← Product Code [LLZ](/productcode/LLZ) · K140797

# IPACS MEDICAL IMAGE MANAGEMENT SYSTEM (K140797)

_Visbion Limited · LLZ · Dec 12, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140797

## Device Facts

- **Applicant:** Visbion Limited
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Dec 12, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

IPACS Medical Image Management Software is used for viewing medical images. IPACS Medical Image Management Software receives, stores and archives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). IPACS Medical Image Management Software can be integrated with an institutions existing Hospital Information System (HIS) or Radiology Information System (RIS) for fully integrated electronic patient records. IPACS Medical Image Management Software can be used to retrieve, process and display medical images and patient information from any connected workstation. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in pre-operative planning and annotations and measurements to aid in diagnosis.

## Device Story

IPACS is a vendor-neutral PACS software platform for enterprise-wide medical image management. It ingests digital images from diverse modalities (CT, MR, US, RF, CR/DR) via DICOM, HL7, and IHE standards. The system archives data and provides a web-based interface for clinicians, radiologists, and technologists to retrieve, view, and process images on standard workstations. Users perform interactive image adjustments (positioning, sizing, rotation, contrast/brightness) and utilize measurement/annotation tools and pre-operative planning templates to assist in diagnosis. The device functions as a clinical repository, integrating with existing HIS/RIS systems. It does not contact patients or control life-sustaining equipment; a qualified physician interprets the displayed output to inform clinical decision-making. Benefits include centralized access to diagnostic images and patient records across hospital departments and remote locations.

## Clinical Evidence

No clinical data provided. Bench testing only.

## Technological Characteristics

Stand-alone software package comprising four elements: Image Archive, Image Viewer, Image Web, and Image Importer. Operates on standard PC hardware platforms. Connectivity via DICOM 3.0, HL7, and IHE standards. Supports JPEG compression. No patient-contacting components. Software-based image processing for display and measurement.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- eFilm Workstation with Modules ([K020995](/device/K020995.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 12, 2014

Visbion Limited % Mr. Thomas Falcon Regulatory Manager Visbion House, Gogmore Lane Chertsey, Surrey KP16 9AP UNITED KINGDOM

Re: K140797

Trade/Device Name: IPACS Medical Image Management Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 13, 2014 Received: November 17, 2014

Dear Mr. Falcon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K140797

### Device Name

IPACS (Integrated Picture Archiving and Communications System) Medical Image Management Software.

### Indications for Use (Describe)

IPACS Medical Image Management Software is used for viewing medical images.

IPACS Medical Image Management Software receives, stores and archives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, scanners, imaging gateways or imaging sources). IPACS Medical Image Management Software can be integrated with an institutions existing Hospital Information System (HIS) or Radiology Information System (RIS) for fully integrated electronic patient records.

IPACS Medical Image Management Software can be used to retrieve, process and display medical images and patient information from any connected workstation. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on aid in pre-operative planning and annotations and measurements to aid in diagnosis.

Typical users of IPACS Medical Imagement Software are trained medical professionals, including but not limited to radiologists, technologists and clinicians.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Visbion Limited IPACS Medical Image Management Software 510(k) K140797 Response to Al e-mail December 3, 2014

# Revised 510(k) Summary of Safety and Effectiveness

# (in accordance with 21 CFR 807.87(h) and 21 CFR 807.92)

#### 1. Submitter's name and address:

Visbion Ltd Visbion House Gogmore Lane Chertsey, Surrey KT16 9AP United Kingdom

FDA Establishment Registration No. - To be applied for after marketing clearance is given.

