The DCAS i allows the user to look up, keep, and manage all the medical images used in cardiology with Angio, ECHO, and IVUS images as the basic ones. It allows the user to view, arrange, and manage the needed image by only controlling it in a simple way. Also, as the user can arrange images as he or she wants regardless of the sort of the images, the user can compare and analyze the images of all the equipments of cardiology on this system. The user can record the images of the patient wanted on a CD or DVD. Here, the exclusive viewer allows the user to check the recorded images of the patient in all sorts of environment. In addition, the user can edit and convert the DICOM Cine images into AVI and static images to use them for editing various images and framing a report. Only the user authorized by using the Internet can use this system based on the Web, and in an environment with a wireless network built, the user can look up data through a mobile device. DCAS i images can be used for reference purposes only and cannot be used for precise readings.
Device Story
DCAS i is a Picture Archiving and Communication System (PACS) for cardiology. It ingests DICOM-compliant images (Angio, ECHO, IVUS) via network from acquisition devices. The system uses a multi-threading image processing engine to display, compare, and manage images. It includes a server for database management and SQL-based patient information storage, a CS View desktop application, a Web View for remote access, and a Mobile View for tablets. Clinicians use the system to view, arrange, and edit images (convert DICOM Cine to AVI/static), frame reports, and record data to CD/DVD. The system facilitates clinical decision-making by allowing side-by-side comparison of cardiology images. It is intended for use in clinical environments by authorized personnel. Benefits include improved workflow, remote access to patient data, and enhanced image management capabilities.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
PACS software utilizing a multi-threading image processing engine and Device Independent Bitmap (DIB) engine. Supports DICOM 3.0 standard. Connectivity via GIGABIT network, web-based access, and wireless mobile integration. Data storage in SQL database and FTP server. Image formats supported: DCM, BMP, JPG. Lossless compression method.
Indications for Use
Indicated for use by clinicians to store, manage, view, and analyze cardiology medical images (Angio, ECHO, IVUS). Intended for reference purposes only; not for precise readings.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K140869/
Page 1 of 7
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Mar. 15, 2014
1. Company and Correspondent making the submission:
Name - GP&P Co., Ltd.
Address - 706, Samseong-ro, Gangnam-gu, Seoul, 135-953, Korea Telephone - +82 2 3445 9336
Fax - +82-2-3445-9330
Contact - Mr. SungChul Ahn / RA Manager
- 2. Device :
| Trade/proprietary name | : DCAS i |
|------------------------|-----------------------------------------------|
| Common Name | : Picture archiving and communications system |
| Classification Name | : System, image processing, radiological |
- 3. Predicate Devices :
| Manufacturer: | GE Medical Systems In formation Technologies |
|----------------|----------------------------------------------|
| Device: | Centricity Cardiology CA1000 |
| 510(k) Number: | K042525 (Decision Date - 10/01/2004) |
- 4. Classifications Names & Citations :
21CFR 892.2050, LLZ, Picture archiving and communications system, Class2
- 5. Description :
- 1) General Description
The DCAS i is an equipment for providing medical treatment services of a good quality in cardiology and developing into an advanced medical institution through innovative work improvement, in order to keep up with the rapidly changing era of medical information. The DCAS i, which has an improved high-speed image processing engine,
{1}------------------------------------------------
processes in the method of multi-threading to display multiple DICOM Cine images in a simultaneous way. It allows the user to look up, keep, and manage all the medical images used in cardiology with XA, ECHO, and IVUS images as the basic ones. The DCAS i sends even high-capacity images within only a few seconds and expresses them on the basis of the GIGABIT network environment of its own. The enhanced convenience of its user interface allows the user to view, arrange, and manage the needed image by only controlling it in a simple way. Also, as the user can arrange images as he or she wants regardless of the sort of the images, the user can compare and analyze the images of all the equipments of cardiology on this system. The user can record the images of the patient wanted on a CD or DVD by only controlling a few times. Here, the exclusive viewer allows the user to check the recorded images of the patient in all sorts of environment. In addition, the user can edit and convert the DICOM Cine images into AVI and static images, and this makes it easier to edit various images and frame a report. Only the user authorized by using the Internet can use this system based on the Web as needed, and in an environment with a wireless network built, the user can look up data through a mobile device.
