The Osseoflex® SB is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal polymethymethacrylate (PMMA) bone cements for use during percutaneous vertebral augmentation, such as kyphoplasty.
Device Story
Osseoflex SB is an inflatable bone tamp used in balloon kyphoplasty. Device consists of a balloon catheter with an articulating/steering feature controlled by a knob; allows clinician to direct the distal end through a pre-existing access channel in the vertebral body. Operated by a physician under fluoroscopic observation. Input: manual manipulation of the steering knob. Transformation: inflation of the distal balloon with 60% contrast medium to create a cavity and reduce vertebral fractures. Output: a void in cancellous bone for subsequent filling with PMMA bone cement. Benefits: fracture reduction and cavity creation for cement interdigitation in spinal procedures.
Clinical Evidence
No clinical data; bench testing only. Performance testing included balloon profile, working length, compliance, maximum pressure, bond tensile strength, maximum volume, fatigue (20 cycles), and inflation/deflation time. All tests met acceptance criteria.
Technological Characteristics
Inflatable bone tamp; polyurethane balloon; 8G cannula; 60% contrast medium inflation; 400 psi max pressure; 15mm max diameter; 16mm length. Articulating/steering mechanism via manual knob. Sterilization not specified. No software/electronics.
Indications for Use
Indicated for patients requiring reduction of fractures and/or creation of a void in cancellous bone in the spine during percutaneous vertebral augmentation (kyphoplasty).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "CSSEA" in a stylized font. The letters are thick and outlined, with a horizontal line running beneath them. The letters are evenly spaced and appear to be part of a logo or brand name.
JUN 1 9 2014
### Attachment 4:
# 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.
# 510(k) Owner
Osseon LLC 2330 Circadian Way Santa Rosa, CA 95407 Phone: 707-636-5940 Fax: 707-636-5941
### Official .Contact
Keith Burger Director of Research and Development
#### Device Information
Trade or Proprietary Name:
Common Name:
Osseoflex® SB
Common Name:
L. S. Ashworth
Classification Name:
Classification Panel:
Regulation:
Product Code(s)
Legally marketed device(s) to which equivalence is claimed Reason for 510(k)
Device Description
Inflatable Bone Tamp
Primary: Arthroscope Secondary: Cement, Bone Vertebroplasty
Orthopedic
Class II per 21CFR §888.1100 Class II per 21CFR §888.3027
HRX; NDN
Osseoflex SB Inflatable Bone Tamp K 122533
New Device
The Osseoflex® SB is designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which
Osseon LLC Osseoflex SB, 2ml Special 510(k)
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "CSSA01" in a stylized font. The letters are outlined in black and have a three-dimensional appearance. The word is underlined with a black line. There is a registered trademark symbol to the right of the number 1.
the physician can inflate the balloon at the distal end of the catheter. After the bone is disrupted, PMMA is injected through an Osseoflex® needle to fill the previously created void(s).
An access channel is required for Osseoflex® SB · placement. The Osseoflex® SB device does not create an access channel; the Osseoflex® SB is designed to follow a pre-existing channel created by an access channel device. The articulating or steering feature of the device assists the clinician in directing the device to the pre-existing channel. The Osseoflex® SB knob can be turned clockwise to aid in directing the distal portion of the device. Turning the knob counter-clockwise will relax the device and allow the device to be returned to its start position. The device should be manipulated only while under fluoroscopic observation with radiographic equipment that provides high quality images.
