K140703 · Ciden Technologies, LLC · MED · Aug 19, 2014 · General Hospital
Device Facts
Record ID
K140703
Device Name
OPA 30 SOLUTION
Applicant
Ciden Technologies, LLC
Product Code
MED · General Hospital
Decision Date
Aug 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6885
Device Class
Class 2
Indications for Use
OPA 30 Solution is a high level disinfectant for reprocessing heat-sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used or reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20°C (68°F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25°C), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an immersion time of at least 5 minutes at a minimum of 25°C (77°F), for a reuse period not to exceed 14 days.
Device Story
OPA 30 Solution is a liquid chemical high-level disinfectant containing 0.60% ortho-Phthalaldehyde (OPA) in an aqueous base with buffers, chelating agents, and corrosion inhibitors. It is used to reprocess heat-sensitive semi-critical medical devices. The solution is applied via manual immersion or in legally marketed automatic endoscope reprocessors. Users verify the Minimum Recommended Concentration (MRC) of 0.3% using Opaciden OPA Reagent Strips before use. The device is intended for clinical settings where medical instruments require disinfection. By effectively eliminating vegetative organisms, fungi, viruses, and spores, the solution ensures instrument safety for subsequent patient use, reducing the risk of cross-contamination.
Clinical Evidence
No clinical data included in this submission. Performance is supported by non-clinical bench testing for stability, reuse, and material compatibility.
Technological Characteristics
Clear, pale blue liquid, pH 7.5. Active ingredient: 0.60% ortho-Phthalaldehyde. Contains buffers, chelating agents, and corrosion inhibitors. Compatible with Olympus and Pentax endoscopes and Acrylonitrile Butadiene Styrene (ABS). Stability: 24-month shelf life. Manual reuse: 30 days; Automated reuse: 14 days. Verification via Opaciden OPA Reagent Strips.
Indications for Use
Indicated for high-level disinfection of heat-sensitive semi-critical medical devices where sterilization is not suitable. Used in manual reprocessing (min 12 min at 20°C) or automatic endoscope reprocessors (min 5 min at 25°C).
Regulatory Classification
Identification
A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.
Special Controls
*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
August 19, 2014
CIDEN Technologies, LLC c/o Courtney Clark, CBA, RAC Regulatory Affairs Manager of Submissions Sybron Dental Specialties 1717 W. Collins Ave Orange, CA 92867
Re: K140703
> Trade/Device Name: OPA 30 Solution Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid chemical sterilant/high level disinfectants Regulatory Class: II Product Code: MED Dated: July 22, 2014 Received: July 23, 2014
Dear Ms. Clark,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140703
Device Name OPA 30 Solution
### Indications for Use (Describe)
OPA 30 Solution is a high level disinfectant for reprocessing heat sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used of reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20°C (68°F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25°C), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an immersion time of at least 5 minutes at a minimum of 25℃ (77ºF), for a reuse period not to exceed 14 days.
### Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 5. 510(k) SUMMARY for OPA 30 Solution (K140703)
- 1. Submitter Information: Ciden Technologies, LLC 360 Cold Spring Avenue West Springfield, MA 01089
Contact Person: Courtney Clark Telephone Number: 714-516-7426 Fax Number: 714-516-7472
Date Prepared: 22 July 2014
2. Device Name:
| Proprietary Name: | OPA 30 Solution |
|------------------------|---------------------------------------------------|
| Classification Name: | Liquid chemical germicide/high level disinfectant |
| CFR Number: | 21 CFR 880.6885 |
| Device Class: | II |
| Product Code: | MED |
| FDA review panel code: | INCB |
#### 3. Predicate Device:
The legally marketed predicate device is Opaciden® Solution (K070627), Product Code MED, manufactured by Ciden Technologies LLC and determined to be substantially equivalent to a legally marketed device on August 2, 2007.