#### 2) Main Function
#### (1)Server
The Server controls each module and communication and saves the data files and the database.
Before linking each module to the Server, the communication test gets done if each module can be connected to the Server; and when the connection is available, the user gets checked if he or she has the authority to access the Server. When a user with no authority to access the Server tries to access or a user tries to access on another system with an ID which is already accessed to the Server, the user gets forced to log out.
The Server saves the files sent from the DICOM Acquisition and also
{2}------------------------------------------------
# GP&P Co., Ltd.
saves the data information and patient information in the SQL.
Also, when the device for lookup requests for data files and data information, the Server provides the pertinent files and information to the device for lookup that has requested for them.
#### (2) Acquisition
The DICOM files get sent from the DICOM output device (Angio, ECHO, IVUS) to the DICOM Acquisition by using the DICOM network based on the DICOM 3.0 standard. The DICOM files get sent from the DICOM Acquisition to the DATA Server. When the DICOM files get sent, the patient information from among the DICOM header information gets saved in the SQL. DB, and the image files get saved in the FTP of the DATA Server. When image files get saved, they get converted into JPG files to be outputted on the Web and the Mobile; also, the Thumbnails to be shown on the list get saved together.
#### (3) CS View
Dental Reformat makes it possible to reconstruct Panoramic and Cross sectional images. Conventional imaging solutions are supported, and also new imaging solutions such as Volume Rendering, 3D Scout are supported. The layout and images are optimized for Reporting
The CS View is a special viewer that perfectly materializes DICOM. which is an international standard of medical images. It allows the user to rapidly and accurately look up and compare images of a variety of format of various medical devices in high definition. It rapidly provides images to the user with the Device Independent Bitmap (DIB) engine, and also allows the user to compare several DICOM images at the same time by employing the 'multi-threading' system.
The user can look up and search the patients registered on the server and the data saved. The CS View provides simple measuring functions for the images looked up.
The CS View supports CD Recording function and the Export function of a variety of format. It also supports a codec to allow the user to
{3}------------------------------------------------
## GP&P Co., Ltd.
convert images into high-quality image files.
#### (4) WEB View
The Web View uses a UI which is consistent with those of the DCAS i cs and the DCAS i mobile for the user to easily adapt to. It allows the user to search and look up images not only within the hospital but anywhere with a PC connected to the Internet. If you have a PC with the Internet and Web surfing available, you can look up and check patients' images on the Web wherever in the same way you use the DCAS i cs. As the user accesses through a Web browser, there is no issue of hardwares or softwares occurring unexpectedly while the user is installing the software.
#### (5) Mobile View
The Mobile View is made of the most frequently used functions in the DCAS i cs. As it is optimized for latest mobile devices, you can actively utilize it while you are having a conversation with a patient in the environment with a wireless network built. Also, the doctors can search items on behalf of the patient who is not able to move freely to explain to the patient..
In addition, the user can check and comment on the images whenever wherever, even overseas or out of the hospital, if only communication is available to convey accurate information and give instructions to the doctor in charge in the field.
#### 3) Information of the image format
DCAS i can load only DCM files and save results as DCM. BMP and JPG files.
- DCM : DICOM (Digital Image Communication in Medicine) is a Standard Protocol to exchange and transfer the data acquired by Medical Image devices such as a CT, MR, 3D US, etc. It is designated as a Standard Protocol by ACR-NEMA (American College of Radiology-National Electrical Manufacturers Association) and now adopted by most Medical Imaging Devices. DCAS i 2.0 is adaptable technically for all data of DICOM 3.0.
{4}------------------------------------------------
Reference : Digital Imaging and Communications in Medicine (DICOM) ACR-NEMA Standards Publication PS 3.1~PS 3.16 2003.
- BMP : The standard bit-mapped graphics format used in the Windows environment. By convention, graphics files in the BMP format end with a BMP extension. BMP files store graphics in a format called device-independent bitmap (DIB).
- JPG/JPEG : Short for "Joint Photographic Experts Group", the original name of the committee that wrote the standard. JPG is one of the image file formats supported on the Web. JPG images support 16 million colors and are best suited for photographs and complex graphics.