#### 2. Submitter's telephone number and fax number: Tel: 011 44 870 850 3486 Fax: 011 44 870 850 3487

#### 3. Contact person: Mr. Tom Falcon - Requlatory Manager and FDA Official Correspondent

- Date this 510(k) summary prepared: ব December 12, 2014
- 5. Trade/proprietary name of the device: IPACS Medical Image Management Software
- 6. Classification name and number of the device: FDA Class - II FDA Classification Name - Picture Archiving and Communications System (PACS) FDA Regulation Number - 21 CFR 892.2050
- 7. Legally marketed predicate device to which substantial equivalence is claimed:

eFilm Workstation with Modules: FDA 510(k) No: K020995 Clearance to market this device was given by FDA on April 12, 2002 FDA Device Classification: Class 2 FDA Regulation Number: 21 CFR 892.2050 FDA Product Code: LLZ

#### 8. Description of the device that is the subject of this premarket notification:

The Visbion IPACS (Integrated Picture Archiving and Communications System) Medical Image Management Software is a vendor-neutral PACS (Picture Archiving and Communication System) solution delivering enterprise-wide diagnostic images and patient reports, available for review at any time, from any location using a standard web browser.

IPACS is a single healthcare platform that allows the seamless acquisition of images from multiple types of imaging devices from a range of different equipment suppliers. It

> Attachment 2 Page 1 of 4

Image /page/3/Picture/17 description: The image shows the logo for Visbion. The logo consists of the word "visbion" in lowercase letters, with a blue curved line above the word. The word "visbion" is in a light gray color. The logo is simple and modern.

{4}------------------------------------------------

### Visbion Limited IPACS Medical Image Management Software 510(k) K140797 Response to Al e-mail December 3, 2014

is a true clinical repository with the ability to harmonise all departments within a hospital. The approach to connectivity is based on three key standards; Digital Imaging and Communications in Medicine (DICOM) 3.0, Health Level Seven International (HL7) and Integrating the Healthcare Enterprise (IHE).

Using a standard web browser registered users have the ability to log in thus provide secure access to images from diagnostic workstation in hospitals and consulting rooms plus from off-site storage facilities. 24 hours a day.

The device is constructed from the four software elements. These elements are combined according to the configuration required by the user. These terms are used by the user, IPACs being a general term to consolidate the following four elements:

Image Archive lmage Viewer Image Web Image Importer

#### 9. Intended use and indication for use:

IPACS Medical Image Management Software is used for viewing medical images.

IPACS Medical Image Management Software receives, stores and archives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). IPACS Medical Image Management Software can be integrated with an institutions existing Hospital Information System (HIS) or Radiology Information System (RIS) for fully integrated electronic patient records.

IPACS Medical Image Management Software can be used to retrieve, process and display medical images and patient information from any connected workstation. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in pre-operative planning and annotations and measurements to aid in diagnosis.

Typical users of IPACS Medical Image Management Software are trained medical professionals, including but not limited to radiologists, technologists and clinicians.

#### 10. Technoloqical characteristics:

IPACS is a stand-alone software package comprising four components that can be installed on a number of different hardware platforms, providing that the minimum hardware specification requirements are met. The system can transmit images to remote viewing workstations over a medical imaging network. None of the software components makes contact with the patient, nor do they control any life-sustaining devices. A qualified physician will review and interpret the images and information displayed in order to make clinical decisions.

The following comparison of the technological characteristics of the IPACS Medical Image Management Software and the predicate device, eFilm Workstation with Modules, illustrates that the characteristics are the same.

> Attachment 2 Page 2 of 4

Image /page/4/Picture/14 description: The image shows the word "visbion" in a sans-serif font, with a curved line above it. The word is in lowercase letters and is a light gray color. The curved line above the word is a dark blue color. The logo is simple and modern.