#### 4. Description of Device:
The OPA 30 Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.60% ortho-Phthalaldehyde in an aqueous base containing buffers, chelating agents and a corrosion inhibitor. It is stable at 15-30° C (59-86° F) for its labeled shelf life of 24 OPA 30 Solution may be used or reused for manual (up to 30 days) or months. automated reprocessing (up to 14 days), according to the Directions for Use. OPA 30 Solution must be used at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with the immersion time and temperature specified for the disinfection method.
- న. Statement of Intended Use:
OPA 30 Solution is a high level disinfectant for reprocessing heat-sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used or reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20° C (68° F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25℃), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an
{4}------------------------------------------------
immersion time of at least 5 minutes at a minimum of 25°C (7°P), for a reuse period not to exceed 14 days.
#### 6. Description of Safety and Substantial Equivalence:
## Technological Characteristics
The chemical formulation and technical characteristics of the proposed OPA 30 Solution are identical to those of the predicate Opaciden® Solution (K070627). The indications for use, material compatibility and shelf life have been expanded for the proposed OPA 30 Solution due to the results of additional performance testing. There are no substantial technical or functional differences between the proposed OPA 30 Solution and the predicate (K070627) device. Based on this information and the results of the performance testing, it is expected that the proposed OPA 30 Solution will be as safe and effective as the predicate Opaciden® Solution (K070627) when used as directed.
| ELEMENT | PREDICATE DEVICE: | PROPOSED DEVICE: |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Opaciden® Solution<br>(K070627) | OPA 30 Solution |
| CLASSIFICATION NAME<br>(CFR; Product code) | Liquid chemical germicide/high<br>level disinfectant (21 CFR §<br>880.6885, Product Code MED) | Liquid chemical germicide/high<br>level disinfectant (21 CFR §<br>880.6885, Product Code MED) |
| INDICATIONS FOR USE | Opaciden® Solution (K070627)<br>is a high level disinfectant for<br>reprocessing heat-sensitive semi-<br>critical medical devices for<br>which sterilization is not suitable,<br>and when used according to the<br>Directions for Use. Opaciden®<br>Solution may be used or reused<br>at or above its Minimum<br>Recommended Concentration<br>(MRC) of 0.3%, as determined<br>by Opaciden OPA Reagent<br>Strips, in manual reprocessing<br>with an immersion time of at<br>least 12 minutes at a minimum of<br>20°C (68°F), for a reuse period<br>not to exceed 14 days.<br>Opaciden® Solution may also<br>be used or reused in a legally<br>marketed automatic endoscope<br>reprocessor (that can be set to a<br>minimum 25°C), at or above its<br>Minimum Recommended<br>Concentration (MRC), as<br>determined by Opaciden OPA<br>Reagent Strips, with an<br>immersion time of at least 5<br>minutes at a minimum of 25°C<br>(77°F), for a reuse period not to<br>exceed 14 days. | OPA 30 Solution is a high level<br>disinfectant for reprocessing<br>heat-sensitive semi-critical<br>medical devices for which<br>sterilization is not suitable, and<br>when used according to the<br>Directions for Use. OPA 30<br>Solution may be used or reused<br>at or above its Minimum<br>Recommended Concentration<br>(MRC) of 0.3%, as determined<br>by Opaciden OPA Reagent<br>Strips, in manual reprocessing<br>with an immersion time of at<br>least 12 minutes at a minimum of<br>20°C (68°F), for a reuse period<br>not to exceed 30 days. OPA 30<br>Solution may also be used or<br>reused in a legally marketed<br>automatic endoscope reprocessor<br>(that can be set to a minimum of<br>25°C), at or above its Minimum<br>Recommended Concentration<br>(MRC), as determined by<br>Opaciden OPA Reagent Strips,<br>with an immersion time of at<br>least 5 minutes at a minimum of<br>25°C (77°F), for a reuse period<br>not to exceed 14 days. |
