K140683 · Arcoma AB · KPR · Jun 3, 2014 · Radiology
Device Facts
Record ID
K140683
Device Name
0180 INTUITION, 0072 PRECISION
Applicant
Arcoma AB
Product Code
KPR · Radiology
Decision Date
Jun 3, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
These are stationary x-ray systems intended for obtaining radiographic images of various portions of the human body in a clinical environment. The devices are not intended for mammography
Device Story
Stationary digital X-ray system for clinical diagnostic imaging; utilizes digital X-ray panels and high-frequency generators to capture radiographic images. Features include ceiling tube suspension, motorized movements, auto-positioning, and an elevating 6-way table (550 lbs capacity). Operated by clinical staff in a hospital or clinic setting. System includes a tube-mounted 10" touch screen and a Windows-compatible workstation for image acquisition. Precision model supports automatic stitching for enhanced workflow. Output consists of digital radiographic images for physician review to support clinical decision-making. Benefits include improved ergonomic positioning, efficient patient throughput, and high-quality diagnostic imaging.
Clinical Evidence
Clinical images obtained in accordance with FDA Guidance Document on Solid State Imaging Devices. Evaluated by professional radiologists and determined to be of good diagnostic quality. No quantitative clinical metrics (sensitivity/specificity) reported; bench testing and software integration validation also performed.
Technological Characteristics
Stationary digital X-ray system; high-frequency generator (50, 65, 80 kW); digital receptor panels (Canon 401C, 70C wireless); motorized ceiling tube suspension and wall stand; 6-way elevating table. Connectivity via DICOM. Safety standards: UL/CSA listings, IEC 60601-1, IEC 60601-1-2. Software-based image acquisition on Windows workstation.
Indications for Use
Indicated for obtaining radiographic images of various portions of the human body in a clinical environment. Not intended for mammography. No specific age or gender restrictions.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Page 1 of 2
JUN 0 3 2014
510(k) Summary, Device Modification, K14 0 68 3
- 1. Submitter: Arcoma AB Annavagen 1 SE-352 46 Vaxjo Sweden Phone: +46 470 70 69 00 Contact/Prepared by: Mikael Larsson, CEO. Date prepared: 17 February 2014
- 2. Identification of the Device: 0180 Intuition & 0072 Precision Stationary Digital X-Ray Systems Classification Name: Stationary x-ray system Common/Usual Name: Stationary digital x-ray system Regulation Number: 21 CFR 892.1680 Product Codes: KPR and MQB
- 3. Predicate Device: Arcoma Intuition, K073632
- Modification Summary: (1) Newer versions of Canon digital panels will be supplied. (2) A 4. different brand of high frequency generator will be supplied. (3) The overhead tube crane has been motorized. (4) The wall stand now has motorized vertical movement.
- 5. A description of the device: This is a complete stationary diagnostic x-ray system employing digital x-ray panels coupled with image acquisition software. The key features are:
- Ceiling Tube Suspension a.
- Motorized Movements, Auto-Positioning ﻗ
- c. Tube Mounted 10" Touch Screen
- Elevating 6 Way Table Patient Load: 5501bs ರ.
- Single User Interface e.
- High Frequency Generator 。
This ergonomic design incorporates automatic motorized positioning, creating an efficient workflow increasing the patient throughput. Precision offers automatic stitching enhancing the workflow further. Tube, vertical wall stand and table have extended range of motion for easy patient positioning. The acquisition software is installed on a Windows compatible workstation. The high frequency generator is available in different power ratings. All components are either 510(k) exempt or previously cleared. Generators are available in 50, 65, 80 kW (High Frequency) models.
- 6. These are stationary x-ray systems intended for obtaining radiographic Indications for use: images of various portions of the human body in a clinical environment. They are not intended for mammography.
- 7. The 0180 Intuition & 0072 Precision Stationary Digital X-Ray Systems have essentially the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device Arcoma Intuition, K073632. See the comparison table below. There are really only a few main differences: The digital panels, x-ray generator, and tube stands have been modified but have functionally identical capabilities. The proposed new digital x-ray panels each have their own 510(k) clearances.
{1}------------------------------------------------
K140683
Page 2 of 2
| Characteristic | Arcoma Intuition, K073632 | 0180 Intuition 0072 Precision |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | The 0170 Intuition is a stationary x-ray<br>system intended for obtaining<br>radiographic images of various portions<br>of the human body in a clinical<br>environment. The 0170 Intuition is not<br>intended for mammography | The 0180 Intuition and 0072 Precision are<br>stationary digital x-ray systems intended<br>for obtaining radiographic images of<br>various portions of the human body in a<br>clinical environment. They are not<br>intended for mammography |
| Digital Receptor<br>Panel | Canon 40C/G K031633 and<br>and Canon 50G K031447 | Canon 401C K103591<br>and Canon 70C wireless K102012 |
| Panel Resolution | Canon 40C/G<br>Imaging area 43 x 43 cm<br>Pixel Pitch 160 $\mu m$<br>Image matrix size 2,688 x 2,688 pixels<br>Canon 50G<br>Imaging area 35 x 43 cm.<br>Pixel Pitch 160 $\mu m$<br>Image matrix size 2,208 x 2,688 pixels | Canon 70C<br>Imaging area 35 X 43 cm<br>Pixel Pitch: 125 $\mu m$<br>Image matrix size: 2,800 x 3,408 Pixels<br>Canon 401C<br>Effective imaging area: 41.5 × 42.6 cm<br>Pixel pitch: 125 $\mu m$<br>Image matrix size: 3320 × 3408 pixels |
| Panel Connection | Wired | Wired or Wireless (WiFi) |
| DICOM | Yes | Yes |
| Tube Stand | Ceiling Mount | Same |
| Collimator | Siemens or Ralco | Siemens |
| Generator | EMD EPS 50, 65, 80 kW (High Frequency) | CPI, CMP 200 50, 65, 80 kW (High<br>Frequency) |
| Safety | UL Listings and IEC Standards IEC<br>60601-1 and IEC 60601-1-2, US<br>Performance Standards | UL/CSA Listings and IEC Standards IEC<br>60601-1 and IEC 60601-1-2, US<br>Performance Standards |
| Photo | Image: Arcoma Intuition, K073632 | Image: 0180 Intuition 0072 Precision |
- Description of non-clinical tests. The modified unit has undergone individual and system 8. performance tests, electrical safety and electromagnetic compatibility testing, as well as software integration validation and risk analysis. The unit meets IEC safety and EMC standards.
- 9. Description of clinical tests. Clinical images were obtained in accordance with the FDA Guidance Document on Solid State Imaging Devices. They were evaluated by professional radiologist and found to be of good diagnostic quality.
- 10. Conclusions drawn: The nonclinical and clinical tests that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph 3, above.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three wing-like shapes extending from its body.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Asenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3. 2014
Arcoma AB % Daniel Kamm, P.E. Kamm & Associates 8870 Ravello Court NAPLES FL 34114
Re: K140683
Trade/Device Name: 0180 Intuition and 0072 Precision Stationary Digital X-Ray Systems Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: April 29, 2014 Received: May 05, 2014
## Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Kamm
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fdg.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Janine M Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
### 510{k) Number (if known) K140683
#### Device Name
0180 Intuition and 0072 Precision Digital Radiography Systems
#### Indications for Use (Describe)
These are stationary x-ray systems intended for obtaining radiographic images of various portions of the human body in a clinical environment. The devices are not intended for mammography
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (1/14)
PSL" Publiching Seri nees 1 2011 443-4740 H
Page 17 of 3587
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.