The Fingertip Pulse Oximeter MD300C29-H is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent and child patient in hospital.
Device Story
Battery-powered, portable fingertip pulse oximeter; uses dual-wavelength LED (660nm red, 905nm infrared) and photodetector to measure light absorption in pulsating arteriolar vascular bed. Device calculates SpO2 based on ratio of light absorption during systole and diastole. Displays SpO2, pulse rate, pulse bar graph, and plethysmographic waveform on OLED screen. Operated by healthcare professionals in hospital settings for spot-checks. Provides real-time physiological data to assist clinicians in assessing patient oxygenation and heart rate status. Not for life-support or continuous monitoring.
Clinical Evidence
Clinical testing conducted per ISO 80601-2-61:2011. Study evaluated SpO2 accuracy under steady-state, non-motion conditions in the 70-100% range. Results confirmed device performance meets claimed accuracy specifications.
Technological Characteristics
Battery-powered (2x AAA). Components: LED emitter (660nm/905nm), photodetector, signal amplifier, CPU, OLED display. Materials: Medical-grade silicone gel, plastic enclosure. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 80601-2-61 (pulse oximeter performance), ISO 10993 (biocompatibility). Connectivity: Standalone. Software: Embedded firmware.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate in adult, adolescent, and pediatric patients in hospital settings. Not for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
### JUL 1 7 2014
# Section III 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(K) number is _ K140682
### 3.1 Date of Submission: June 11, 2014
3.2 Sponsor Information
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 320, West Building 4, No.83 Fuxing Road, Beijing 100039, P.R.China
### Contact Person:
Mr. Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No.9 Shuangyuan Road, Badachu Hi -tech Zone, Shijingshan District, Beijing, P.R. China, 100041 Phone: +86-10-88798300 Ext 6020 Fax: 215-4052545
### 3.3 Proposed Device Information
Device Common or Usual Name: Pulse Oximeter Device Trade or Proprietary Name: Fingertip Pulse Oximeter Model: MD300C29-H Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700 Panel: Anesthesiology Class: II
### 3.4 Predicate Device
510(k) Number: K070371 Common Name: Oximeter Device Trade or Proprietary Name: Fingertip Pulse Oximeter Model: MD300C Classification Name: Oximeter
{1}------------------------------------------------
Device Class: II Product Code: DQA Regulation Number: 870.2700 Review Panel: Anesthesiology
Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
Intended Use : Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
### 3.5 Device Description
The proposed device Fingertip Pulse Oximeter MD300C29-H is a battery powered device, which can detect and display the measured %SpO2 and pulse rate value, pulse bar graph and SpO2 waveform. The device is normally applied to adult, child and adolescent patient in the hospital.
The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit.
The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization or re-sterilization.
The device is for prescription.
The device does not contain drug or biological products.
The device is software -driven and the software validation is provided in Software
### 3.6 Intended Use
The Fingertip Pulse Oximeter MD300C29-H is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent and child patient in hospital.
{2}------------------------------------------------
| 1 |
|--------|
| 82 |
| 9<br>0 |
| T |
| V |
| > |
Page 3 of 7
Premarket Notification 510(k) Submission - Section III 510(k) Summary
Fingertip Pulse Oximeter MD300C29-H
# 3.7 Comparison with the Predicate Device
| Comparison with the Predicate Device | | | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Comparison Elements | Applicant Device | Predicate Device | |
| Device Name | Fingertip Pulse Oximeter MD300C29-H | MD300C Fingertip Pulse Oximeter (K070371) | |
| Model | MD300C29-H | MD300C | |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | |
| Classification | II | II | |
| Classification Name | Oximeter | Oximeter | |
| Product Code | DQA | DQA | |
| Indented Use | The Fingertip Pulse Oximeter MD300C29-H is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, child and adolescent patient in hospital. | Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring. | |
| Comparison Statement | The proposed devices have the similar intended use and classification. | | |
| Components | The applicant device consists of detector and emitter LED, signal amplify unit, CPU, data display unit and power unit | detector and emitter LED, signal amplify unit, CPU, data display unit and power unit | |
| Design Principle | Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. | Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. | |
| | | | |
| Measurement<br>wavelength | Red | $660\pm3nm$ | $660\pm2nm$ |
| | Infrared | $905\pm8nm$ | $940\pm10nm$ |
| | Comparison Statement | The proposed device has the same design principle and similar components. The only difference is the<br>wavelength of the infrared LED emitter, and we can verify that which will not effect the basic safety<br>and the essential performance of the proposed device. | |
| | Display Type | OLED | OLED |
| | Working time | Approximately 25 hours of continuous operation | Approximately 30 hours of continuous operation |
| | Brightness of backlight | Adjustable | Adjustable |
| | User Interface | 6 directions for display | 6 directions for display |
| | Power supply | 2*AAA alkaline battery | 2*AAA alkaline battery |
| | Display Data | SpO2; PR | SpO2, PR |
| | SpO2 display range | 35%~100% | 0~100% |
| | SpO2 measurement range | 70~100% | 70~100% |
| Device Specification | SpO2 Accuracy | 70%~100%; $\pm$ 2%<br>0%~69% no definition | 70%~100%; $\pm$ 3%<br>0%~69% no definition |
| | SpO2 resolution | 1% | 1% |
| | PR Display Range | 30-250bpm | 0-254bpm |
| | PR Measurement Range | 30~250bpm | 30~235bpm |
| | PR Accuracy | $\pm$ 2bpm (30-99bpm) and $\pm$ 2%(100-250bpm) | $\pm$ 2bpm (30-99bpm) and $\pm$ 2% (100-235bpm) |
| | PR resolution | 1bpm | 1bpm |
| | Operating temperature | 5℃~40℃ | 5°C~40°C |
| | Relative humidity | ≤80% , no condensation (operating)<br>≤93% no condensation (storage) | ≤80% , no condensation (operating)<br>≤93% no condensation (storage) |
| | Atmosphere pressure | 86kPa~106kPa | 86kPa~106kPa |
| | Comparison Statement | The proposed device has similar device specifications as the predicate device. | |
III-3
{3}------------------------------------------------
.
