ENDO VAC APICAL NEGATIVE PRESSURE IRIGATION SYSTEM
Applicant
Sybron Endo
Product Code
NYL · Dental
Decision Date
Jul 2, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The EndoVac® system is intended for the delivery and evacuation of endodontic irrigation solutions during root canal procedures.
Device Story
EndoVac system provides negative pressure irrigation for root canal therapy. Components include Multi-Port Adapter (connects to dental office Hi-Vac system), Master Delivery Tip (simultaneous irrigation/evacuation), MacroCannula (removes coarse debris), and MicroCannula (removes apical debris via 12 laser-drilled 100 um holes). Device operates by drawing fluid apically via vacuum suction, moving irrigants from pulp chamber to apical termination and into Hi-Vac system. Used in dental clinics by practitioners. Benefits include effective apical irrigation and reduced risk of irrigant overflow compared to positive pressure systems.
Clinical Evidence
No clinical data. Safety and efficacy supported by bench testing and review of published literature.
Technological Characteristics
Negative pressure irrigation system. Components: Multi-Port Adapter, Master Delivery Tip, MacroCannula, MicroCannula (12 laser-drilled 100 um holes). Materials comply with ISO 10993. Sterilization via moist heat (ISO 17665-1, ANSI/AAMI ST79). Reprocessing per AAMI TIR 12/39. Non-powered, vacuum-driven.
Indications for Use
Indicated for patients undergoing root canal procedures requiring delivery and evacuation of endodontic irrigation solutions.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 2, 2014
SybronEndo C/O Ms. Courtney Clark, CBA, RAC Regulatory Affairs Manager, Submissions Sybron Dental Specialties, Incorporated 1717 W. Collins Avenue Orange. California 92867
Re: K140685
Trade/Device Name: Endo Vac® Apical Negative Pressure Irrigation System Regulation Number: 21 CFR 872.4200 Regulation Name: Handpiece, Air-Powered, Root Canal Irrigation Regulatory Class: I Product Code: NYL Dated: May 30, 2014 Received: June 2, 2014
Dear Ms. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Clark
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# SECTION 4. INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K140685
Device Name: EndoVac® Apical Negative Pressure Irrigation System
Indications for Use:
The EndoVac® system is intended for the delivery and evacuation of endodontic irrigation solutions during root canal procedures.
Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sheena A. Green -S 2014.07.02 08:42:50 -04'00'
{3}------------------------------------------------
### SECTION 5. 510(k) SUMMARY
Image /page/3/Picture/1 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized, sans-serif font, with a curved line above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are written in a smaller, sans-serif font. The color of the text and the curved line is a dark blue.
#### EndoVac® Apical Negative Pressure Irrigation System
- Submitter Information: 1.
Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687
| Contact Person: | Courtney Clark |
|-------------------|----------------|
| Telephone Number: | 714-516-7426 |
| Fax Number: | 714-516-7472 |
| Date Prepared: | 17 March 2014 |
#### 2. Device Name:
.
- EndoVac® Negative Pressure Irrigation System Proprietary Name: ●
- Classification Name: Handpiece, air-powered, root canal irrigation
- Regulation Description: Dental handpiece and accessories ●
- CFR Number: 872.4200
- Device Class: Class I reserved ●
- . Product Code: NYL
- 3. Predicate Device:
EndoVac® Apical Negative Pressure Irrigation System is substantially equivalent to the legally marketed device RinsEndo, manufactured by Air Techniques, K052271 cleared on May 26, 2006, product code NYL.
- 4. Description of Device:
The EndoVac® System is intended for the delivery and evacuation of endodontic irrigation solutions during root canal procedures. EndoVac® Apical Negative Pressure Irrigation System is used in root canal therapy and is comprised of four main components. The Multi-Port Adapter is used to connect the Master Delivery Tip to the dental office Hi-Vac system. The Master Delivery Tip provides a constant flow of irrigant that lowers the risk of overflow due to continuous suction. Additional tips include the MacroCannula which is used to remove coarse debris after instrumentation, and the MicroCannula which features microscopic evacuation holes allowing the irrigant to flow to the apical termination.
{4}------------------------------------------------
- న్. Statement of Intended Use:
The EndoVac® System is intended for the delivery and evacuation of endodontic irrigation solutions during root canal procedures.
