1.8 MM PHOENIX ATHERECTOMY CATHETER, 2.2 MM PHOENIX ATHERECTOMY CATHETER, PHOENIX HANDLE, PHOENIX WIRE SUPPORT CLIP
K140944 · Atheromed, Inc. · MCW · Aug 22, 2014 · Cardiovascular
Device Facts
Record ID
K140944
Device Name
1.8 MM PHOENIX ATHERECTOMY CATHETER, 2.2 MM PHOENIX ATHERECTOMY CATHETER, PHOENIX HANDLE, PHOENIX WIRE SUPPORT CLIP
Applicant
Atheromed, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Aug 22, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Phoenix® Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Device Story
Peripheral atherectomy system; consists of sterile, single-use Phoenix Catheter and battery-powered Phoenix Handle. Catheter inserted over 0.014" guidewire; handle motor rotates distal cutter to debulk plaque. Excised debris continuously captured and cleared proximally through catheter into external collection reservoir. Used in peripheral vasculature; operated by physician. Models include non-deflecting (1.8mm, 2.2mm) and deflectable (2.4mm) versions. Deflectable model allows eccentric debulking. Device provides mechanical plaque removal to improve vessel patency.
Clinical Evidence
Bench testing only. Evaluations included dimensional/visual inspection, simulated use, comparative predicate testing, torque-to-failure, drive train stress, cutter stall, temperature rise, corrosion, kink bend radius, guidewire/sheath compatibility, and extensive biocompatibility testing (cytotoxicity, sensitization, reactivity, systemic toxicity, hemolysis, coagulation assays, complement activation, and pyrogenicity).
Technological Characteristics
Materials: Multi-strand stainless steel torque shaft, stainless steel outer shaft, Teflon sheath. Energy: Battery-powered motor (10,000-12,000 RPM). Dimensions: 1.8mm, 2.2mm, 2.4mm tip diameters; 130cm working length. Connectivity: None. Sterilization: Ethylene Oxide. Single-use. Mechanical debulking via rotating cutter; continuous debris conveyance.
Indications for Use
Indicated for atherectomy of peripheral vasculature. Not for use in coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a central emblem. The emblem consists of three stylized human profiles facing right, with flowing lines extending from the bottom of the profiles.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2014
AtheroMed, Inc. Ms. Jean Chang Vice President, Operations 1455 Adams Drive Suite 1120 Menlo Park, CA 94025
K140944 Re:
Trade/Device Name: Phoenix® Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: July 24, 2014 Received: July 25, 2014
Dear Ms. Chang.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Fernando Aguel Fernando Aguel -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140944
Device Name
Phoenix® Atherectomy System
Indications for Use (Describe)
The Phoenix® Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Type of Use (Select one or both, as applicable)
ال Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 9 510(k) Summary
# Submitter Information:
| Date of 510(k) Summary Preparation: | July 24, 2014 |
|-------------------------------------|------------------------------------------------------------------------------------------|
| Name and Address of Manufacturer: | AtheroMed, Inc.<br>1455 Adams Dr.<br>Menlo Park, CA 94025 |
| Contact Person: | Jean Chang<br>Vice President, Operations<br>Phone: (650) 473-6846<br>Fax: (650) 473-9927 |
| Subject Device: | |
| Device Trade Name: | Phoenix® Atherectomy System |
| Common Name: | Peripheral Atherectomy Catheter |
| Regulation Description: | Intraluminal Artery Stripper |
| Regulation Number: | 21 CFR 870.4875 |
| Product Code: | MCW |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
# Predicate Device:
| Trade Name: | Phoenix Atherectomy System |
|----------------|----------------------------|
| 510(k) Number: | K132682 |
| Manufacturer: | AtheroMed, Inc. |
## Device Description:
The AtheroMed Phoenix Atherectomy System is a sterile, single-use device designed for atherectomy of the peripheral vasculature. The Phoenix Atherectomy System has two main components: the Phoenix Catheter and the Phoenix Handle.
{4}------------------------------------------------
The Phoenix Catheter is a flexible, over-the-wire (OTW), front-cutting Catheter that continuously captures and clears debulked plaque proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectomy Catheter at its specified rotational speed. The device is activated by an ON/OFF slider switch on the top of the Handle. An optional Wire Support Clip can also be used to clip a guidewire torque device in a fixed position relative to the Handle. The Catheter, Handle, and Wire Support Clip are each packaged separately as sterile, single-use components of the Phoenix Atherectomy System.
There are three different models of the Phoenix Catheter. Two Phoenix Catheter models track directly over the guidewire with no tip deflection capability. These models are available in 1.8mm and 2.2mm tip diameter sizes. A third, 2.4mm tip diameter model has a Catheter tip design that can be deflected and rotated by the user so that the cutter can eccentrically debulk to a larger diameter than the Catheter's 2.4mm distal cutter. The controls for deflection and rotation are housed in the Phoenix Handle when the Catheter is inserted into the Handle. All three Phoenix Catheter models are compatible with commercially available 0.014" exchange length (260cm or greater) guidewires, and all use the same Phoenix Handle.
