Published medical literature (CONFIRM study by Narayan et al., 2012)
The sponsor utilized data from the published CONFIRM study to support the safety of the Constellation Catheter when used in left atrial electrophysiology procedures.
Left atrial procedures; Safety data; Published literature
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
CONFIRM (Conventional Ablation With or Without FIRM) study; Retrospective analysis of clinical outcomes; Study Period: 2012 (publication year)
Patients undergoing left atrial mapping; Sample Size: 92 patients
Not applicable for this study
Safety (incidence of strokes, TIAs, and unanticipated complications)
Indications for Use
For use in right and left atrial electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The Constellation Multiple Electrode Recording and Pacing Catheter System may also be used for delivery of externally generated pacing stimuli.
Device Story
Constellation Catheter is a sterile, single-use diagnostic device for cardiac electrophysiology. Distal expandable 'basket' assembly features 32 or 64 electrodes on eight resilient splines; available in 31-75mm sizes. Device is inserted into the heart to detect endocardial electrical potentials and deliver pacing stimuli. Used in clinical electrophysiology labs by physicians. Output consists of electrical signals used for mapping complex arrhythmias; assists clinicians in identifying cardiac electrical activity. Benefits include improved diagnostic capability for complex arrhythmias compared to conventional linear catheters.
Clinical Evidence
Safety and effectiveness for left atrial procedures supported by cumulative animal and human data. Human data includes original IDE study (G940612) and published literature, specifically the CONFIRM study (Narayan et al., 2012). CONFIRM study mapped the left atrium in 92 patients using the device; reported no strokes, TIAs, or unanticipated complications related to catheterization or ablation.
Technological Characteristics
Sterile, single-use intracardiac mapping catheter. Features expandable basket assembly with 32 or 64 electrodes on eight splines. Available in 31, 38, 48, 60, and 75mm sizes. Configurations include unipolar and bipolar arrays. Principle of operation involves endocardial electrical potential detection and pacing stimulus delivery. Identical materials and design to predicate Constellation devices.
Indications for Use
Indicated for patients undergoing right and left atrial electrophysiology procedures to assist in diagnosing complex arrhythmias difficult to identify with conventional linear mapping catheters; also indicated for delivery of externally generated pacing stimuli.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
.
Traditional Premarket Notification Constellation® Multitude Electrode Recording and Pacing Catheter
#### 510(k) Summary per 21 CFR §807.92 (c) Section 5
| Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>Electrophysiology<br>150 Baytech Drive<br>San Jose, CA 95134 |
|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name<br>and Information | Michelle Roeding<br>Principal Regulatory Affairs Specialist<br>Tel: (408) 935-4912<br>Fax: (408) 957-6202<br>E-mail: Michelle.Roeding@bsci.com |
| Alternate<br>Contact and<br>Information | Lisa Scott<br>Director, Regulatory Affairs<br>Tel: (408) 935-6382<br>Fax: (408) 957-6202<br>E-mail: Lisa.Scott@bsci.com |
| Date Prepared | March 21, 2014 |
| Trade Name | Constellation® Multiple Electrode Recording and Pacing<br>Catheter |
| Common Name | Catheter, Intracardiac, High Density Array |
| Classification<br>Name | Catheter, Intracardiac, High Density Array (Product Code<br>MTD) has been classified as Class II per 21 CFR 870.1220 |
| Predicate<br>Device | The Constellation Catheter Multiple Electrode Recording and<br>Pacing Catheter is substantially equivalent in design and<br>intended use to the same device legally marketed under<br>K983171, K992777, K000277, K003782 and K021232 and<br>Johnson & Johnson/Biosense-Webster Flower High-Density<br>Mapping Catheter/ PENTARAY™ High-Density Mapping<br>Catheter legally marketed under K050217. |
| Description of<br>Device | The Boston Scientific Constellation® Multiple Electrode<br>Recording and Pacing Catheter (Constellation Catheter) is a<br>sterile, single use advanced heart mapping diagnostic device<br>designed to detect electrical potentials from the endocardial<br>surfaces of the heart and may also be used to deliver<br>externally generated pacing stimuli. The distal expandable<br>'basket' assembly contains an array of 32 or 64 electrodes<br>mounted along eight resilient support structures called<br>'splines.' Several configurations are available, including<br>unipolar (electrodes evenly spaced), bipolar (electrodes<br>evenly distributed into pairs), and lower density arrays. The<br>product is available in 31, 38, 48, 60, and 75mm basket<br>sizes. |
| Intended<br>Use/Indications<br>for Use | For use in right and left atrial electrophysiology procedures to<br>assist in the diagnosis of complex arrhythmias that may be<br>difficult to identify using conventional mapping systems alone<br>(i.e., linear mapping catheters). The Constellation Multiple<br>Electrode Recording and Pacing Catheter System may also<br>be used for delivery of externally generated pacing stimuli. |
