HALO MEDICAL TECHNOLOGIES CATALYST MIDCRYSTL HALO ULTRASOUND SYSTEM
K140899 · Halo Medical Technologies, LLC · IYO · May 5, 2014 · Radiology
Device Facts
Record ID
K140899
Device Name
HALO MEDICAL TECHNOLOGIES CATALYST MIDCRYSTL HALO ULTRASOUND SYSTEM
Applicant
Halo Medical Technologies, LLC
Product Code
IYO · Radiology
Decision Date
May 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
"Catalyst" is a diagnostic ultrasound system designed to be used for investigating disorders of the pelvic floor. An ultrasonographic crystal within the probe records images of the organ, muscle, and tissue structures of the pelvic region. MidCRYSTL and HALO probes allow for ultrasonography of the following: 1) on the surface of the perineum and/or abdomen, 2) endocavity, by inserting the endovaginal probe into the vagina, and 3) endocavity, by inserting the endoanal probe into the anal canal.
Device Story
Portable, single-mode, single-application B-Mode ultrasound imaging system; utilizes mechanical sector ultrasound probes (3.5 MHz to 24.0 MHz). System includes ultrasound generator/receiver, analog-to-digital converter, microcontroller, and USB 2.0 interface. Operator inputs transmit power, image selection, and Time Gain Compensation (TGC) via control panel; system stores settings for recall. Device captures, displays, and records 2D ultrasound images; supports cine function (up to 512 images) and distance/angle measurements. Used in clinical settings for pelvic floor investigation; operated by healthcare professionals. Output viewed on LCD monitor; assists clinicians in visualizing organ, muscle, and tissue structures to inform diagnostic decisions.
Clinical Evidence
Bench testing only. No clinical studies were required to support substantial equivalence. Compliance verified through acoustic output measurement (AIUM/NEMA UD-2/UD-3), electrical/ultrasound safety (IEC 60601-1, IEC 60601-2-37), EMC (IEC/EN 60601-1-2), and biocompatibility (ISO 10993-1, -5, -10, -12).
Technological Characteristics
Mechanical sector ultrasound system; 3.5-24.0 MHz probes. Materials: TPX (Polymethyl Pentene) lens, Ertalyte (PET-P) housing. Energy: Piezo-electric transducer. Connectivity: USB 2.0 to host computer. Display: 2D, 256 gray shades. Features: Digital dynamic receive focusing, cine storage (512 images), distance/angle measurement. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993.
Indications for Use
Indicated for diagnostic ultrasound imaging of the pelvic floor, including perineal/abdominal surface, endovaginal, and endoanal applications for investigation of pelvic floor disorders.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K140899
Page 1 of 6
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
### 1) Submitter Information and Preparation Date
Halo Medical Technologies, LLC
1805 Foulk Road, Suite G
Wilmington, DE 19810
Jennifer L. Kinney
Vice President of Marketing and Regulatory Affairs
Telephone: (505) 231-6452
(505) 989-3507 Fax:
Email j.kinney@halomedtech.com
Prepared: March 6, 2014
## 2) Device Name, Common Name, and Classification Name
### Proprietary Name
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### Common/Usual Name
Diagnostic Ultrasound System and Accessories
### Classification Names
| | CFR Number | Product Code |
|---------------------------------------|------------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
{1}------------------------------------------------
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### 3) Identification of the Predicate Device
The Laborie NuWav (AKA "AquaVU360) (K081781) is a comparable and substantially equivalent device. The Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System Transducer Probes are produced by the same OEM manufacturer, and the system has the same technological characteristics, safety and effectiveness features, design features, and comparable intended uses and basic operating modes.
### 4) Subject Device Description
The Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System is a self contained portable, single-mode, and singleapplication ultrasound imaging system. The system contains an ultrasound generator/receiver, analog to digital converter, microcontroller, control logic, USB 2.0 interface and control offering a full complement of conventional operating modes, software-based parameter controls, and video recording.
User-customized parameter settings for the MidCRYSTL™ Endoanal, HALO™ Endovaginal and HALO™ Surface Ultrasound Probes may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit power, images controls selection, and Time Gain Compensation (TGC). Controls are also provided to select display format and to utilize the cine function.