The Osseoflex® SB is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal polymethymethacrylate (PMMA) bone cements for use during percutaneous vertebral augmentation, such as kyphoplasty.
| Summary of Technological Characteristics of the Additional Size<br>Compared to the Current Size (Predicate) | | |
|-------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------|
| Characteristic | Additional Size | Current Size (Predicate) |
| Trade Name, Model | Osseoflex SB,<br>OF-8222 | Osseoflex SB,<br>OF-0005 |
| Cannula size | 8G | 8G |
| Balloon Inflation Medium | 60% Contrast | 60% Contrast |
| Balloon Material | Polyurethane | Polyurethane |
| Balloon Diameter at nominal volume | 15 mm max | 15 mm max |
| Balloon Length at nominal volume | 16 mm | 15 mm |
| Balloon Shape | Spherical | Cylindrical |
| Max inflation pressure | 400 psi (27 ATM) | 400 psi (27 ATM) |
Osseon LLC Osseoflex SB, 2ml
Special 510(k)
Intended Use
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the letters CSSEA in a stylized font. The letters are all uppercase and are connected by a horizontal line at the bottom. The letters are all outlined in black, and the inside of the letters is white. The last letter, A, has a registered trademark symbol next to it.
| Company of Children And List and A A Marcel Company of Children Company of Children | Angel Build Build Build Build Build Build Artistics | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| September 1998 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 199<br>Max inflation volume | Zmi | 4m |
| Summary of Non-Clinical Tests Conducted for Determination of Substantial<br>Equivalence | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Test Summary | | |
| Test Performed | Acceptance Criteria | |
| Balloon Profile and<br>Tamp (Catheter)<br>Working Length (TM-<br>003) | Balloon profile $\le$ 3.48 mm (0.137 in) Balloon catheter working length > 16.5 cm (length of access cannula) | |
| Balloon Compliance<br>(TM-004): | Balloon working length (L) is 16 mm (reference) at the maximum recommended volume 2 mL. Balloon diameter (D) is 15 mm maximum at the maximum recommended volume 2 mL. The 15 mm diameter maximum specification is to ensure that the diameter of the balloon will not grow large enough to possibly go through the end plates of the vertebrae. | |
| | Maximum Pressure<br>(TM-006) | The inflatable bone tamp exceeds the maximum inflation pressure, 27 atm (~400 psi) without failure. |
| | Bond Tensile<br>Strengths (TM-007) | Bond tensile strength $\ge$ 15 N (3.37 lbf). The tensile force specification was adopted directly from ISO 10555 (Single Use Intravascular Catheters) requirements. This tensile force maybe applied to the device during use when the balloon is deflated and retracted back through the access cannula. |
| | Balloon Maximum<br>Volume (TM-008) | Maximum inflation volume 2 mL with 95% confidence and 90% reliability. |
| | Balloon Fatigue,<br>Unconstrained (TM-<br>009) | Inflate to maximum recommended volume of 2 mL, hold for 30 seconds / deflate; without leaks for 20 cycles. |
| Balloon Inflation,<br>Deflation Time (TM-<br>010) | The Osseoflex SB 2ml samples deflation times to be clinically equivalent to other marketed inflatable bone tamps. | |
| Summary of Clinical Tests Conducted for Determination of Substantial Equivalence | | |
| N/A – No clinical test were conducted for this submission | | |
| Conclusions Drawn from Non-Clinical and Clinical Data | | |
| The results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements, and are substantially equivalent to the predicate device. | | |
. ,
:
.
,
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 19, 2014
Osseon LLC % Mr. Keith Burger Director of Research and Development 2330 Circadian Way Santa Rosa, California 95407
Re: K140937
Trade/Device Name: Osseoflex SB Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: May 30, 2014 Received: June 2, 2014
Dear Mr. Burger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr. I has I statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); Jabeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{4}------------------------------------------------
### Page 2 - Mr. Keith Burger
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017
See PRA Statemant below.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
### Indications for Use
510(k) Number (if known) K140937
Device Name Osseoflex SB
#### Indications for Use (Describe)
The Osseoflex® SB is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during perculancous vertebral augmentation. This system is to be used with cleared spinal polymethymethacrylate (PMMA) bone cements for use during percutancous vertebral augmentation, such as kyphoplasty.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY |
|------------------|
|------------------|
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Laurence D. Coyne -A
(Division Sign-OST) Division of Orthopedic Devices SID(k) Number: K 140937
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *Do Not send your completed form to the pra staff Email Address BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the I he barbon and the services existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
*An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
PSC Publicitang Services (201) 443-604 D
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.