# CELESTEION, PCA-9000A/2 (K140651)

_Toshibamedical Systems Corporation · KPS · Jun 11, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140651

## Device Facts

- **Applicant:** Toshibamedical Systems Corporation
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Jun 11, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images. This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, diagnosis, therapeutic planning and therapeutic outcome assessment of (but not limited to) cancer, cardiovascular disease and brain dysfunction. Additionally, this device can be operated independently as a whole body multi-slice CT scanner.

## Device Story

Celesteion, PCA-9000A/2, is a large-bore, time-of-flight (TOF) PET-CT system. Inputs: X-ray transmission data (CT) and positron-emitting radiopharmaceutical distribution (PET). Operation: CT component performs multi-slice helical scanning; PET component utilizes TOF detectors (450 ps temporal resolution). System integrates PET and CT data for attenuation correction and image fusion. Used in clinical settings by trained healthcare professionals. Output: Cross-sectional CT images, PET images, and fused PET/CT images. Clinical utility: Provides metabolic and functional information to assist in diagnosis, therapeutic planning, and outcome assessment for cancer, cardiovascular disease, and brain dysfunction. System features a 900 mm gantry aperture and 700 mm scanning field; supports independent CT-only operation.

## Clinical Evidence

Bench testing only. No clinical data. Performance validated using phantoms to assess spatial resolution, CT number, contrast-to-noise ratio, noise properties, sensitivity, NECR, energy/timing resolution, and PET/CT alignment. Results demonstrate substantial equivalence to predicate devices.

## Technological Characteristics

Large-bore PET-CT system. CT: 16-row solid-state detector, 72 kW max output, 80-135 kV tube voltage. PET: TOF detector, 450 ps temporal resolution. Materials: Standard medical imaging components. Connectivity: Dual-console (scanning/display). Sterilization: N/A. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-44, IEC 60825-1, IEC 62304, IEC 62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26, NEMA NU2.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Aquilion LB Triton, V4.91 ([K123500](/device/K123500.md))
- Gemini Raptor ([K052640](/device/K052640.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K.140651
Page 1 of 4

TOSHIBA AMERICA MEDICAL BYBTEMS, INC. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000

JUN 1 1 2014

## 510(k) SUMMARY

- 1. SUBMITTER'S NAME: Toshiba America Medical Systems, Inc.
- 2. ADDRESS: 2441 Michelle Drive Tustin, CA 92780-2068
- 3. ESTABLISHMENT REGISTRATION: 2020563
- 4. CONTACT PERSON: Paul Biggins Director, Regulatory Affairs (714) 730-5000
- 5. Date Prepared: March 12, 2014

.

.

- 6. TRADE NAME(S): Celesteion, PCA-9000A/2
- 7. COMMON NAME: System, X-ray, Computed Tomography System, Positron Emission Tomography
- 8. DEVICE CLASSIFICATION: Class II (per 21 CFR §892.1750 and 21 CFR §892.1200)

# 9. PRODUCT CODE / DESCRIPTION:

JAK - System, Computed Tomography KPS - System, Positron Emission Tomography

# 10. PERFORMANCE STANDARD:

This device conforms to applicable Performance Standards for lonizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]

{1}------------------------------------------------

## 11. PREDICATE DEVICE:

| Product                   | Marketed by                     | 510(k) Number | Clearance Date  |
|---------------------------|---------------------------------|---------------|-----------------|
| Aquilion LB Triton, V4.91 | Toshiba America Medical Systems | K123500       | May 2, 2013     |
| Gemini Raptor             | Philips Healthcare              | K052640       | October 7, 2005 |

#### 12. REASON FOR SUBMISSION:

New device

## 13. DEVICE DESCRIPTION:

Celesteion, PCA-9000A, is a large bore, TOF, PET-CT system, which combines a high-end CT system with a high-throughput PET system. The high-end CT system is a multi-slice helical CT scanner with a gantry aperture of 900 mm and a maximum scanning field of 700 mm. The high-throughput PET system has a time of flight (TOF) detector with temporal resolution of 450 ps. Celesteion, PCA-9000A, is intended to acquire PET images of any desired region of the whole body and CT images of the same region (to be used for attenuation correction or image fusion), to detect the location of positron emitting radiopharmaceuticals in the body with the obtained images. This device is used to gather the metabolic and functional information from the distribution of radiopharmaceuticals in the body for the assessment of metabolic and physiologic functions. This information can assist research, diagnosis, therapeutic planning, and therapeutic outcome assessment. This device can also function independently as a whole body multi-slice CT scanner.

### 14. INDICATIONS FOR USE:

The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images.

This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, diagnosis, therapeutic planning and therapeutic outcome assessment of (but not limited to) cancer, cardiovascular disease and brain dysfunction. Additionally, this device can be operated independently as a whole body multi-slice CT scanner.

#### 15. SUBSTANTIAL EQUIVALENCE:

Celesteion, PCA-9000A/2, is substantially equivalent to the Gemini Raptor which was cleared via Pre-Market Notification 510(k), K052640, marketed by Philips Healthcare as Gemini TF Big Bore.

The CT portion of the Celesteion, PCA-9000A/2, is substantially equivalent to the Aquilion LB Triton system which received 510(k) clearance on May 2, 2013, per K123500. Some minor differences include the use of a moving table base and longer couch top in order to allow scanning in both CT and PET portions and the use of two consoles, one for scanning and one for display, instead of one console that performs both functions. The imaging chain including but not limited to the X-ray tube, detector and data acquisition system of the CT portion is identical to that of the Aquilion LB Triton.