{5}------------------------------------------------

# Visbion Limited IPACS Medical Image Management Software - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

# Comparison of Technological Characteristics:

| Ref.<br>No. | Technological<br>Characteristic                                                | Candidate Device<br>IPACS Medical Image<br>Management Software                                                                                                                                                                                                                | Predicate Device<br>eFilm Workstation<br>with Modules                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1           | Type of Device                                                                 | Software only product<br>residing on a PC<br>Workstation or server.                                                                                                                                                                                                           | Software only product<br>residing on a PC<br>Workstation or server.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 2           | Type and Source<br>of Images                                                   | Digital images received<br>from various modalities<br>and digitised images via a<br>network or the internet.                                                                                                                                                                  | Digital images received from<br>various modalities and<br>digitised images via a<br>network or the internet.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| 3           | Scope of Image<br>Processing                                                   | Displays files received as<br>images from any third<br>party modality where the<br>third party manufacturer<br>has provided appropriate<br>DICOM drivers                                                                                                                      | Displays files received as<br>images from any third party<br>modality where the third<br>party manufacturer has<br>provided appropriate<br>DICOM drivers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 4           | Display of an<br>Image                                                         | On a visual display unit. A<br>hardcopy can be sent to a<br>suitable third party laser<br>printer.                                                                                                                                                                            | On a visual display unit. A<br>hardcopy can be sent to a<br>suitable third party laser<br>printer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| 5           | Processing of an<br>Image                                                      | Permits interactive<br>positioning of an image.<br>Permits interactive sizing<br>of an image (length, angle,<br>area etc).<br>Permits interactive rotation<br>of an image.<br>Permits interactive<br>adjustment of contrast and<br>brightness of an image.<br>Not applicable. | Permits interactive<br>positioning of an image.<br>Permits interactive sizing of<br>an image (length, angle,<br>area etc).<br>Permits interactive rotation<br>of an image.<br>Permits interactive<br>adjustment of contrast and<br>brightness of an image.<br>Permits mechanical linking<br>of components associated<br>with prosthetic templates<br>and trauma templates<br>(fixation device) - via third<br>party accessories - see<br>below.<br>Permits integration of third<br>party accessories such as<br>orthopaedic templating and<br>trauma processing.<br>Permits interactive display<br>or three dimensional<br>images.<br>Permits interactive<br>rendering on an image.<br>Supports Hanging<br>Protocols. |
| Ref.<br>No. | Feature                                                                        | Candidate Device<br>IPACS Medical Image<br>Management Software                                                                                                                                                                                                                | Predicate Device<br>eFilm Workstation<br>with Modules                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 6           | Means of<br>Collecting Data                                                    | From pre-obtained digital<br>images.                                                                                                                                                                                                                                          | From pre-obtained digital<br>images.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 7           | Processing of<br>Data                                                          | The software processes<br>data and provides an<br>indication of the extent of<br>quality loss. Images are<br>stored lossless.                                                                                                                                                 | The software processes<br>data and provides an<br>indication of the extent of<br>quality loss. Images are<br>stored lossless.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| 8           | Standards<br>Compliance                                                        | Digital Imaging and<br>Communications in<br>Medicine (DICOM) 3.0<br>Joint<br>Photographic<br>Experts<br>Group (JPEG)<br>Standard<br>Health Level Seven<br>International (HL7)<br>Integrating the Healthcare<br>Enterprise (IHE)                                               | Digital Imaging and<br>Communications in<br>Medicine (DICOM) 3.0<br>Joint Photographic Experts<br>Group (JPEG) Standard<br>Health Level Seven<br>International (HL7)<br>Integrating the Healthcare<br>Enterprise (IHE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 9           | Stand-alone<br>Software                                                        | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 10          | Can be used on<br>Multiple Hardware<br>Platforms                               | Yes - provided that stated<br>minimum hardware<br>requirements are met.                                                                                                                                                                                                       | Yes - provided that stated<br>minimum hardware<br>requirements are met.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 11          | Transmit Images<br>to Remote<br>Viewing Stations<br>Over an Imaging<br>Network | Yes                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

Image /page/5/Picture/4 description: The image shows the logo for Visbion. The logo consists of the word "visbion" in lowercase, sans-serif font, with a curved line above the word. The word and the curved line are in a light gray color. The curved line is slightly thicker in the middle and tapers off towards the ends.

{6}------------------------------------------------

## Visbion Limited IPACS Medical Image Management Software 510(k) K140797 Response to Al e-mail December 3, 2014

From the above information it is concluded that the IPACS Medical Image Management Software is substantially equivalent to the eFilm Workstation with Modules predicate device.

This concludes the 510(k) Summary.

Image /page/6/Picture/4 description: The image shows the logo for Visbion. The logo consists of the word "visbion" in lowercase letters, with a blue curved line above the word. The word "visbion" is in a light gray color. The logo is simple and modern.

---

**Source:** [https://fda-staging.innolitics.com/device/K140797](https://fda-staging.innolitics.com/device/K140797)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
