LEADPOINT FOCUS WORKSTSTION, LEADPOINT FOCUS NOTEBOOK SYSTEM
Applicant
Alpine Biomed Aps
Product Code
GZL · Neurology
Decision Date
Dec 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1330
Device Class
Class 2
Indications for Use
The Leadpoint Focus microelectrode targeting system is intended to assist in neurosurgical procedures where recording of neuronal activity and stimulation of brain neurons will aid in the placement of depth electrodes. The Leadpoint Focus microelectrode targeting system is intended to be used in the operating room by a neurosurgeon, neurologist, or clinical neurophysiologist.
Device Story
Leadpoint Focus is a microelectrode recording (MER) system used during stereotactic neurosurgery. It captures neuronal activity via microelectrodes (not included) and provides macro stimulation to aid target localization. System components include software, hardware, macro stimulator, MER amplifier, and sterile interconnection cables. Operated by neurosurgeons, neurologists, or clinical neurophysiologists in the operating room. The system processes electrical signals from the brain, displaying them on a digital monitor and providing audio feedback via loudspeakers. Clinicians use the visual and auditory output to guide the precise placement of depth electrodes. The device facilitates intraoperative target localization, potentially improving surgical accuracy and patient outcomes.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through software testing per FDA guidance, electrical safety testing (IEC 60601-1), and electromagnetic compatibility testing (EN/IEC 60601-1-2, 60601-2-40, 61000-3-2, 61000-3-3).
Technological Characteristics
PC-based workstation or portable notebook system. 6-channel isolated amplifier; 24-bit ADC at 48kHz. Constant current macro stimulator (0-50V, 10mA max). Connectivity via USB. Safety isolation Type BF. Software-driven interface with digital display and audio monitoring. Sterilization for consumable cables.
Indications for Use
Indicated for neurosurgical procedures requiring recording of neuronal activity and stimulation of brain neurons to assist in depth electrode placement. Intended for use by neurosurgeons, neurologists, or clinical neurophysiologists in the operating room.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 12, 2014
Alpine Biomed ApS Seamus O'Connor Tonsbakken 16-18 DK-2740 Skovlunde, Denmark
Re: K140680
Trade/Device Name: Leadpoint Focus Regulation Number: 21 CFR 882.1330 Regulation Name: Depth electrode Regulatory Class: Class II Product Code: GZL Dated: November 10, 2014 Received: November 12, 2014
Dear Mr. O'Connor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K140680
Device Name Leadpoint Focus
Indications for Use (Describe)
The Leadpoint microelectrode targeting system is intended to assist in neurosurgical procedures where recording of neuronal activity and stimulation of brain neurons will aid in the placement of depth electrodes.
The Leadpoint microelectrode targeting system is intended to be used in the operating room by a neurosurgeon, neurologist, or clinical neurophysiologist.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5 510(k) Summary
Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
12/10/2014
5 510(k) Summary - 1 of 8
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for Alpine BioMed. The logo is divided into two sections, with the left section containing a green and blue circular design with a white line running through it. The right section contains the words "ALPINE" in blue, "bioMed" in green, and a green line separating the two words. Below the logo, the words "a division of natus" are written in a smaller font.
## 510(k) Summary
510(k) Number:
K140680
O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark
Phone +45 44 57 90 00 +45 44 57 90 10
10 December 2014
#### Submitter's name:
Alpine Biomed ApS Tonsbakken 16-18, DK-2740 Skovlunde, Denmark Tel.: +45 4457 9000
#### Contact Person:
Seamus O'Connor, Email seamus.oconnor@natus.com Tel .: +353-(0)87-6330532
Trade name: Leadpoint Focus
Device Class Class II
Classification name:
Depth electrode.
#### Product code and Regulation:
21CFR 882.1330 GZL
#### Predicate devices:
The Leadpoint Focus is substantially equivalent to the predicate device K071364 Guideline 4000 from FHC, Inc.
#### Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
12/10/2014
5 510(k) Summary - 2 of 8
{5}------------------------------------------------
Image /page/5/Picture/20 description: The image is a logo for Alpine BioMed. The logo is in a blue box with a green and blue symbol on the left. To the right of the symbol is the text "ALPINE" in blue, with a green line underneath it. Below that is the text "bioMed" in green and blue. Below the logo is the text "a division of natus."
