K140865 · Pro Medical Supplies, Inc. · ILY · Sep 3, 2014 · Physical Medicine
Device Facts
Record ID
K140865
Device Name
INFRARED HEATING DEVICE
Applicant
Pro Medical Supplies, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Sep 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Infrared Heating Device, Model PM-750, temporarily increases local blood circulation where applied, and temporarily relieves minor muscle and joint aches and pains.
Device Story
The Infrared Heating Device, Model PM-750, is a non-invasive, 110V AC-powered therapeutic device. It utilizes a tungsten infrared bulb to emit far-infrared (FIR) radiation at a 7,000 nm wavelength. The device features a handle housing electronics and controls (High/Off/Low settings) and a light filter designed for direct skin contact. Operated by the user in a home or clinical setting, the device provides continuous heat therapy to a 18cm² area. By maintaining a therapeutic skin temperature range of 40-45°C, the device increases local blood circulation, which helps alleviate minor muscle and joint aches and pains. The healthcare provider or user monitors the treatment duration (recommended 15-20 minutes) to ensure safe application. The device is constructed with an ABS housing.
Clinical Evidence
Bench testing only. Performance testing conducted on 6 participants across different body parts to measure heat output. Results confirmed the device maintains a therapeutic temperature above 40°C for at least 10-15 minutes of a 30-minute cycle, without exceeding 45°C at the skin surface. Biocompatibility testing (ISO 10993-1, -5, -10) performed on ABS housing and light filter materials. Electrical safety and EMC testing performed per AAMI ANSI ES60601-1 and IEC 60601-1-2.
Technological Characteristics
110V AC/60Hz powered infrared lamp; 7.5W max output; 1 tungsten infrared bulb; 7,000 nm wavelength; ABS housing; 200g weight; 18cm² treatment area. Standards: AAMI ANSI ES60601-1, IEC 60601-1-2, ISO 10993-1, ISO 10993-5, ISO 10993-10. Manual High/Off/Low power control.
Indications for Use
Indicated for temporary increase of local blood circulation and temporary relief of minor muscle and joint aches and pains in patients requiring heat therapy.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 3, 2014
Pro Medical Supplies Inc. c/o Ms. Doris Dong Shanghai CV Technology Co., Ltd. Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai. China 201600
Re: K140865
> Trade Name: Infrared Heating Device, Model PM-750 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: July 17, 2014 Received: July 28, 2014
Dear Ms. Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Felipe Aguel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K140865
Device Name Infrared Heating Device, Model PM-750
Indications for Use (Describe)
The Infrared Heating Device, Model PM-750, temporarily increases local blood circulation where applied, and temporarily relieves minor muscle and joint aches and pains.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# )14 (19 2 -04 00
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Section 5 510(k) Summary [As required by 21 CFR 807.92]
#### 1. Submission Information:
| 510(k) Number: | K140865 |
|------------------------------|---------------------------------------------------------------------------------|
| Date: | July 15th, 2014 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Applicant/US importer: | Pro Medical Supplies Inc. |
| | 1800 Byberry Rd. Unit 905, Huntingdon Valley, PA 19006, USA |
| Manufacturer: | Shenzhen Yizekang Technologies Co., Ltd. |
| | 2nd Floor, No. 6 Factory, Hekan Road, Hekan Village, |
| | Bantian Sub-District, Longgang District, Shenzhen 518131, China |
| Contactor: | Doris Dong (Consultant) |
| | Shanghai CV Technology Co., Ltd. |
| | Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, China 201600 |
| | E-mail: doris_d@126.com |
| 2. Device Description: | |
| Proprietary Name: | Infrared Heating Device, Model PM-750 |
| Common Name: | Infrared lamp |
| Classification Name: | Infrared lamp |
| Product Code: | ILY |
| Device Class: | 2 |
| Review Panel: | Physical Medicine |
| Regulation Number: | 890.5500 |
| Indications for use: | The Infrared Heating Device, Model PM-750, temporarily increases local |
| | blood circulation where applied, and temporarily relieves minor muscle |
| | and joint aches and pains. |
| Device Description: | The Infrared Heating Device, Model PM-750, is 110V AC powered, |
| | non-invasive, therapeutic medical device that provides continuous heat |
| | therapy through the use of infrared bulb. The device consists of a handle |
| | that houses the electronics and controls, and tungsten infrared bulb with |
| | light filter. The light filter is intended to be placed directly on the skin to |
| | provide topical heating. The infrared bulb is replaceable. The housing |
| | material is ABS. |
| Standards: | AAMI ANSI ES60601-1, IEC 60601-1-2, ISO 10993-1, ISO 10993-5, ISO |
| | 10993-10 |
| | |
## 3. Predicate Device Identification
| 510k Number: | K990233 |
|---------------|--------------------------------|
| Product Code: | ILY |
| Device Name: | ST-302 Infrarex |
| Manufacturer: | Skylark Device Company Limited |
## 4. Substantial Equivalence:
Detailed comparison data is included in Section 9 of "Substantial Equivalence Discussion" of this 510(k) submission.
