K140814 · Smith & Nephew, Inc. · HRS · May 7, 2014 · Orthopedic
Device Facts
Record ID
K140814
Device Name
EVOS MINI-FRAGMENT PLATING SYSTEM
Applicant
Smith & Nephew, Inc.
Product Code
HRS · Orthopedic
Decision Date
May 7, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bone plates and bone screws from the EVOS Mini-Fragment Plating System are intended for use in internal fixation of small bones and fragments and as non-load bearing stabilization and fixation of bone fragments.
Device Story
Internal fixation system comprising 316L stainless steel flat locking bone plates (4-20 holes) and compatible locking/non-locking screws and washers. Used by orthopedic surgeons for small bone fracture fixation, arthrodesis, and reconstruction. Provides mechanical stabilization of bone fragments to facilitate healing. System components are single-use. Clinical benefit derived from restoring structural integrity to fractured or compromised bone segments.
Clinical Evidence
Bench testing only. Finite Element Analysis (FEA) identified worst-case plates. Four-point bend fatigue testing on plates and 2.0mm/2.4mm/2.7mm screws demonstrated non-inferiority to predicates. Torque-to-failure testing on cortex and osteopenia screws met acceptance criteria and ASTM F543 minimums.
Technological Characteristics
Implant-grade 316L stainless steel. Locking and non-locking bone plates and screws. Mechanical fixation via locking mechanism. Dimensions include 4-20 hole plate configurations and 2.0mm, 2.4mm, 2.7mm, and 4.0mm screw diameters. Sterilization: sterile or non-sterile packaging. Testing per ASTM F543.
Indications for Use
Indicated for adolescent (12-18 years), transitional adolescent (18-21 years), and adult patients, including those with osteopenic bone. Used for fracture fixation, arthrodesis, reconstruction, replantation, or reduction of small bones and fragments, and non-load bearing stabilization/reduction of long bone fragments.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Variable-Angle Locking Mini-Fragment Plating System (K132886)
VLP FOOT Plating, Screw System and Accessories (K090675)
Submission Summary (Full Text)
{0}------------------------------------------------
## > We are smith&nephew
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | May 2, 2014 |
| Contact Person and Address: | Samantha Staubach<br>Regulatory Affairs Specialist I<br>T 901-399-6132<br>F 901-566-7596 |
| Name of Device: | EVOS Mini-Fragment Plating System |
| Common Name: | Bone Plates, Screws and Washers |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br>Class II<br>21 CFR 888.3040 Smooth or threaded metallic bone fixation<br>fastener<br>Class II |
| Panel Code: | Orthopedics/87 |
| Product Code: | HRS/HTN/HWC |
#### Device Description
The subject EVOS Mini-Fragment Plating System is a modification to a subset of the devices previously cleared Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (K132886). It is an internal fixation plating system comprised of assorted implantable, small utility "flat" locking bone plates and compatible locking and non-locking bone screws and washers to be used on various small bones and fragments. Plates consist of groups of devices with a flat cross-section where the number of holes in plates will range from 4 through 20. All described implant devices are manufactured from implant-grade 316L Stainless Steel material and will be available in a sterile or non-sterile packaged condition.
#### Intended Use
Bone plates and bone screws from the EVOS Mini-Fragment Plating System are intended for use in internal fixation of small bones and fragments and as non-load bearing stabilization and fixation of bone fragments.
## Indications for Use
The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The Smith & Nephew Variable-Angle Locking Mini
{1}------------------------------------------------
Fragment Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and fragments. This system is also indicated for nonload bearing stabilization and reduction of bone fragments in long bones.
Components in the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System are for single use only.
## Technological Characteristics
Device comparisons described in this premarket notification demonstrate that the proposed bone plates and bone screws are substantially equivalent to the legally marketed predicate devices (listed below in Toble 1) with regard to intended use, indications for use, and performance characteristics.
#### Substantial Equivalence Information
When compared to the predicate devices listed below, substantial equivalence is based on similar function, intended use, indications for use, and overall design to the devices listed in the table below.
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|----------------------|---------------------------------------------------------|----------------------|----------------|
| Smith & Nephew, Inc. | Variable-Angle Locking Mini-<br>Fragment Plating System | K132886 | 02-04-2014 |
| Smith & Nephew, Inc. | Plate and Screw Instruments | K123055 | 01-25-2013 |
| Smith & Nephew, Inc. | VLP FOOT Plating, Screw<br>System and Accessories | K090675 | 06-04-2009 |
#### Table 1: Substantially Equivalent Predicates to the EVOS Mini-Fragment Plating System
#### Summary of Pre-Clinical Testing
To further support a determination of substantial equivalence, non-clinical bench (mechanical) testing was conducted to support the proposed bone plates and bone screws. Test results demonstrated that the proposed devices are substantially equivalent to one or more of the previously cleared predicate devices described above. The specific types of non-clinical testing conducted are listed below.
- . Finite Element Analysis (FEA) was conducted on all proposed bone plates to predict the worst-case plate to be used in subsequent non-clinical bench (mechanical) testing.
- Four-point bend fatigue testing was conducted on the worst-case bone plates . identified through FEA. Results of testing concluded that, when compared to
{2}------------------------------------------------
predicates tested at the same bending moment, the number of log cycles to failure met the acceptance criteria in that they were found to be non-inferior to the number of log cycles to failure achieved by the predicates.
- Four-point bend fatigue testing was also performed on the proposed 2.0mm, 2.4mm, . and 2.7mm bone screws. Results of the testing concluded that, compared to the predicates and when tested at the same bending moment, the proposed bone screws met the acceptance criteria in that they were found to be non-inferior to the number of log cycles to failure achieved by the predicates.
- Torque-to-failure testing of the proposed 2.0mm, 2.4mm, and 2.7mm cortex screws . and 4.0mm osteopenia screws was conducted. Results concluded that the average values for torsional strength of the proposed devices met the acceptance criteria in that they were non-inferior to the predicates and non-inferior to the minimum values cited in ASTM F543.
### Conclusion
As previously noted, this Special 510(k) Premarket Notification is being submitted to request clearance for the EVOS Mini-Fragment Plating System. Based on the similarities to the predicate components and a review of the mechanical testing performed, the devices are substantially equivalent to the predicate plating systems listed above in Table 1.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus, a traditional symbol of medicine, with three wavy lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 7, 2014
Smith & Nephew, Incorporated Ms. Samantha Staubach Regulatory Affairs Specialist I 1450 Brooks Road Memphis, Tennessee 38116
Re: K140814
Trade/Device Name: EVOS Mini-Fragment Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HTN, HWC Dated: April 10, 2014 Received: April 11, 2014
Dear Ms. Samantha Staubach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Samantha Staubach
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Premarket Notification Indications for Use Statement
K140814 510(k) Number (if known): _
Device Name: EVOS Mini-Fragment Plating System
Indications for Use:
The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and fragments. This system is also indicated for non-load bearing stabilization and reduction of bone fragments in long bones.
Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System are for single use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
**Casey Hanley -S**
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K140814
Page 1 of 1
4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.