K140723 · Aribex, Inc. · IZL · Aug 19, 2014 · Radiology
Device Facts
Record ID
K140723
Device Name
NOMAD MD 75K V HANDHELD X-RAY SYSTEM
Applicant
Aribex, Inc.
Product Code
IZL · Radiology
Decision Date
Aug 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The NOMAD MD is a handheld and portable general purpose X-ray system. The device uses a fixed tube current and voltage (kVp) and, therefore, is limited to taking diagnostic X-rays of extremities. It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.
Device Story
Handheld, portable general-purpose X-ray system; operates on 14.4V DC rechargeable NiCd battery. Integrates X-ray tubehead, controls, and power source in single enclosure. Features fixed 75 kVp and 2.0 mA output; utilizes two-stage triggering mechanism and Class II laser for distance indication. Operated by clinicians in clinical settings for extremity imaging. Includes internal bismuth-blend shielding to minimize operator leakage radiation. Emits audible/visual signals for exposure status and battery levels. Requires operator use of lead apron/thyroid collar due to lack of external backscatter shield. Output images used by clinicians for diagnostic assessment of extremities (fingers to thighs). Benefits include portability for point-of-care extremity radiography.
Clinical Evidence
No clinical data conducted. Evidence consists of bench testing and phantom image comparison. Radiographic images of foot, knee, elbow, and hand phantoms were acquired using both subject and predicate devices and reviewed by a board-certified radiologist, confirming diagnostic quality equivalence. Electrical safety and EMC testing performed per IEC 60601 series.
Technological Characteristics
Handheld X-ray system; 14.4V DC NiCd battery powered. Fixed 75 kVp, 2.0 mA. 0.4 mm focal spot. Internal bismuth-blend shielding. Class II laser distance indicator. LED light field center indicator. Tactile membrane interface. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54, IEC 60825, IEC 62366, IEC 60601-1-2, 21 CFR 1020.30/31.
Indications for Use
Indicated for diagnostic X-ray imaging of extremities in adult and pediatric patients by qualified, trained clinicians.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a wing-like shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 19, 2014
Aribex, Inc. % Mr. Sanjay Ahuja Director, Regulatory Affairs 744 South 400 East OREM UT 84097
Re: K140723 Trade/Device Name: Nomad MD Handheld X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: June 5, 2014 Received: July 21, 2014
Dear Mr. Ahuja:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh. 7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140723
Device Name NOMAD MD Handheld X-ray System
#### Indications for Use (Describe)
The NOMAD MD is a handheld and portable general purpose X-ray system. The device uses a fixed tube current and voltage (LVp) and, therefore, is limited to taking diagnostic X-rays of extremities.
It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# NOMAD MD Handheld X-ray System
#### Submitter:
Aribex, Inc. 744 South 400 East Orem, UT 84097 Phone No: 801-226-5522 FAX No:
## Date of Summary:
March 14, 2014
## Contact Person:
Christopher Dodge, Senior Regulatory Affairs Specialist or Sanjay Ahuja, Ph.D., Director, Regulatory Affairs
#### Device Name:
●
- Trade Name: NOMAD MD Handheld X-ray System ●
- . Common Name:
- Nomad MD
- Classification Name: Mobile X-ray system 21CFR 892.1720
- Regulation:
- Product Code: . IZL
# Predicate Device for Which Substantial Equivalence is Claimed:
MinXray HF 100H+ Portable X-ray Unit (K052721)
#### Device Description:
The NOMAD MD is a handheld, general purpose X-ray system that operates on 14.4 V DC supplied by a rechargeable NiCd battery pack. The X-ray tubehead, X-ray controls, and power source are assembled into a single hand-held enclosure. The package includes a spare battery, a Charging Cradle, and an AC-to-DC Power Supply.
Consideration was given to three FDA guidance documents, "Pediatric Information for X-ray Imaging Device Premarket Notifications" issued on May 10,
{5}------------------------------------------------
2012, "Radiation Safety Considerations for X-Ray Equipment for Hand-Held Use," issued on December 24, 2008, and "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices," dated August 6,1999. Aribex determined that the solid state quidance did not apply to the NOMAD MD because it is an X-ray emitter and the quidance document is written for X-ray image recording systems, not for X-ray emitters.
#### Indications for Use/Intended Use:
The NOMAD MD is a handheld and portable general purpose X-ray system. The device uses a fixed tube current and voltage (kVp) and, therefore, is limited to taking diagnostic X-rays of extremities.
It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp) which may be required for full optimization of image quality and radiation exposure for different exam types.