Alpine Biomed ApS DK-2740 Skovlunde
+45 44 57 90 10
#### Intended use / Indications for use:
The Leadpoint Focus microelectrode targeting system is intended to assist in neurosurgical procedures where recording of neuronal activity and stimulation of brain neurons will aid in the placement of depth electrodes.
The Leadpoint Focus microelectrode targeting system is intended to be used in the operating room by a neurosurgeon, neurologist, or clinical neurophysiologist.
#### Device Description:
Intraoperative microelectrode recording (MER) has been introduced to improve the target localization during stereotactic surgery. The Leadpoint Focus is a MER system which consists of the Leadpoint Focus Software, Focus Hardware, a Macro Stimulator, a MER amplifier and a sterile interconnection cable (MER Cable) that connects the system to a Microelectrode. The Microelectrode is supplied by another manufacturer and is not part of the Leadpoint Focus system.
The recognized upper limit for safe stimulation is 30μC/cm2 per phase.
#### Microelectrode Connection
A MER Cable is an electrode cable composed of strands of insulated electrical conductors intended to connect a Microelectrode from a patient to the Leadpoint Focus system.
The MER Cable is an optional consumable part.
The MER cable is for single use only and delivered sterile.
#### Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
5 510(k) Summary - 3 of 8
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for Alpine BioMed, a division of Natus. The logo is blue and green, and it features a stylized mountain range with a heartbeat line in the negative space. The text "ALPINE bioMed" is written in blue and green, with "bioMed" in a smaller font. Below the logo, the text "a division of natus." is written in a smaller font.
O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark
□ Phone +45 44 57 90 00 +45 44 57 90 10 Fax
#### Summary of Technological Characteristics comparison
| Characteristic | Alpine Biomed ApS<br>Leadpoint Focus<br>(This submission) | FHC, Inc.<br>Guideline 4000<br>(K071364) | Discussion of<br>Differences |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Indications for Use | | | |
| 1.1 Intended Use /<br>Indications for Use | The Leadpoint Focus<br>microelectrode targeting system is<br>intended to assist in neurosurgical<br>procedures where recording of<br>neuronal activity and stimulation of<br>brain neurons will aid in the<br>placement of depth electrodes.<br>The Leadpoint Focus<br>microelectrode targeting system is<br>intended to be used in the operating<br>room by a neurosurgeon,<br>neurologist, or clinical<br>neurophysiologist. | The microTargeting® Guideline<br>4000 is intended to be used by a<br>neurosurgeon, neurologist or clinical<br>neurophysiologist to accurately<br>position depth electrodes during<br>functional neurosurgical procedures.<br>The Guideline 4000 is intended to<br>be used to accurately position depth<br>electrodes during functional<br>neurosurgical procedures. | Similar |
| 1.2 Clinical Users | The Leadpoint microelectrode<br>targeting system is intended to be<br>used by a neurosurgeon,<br>neurologist, or clinical<br>neurophysiologist. | The Guideline 4000 is intended to<br>be used by a neurosurgeon,<br>neurologist or clinical<br>neurophysiologist. | Similar |
| 1.3 Clinical<br>Environment | The Leadpoint microelectrode<br>targeting system is intended to be<br>used for neurosurgical procedures<br>in the operating room. | The Guideline 4000 is intended to<br>be used during functional<br>neurosurgical procedures. | Similar |
### Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for Alpine BioMed, a division of Natus. The logo features a blue square with a green and blue design inside. The words "ALPINE" and "bioMed" are written in blue, with a green line separating the two words. The words "a division of natus" are written in black below the square.