| Device<br>Technological<br>Characteristics<br>and<br>Comparison to<br>Predicate<br>Device | The Constellation Catheter indicated for use in left atrial<br>electrophysiology procedures is the same device indicated<br>for use in right atrial procedures. Design, principle of<br>operation, and materials are identical to the Constellation<br>Catheter cleared under K983171, K992777, K000277,<br>K003782 and K021232 and equivalent to the Johnson &<br>Johnson/Biosense-Webster Flower High-Density Mapping<br>Catheter/ PENTARAY™ High-Density Mapping Catheter<br>legally marketed under K050217. |
| | Expanding the indication of the Constellation Catheter does<br>not involve incorporation of new design features, which are<br>identical to the predicate Constellation Catheter. In support<br>of substantial equivalence, Boston Scientific has compared<br>and evaluated technological characteristics of the Johnson &<br>Johnson/ Biosense-Webster Flower/PENTARAY High-<br>Density Mapping Catheter to the Constellation Catheter. The<br>Constellation Catheter is unique only in its geometrical<br>configuration of splines and electrodes. Otherwise, the<br>devices are similar in all other technical aspects. |
| Non-Clinical<br>Performance<br>Data | Device design specifications are identical for both the<br>predicate and subject Constellation Catheter. Therefore the<br>non-clinical testing for the Constellation Catheter that was<br>submitted in previous submissions (K983171, K992777,<br>K000277, K003782, and K021232) remains relevant. |
| Clinical<br>Performance<br>Data | Safety and effectiveness of the Constellation Catheter with<br>expanded indications for use to include the left atrial<br>electrophysiology procedures is based on cumulative animal<br>and human data. Human clinical data include safety and<br>effectiveness data from the original Constellation Catheter<br>IDE (G940612) and safety data for use in the left atrium as<br>evidenced from published medical literature, and in<br>particular, from the CONFIRM [Conventional Ablation With or<br>Without FIRM (Focal Impulse and Rotor Modulation)] study<br>reported by Narayan et al (2012). In the CONFIRM study, the<br>Constellation Catheter was advanced via an 8.5 F sheath to<br>map the left atrium in 92 patients. There were no strokes or<br>TIAs (transient ischemic attack) reported nor any<br>unanticipated complications related to cardiac catheterization<br>and/or ablation. |
| Conclusion | The Constellation Catheter with expanded indications for use<br>to include left atrial electrophysiology procedures is<br>substantially equivalent to the predicate Constellation<br>Catheter (K021232) and the Flower/PENTARAY High<br>Density Mapping Catheter (K050217) based on comparison<br>of the devices. The summation of the existing animal data,<br>data from the IDE study (G940612) and the CONFIRM study<br>(Narayan et al, 2012), along with the additional medical<br>literature, support continued safety and effectiveness of the<br>Constellation Catheter as it relates to expanding the current<br>indications for use. |
{1}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.
{2}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
and the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of the co
-
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA". Inside the circle is a stylized eagle or bird-like symbol with three curved lines representing its wings or body.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
April 28, 2014
Boston Scientific Corporation Michelle Roeding Principal, Regulatory Affairs 150 Baytech Drive San Jose. CA 95134
Re: K140733
Trade/Device Name; Constellation Multiple Electrode Recording and Pacing Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode recording catheter or electrode recording probe Regulatory Class: Class II Product Code: MTD Dated: March 25, 2014 Received: March 26, 2014
Dear Ms. Michelle Roeding:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Michelle Roeding
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
.........
### 510(k) Number (if known)
K140733
### Device Name
Constellation® Multiple Electrode Recording and Pacing Catheter
- - -
Indications for Use (Describe)
For use in right and left atrial electrophysiology procedures to assist in the diagnosis of complex arrhytmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The Constellation Multiple Electrode Recording and Pacing Catheter System may also be used for delivery of externally generated pacing stimuli.
Type of Use (Selact one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR HOLD HOW HO HOL
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/5/Picture/15 description: The image shows a logo with the letters FDA. Next to the logo is the word "Date:" followed by the date "2014.04.28". Below the date is the time "20:35:56 -04'00'".
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden lime for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda:hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number.
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.