#### A. Basic technological characteristics & Specifications, New device VS. Predicate device:
{5}------------------------------------------------
Pro Medical Supplies Inc. 1800 Byberry Rd. Unit 905, Huntingdon Valley, PA 19006, USA
| | Parameters | New Device | Predicate Device |
|----|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510(k) Number: | K140865 | K990233 |
| 2 | Marketing clearance date: | -- | Oct 27, 1999 |
| 3 | Proprietary Name: | Infrared Heating Device, Model<br>PM-750 | ST-302 Infrarex |
| 4 | Manufacturer: | Shenzhen Yizekang Technologies Co.,<br>Ltd. | Skylark Device Company Limited |
| 5 | Indications for use: | The Infrared Heating Device, Model<br>PM-750, temporarily increases local<br>blood circulation where applied, and<br>temporarily relieves minor muscle and<br>joint aches and pains. | The Infrarex temporarily increases local<br>blood circulation where applied, and<br>temporarily relieves minor muscle and<br>joint aches and pains. |
| 6 | Power source | 110V AC/60Hz | 110V AC/60Hz |
| 7 | Maximum output power | 7.5W | 7.5W |
| 8 | Power switch function | High/Off/Low | High/Off/Low |
| 9 | Type of heat source and number | 1 tungsten infrared bulb | 1 tungsten infrared bulb |
| 10 | Design Principle | FIR (far-infrared) therapy | FIR (far-infrared) therapy |
| 11 | Wavelength | 7 000 nanometers | 7 000 nanometers |
| 12 | Therapeutic temperature range | 40-45 °C | 40-45 °C |
| 13 | Approximate maximum skin<br>temperature | Low Level: 40.5°C<br>High Level: 44.5°C | Low Level: 40.5°C<br>High Level: 44.5°C |
| 14 | Time to reach therapeutic<br>temperature range | Low Level: 5~10 minutes<br>High Level: 3~5 minutes | Low Level: 5~10 minutes<br>High Level: 3~5 minutes |
| 15 | Recommended treatment time | Used for 15-20 minutes on target area | Used for 15-20 minutes on target area |
| 16 | Weight | 200g | 200g |
| 17 | Treatment area | 18cm² | 18cm² |
| 18 | Performance standards | ISO 10993-1, ISO 10993-5, ISO<br>10993-10, AAMI ANSI ES60601-1,<br>IEC 60601-1-2 | ISO 10993-1, ISO 10993-5, ISO<br>10993-10, IEC 60601-1, IEC 60601-1-2 |
## B. Substantial Equivalence Discussion
| Similarities between New device | The new device and the predicate device are produced by different |
|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| and Predicate Device: | manufacturers. |
| Differences between New device | Same intended use, design, materials, technical parameters, therapeutic |
| and Predicate Device: | parameters, and performance standards. |
| Conclusion: The two devices have exactly same intended use, design, materials, technical parameters, | |
therapeutic parameters and performance standards.
The new device is as safe and effective as the predicate device.
## 5. Safety and Effectiveness of the device
Testing of Infrared Heating Device, Model PM-750, includes electrical safety testing and performance testing.
The device is manufactured to comply with the following international standards:
AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests The housing material and the material of the light filter of PM-750, ABS, was tested and found to meet the biocompatibility standards of:
{6}------------------------------------------------
* ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
* ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity, and * ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity
In addition, two devices were tested for heat output, with the infrared light filter contacting the skin. Temperatures were recorded every 1 minute, up to 30 minutes, using a digital temperature sensor, through 6 participants on different body part. The test results concluded that the Infrared Heating Device, Model PM-750 was able to maintain temperature above 40℃ (the desired effective temperature) at least 10~15 minutes of the 30 minutes cycle time. The tests also showed that the temperature at the sensor (skin) never exceeded a temperature of 45℃. Both devices used for this test tracked each near 100%.
The conclusion drawn from the testing is that the device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.