O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark
□ Phone +45 44 57 90 00 +45 44 57 90 10 Fax
| General Design | | | |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| 2.1 General systems<br>approach | Standard PC based equipment with<br>dedicated hardware peripherals /<br>components. | Standard PC based equipment with<br>dedicated hardware peripherals /<br>components. | Identical |
| 2.2 Configurations | Workstation on cart<br>Portable notebook PC based<br>system | Workstation on cart: Guideline<br>4000TM<br>Portable notebook PC based<br>system: Guideline 4000 LP+TM | Identical |
| 2.3 User input device | Window mouse/keyboard driven<br>graphic interface with dedicated<br>control panel. | Window mouse/keyboard driven<br>graphic interface with dedicated<br>control panel. | Identical |
| 2.4 User output<br>device | Digital color display, loudspeakers<br>and commercial printers. | Digital color display, loudspeakers<br>and commercial printers. | Identical, both<br>devices uses dual<br>loudspeakers for high<br>fidelity audio<br>monitoring. |
| 2.5 Use of standard<br>software platform<br>(Operating System) | Microsoft Windows. | Microsoft Windows. | Identical |
| 2.6 System<br>components | Main Unit including power supply,<br>speakers and control panel<br>Amplifier<br>Macro Stimulator<br>Additional for notebook based<br>system: Notebook PC including<br>monitor, keyboard, and mouse<br>Additional for Workstation: PC,<br>monitor, keyboard, mouse, cart and<br>isolating transformer | Main Unit including power supply,<br>speakers and control panel<br>Amplifier<br>Macro Stimulator<br>Notebook PC including monitor,<br>keyboard, and mouse.<br>Additional for Workstation: PC,<br>monitor, keyboard, mouse, cart and<br>isolating transformer. | Similar |
| 2.7 System –<br>computer interface | USB | USB | Identical |
| 2.8 System power<br>supply | Mains (100-120V~50/60Hz or 200-<br>240V~50/60Hz) through an isolating<br>transformer. | 100/120/220/240VAC~50/60Hz.<br>400VA | Similar |
| 2.9 Display: Size /<br>Resolution | Workstation: 22" / 1680 x 1050<br>Notebook based system: 15.6" /<br>1920 x 1080 | Notebook based system: 17" /<br>UXGA (1600 x 1200) | Similar |
| 2.10 Safety Isolation | Type BF | Type BF | Identical |
## Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
5 510(k) Summary - 5 of 8
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Alpine BioMed, which is enclosed in a blue box. The logo includes a green and blue symbol on the left, followed by the text "ALPINE" in blue, a green line, and "bioMed" in blue. Below the logo, the text "a division of natus." is written in a smaller font.
O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark
□ Phone +45 44 57 90 00 +45 44 57 90 10 Fax
| Microeletrode Recording | | | |
|---------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3.1 Patient inputs | 6 channel amplifier, isolated<br>All active channels continuously displayed. | 2 channel amplifier, isolated.<br>Expandable recording channels to a maximum of ten.<br>All active channels continuously displayed. | Standard single sided MER procedures requires simultaneous recording from 1-5 channels.<br>This can be done with both systems. |
| 3.2 Bilateral procedure | Supported with up to 2 x 3 channels | Supported with up to 2 x 5 channels. | Standard MER recordings for bilateral procedures are performed with 2 x 1 channels.<br>This can be performed with both systems |
| 3.5 Input Impedance | >1000Mohm//30pF with electrode cable capacitance reduction | $10^{13}\Omega$ // 0.2 pF AC, 200M $\Omega$ DC Differential<br>$10^{13}\Omega$ // 7.0 pF AC, 200M $\Omega$ DC Common Mode | Both within the range required for MER. Leadpoint Focus through use of electrode capacitance reduction and the predicate device through unusual high impedances |
| 3.6 Filters | High Pass: 0.01 – 3000 Hz (16 steps)<br>Low Pass: 20 Hz - 13 kHz (12 Steps) | High Pass: Analog: 0 or 240 Hz.<br>Digital: 0 – 1000 Hz<br>Low Pass: Analog: 20 kHz. Digital: 3 - 20 kHz<br>The digital filters are continually adjustable. | Both within the range required for MER. Both systems secure minimal noise: Leadpoint Focus through many steps in the filters and the predicate device via continually adjustable filters. |