More detailed explanations of these functions and controls are included in the User Manuals, and in the software/firmware documentation included in this 510(k) Notification. Patient contact materials have been used in accordance to their intended use and are described below for each individual transducer. The transducers were previously cleared for use on other Systems (K951976 and
510(k) Summary of Safety and Effectiveness
Page 2 of 6
{2}------------------------------------------------
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
referenced in K070907).
The Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System is a B-Mode ultrasound scanner which provides high resolution, high penetration performance. Probes are supported in frequencies from 3.5 MHz to 24.0 MHz. The probes are indicated expressly for endoanal, endovaginal, and/or abdominal/transperineal (surface) application.
The Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System provides various measuring functions. It can measure distances and calculate angles. Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System supports the Cine function (capable of storing up to 512 sequential images). Management of patient history is possible by image-storage function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing. The same clinical uses were cleared for the predicate device(s), Laborie NuWAV (K081781).
### 5) Intended Use
"Catalyst" is a diagnostic ultrasound system designed to be used for investigating disorders of the pelvic floor. An ultrasonographic crystal within the probe records images of the organ, muscle, and tissue structures of the pelvic region. MidCRYSTL and HALO probes allow for ultrasonography of the following: 1) on the surface of the perineum and/or abdomen, 2) endocavity, by inserting the endovaginal probe into the vagina, and 3) endocavity, by inserting the endoanal probe into the anal canal.
6) Technological Characteristics
{3}------------------------------------------------
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
This device has the same technological characteristics (i.e., materials, image display, energy source, design, and acoustic output limits) as the predicate device.
# 6.a.) Technological Comparison Summary of Subject and Predicate Devices
| Technological<br>Characteristics | Subject Device:<br>Halo Medical Technologies,<br>Catalyst™™, MidCRYSTL™, HALO™<br>Ultrasound System | Predicate Device:<br>Laborie Medical Technologies, NuWav<br>Ultrasound System (K081781) | |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Lens | TPX brand Polymethyle Pentene(PMP) | TPX brand Polymethyle Pentene(PMP) |
| Materials | Housing | Ertalyte brand Polyetheylene<br>Terephthalate (PET-P) | Ertalyte brand Polyetheylene<br>Terephthalate (PET-P) |
| | Display Mode | 2-D, LCD monitor (256 gray shades) | 2-D, LCD monitor (256 gray shades) |
| Measurements | | Distance (mm), Angle (%) | Distance (mm), Angle (%) |
| Principle of Operation | | Apply high voltage bursts to Piezo-<br>electric material in the transducer and<br>detect the reflected echo to construct<br>2D images for diagnostic purposes. | Apply high voltage bursts to Piezo-<br>electric material in the transducer and<br>detect the reflected echo to construct 2D<br>images for diagnostic purposes. |
| Transducer Probe<br>Design | | Mechanical sector ultrasound imaging<br>probe that connects directly to host<br>computer via Universal Serial Bus<br>(USB). Host computer forms real-time<br>ultrasonic images of human tissue<br>without need for additional electronics,<br>power supplies, or support devices of<br>any kind. | Mechanical sector ultrasound imaging<br>probe that connects directly to host<br>computer via Universal Serial Bus (USB).<br>Host computer forms real-time ultrasonic<br>images of human tissue without need for<br>additional electronics, power supplies, or<br>support devices of any kind. |
| Acoustic Output Limits:<br>All Applications | | ISPTA.3 94 mW/cm2 (Maximum)<br>MI 1.9 (Maximum) | ISPTA.3 94 mW/cm2 (Maximum)<br>MI 1.9 (Maximum) |
## 6.b.1.) Summary of Nonclinical Tests and Standards Used
Substantial equivalency to the predicate device has been demonstrated through the following nonclinical tests and standards referenced in this 510(k) submission:
{4}------------------------------------------------
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
- Acoustic Output Measurement, conducted in accordance with the . AIUM/NEMA UD-2 and UD-3 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- Product Safety and EMC Requirements for Medical Equipment: .