#### Summary of Technological Characteristics:
The NOMAD MD has been demonstrated to be substantially equivalent to the predicate device, the MinXray HF 100H+ Portable X-ray Unit (cleared under K052721). The proposed device has been verified and validated to satisfy the requirements derived from the indications for use and the substantial equivalence table.
NOMAD MD shares many of the same components and features as the predicate device including:
- Portable X-ray source
- . Collimation of X-ray
- Two stage triggering mechanism .
- Source to skin guard to limit distance between X-ray source and patient .
- Constant potential waveform
- Similar user interface with tactile membrane pad with up-down push ● buttons for making selections
- . Similar alerts and alarms that indicate equipment state of readiness to use
The MinXray operator instructions require that the operator "stand as far as possible from the X-ray unit while holding the exposure switch and pressing the button" in order to avoid being exposed to excess leakage and backscatter radiation.
The following is a comparison between Nomad MD (proposed) and the predicate device, the MinXray HF 100H+ Portable X-ray Unit (cleared under K052721).
{6}------------------------------------------------
| | MinXray HF100H+<br>(K052721) | Nomad MD<br>X-ray System (K140723) |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | | |
| Regulation No. | 21 CFR 892.1720 | 21 CFR 892.1720 |
| Regulation Class | II | II |
| Product Code | IZL | IZL |
| Indications for<br>Use/Intended Use: | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic x-rays. | The NOMAD MD is a<br>handheld and portable<br>general purpose X-ray<br>system. The device uses a<br>fixed tube current and<br>voltage (kVp) and,<br>therefore, is limited to<br>taking diagnostic X-rays of<br>extremities. |
| | | It is intended to be used by<br>a qualified and trained<br>clinician on both adult and<br>pediatric patients. It is not<br>intended to replace a<br>radiographic system with<br>variable tube current and<br>voltage (kVp) which may be<br>required for full optimization<br>of image quality and<br>radiation exposure for<br>different exam types. |
| Principle of Operation | General Purpose Diagnostic X-<br>Ray | General Purpose<br>Diagnostic X-Ray |
| TECHNOLOGICAL: | | |
| Size: Body | 8.75"Hx9.5"Wx15.35"L<br>excluding skin guards | 9.5"H x 5.25"Wx10"L<br>(excluding Source Skin<br>guard) |
| Weight | 40.9 lbs. | 11.0 lbs. |
| Source to skin distance | 30 cm | 30 cm |
| Focal Spot | 1.2 mm | 0.4 mm |
| Collimator | Four manually and steplessly<br>adjustable shutters with light<br>beam type central x-ray<br>indicator (Advantech R72) | Four manually and<br>steplessly adjustable<br>shutters with LED Light<br>Field Center Indicator |
| Triggering Mechanism | Two stage triggering | Two stage triggering |
| User Interface | Up-down push buttons for kVp<br>selections and exposure time<br>selections with LED indicators<br>and mAs indicators. | Up-down buttons for<br>exposure time selection,<br>with timer display. |
| Energy Source | 100-140 VAC 50/60 Hz | Rechargeable 14.4 V DC<br>NiCd battery pack |
| Exposure Time | 0.03-0.2 sec (in 0.01 sec Steps)<br>0.2-0.4 sec (in 0.02 sec Steps)<br>0.4-1.0 sec (in 0.05 sec Steps) | 0.02 - 0.99 seconds in 0.01<br>increments |
| Feature | MinXray HF100H+<br>(K052721) | Nomad MD<br>X-ray System (K140723) |
| | 1.0-4.0 sec (in 0.1 sec Steps) | |
| mA | 30 mA @ 40-60 kV<br>25 mA @ 62-80 kV<br>20 mA @ 82-100 kV | 2.0 mA fixed |
| kVp | 40-100kVp | 75 kVp fixed |
| Waveform | Constant Potential | Constant Potential |
| Alerts and Alarms | As least two specific alerts and<br>alarms indicate equipment state<br>of readiness to use | Seven specific alerts and<br>alarms indicating<br>equipment's state of<br>readiness to use |
| Class II Laser | Not a feature on this unit | Used to indicate the<br>distance from the x-ray<br>source to the target to be<br>diagnosed. |
| User Interface Ease | Tactile membrane pad with<br>audible signals indicates<br>settings.<br>Large illuminated digital<br>exposure setting screen. | Tactile membrane pad with<br>audible signals indicates<br>settings.<br>Large illuminated digital<br>exposure setting screen. |
| Safety | Exposure switch button<br>illuminates if unit misfires. The<br>unit will lock up and needs to<br>be restarted from the off<br>position<br>During the exposure an audible<br>signal and the X-ray indicator<br>illuminates. | Locked upon powering on<br>with LC indicator, along with<br>elongated audible signal.<br>Unit will power down after<br>2.5 minutes of inactivity.<br>After 10 seconds of being<br>enabled, the unit will lock if<br>exposure does not occur.<br>Continuous audible signal is<br>emitted along with indicator<br>lights indicating that<br>exposure is completed. |