### 4) Compression
- Compression Method : Lossless Compression
#### 6. Indication for use :
The DCAS i allows the user to look up, keep, and manage all the medical images used in cardiology with Angio, ECHO, and IVUS images as the basic ones. It allows the user to view, arrange, and manage the needed image by only controlling it in a simple way. Also, as the user can arrange images as he or she wants regardless of the sort of the images, the user can compare and analyze the images of all the equipments of cardiology on this system. The user can record the images of the patient wanted on a CD or DVD. Here, the exclusive viewer allows the user to check the recorded images of the patient in all sorts of environment. In addition, the user can edit and convert the DICOM Cine images into AVI and static images to use them for editing various images and framing a report. Only the user authorized by using the Internet can use this system based on the Web, and in an environment with a wireless network built, the user can look up data through a mobile device.
DCAS i images can be used for reference purposes only and cannot be used for precise readings.
#### 7. Comparison with predicate device :
GP&P Co., Ltd., ensures that the DCAS i, Picture archiving and communications
{5}------------------------------------------------
system is substantially equivalent to Centricity Cardiology CA1000 of GE Medical Systems In formation Technologies.
The DCAS i described in this 510(k) has the same intended use and similar t echnical characteristics as the Centricity Cardiology CA1000 of GE Medical Sys tems In formation Technologies. The similanties and differences between these systems are described in the table shown above.
The similarities are as follows.
- 1. Similarity of DICOM-Compliant, Architecture, Server, Security, Storage.
- 2. Similarity of System failure alert, Database Management.
- 3. Similarity of Web-Based Access, Thumbnail, Image Compression, Multi-frame support, Image Control, Play Control, Zoom
- 4. Similarity of Search Patient, Search Case Date, Search Modality, Sort.
- 5. Similarity of Common Measurement XA Analysis, Us Analysis.
The difference is as follows.
DCAS i provides mobile applications for iPads and Android Tablets, named "mobile View".
In summary, The DCAS i does not introduce any new potential safety risks an d is substantially equivalent to and performs as well as the predicate device.
In conclusion, the DCAS i is substantially equivalent to Centricity Cardiology CA1000 of GE Medical Systems In formation Technologies.
- 8. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification GP&P Co., Ltd. concludes that DCAS i is safe and effective and substantially equivalent to predicate devices as described herein.
- 9. GP&P Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
{6}------------------------------------------------
GP&P Co., Ltd.
ﺮ
・・・・・・・・・・・・・・・・・・・・・・・・・・・
・
END
」
.
to the same of the manage the same the state with and states of the states of the states of the states of the status and the status of the status the status and the many of t
706, Samseong-ro, Gangnam-gu, Seoul, 135-953, Korea
·
.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized representation of an abstract human figure with three lines extending from the figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the figure.
Public 1 lealth Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W(166-G609 Silver Spring, MD 20993-0002
July 3, 2014
GP&P Co., Ltd. % Mr. Charlie Mack, Principal Engineer International Regulatory Consultants 12226 Washington Lane PARKER AZ 85344
Re: K140869
Trade/Device Name: DCAS i Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 16, 2014 Received: June 24, 2014
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041
{8}------------------------------------------------
Page 2-Mr. Mack
or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use
510(k) Number(if known): K140869
Device Name: Picture Archiving Communications System / DCAS i
Indications for Use:
The DCAS i allows the user to look up, keep, and manage all the medical images used in cardiology with Angio, ECHO, and IVUS images as the basic ones. It allows the user to view, arrange, and manage the needed image by only controlling it in a simple way. Also, as the user can arrange images as he or she wants regardless of the sort of the images, the user can compare and analyze the images of all the equipments of cardiology on this system. The user can record the images of the patient wanted on a CD or DVD. Here, the exclusive viewer allows the user to check the recorded images of the patient in all sorts of environment. In addition, the user can edit and convert the DICOM Cine images into AVI and static images to use them for editing various images and framing a report. Only the user authorized by using the Internet can use this system based on the Web, and in an environment with a wireless network built, the user can look up data through a mobile device. DCAS i images can be used for reference purposes only and cannot be used for
Prescription Use (Part 21 CFR 801 Subpart D)
precise readings.
Over-The-Counter Use_ (Part 21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Michael D. O'Hara
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.