The SOMNOtouch RESP is a portable physiological signal recorder. It is indicated for use in the recording, displaying, monitoring, printing, and storage of biophysical parameters for the purpose of assisting in the diagnosis of sleep disorders and sleep related respiratory disorders of adult patients. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home. This device is NOT designed to be used in life support situations.
Device Story
Portable physiological signal recorder; worn on thorax via belt. Inputs: thoracic/abdominal respiratory effort (belts), SpO2/pulse rate (finger probe), respiratory flow/snore (nasal cannula), body position/motion (internal accelerometer). Data stored on internal 512MB flash memory; transferred via USB docking station to PC. Used in physician office, sleep lab, or home; operated by physician or patient. DOMINOlight software on PC displays/analyzes data for clinician review. Assists diagnosis of sleep-related respiratory disorders; does not provide automatic diagnosis. Benefits patient by enabling ambulatory sleep monitoring.
Clinical Evidence
Clinical study compared SOMNOtouch RESP to predicate (SOMNOscreen) in a professional sleep lab. Evaluated ability to detect apnea/hypopnea events. Primary endpoint: Apnea-Hypopnea Index (AHI). Analysis included regression, sensitivity, and specificity calculations. Results demonstrated substantial equivalence to predicate; no adverse effects or complications reported.
Technological Characteristics
Portable device, 84x55x18mm, 64g. 10 internal channels, 1 external channel. 12-bit ADC, sampling rates 1/60s to 512/s. Powered by 3.7V Li-Ion rechargeable battery. Connectivity: USB docking station. Standards: DIN EN 60601-1, DIN EN 60601-1-2, DIN EN 60601-1-11, DIN EN 62366, DIN EN ISO 14971, DIN EN IEC 62304, ISO 10993-1. IP22 rating.
Indications for Use
Indicated for adult patients suspected of having sleep disorders or sleep-related respiratory disorders. Not for use in life support situations.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black against a white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 30, 2015
SOMNOmedics GmbH Mr. Matthias Broenner Am Sonnenstuhl 63 97236 Randersacker GERMANY
Re: K140861
Trade/Device Name: SOMNOtouch™ RESP Ventilatory Effort Recorder Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: II Product Code: MNR Dated: December 31, 2014 Received: January 29, 2015
Dear Mr. Broenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K140861
Device Name: SOMNOtouch™ RESP
Indications For Use:
The SOMNOtouch RESP is a portable physiological signal recorder. It is indicated for use in the recording, displaying, monitoring, printing, and storage of biophysical parameters for the purpose of assisting in the diagnosis of sleep disorders and sleep related respiratory disorders of adult patients. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.
This device is NOT designed to be used in life support situations.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
# Section 5: 510(k) Summary of Safety and Effectiveness
# The following information is in conformance with 21 CFR 807.92.
# Submitter's Information: 21 CFR 807.92(a)(1)
SOMNOmedics GmbH Am Sonnenstuhl 63 97236 Randersacker Germany
| Phone: | +49 931 35 90 94 0 |
|-----------------|---------------------|
| Fax: | +49 931 35 90 94 49 |
| Contact Person: | Dr. Gert Küchler |
| Date Prepared: | April, 1st 2014 |
# Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | SOMNOtouch™ RESP |
|----------------------|-----------------------------|
| Common Name: | Ventilatory Effort Recorder |
| Classification Name: | Breathing frequency monitor |
| Product code: | MNR |
| Regulation Number: | 21 CFR 868.2375 |
### Predicate Device: 21 CFR 807. 92(a)(3)
FDA has classified the predicate device (K060708) as Class II, CFR 868.2375, MNR. It is our understanding that SOMNOtouch™ RESP device falls under the same classification as the predicate device. Predicate device details are as follows:
| Device Classification Name: | Breathing frequency monitor |
|------------------------------------|------------------------------------------------------------------------|
| 510(k) Number: | K060708 |
| Regulation Number: | 21 CFR 868.2375 |
| Device Name: | SOMNOscreen |
| Applicant: | SOMNOmedics GmbH<br>Am Sonnenstuhl 63<br>97236 Randersacker<br>Germany |
| Classification Product Code: | MNR |
| Decision Date: | 01/24/2007 |
| Classification Advisory Committee: | Anesthesiology |
{4}------------------------------------------------
# Device Description: 21 CFR 807 92(a)(4)
The SOMNOtouch™ RESP is a small, portable physiological signal recording system intended to be used to record, display, monitor, print and store biophysical parameters to aid in the diagnosis of sleep disorders. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.
The SOMNOtouch™ RESP consists of the following:
- 1. The recording device (worn on the thorax),
- the finger probe, which is used to detect SpO2, 2.
- 3. the external effort sensor,
- two effort belts to measure thoracic and abdominal expansion, 4.
- న్. the nasal cannula.
- 6. the Software DOMINOlight for visualization of the recorded data.