| Low Battery Indicator<br>and Audible Signal | Not a feature on this unit<br>because it is AC/DC powered. | LED illuminates and flashes<br>when battery charge is<br>needs or needs recharging<br>Audible signal indicates low<br>battery signal |
| Electrical Safety<br>Standards | UL 2601, IEC60601 | IEC60601-1, EN60601-2 |
| Performance Standard | 21 CFR 1020.30 | 21 CFR 1020.30, 1020.31<br>IEC60601-3 |
{7}------------------------------------------------
Per the recommendations of the FDA Guidance document on "Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use," Aribex has incorporated internal shielding into the design of the NOMAD MD. This shielding, made from a proprietary bismuth blend, encapsulates the X-ray source assembly and prevents the operator of the device from being exposed to any appreciable amounts of leakage radiation. Test results indicate that the level of exposure is
{8}------------------------------------------------
well below the requirements of 21 CFR 1020.30(k). Evidence of the effectiveness of this internal shielding can be found in Section 19 NOMAD MD Performance Testing - Bench. Specifically, TP-0223, Test Case Scatter Radiation NOMAD MD, identifies the operator exposure dose within the significant zone of occupancy. See also QR-0200 Risk Analysis. NOMAD MD - Risk ID R009. R017.
Since the NOMAD MD, unlike its dental predecessor, is held on average 28 inches from the patient, it was not considered practical to incorporate an external backscatter shield into the design of the device. The effective size of the backscatter shield would need to be about three feet in diameter. For this reason, Aribex recommends in its operator manual (MP-0221, Section 3.3, Section 4.7[6], and Section 7.5) the use of a lead apron and thyroid collar when operating the NOMAD MD. Also see QR-0200 Risk Analysis, NOMAD MD - Risk ID R061.
As evidenced through Design Verification and Validation, NOMAD MD can produce diagnostic images of adult and pediatric extremities and they are of equivalent diagnostic quality as the MinXray device.
An extremity is generally defined as the distal or terminal portion of any bodily limb, or any portion of bodily limbs such as arms and legs. The NOMAD MD has been validated (see Section 19, TP-0302, Extremity Thickness Performance, NOMAD MD) to capture diagnostic quality images of body part extremities, such as fingers (from 1.5cm) and thighs (up to 21cm).
Based upon an analysis of the technological differences, and the substantiation through design verification and validation. Aribex determines that the safety and effectiveness of the proposed NOMAD MD is substantially equivalent to the MinXray predicate device and does not raise any new concerns.
#### Non-Clinical Test Data:
Performance bench testing was conducted as part of design control to ensure the safety and effectiveness of the NOMAD MD is equivalent to the predicate device. The safety and effectiveness of NOMAD MD for the proposed indications for use is substantiated through verification and validation testing on diagnostic images.
An image comparison study was also conducted by taking radiographic images of foot, knee, elbow, and hand phantoms using both the NOMAD MD and the predicate device. The images were then reviewed by a board-certified radiologist who determined that images from both devices were of diagnostic quality and, therefore, substantially equivalent.
Software documentation has been provided in Section 17 according to the FDA's requirement for software that meets the moderate level of concern.
{9}------------------------------------------------
EMC and Electrical Safety testing on the NOMAD MD was performed and found to meet all the requirements in standards IEC 60601-1: 2005, IEC 60601-1-3:2008, IEC 60601-1-6:2010, IEC 60601-2-28:2010, IEC 60601-2-54:2009, IEC 60825:2007, IEC 62366:2007, and IEC 60601-1-2:2007.
#### Clinical Test Data:
Clinical testing was deemed not required to support the indications for use and therefore, it has not been conducted on the Nomad MD proposed product.
## Conclusion:
Based on the comparison of the indications for use/intended use, the technological characteristics, and the operation of the device, Aribex concludes that NOMAD MD is substantially equivalent to the predicate device. In addition, based upon the review of the performance, standards analysis, and labeling, the differences between NOMAD MD device and the MinXray predicate device do not raise new concerns on safety and effectiveness for the proposed indications of use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.