49619 | 0.03 – 0.12 | | Reference | Manual (Vizion) | AutoRead | | | | ≤ 0.015 | 0 | 0 | 0 | | | | 0.03 | 26 | 8 | 19 | | | | 0.06 | 34 | 51 | 40 | | | | 0.12 | 0 | 1 | 1 | | | | 0.25 | 0 | 0 | 0 | | | | 0.5 | 0 | 0 | 0 | | | | 1 | 0 | 0 | 0 | {6} Growth Failure Rate: All isolates tested during the clinical studies grew in both the reference panel and in the dried Sensititre panels. Purity Check Plates were performed to detect contamination during the clinical testing at the clinical sites. Inoculum Density Check: All organism suspensions were standardized spectrophotometrically. Turbidity meter readings were recorded each day of use and the inoculum density of the QC isolate was determined using traditional colony counting techniques. The mean inoculum densities were 4.2 X 10⁵, 6.5 X 10⁵, and 1.8 X 10⁵, for sites 1, 2 and 3, respectively, and were within the recommended ranges. The quality control results are acceptable. d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: Results obtained with the Thermo Scientific Sensititre dried MIC susceptibility panels with clindamycin were compared to results obtained using a frozen broth microdilution panel prepared according to CLSI M07-A9 guidelines. All isolates were tested using seven dilutions of clindamycin. Dilutions tested were appropriate for the interpretive breakpoints established for the drug. Test inocula were standardized using a spectrophotometric method; Sensititre panels were inoculated using the Autoinoculator and read using the both the manual read method (Vizion) and the AutoRead method. Reference panels were inoculated as outlined in the CLSI M07-A9 document and were read manually. The device labeling indicates that the Haemophilus influenzae/Streptococcus pneumoniae MIC panels can be manually inoculated. However, this procedural option was not utilized for testing either the clinical isolates or the challenge isolates. Therefore, the sponsor was asked to include a limitation in the device package insert stating, "The performance of clindamycin with Streptococcus spp. was performed 7 {7} using the AIM autoinoculator. The use of an alternative inoculation system when testing clindamycin has not been evaluated." Seventy-five clinical isolates of Streptococcus were tested at each of three sites for a total of 225 Streptococcus isolates. The species tested included S. agalactiae (90 isolates), S. pyogenes (90 isolates) and S. anginosus (45 isolates). All isolates were fresh clinical isolates; no stock isolates were tested during the clinical studies. A total of 60 challenge isolates were tested at one site. Isolates included S. pyogenes (25 isolates), S. agalactiae (25 isolates) and S. anginosus (10 isolates). Organism selection for the challenge isolates was based on the intended use of clindamycin for this submission. Expected results for the challenge isolates were determined using a frozen reference panel prepared using CLSI M07-A9 guidelines. Challenge isolates were tested using inocula prepared using the AutoInoculator and were read both manually and with AutoRead. Combined results from clinical and challenge studies demonstrated an overall EA of 99.6% and an overall CA of 100% for both manual read and AutoRead. The performance evaluation summary of essential and categorical agreement results for challenge and clinical isolates (all results combined and results by species) are shown in the tables 5 - 8 below. EA = Essential Agreement R = Resistant Isolates maj = major discrepancies CA = Category Agreement min = minor discrepancies vmj = very major