| 3.7 Connection Types | 1.5mm Touch Proof / DIN socket / Micro D-sub | 5 pin DIN socket | Both within the range required for MER. |
| 3.8 Signal acquisition | 24 bit analog to digital conversion at 48kHz sample rate | 16 bit analog to digital conversion at 48kHz sample rate | Both within the range required for MER. |
| 3.9 Display | Up to six analysis windows. | Up to five analysis windows. | Both within the range required for MER. |
| 3.10 Impedance test | Integrated 220 Hz sine wave, max. 1V / max. 0.1μΑ | Adjustable 100 Hz - 5 kHz sine wave. | Both within the range required for MER. |
| 3.11 Impedance test range | 10 ΜΩ | Auto ranging 100 kΩ, 1 ΜΩ or 10 ΜΩ. | Both within the range required for MER. |
| 3.12 Epoch Recording | Max. 80 s per Epoch | 2, 10 or 30 sec | Both within the range required for MER. |
| 3.13 Audio Modes | Raw or clipped sound | Raw or clipped sound | Identical |
| 3.14 Stereotactic systems | Most common | Most common | Identical |
### Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
5 510(k) Summary - 6 of 8
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image is a logo for Alpine BioMed. The logo features a blue square with a green circle inside. Inside the green circle is a white line that resembles a heartbeat. To the right of the square is the text "ALPINE bioMed" in blue and green. Below the logo is the text "a division of natus".
Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark
Phone +45 44 57 90 00 +45 44 57 90 10
| Macro Stimulation | | | |
|--------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------|
| 4.1 Type | Constant Current (CC) | Constant Current (CC) | Identical |
| 4.2 Max output intensity | 10 mA | 10 mA | Identical |
| 4.3 Voltage Range | 0-50 V | 0-50 V | Identical |
| 4.4 Intensity resolution | Fine setting: 0.02 mA<br>Coarse setting: 0.1 mA | 0.1 mA | Both within the range required for MER |
| 4.5 Stimulus Duration | 20 µs – 1 ms (9 steps) | 50 µs – 1 ms (8 steps) | Both within the range of required stimulation frequencies (typical 50 µs) |
| 4.6 Repetition Rate | 0.1 – 200 Hz (19 steps) | 10 - 300Hz (12 steps) | The added range from 0.1 to 10(0) Hz does not add any new risks. |
| 4.7 Stimulus Polarity | Monophasic or biphasic stimulation | Monophasic or biphasic | Identical to the Leadpoint Focus system |
| 4.8 Remote control | Yes, handheld device with controls for stimulation On/Off and intensity | Yes, handheld device with controls for stimulation On/Off and intensity | Identical |
#### Performance data.
#### Software Test.
The Leadpoint Focus software was tested in accordance with:
FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05;
FDA quidance: Off-the-shelf software use in medical devices, 09 Sep 99.
#### Electrical Safety and Electromaqnetic Compatibility (EMC) Test.
The Leadpoint Focus was tested for electrical safety. Test results demonstrated that the Leadpoint Focus complies with:
IEC 60601-1:1988 + A1:1991 + A2:1995.
Electromagnetic Compatibility (EMC) testing was conducted on the Leadpoint Focus according to the applicable standard. Test results indicated that the system complies with the following:
EN/(IEC) 60601-1-2:2007 (3rd edition)
EN/(IEC) 60601-2-40:1998, section 5
EN/(IEC) 61000-3-2:2006+A1+A2
EN/(IEC) 61000-3-3:2008
#### Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
12/10/2014
Tel: +45 44 57 90 00
5 510(k) Summary - 7 of 8
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image is a logo for Alpine BioMed, a division of Natus. The logo features a blue square with a green circle inside, and a white line that resembles a heartbeat. The words "ALPINE" and "bioMed" are written in blue and green, respectively, above the words "a division of natus" in black. The logo is simple and professional, and it conveys a sense of trust and reliability.
O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark
O Phone +45 44 57 90 00 +45 44 57 90 10 Fax
### Conclusion.
The comparison to the predicate devices demonstrates that the Leadpoint Focus system is safe and effective for its intended use and is substantially equivalent to the predicate devices
#### Company Confidential
Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde
Tel: +45 44 57 90 00
12/10/2014
5 510(k) Summary - 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.