- Electrical Safety, IEC 60601-1 o
- Electromagnet Compatibility, IEC/EN 60601-1-2 o
- o Ultrasound Equipment Safety, IEC 60601-2-37
- Includes Thermal Safety Validation .
- Biological Safety .
- o ISO 10993-1 Compliant, Selection of Tests
- o ISO 10993-5 Compliant, Cytotoxicity Study using ISO Elution Method
- o ISO 10993-10 Compliant, ISO Vaginal Irritation Study and ISO Maximum Sensitization Study - Extract
- o ISO 10993-12 Compliant, Sample Preparation
### 6.b.2.) Summary of Clinical Tests
The subject of this premarket submission, the Halo Medical Technologies Catalyst™, MidCRYSTL™, HALO™ Ultrasound System, did not require clinical studies to support substantial equivalence.
### 6.b.3.) Conclusions
Halo Medical Technologies considers the Catalyst™, MidCRYSTL™, HALO™ Ultrasound System to be as safe and effective as the predicate device(s). The performance of the subject device is substantially equivalent to the predicate device(s).
Intended uses and other key features are consistent with traditional clinical
{5}------------------------------------------------
K140899
Page 6 of 6
Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with medical device industry standards. The device conforms to applicable medical device
safety standards and compliance is verified through independent evaluation. Diagnostic ultrasound has accumulated a long history of safe and effective performance.
Therefore, it is the opinion of Halo Medical Technologies that the Catalyst™, MidCRYSTL™, HALO™ Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
May 5, 2014
HALO Medical Technologies, LLC % Mr. Mark Job Responsible Third Party Official 1394 25th Street NW BUFFALO MN 55313
Re: K140899
Trade/Device Name: Halo Medical Technologies Catalyst™, MidCRYSTI,"", HALO"* Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: April 7, 2014 Received: April 8. 2014
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Halo Medical Technologies Catalyst™, HALO™ Ultrasound System, as described in your premarket notification:
Transducer Model Number
ER/12 MHz/ES 12MHz
GP 3.5 MHz/AB 3.5 MHz. EC 7.5 MHz/EB 7.5 MHz
If your device is classified (see above) into either class II (Special Controls) or class !!! (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Paris 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2-Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda,gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### 510{k) Number (if known)
### K140899
#### Device Name
Halo Medical Technologies, Catalys1™, MidCRYSTL™, HALO™ Ultrasound System
#### Indications for Use (Describe)
"Catalyst" is a diagnostic ultrasound system designed to be used for investigating disorders of the pelvic floor. An ultrasonographic crystal within the probe records images of the organ, muscle, and tissue structures of the pelvic region. MidCRYSTL and HALO probes allow for ultrasonography of the following: 1) on the surface of the perineum and/or abdomen, 2) endocavity, by inserting the endovaginal probe into the vagina, and 3) endocavity, by inserting the endoanal probe into the anal canal.
### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7).
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
{9}------------------------------------------------
### Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### Diagnostic Ultrasound Indications for Use Form
#### Halo Medical Technologies, Catalyst™ Ultrasound System System:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|--------------------------|--------------------|---------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks 1&3) | B | M | PWD | CWD | Color<br>Doppler<br>(CD) | Combined<br>(B+CD) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | N | | | | | | Note 1 |
| | Intra-Operative (Specify) | | | | | | | |
| | Intra-Operative Neurological | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric<br>(excluding transcranial<br>& neonatal) | | | | | | | |
| | Small Organ (e.g. testicles,<br>lymph nodes, thyroid) | N | | | | | | Note 2 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Trans-rectal | N | | | | | | |
| & Other | Trans-vaginal | N | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-card.) | | | | | | | |
| | Muscular-Skeletal (Conventional) | | | | | | | |
| | Muscular-Skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Invasive Diagnostic or therapeutic<br>procedures (e.g. biopsies, punctures,<br>free fluid detection, regional<br>anesthesia, vascular access) | | | | | | | |
| | Retroperitoneum | | | | | | | |
| | Female reproduction system and<br>fetus (transcutaneous) | | | | | | | |
| | Superficial structures & pathologies | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N=New Indication
Note 1: Abdominal, Solid Organs, Aneurysms, Bladder
Note 2: Small Organ, Testes, Prostale