The SOMNOtouch™ RESP typically will be worn at the thorax, attached by a thorax belt. It contains a sensor to measure the respiratory effort signal of the thorax. The device has an internal accelerometer, measuring body position and motion activity. It provides a connector to attach a nasal cannula, which allows the recording of respiratory flow and snore signals with the internal pressure sensor. Arterial oxygen saturation (SpO2) and pulse rate can be determined via a pulse oximetric finger sensor. Abdominal respiratory effort is measured with an external sensor, attached with a belt to the abdomen. This information is stored in the internal memory of the device.
The system provides up to 10 internal channels for data acquisition, Pressure/Flow, Thoracic Effort, SpO2, Pulserate, Snoring, Finger Plethysmogram, Body Position, Movement, Patient Marker, CPAP, and 1 external Abdominal Effort Sensor. The data from all channels can be recorded separately or in any combination with the other channels.
Information is stored on an internal 512 MB flash memory and can be transferred to a PC via a USB docking station. The DOMINOlight software retrieves the data from the SOMNOtouch™ RESP, displays and analyzes the data, and can store data for future reference and comparison. The SOMNOtouch RESP does not provide automatic diagnosis and is not designed to be used in Life Support situations.
## Indications for Use: 21 CFR 807 92(a)(5)
"The SOMNOtouch RESP is a portable physiological signal recorder. It is indicated for use in the recording, displaying, monitoring, printing, and storage of biophysical parameters for the purpose of assisting in the diagnosis of sleep disorders and sleep related respiratory disorders of adult patients. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.
This device is NOT designed to be used in life support situations."
{5}------------------------------------------------
# Technological Characteristics: 21 CFR 807 92(a)(6)
CHANNELS
10 Internal Channels:
(Body Position, Movement, Patient Marker, Thoracic Effort, SpO2, Plethysmography, Pulse Rate, Pressure/Flow sensor (CPAP) Snoring)
1 External Channel:
(Abdominal Effort)
DATA PROCESSING
12 Bit ADC
Different Sampling Rates Adjustable (1/60s - 512/s)
Technological Characteristics: Comparison/Discussion Table:
| Features/Technical<br>Information | SOMNOmedics<br>SOMNOtouch™ RESP | SOMNOmedics<br>SOMNOscreen | Discussion of differences |
|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | Not assigned | K060708 | n/a |
| Product code | MNR | MNR | n/a |
| Indications for Use | The SOMNOtouch RESP is a<br>portable physiological signal<br>recorder. It is indicated for<br>use in the recording,<br>displaying, monitoring,<br>printing, and storage of<br>biophysical parameters for<br>the purpose of assisting in the<br>diagnosis of sleep disorders<br>and sleep related respiratory<br>disorders of adult patients.<br>The device is intended to be<br>prescribed for use by a<br>physician in the office, sleep<br>laboratory or patient's home.<br>This device is NOT designed<br>to be used in life support<br>situations. | The SOMNOscreen is a<br>non-life-supporting portable<br>physiological signal<br>recording device intended<br>to be used for testing adult<br>patients suspected of having<br>sleep-related breathing<br>disorders. | Indications for use and<br>Intended Use sections<br>have been merged.<br>The use in life support<br>situations was clarified.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Features/Technical<br>Information | SOMNOmedics<br>SOMNOtouch™ RESP | SOMNOmedics<br>SOMNOscreen | Discussion of differences |
| Intended Use | The SOMNOtouch RESP is<br>a portable physiological<br>signal recorder. It is<br>indicated for use in the<br>recording, displaying,<br>monitoring, printing, and<br>storage of biophysical<br>parameters for the purpose<br>of assisting in the diagnosis<br>of sleep disorders and sleep<br>related respiratory disorders<br>of adult patients. The device<br>is intended to be prescribed<br>for use by a physician in the<br>office, sleep laboratory or<br>patient's home.<br>This device is NOT designed<br>to be used in life support<br>situations. | The SOMNOscreen is<br>indicated for use in the<br>recording, displaying,<br>monitoring, printing, and<br>storage of biophysical<br>parameters for the purpose<br>of assisting in the diagnosis<br>of Neurological and Sleep<br>Disorders. | Indications for use and<br>Intended Use sections<br>have been merged.<br>The SOMNOscreen<br>provides sensors for<br>additional neurological<br>measurement.<br>The use in life support<br>situations was clarified.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Intended Patient<br>Population | Adult Patients | Adult Patients | n/a |
| Intended<br>Environment | The intended environment are<br>the physician's office, a sleep<br>laboratory or the patient's<br>home | The intended environment<br>are the physician's office, a<br>sleep laboratory or the<br>patient's home | n/a |
| Prescription Use | Yes | Yes | n/a |