discrepancies Table 5: Overall performance of clinical and challenge isolates, AutoRead method | | Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 225 | 224 | 99.6 | 176 | 175 | 99.4 | 225 | 100 | 42 | 0 | 0 | 0 | | Challenge | 60 | 60 | 100 | 54 | 54 | 100 | 60 | 100 | 6 | 0 | 0 | 0 | | Combined | 285 | 284 | 99.6 | 230 | 229 | 99.6 | 285 | 100 | 48 | 0 | 0 | 0 | {8} Table 6: Performance of clinical and challenge isolates by species, AutoRead method | | EA TOT | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | No. R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | S. agalactiae | | | | | | | | | | | | | | Clinical | 90 | 90 | 100 | 60 | 60 | 100 | 90 | 100 | 32 | 0 | 0 | 0 | | Challenge | 25 | 25 | 100 | 22 | 22 | 100 | 25 | 100 | 3 | 0 | 0 | 0 | | Combined | 115 | 115 | 100 | 82 | 82 | 100 | 115 | 100 | 35 | 0 | 0 | 0 | | S. pyogenes | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 90 | 90 | 100 | 87 | 87 | 100 | 90 | 100 | 2 | 0 | 0 | 0 | | Challenge | 25 | 25 | 100 | 25 | 25 | 100 | 25 | 100 | 0 | 0 | 0 | 0 | | Combined | 115 | 115 | 100 | 112 | 112 | 100 | 115 | 100 | 2 | 0 | 0 | 0 | | S. anginosus | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 45 | 44 | 97.8 | 29 | 28 | 96.6 | 45 | 100 | 8 | 0 | 0 | 0 | | Challenge | 10 | 10 | 100 | 7 | 7 | 100 | 10 | 100 | 3 | 0 | 0 | 0 | | Combined | 55 | 54 | 98.2 | 36 | 35 | 97.2 | 55 | 100 | 11 | 0 | 0 | 0 | Table 7: Overall performance of clinical and challenge isolates, manual read method | | Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 225 | 224 | 99.6 | 189 | 189 | 100 | 225 | 100 | 42 | 0 | 0 | 0 | | Challenge | 60 | 60 | 100 | 54 | 54 | 100 | 60 | 100 | 6 | 0 | 0 | 0 | | Combined | 285 | 284 | 99.6 | 243 | 243 | 100 | 285 | 100 | 48 | 0 | 0 | 0 | Table 8: Performance of clinical and challenge isolates by species, manual read method | | EA TOT | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | No. R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | S. agalactiae | | | | | | | | | | | | | | Clinical | 90 | 90 | 100 | 70 | 70 | 100 | 90 | 100 | 32 | 0 | 0 | 0 | | Challenge | 25 | 25 | 100 | 22 | 22 | 100 | 25 | 100 | 3 | 0 | 0 | 0 | | Combined | 115 | 115 | 100 | 92 | 92 | 100 | 115 | 100 | 35 | 0 | 0 | 0 | | | | | | | | | | | | | | | | S. pyogenes | | | | | | | | | | | | | | Clinical | 90 | 89 | 98.9 | 87 | 87 | 100 | 90 | 100 | 2 | 0 | 0 | 0 | | Challenge | 25 | 25 | 100 | 25 | 25 | 100 | 25 | 100 | 0 | 0 | 0 | 0 | | Combined | 115 | 114 | 99.1 | 112 | 112 | 100 | 115 | 100 | 2 | 0 | 0 | 0 | | | | | | | | | | | | | | | | S. anginosus | | | | | | | | | | | | | | Clinical | 45 | 45 | 100 | 32 | 32 | 100 | 45 | 100 | 8 | 0 | 0 | 0 | | Challenge | 10 | 10 | 100 | 7 | 7 | 100 | 10 | 100 | 3 | 0 | 0 | 0 | | Combined | 55 | 55 | 100 | 39 | 39 | 100 | 55 | 100 | 11 | 0 | 0 | 0 | {9} b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Table 9: FDA interpretive criteria for clindamycin | Organism | Susceptibility Interpretive Criteria (MIC in μg/mL)*† | | | | --- | --- | --- | --- | | | S | I | R | | S. pneumoniae, Streptococcus species (S. anginosus, S. pyogenes, S. agalactiae) | ≤0.25 | 0.5 | ≥1 | * These interpretive standards for S. pneumoniae and other Streptococcus spp. are applicable only to tests performed by broth microdilution using cation-adjusted Mueller Hinton broth with 2 to $5\%$ lysed horse blood inoculated with a direct colony suspension and incubated in ambient air at $35^{\circ}\mathrm{C}$ for 20 to 24 hours. † S, Susceptible; I, Intermediate; R, Resistant # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...