.
'
Page 4 of 7
IJI-4
{4}------------------------------------------------
| 82 |
|-----|
| |
| |
| 406 |
| 1 |
| X |
| |
Page 5 of 7 .
Premarket Notification 510(k) Submission - Section III 510(k) Summar
Fingertip Pulse Oximeter MD300C29-H
| | ABS | Laser Etching Medical Silicon gel | | Inks | The contacting materials of Proposed device are same to those of the predicate device except Fingertip Cushion and<br>Power Button. | The bench tests include SpO2 and PR accuracy Test, Weak<br>Perfusion Test, High and Low Temperature and Humidity<br>Test, Performance Test After Disinfection and ISO<br>80601-2-61. | Meet the requirements of FDA Guidance | Conformed to ISO 9919 | Conformed to ISO 9919 & ISO80601-2-61 | Conformed to IEC60601-1 | Conformed to IEC60601-1 | Conformed to IEC60601-1-2. | Conformed to IEC60601-1-2. | Moderate Level of Concern | Moderate Level of Concern | Compliance with FDA Guidance for the Content of<br>Premarket Submissions for Software Contained in Medical<br>Devices. |
|---------------|-----------|-----------------------------------|--------------|---------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|-----------------------|---------------------------------------|-------------------------|----------------------------|----------------------------------|----------------------------|---------------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------|
| Battery cover | Enclosure | Fingertip<br>Cushion | Power button | Coating | Comparison Statement | Bench Test | | Clinical Test | | Electrical Safety | | Electromagnetic<br>Compatibility | | Software | | |
| | | Contacting<br>Material | | | | Performance Testing | | Clinical Test | | | EMC and<br>Electrical Safe | | | | | |
ાત-ર
{5}------------------------------------------------
K140682
Page 6 of 7
Premarket Notification 510(k) Submission - Section III 510(k) Summan
Fingertip Pulse Oximeter MD300C29-H
| | | Risk Management in Compliance with ISO14971:2007 | | Risk Management in Compliance with ISO14971:2007 |
|----------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------|------------------------------------------------------------------------|
| | In Vitro Cytotoxicity | No cytotoxic potential | In Vitro Cytotoxicity | No cytotoxic potential |
| Biocompatibility | Skin Irritation Test | No evidence of causing sensitization | Skin Irritation Test | No evidence of causing sensitization |
| | Animal skin irritation test | No evidence of significant irritation from the test extract to rabbits | Animal skin irritation test | No evidence of significant irritation from the test extract to rabbits |
| | Medical silicone gel | | | |
| | | | | |
| Comparison Statement | Compliance with the ISO 10993 | | | |
| Label and Labeling | Compliance with the Guidance of pulse oximeter-premarket notification submission issued on March 4, 2013 | | | Compliance with FDA guidance |
:
{6}------------------------------------------------
### 3.8 Test Conclusion
### Non-clinical Test
The Fingertip Pulse Oximeter MD300C29-H is designed and tested and will be manufactured in accordance with the following standards, including:
- . IEC 60601-1:2005 Medical Electrical Equipment - Partl: General requirements for safety.
- . IEC 60601-1-2:2007 Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
- � 80601-2-61:2011 Medical electrical equipment - part2-61:Particular ISO requirements for the basic safety and essential performance of pulse oximeter equipment.
The Software Validation is in compliance with FDA Guidance for the Content of PreMarket Submissions for Software Contained in Medical Devices.
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".
### Clinical Test
The Clinical Test were conducted following the testing described in clause 201.12.1 of ISO 80601-2-61:2011, Medical electrical equipment- Part 2-61 Particular reguirements for basic safety and essential performance of pulse oximeter equipment.
The results of the study provide supporting evidence that the Fingertip Pulse Oximeter MD300C29-H is compliance to the accuracy specification claimed by the manufacturer. The Fingertip Pulse Oximeter can be used under steady state / non-motion conditions for the range 70-100%.
### 3.9 Determination of substantial equivalence
The proposed device has the same classification information, similar intended use, same design principle, similar product design and specifications, same performance effectiveness. performance safety as the predicate device. The main difference is the wavelength of the infrared LED emitter, and we can verify that which will not effect the basic safety and the essential performance of the proposed device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all capital letters and is smaller than the central symbol.
Public Flealth Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### July 17, 2014
Beijing Choice Electronic Technology Co., Ltd. Mr. Lei Chen Quality Director North Building 3F, No.9 Shuangyuan Road Badachu Hi -tech Zone, Shijingshan District. Beijing, P.R. China, 100041
Re: K140682
Trade/Device Name: Fingertip Pulse Oximeter MD300C29-H Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA Dated: June 11, 2014 Received: June 16, 2014
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Lei Chen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21) CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
在线视频 | Tejasbri Purobit-Sheth, M.D. Tejashri Purofiit-Sheth, M.D. Clialcal Deputy Director A CORPORATION DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
## Section II Indication for Use Statement
### Indication for Use
510(k) Number (if known):
Device Name: Fingertip Pulse Oximeter MD300C29-H
Indications for Use:
The Fingertip Pulse Oximeter MD300C29-H is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent and child patient in hospital.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Todd D. Courtney -S 2014.07.16 14:01:12 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.