| Number of<br>patients monitored<br>simultaneously | 1 per Unit | 1 per Unit | n/a |
| Number of<br>Channels | 11 Channels | 28 Channels | The 11 channels are<br>sufficient to fulfill the<br>intended use. |
| Features/Technical<br>Information | SOMNOmedics<br>SOMNOtouch™ RESP | SOMNOmedics<br>SOMNOscreen | Discussion of differences |
| Functions | Recording and processing of<br>Pressure Flow,<br>Thorax/Abdominal<br>Respiratory Effort, SPO2,<br>Snore, Pulse Rate,<br>Plethysmogram, Body<br>Position, Movement and<br>Patient Marker for basic<br>respiration screening or<br>CPAP-titration. | Recording and processing<br>of Pressure Flow.<br>Thorax/Abdominal<br>Respiratory Effort, SPO2,<br>Snore, Pulse Rate,<br>Plethysmogram, Body<br>Position, Brightness and<br>Patient Marker for basic<br>respiration screening or<br>CPAP-titration.<br>Optional:<br>Additional signals for full<br>PSG recording with or<br>without video, 24-hour<br>ambulatory EEG recording.<br>stationary EEG-monitoring<br>with video control or 24-<br>hour ECG and pulse<br>oximetry. | The 11 channels of the<br>SOMNOtouch RESP are<br>sufficient to fulfill the<br>intended use and already<br>extend the basic<br>requirements for a Type<br>III or Type IV Home<br>Sleep Testing device<br>The additional channels<br>are not required for the<br>diagnosis of sleep-related<br>respiratory disorders.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Portable Design | Yes | Yes | n/a |
| Size | 84 x 55 x 18 mm3<br>( 3.3" x 2.2" x 0.7" ) | 140 x 70 x 28 mm3<br>( 5.5" x 2.8" x 1.1" ) | The SOMNOtouch RESP<br>is smaller, reducing the<br>disturbance of the<br>patient's sleep when using<br>the device.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Weight | 64 grams including battery<br>(2.26 oz.) | 220 grams including battery<br>(7.76 oz.) | The reduced weight does<br>not raise concerns in<br>safety or effectiveness. |
| Device Application | Attachment to patient's<br>thorax using an effort belt | Attachment to patient's<br>thorax or waist using an<br>effort belt | Application on the thorax<br>is the preferred application<br>position. The restriction to<br>this position does not raise<br>concerns in safety or<br>effectiveness. |
| Device User<br>Interface | Power and control buttons<br>Color Touch Display<br>Status Indication LEDs | Power and control buttons<br>Dot Matrix display<br>Status Indication LEDs | The touch display allows a<br>more intuitive interaction<br>with the user interface.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Data collection | Yes | Yes | n/a |
| Features/Technical<br>Information | SOMNOmedics<br>SOMNOtouch™ RESP | SOMNOmedics<br>SOMNOscreen | Discussion of differences |
| Selectable<br>Montage<br>Configuration | Yes | Yes | n/a |
| Calibration Check | No calibration required.<br>Signal quality check on<br>device display possible. | No calibration required.<br>Signal quality check on<br>device display possible. | n/a |
| Sampling method | Analogue to digital<br>conversation, 12 Bit | Analogue to digital<br>conversation, 16 Bit | The 12-Bit A/D-converter<br>used to transform the<br>analogue signals from<br>effort sensors and the<br>pressure sensor is<br>sufficient to fulfill all<br>specifications and allows a<br>smaller device size, as it is<br>integrated in the<br>microcontroller housing.<br><br>The change does not affect<br>the measurement quality<br>compared with the<br>predicate device, as also<br>confirmed within the<br>validation described in<br>section 20: Performance<br>Testing - Clinical. |
| Sampling rates | Up to 512 Hz | Up to 512 Hz | n/a |
| Raw Data Storage | Internal 512MB flash-memory | Compact Flash Card with a<br>capacity up to 512 MB | The integrated memory<br>card, allows a reduced<br>size, minimizes errors in<br>handling with the card and<br>is an improved protection<br>against unauthorized<br>access to data stored on<br>the memory card.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Radio LAN<br>Capabilities for<br>Remote Live View<br>of Data | No | Yes | The Live View of Data is<br>required for fully attended<br>sleep testing in the sleep<br>lab. For basic home sleep<br>testing the live view is not<br>required and not relevant<br>for the predicate device in<br>that field of usage as well.<br>The differences do not<br>raise concerns in safety or<br>effectiveness. |
| Features/Technical<br>Information | SOMNOmedics<br>SOMNOtouch™ RESP | SOMNOmedics<br>SOMNOscreen | Discussion of differences |
| Data interface | Serial data transfer | Serial data transfer | n/a |
| Transfer method | Docking station | External card reader on PC | The docking station allows<br>a direct communication<br>with the device, prevents<br>handling errors with the<br>memory card, allows to<br>check correct transfer of<br>montage and patient data<br>to the SOMNOtouch<br>RESP and allows status<br>indication (i.e. battery<br>state) directly during<br>device initialization.<br>The differences do not<br>raise…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.