图 Prescriptive Use (Part 21 CFR 80) Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Concurrence of CDRH. Office of in Viro Diagnostic Devices (OTVD)
{10}------------------------------------------------
### Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### Diagnostic Ultrasound Indications for Use Form
#### Transducer: ER/12 MHz/ES 12MHz Mechanical Sector Probe
#### Intended Use: Diagoostic ultrasound imaging or Nuid Now analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|--------------------------|--------------------|---------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler<br>(CD) | Combined<br>(B+CD) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-Operative (Specify) | | | | | | | |
| | Intra-Operative Neurological | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric<br>(excluding transcranial<br>& neonatal) | | | | | | | |
| | Small Organ (e.g. testicles,<br>lymph nodes, thyroid) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-card.) | | | | | | | |
| | Muscular-Skeletal (Conventional) | | | | | | | |
| | Muscular-Skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Invasive Diagnostic or therapeutic<br>procedures (e.g. biopsies, punctures,<br>free fluid detection, regional<br>anesthesia, vascular access) | | | | | | | |
| | Retroperitoneum | | | | | | | |
| | Female reproduction system and<br>fetus (transcutaneous) | | | | | | | |
| | Superficial structures & pathologies | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N=New Indication
图 Prescriptive Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
.
{11}------------------------------------------------
### Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### Diagnostic Ultrasound Indications for Use Form
#### Transducer: GP 3.5 MHz / AB 3.5 MHz Mechanical Sector Probe
#### Intended Use: Diagnostic ultrasound imaging or Nuld flow analysis of the homan body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|--------------------------|--------------------|---------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks 1&3) | B | M | PWD | CWD | Color<br>Doppler<br>(CD) | Combined<br>(B+CD) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | N | | | | | | Note 1 |
| | Intra-Operative (Specify) | | | | | | | |
| | Intra-Operative Neurological | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric<br>(excluding transcranial<br>& neonatal) | | | | | | | |
| | Small Organ (e.g. testicles,<br>lymph nodes, thyroid) | N | | | | | | Note 2 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-card.) | | | | | | | |
| | Muscular-Skeletal (Conventional) | | | | | | | |
| | Muscular-Skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Invasive Diagnostic or therapeutic<br>procedures (e.g. biopsies, punctures,<br>free fluid detection, regional<br>anesthesia, vascular access) | | | | | | | |
| | Retroperitoneum | | | | | | | |
| | Female reproduction system and<br>fetus (transcutaneous) | | | | | | | |
| | Superficial structures & pathologies | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N=New Indication
Note I : Abdominal, Solid Organs, Aneurysms, Bladder
Note 2: Small Organ, Testes, Prostate
图 Prescriptive Use (Part 2) CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{12}------------------------------------------------
### Halo Medical Technologies, Catalyst™, MidCRYSTL™, HALO™ Ultrasound System
### Diagnostic Ultrasound Indications for Use Form
#### Transducer: EC 7.5 MHz / EB 7.5 MHz Mechanical Sector Probe
#### Intended Use: Diagnostic ultrasound imaging or Nuid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|--------------------------|--------------------|---------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks 1&3) | B | M | PWD | CWD | Color<br>Doppler<br>(CD) | Combined<br>(B+CD) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-Operative (Specify) | | | | | | | |
| | Intra-Operative Neurological | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric<br>(excluding transcranial<br>& neonatal) | | | | | | | |
| | Small Organ (e.g. testicles,<br>lymph nodes, thyroid) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Trans-rectal | | | | | | | |
| | Trans-vaginal | N | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-card.) | | | | | | | |
| | Muscular-Skeletal (Conventional) | | | | | | | |
| | Muscular-Skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Invasive Diagnostic or therapeutic<br>procedures (e.g. biopsies, punctures,<br>free fluid detection, regional<br>anesthesia, vascular access) | | | | | | | |
| | Retroperitoneum | | | | | | | |
| | Female reproduction system and<br>fetus (transcutaneous) | | | | | | | |
| | Superficial structures & pathologies | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N¤New Indication
🇿 Prescriptive Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITT: BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
#### Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.