| ACTIVE INGREDIENT | ortho-Phthalaldehyde 0.60% | ortho-Phthalaldehyde 0.60% |
| ELEMENT | PREDICATE DEVICE:<br>Opaciden® Solution<br>(K070627) | PROPOSED DEVICE:<br>OPA 30 Solution |
| INERT INGREDIENTS | 99.40% | 99.40% |
| PHYSICAL<br>PROPERTIES | Clear, pale blue liquid with a pH<br>of 7.5. Contains 0.60% OPA in<br>an aqueous base containing<br>buffers, chelating agents and a<br>corrosion inhibitor. It is stable at<br>15-30° C (59-86° F) for its<br>labeled shelf life. | Clear, pale blue liquid with a pH<br>of 7.5. Contains 0.60% OPA in<br>an aqueous base containing<br>buffers, chelating agents and a<br>corrosion inhibitor. It is stable at<br>15-30° C (59-86° F) for its<br>labeled shelf life. |
| MINIMUM EFFECTIVE<br>CONCENTRATION (MEC) | 0.3% | 0.3% |
| REUSE PERIOD | 14 Days Manual, 14 days<br>automated | 30 Days Manual, 14 days<br>automated |
| MICROBIOLOGY | Effective against: | Effective against: |
| Vegetative organisms | Staphylococcus aureus,<br>Salmonella choleraesuis,<br>Pseudomonas aeruginosa,<br>Mycobacterium bovis | Staphylococcus aureus,<br>Salmonella choleraesuis,<br>Pseudomonas aeruginosa,<br>Mycobacterium bovis |
| Fungi | Trichophyton mentagrophytes | Trichophyton mentagrophytes |
| Viruses, non-enveloped | Poliovirus Type 1<br>Rhinovirus Type 42<br>Hepatitis A (AER) | Poliovirus Type 1<br>Rhinovirus Type 42<br>Hepatitis A (AER) |
| Viruses, enveloped | Avian influenza<br>Hepatitis B<br>Influenza Type A (Hong Kong)<br>HIV-1 | Avian influenza<br>Hepatitis B<br>Influenza Type A (Hong Kong)<br>HIV-1 |
| Spores | Herpes simplex Type 1<br>Clostridium sporogenes<br>Bacillus subtilis | Herpes simplex Type 1<br>Clostridium sporogenes<br>Bacillus subtilis |
| TOXICITY | Positive for acute oral toxicity,<br>moderate skin irritation.<br>Negative for dermal and<br>cytotoxicity. Probable eye<br>irritant. | Positive for acute oral toxicity,<br>moderate skin irritation.<br>Negative for dermal and<br>cytotoxicity. Probable eye<br>irritant. |
| REUSABLE DEVICE<br>COMPATIBILITY | Olympus and Pentax endoscopes<br>are compatible | Olympus and Pentax endoscopes<br>are compatible |
| CHEMICAL INDICATOR | Opaciden OPA Reagent Strips | Opaciden OPA Reagent Strips |
| SHELF LIFE | 15 month shelf life for a closed<br>container. | 24 month shelf life for a closed<br>container. |
| MATERIAL<br>COMPATIBILITY | Materials previously cleared in<br>the predicate Opaciden®<br>Solution<br>(K070627) | Materials previously cleared in<br>the predicate Opaciden®<br>Solution<br>(K070627)<br>and<br>Acrylonitrile Butadiene<br>Styrene (ABS) |
## Device Comparison Table
{5}------------------------------------------------
{6}------------------------------------------------
## Non-Clinical Performance Data
The OPA 30 Solution has been tested for stability, re-use for manual processing and material compatibility following:
- . Guidance for Industry and FDA Reviewers: "Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants" Section III
Due to the fact that the chemical formulation of the proposed OPA 30 Solution and the predicate Opaciden® Solution (K070627) is identical, additional testing requirements described in the FDA guidance "Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants" have not been repeated. The required testing was previously submitted, reviewed and cleared in K070627. Results for the applicable tests were acceptable and demonstrated performance in accordance with the proposed OPA 30 Solution product labeling.
## Clinical Performance Data
There is no clinical data included in this submission.
## Conclusion as to Substantial Equivalence
The OPA 30 Solution is as safe, as effective, and performs as well as the predicate Opaciden® Solution (K070627), and is therefore determined to be substantially equivalent to Opaciden® Solution (K070627).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.