This 510(k) includes modifications to the cutter and distal assembly of the 1.8mm and 2.2mm tip diameter Phoenix Catheter models, as well as modifications to the Catheter and proximal chassis design, to further optimize the device design and manufacturability of the device. Table 9-1 summarizes the subject modifications relative to the predicate device. Additionally, the Phoenix Guidewire has been added as a compatible guidewire in the product labeling.
# Indications for Use:
The Phoenix® Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
# Testing Summary:
To demonstrate the substantial equivalence of the modified Phoenix Atherectomy System to the predicate Phoenix Atherectomy System, the performance and technological characteristics were evaluated by completion of the following testing:
- . Dimensional and Visual Inspection
- . Simulated Use
- . Comparative Predicate Testing in Simulated Lesion
- . Cutter Torque Chain Torque-to-Failure Test
- Functional Outer Shaft Torque Test
- . Knob to Shaft Testing
- . Catheter Drive Train Stress Test
- Cutter Stall Test
- Temperature Rise of Catheter During Simulated Use
- . Corrosion Test
- Kink Bend Radius Test
- Guidewire Compatibility
- . Sheath Compatibility
{5}------------------------------------------------
### ATHEROMED K140944 AI Response
- Sheath Flow Rate
- . Catheter Trackability in Below-the-Knee Anatomy
- Shelf Life
- . Biocompatibility
- o Cytotoxicity (L-929 MEM Test)
- Sensitization (Kligman Maximization Test, 2 extracts saline and SO) O
- Intracutaneous Reactivity (Injection Test, 2 extracts saline and SO) O
- Acute Systemic Toxicity (2 extracts saline and SO) o
- Hemolysis (Human Blood indirect contact) O
- Hemolysis (Human Blood direct contact) O
- Prothrombin Time Assay (ISO direct contact) O
- Partial Thromboplastin Time Assay (ISO direct contact) O
- In Vitro Platelet Aggregation Assay (ISO direct contact) O
- o Complement Activation (C3a and SC5a-9 direct contact)
- Pyrogenicity (Material Mediated) O
- Pyrogenicity Bacterial Endotoxin/LAL (Kinetic Turbidimetric Assay Pyrogen о Test)
The results from this testing demonstrate that the performance and technological characteristics of the modified Phoenix Atherectomy System meet defined design requirements and that the modified Phoenix Atherectomy System performs in a manner equivalent to the predicate Phoenix Atherectomy System with the identical intended use.
{6}------------------------------------------------
| Table 9-1: Summary of Technological Characteristics for the Modified Phoenix<br>Atherectomy System | | |
|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristic | Predicate Phoenix Atherectomy<br>System (K132682),<br>5F (FG1105) and 6F (FG1493)<br>Phoenix Catheters | Modified Phoenix Atherectomy<br>System (Subject Device),<br>5F (FG1847) and 6F (FG1984)<br>Phoenix Catheters |
| Rotational Speed | 10,000-12,000 RPM | Identical |
| Guidewire Exchange | Over-the-wire | Identical |
| Guidewire Compatibility | 0.014" | Identical |
| Sheath Compatibility | 5F - 6F | Identical |
| Catheter Working Length | 130 cm | Identical |
| Catheter Torque Shaft | Multi-Strand Stainless Steel (SS) | Multi-Strand Stainless Steel (SS) |
| Catheter Outer Shaft | Stainless Steel Outer Shaft and<br>Teflon sheath | Stainless Steel Outer Shaft and thicker<br>Teflon sheath |
| Catheter Shaft Diameter | 1.7mm | Identical |
| Distal Cutter Flute<br>Maximum Diameter | 1.6mm (FG1105)<br>2.0mm (FG1493) | 1.8mm (FG1847)<br>2.2mm (FG1984)<br>(Modified Distal Cutter Flute is flush<br>with Cutter Housing) |
| Tip Diameter and<br>Crossing Profile | 1.8mm (FG1105)<br>2.2mm (FG1493) | Identical (FG1847)<br>Identical (FG1984) |
| Cutting Tip Port | None | Single exit port added to convey<br>excised debris from the inner guidewire<br>lumen into the Distal Cutting Flute<br>channel |
| Second Stage<br>Maceration within<br>Housing | Yes | Yes, with increased capacity/volume |
| Cutter Housing | Coated | No Coating |
| Distal Tip Assembly<br>Coating | Coated | Coated |
| Catheter Chassis<br>Material Changes | Bearing Lubricant - Aerospace Grade<br>Silicone O-Ring | Bearing Lubricant - Food Grade<br>Silicone O-Ring - Medical Grade |
| Minimum Vessel Size for<br>Device Use | 2.5mm (FG1105)<br>3.0mm (FG1493) | Identical (FG1847)<br>Identical (FG1984) |
| Debris Collection &<br>Removal | Continuous collection and removal of<br>excised debris by mechanical<br>conveyance | Identical |
| Steering (Directional)<br>mechanism | Rotation of knob on handle steers<br>distal tip and cutter by torqueing<br>catheter shaft | Identical |
| Catheter Coating | No | Identical |
| Sterilization | Ethylene Oxide | Identical |
| Single-use only | Yes | Identical |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.