- 6. Description of Safety and Substantial Equivalence:
### Technological Characteristics
The EndoVac® System is intended for the delivery and evacuation of endodontic irrigation solutions during root canal procedures. EndoVac® Apical Negative Pressure Irrigation System presents a method to effectively irrigate during root canal treatments. Unlike positive pressure systems which use a cannula or side-port needles to deliver irrigants to the canal, the EndoVac® is a negative pressure system that draws fluid apically by way of vacuum suction. Because the system utilizes negative pressure irrigation, irrigants are suctioned from the pulp chamber, apically, to the apical termination and then away into the office Hi-Vac system. The EndoVac® Apical Negative Pressure Irrigation System is comprised of a Multi-Port Adapter which attaches to the dental Hi-Vac system, the Master Delivery Tip which allows for simultaneous irrigation and evacuation, the MacroCannula which removes coarse debris left in the canal from instrumentation, and the MicroCannula which removes microscopic debris at the apical 1 mm via 12 microscopic, laser-drilled holes, each 100 um in diameter. Other accessories include handpiece and finger-piece attachments, evacuation tubing and syringes.
EndoVac® Apical Negative Pressure Irrigation System is substantially equivalent to the legally marketed RinsEndo dental handpiece manufactured by Airtechniques, Inc. (K052271). The RinsEndo uses a turbine driven dental handpiece to provide hydrodynamic rinsing and flushing of root canals, but has the same intended use as the proposed EndoVac® Apical Negative Pressure Irrigation System. A comparison of the proposed and predicate devices is provided in Table 1.
| Device Name | Proposed EndoVac® System | Predicate RinsEndo (K052271) |
|---------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| FDA Product<br>Code | NYL; Handpiece, air-powered,<br>root canal irrigation | NYL; Handpiece, air-powered,<br>root canal irrigation |
| Classification | Class I reserved | Class I reserved |
| Manufacturer | SybronEndo<br>1332 South Lone Hill Avenue<br>Glendora CA, 91740 | Air Techniques, Incorporated<br>70 Cantiague Rock Road<br>Hicksville, New York 11801 |
| Method of<br>aspiration | Suction through tubing attached<br>to dental office Hi-Vac | Air-driven piston that creates<br>flushing action for rinsing root<br>canals; attached to dental office<br>Hi-Vac |
| Attached to dental<br>handpiece | No | Yes |
#### Table 1: Comparison of Proposed and Predicate Devices
{5}------------------------------------------------
| Indications for use | The EndoVac® system is | The RinsEndo is used for root |
|---------------------|----------------------------------|--------------------------------|
| | intended for the delivery and | canal irrigation/disinfection. |
| | evacuation of endodontic | |
| | irrigation solutions during root | |
| | canal procedures. | |
| Single use | Yes | Yes |
| disposable cannula | | |
| (tips) | | |
| Biocompatibility | Complies with ISO 10993 | Complies with ISO 10993 |
| | standards | standards |
| Sterilization | Autoclavable Master Delivery | Autoclavable handpiece |
| | Tip, Multi-Port Adapter, | |
| | Handpiece and Finger-Piece | |
# Non-Clinical Performance Data.
The performance of EndoVac® Apical Negative Pressure Irrigation System has been verified utilizing the following standards:
- ISO 10993-1 Biological Evaluation of Medical Devices Evaluation and Testing ●
- ISO 10993-5:2009 Biological evaluation of medical devices -- Tests for in vitro ● cytotoxicity
- ISO 10993-10 Biological evaluation of medical devices -- Tests for irritation and ● skin sensitization
- AAMI TIR 12: 2010 Designing, Testing and Labeling Reusable Medical Devices ● for Reprocessing in Health Care Facilities: A Guide for Manufacturers
- . ANSI/AAMI ST:79:2010 Comprehensive guide to steam sterilization and sterility assurance in health care facilities
- AAMI TIR 39:2009 Guidance on selecting microbial challenge and inoculation . sites for sterilization validation of medical devices
- ISO 17665-1:2006 Sterilization of health care products Moist Heat Part 1: . Requirements for the development, validation and routine control of a sterilization process for medical devices
- Internal method for autoclave life testing of reusable products
# Clinical Performance Data.
SybronEndo has not performed human clinical evaluations for EndoVac® Apical Negative Pressure Irrigation System. A summary of published literature has been included in order to support the safety and efficacy of the device.
#### Conclusion as to Substantial Equivalence
The similarities in design, function, safety and intended use of EndoVac® Apical Negative Pressure Irrigation System with the legally marketed device RinsEndo dental handpiece (K052271) provide evidence that these devices are substantially equivalent. The data provided in this submission support the substantial equivalence of the Endo Vac® Apical Negative Pressure Irrigation System to the predicate, RinsEndo dental handpiece (K052271).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.