{2}------------------------------------------------

The PET portions of the Celesteion and the Gemini Raptor operate on the same principles and basic technologies. Both systems acquire PET images and use their respective CT portions for attenuation correction. Additionally, both systems can combine PET and CT images as fused PET/CT images. The main difference between the systems is that the Celesteion requires a longer imaging duration (user selectable) to obtain equivalent data and image quality as compared to the predicate device.

| Item                           | Celesteion<br>PCA-9000A/2                        | Gemini Raptor                                  | Aquilion LB Triton,<br>V4.91                     |
|--------------------------------|--------------------------------------------------|------------------------------------------------|--------------------------------------------------|
| 510(k) Number                  | This submission                                  | K052640                                        | K123500                                          |
| PET Specifications             |                                                  |                                                |                                                  |
| Sensitivity (cps/kBq)          | $\u2265$ 3.6                                     | 6.6                                            | N/A                                              |
| Count rate maximum NECR        | 61 $\u00b1$ 10 kcps                              | 90                                             | N/A                                              |
| System energy resolution       | $\u2264$ 13.7%                                   | 11.7%                                          | N/A                                              |
| System timing resolution       | $\u2264$ 450 ps                                  | 495 ps                                         | N/A                                              |
| Scatter fraction (%)           | $\u2264$ 42.7                                    | 26                                             | N/A                                              |
| Spatial Resolution FWHM at 1cm | $\u2264$ 5.1                                     | 4.7                                            | N/A                                              |
| Performance Specifications     |                                                  |                                                |                                                  |
| Scan Regions                   | Whole body                                       | Whole body                                     | Whole body                                       |
| CT scan FOV                    | 70 cm                                            | 60 cm                                          | 70 cm                                            |
| Scan System                    | CT: 360°<br>continuous rotate/rotate             | CT: 360°<br>continuous rotate/rotate           | CT: 360°<br>continuous rotate/rotate             |
| CT Specifications              |                                                  |                                                |                                                  |
| CT Detection System            | 16 row<br>Solid State Detector                   | 16 row<br>Solid State Detector                 | 16 row<br>Solid State Detector                   |
| Output capacity                | 72 kW (max)                                      | 60 kW                                          | 72 kW (max)                                      |
| X-ray Tube Voltage             | 80, 100, 120 and 135 kV                          | 90, 120, 140 kVp                               | 80, 100, 120 and 135 kV                          |
| X-ray Tube Current             | 10 mA to 600 mA                                  | 20-500 mA                                      | 10 mA to 600 mA                                  |
| X-ray Tube Heat Capacity       | 7.5 MHU                                          | 8 MHU<br>(Anode storage capacity)              | 7.5 MHU                                          |
| X-ray Tube cooling rate        | 1,386 kHU/min (max)<br>1,008kHU/min (actual)     | 1,608 kHU/min (max)                            | 1,386 kHU/min (max)<br>1,008kHU/min (actual)     |
| Focal Spot Size (IEC)          | 0.09mm x 0.8mm (small)<br>1.6mm x 1.4 mm (large) | 0.5mm x 1.0mm (small)<br>1.0mm x 1.0mm (large) | 0.09mm x 0.8mm (small)<br>1.6mm x 1.4 mm (large) |
| Lowest couch height            | 475 mm<br>(includes moving base)                 | Not available                                  | 312 mm (312 mm)                                  |
| Couch-top stroke               | 2390 mm                                          | Not available                                  | 2190 mm (1890 mm)                                |

A complete comparison table is included in this submission. See below for a brief comparison of the technological characteristics between the subject and the predicate devices:

# 16. SAFETY:

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, IEC61223, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26 and NEMA NU2.

{3}------------------------------------------------

K140651
Page 4 of 4

Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.

## 17. TESTING

Risk analysis and verification testing conducted through bench testing are included in this submission which demonstrates that the established specifications for the device have been met. CT image quality metrics were performed, utilizing phantoms, which validated that the subject device is substantially equivalent to the predicate device with regard to spatial resolution, CT number and contrast-to-noise ratio and noise properties. Additionally, PET image quality metrics were performed which validated that the subject device met established specifications for spatial resolution, sensitivity, NECR, energy/timing resolution and PET/CT alignment.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

Additionally, testing of the system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.

#### 18. CONCLUSION

Celesteion, PCA-9000A/2, performs in a manner that is similar to and is intended for the same use as the predicate devices, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate devices.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 11, 2014

Toshiba Medical Systems Corporation % Mr. Paul Biggins Director, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K140651

Trade/Device Name: Celesteion, PCA-9000A/2 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: March 12, 2014 Received: March 13, 2014

Dear Mr. Biggins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's reguirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

> ﺮ ﺗ .

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for

Janine M Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

#### K140651

Device Name

Celesteion, PCA-9000A/2

#### Indications for Use (Describe)

The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces crosssectional images of the body by computer reconstruction of x-ray transmission data. The PET component images the distribution of PBT radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images.

This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, diagnosis, therapeutic planning and therapeutic outcome assessment of {but not limited to) cancer, cardiovascular disease and brain dysfunction. Additionally, this device can be operated independently as a whole body multi-slice CT scanner.

Type of Use (Select one or both, as applicable)

كا Prescription Use (Part 21 CFR 801 Subpart D)

ال Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

# FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Sm.7)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Orug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

**FORM FDA 3881 (1/14)**

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

---

**Source:** [https://fda-staging.innolitics.com/device/K140651](https://fda-staging.innolitics.com/device/K140651)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).