K140608 · Merit Medical Systems, Inc. · DQX · Sep 12, 2014 · Cardiovascular
Device Facts
Record ID
K140608
Device Name
MERIT HEPARIN-COATED INQWIRE DIAGNOSTIC GUIDEWIRE
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Sep 12, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
Merit Medical guide wires are used to facilitate the placement of devices during diagnostic and interventional procedures.
Device Story
Diagnostic guide wire used to facilitate placement of devices during diagnostic and interventional procedures. Device consists of PTFE-coated coil, internal core wire, and safety wire; Benzalkonium Heparin coating applied over PTFE. Core wire provides stiffness; safety wire ensures component integrity. Available in 0.035 or 0.038 inch diameters, various tip configurations (straight/J-tip), and lengths (70cm-180cm). Supplied in loop flush dispenser (hoop) with female Luer connection for hydration. Modification involves replacing PFOA with GenX as processing aid in PTFE coating manufacture. Used by clinicians in interventional settings.
Clinical Evidence
Bench testing only. No clinical data. Testing included surface analysis, coating adherence/integrity (fracture, flexing, coil lubricity), catheter compatibility, and biocompatibility. Results met predetermined acceptance criteria and were comparable to predicate devices.
Technological Characteristics
Materials: PTFE-coated coil, core wire, safety wire, Benzalkonium Heparin coating. Dimensions: 0.035-0.038 inch OD, 70-180cm length. Standards: ISO 11070:1998 (introducers), ISO 11135-1:2007 (sterilization), ASTM F1980-07 (aging), ISO 10993-1:2009 (biocompatibility). Connectivity: None. Energy source: None (mechanical).
Indications for Use
Indicated for use in patients undergoing diagnostic and interventional procedures to facilitate the placement of medical devices.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The figure is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 12, 2014
Merit Medical, Inc. c/o Ms. Siobhan King Regulatory Affairs Specialist II Parkmore Business Park West Galway, Ireland
Re: K140608
Trade/Device Name: InOwire® Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II Product Code: DQX Dated: August 14, 2014 Received: August 18, 2014
Dear Ms. King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section 4 Indications for Use Statement
510(k) Number (if known):
Device Name: Merit Heparin-coated InQwire® Diagnostic Guide Wire
Indications for Use:
Merit Medical guide wires are used to facilitate the placement of devices during diagnostic and interventional procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Merit Medical. The logo consists of a red square with white lines inside, followed by the words "MERIT MEDICAL" in red. The font is sans-serif and the letters are slightly rounded.
## Section 5 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: 1721504<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: 9616662 | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(+353) 91 703700 (3061)<br>(+353) 91 680 104<br>Mark Mullaney<br>Correspondent Name: Merit Medical Ireland Ltd.<br>Parkmore Business Park West<br>Galway, Ireland<br>(+353) 91 703700 (3052)<br>(+353) 91 680 104<br>Siobhan King<br>06/03/2014 |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| Subject<br>Device | Trade Name: | InQwire®<br>Common/Usual Name:Merit Medical Guide Wire<br>Classification Name: 21 CFR 870.1330 Catheter guide wire |
| Predicate | Premarket Notification Predicate#1: | |
| | Trade Name:<br>Premarket Notification: K822586<br>Manufacturer: | InQwire®<br>Classification Name: 21 CFR 870.1330 Catheter guide wire<br>Merit Medical Systems, Inc. |
| Device | Premarket Notification Predicate#2:<br>Trade Name:<br>Premarket Notification: K133230<br>Manufacturer: | InQwire®<br>Classification Name: 21 CFR 870.1330 Catheter guide wire<br>Merit Medical Systems, Inc. |
| Classification | Class II<br>21 CFR 870.1330 Catheter guide wire<br>FDA Product Code: DQX | Review Panel: Division of Cardiovascular Devices |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Merit Medical. The logo consists of a red square with white lines on the left, followed by the word "MERIT" in red and the word "MEDICAL" in white with a red outline. The logo is simple and professional, and it is likely used to represent the company in a variety of contexts.
Merit Heparin-coated InQwire® Diagnostic Guide Wires are intended to facilitate the placement of devices during diagnostic and interventional procedures. The quide wires have a continuous PTFE (Polytetrafluoroethylene) coated coil, inside core wire and a safety wire. The core wire is fixed at the proximal end only and extends to a specified distance from the distal end. The distal segment of the core wire has a profiled taper. The core wire enhances the stiffness of the guide wire. The safety wire extends the full length of the guide wire and is welded at both the distal and proximal end. The safety wire is designed to provide integrity and ensures that the guide wire components remain together. A Benzalkonium Heparin coating is applied over the PTFE coating.
The Heparin-coated InQwire® Diagnostic Guide Wires are offered in 0.035 inch or 0.038 inch outer diameter with different tip configurations (straight or J-tip), single ended or double ended (either end of the wire can be placed into the patient), standard or firm shaft, and are available in lengths from 70cm to 180cm. The wire is placed inside a loop flush dispenser, also referred to as a hoop. The dispenser has a female Luer connection which facilitates solution flushing through the hoop to hydrate the guide wire. The guidewires are secured in the hoop dispenser by a locking J-tip straightener.
## Device
Description
This modification details a supplier process change relating to removal of perfluorooctanoic acid (PFOA), used in the manufacture of the PTFE (Polytetrafluoroethylene) coating. More specifically PFOA is a processing aid used for the polymerization process of PTFE, which is flashed off during the oven curing process and therefore not part of the finished Heparin-coated InQwire® Diagnostic Guidewire. The PTFE coating formulation has changed - PFOA has been removed as a processing aid and replaced with GenX. The oven curing process which cures the coating and flashes off the PFOA remains unchanged. The manufacturing process within Merit has not changed. Due to environmental concerns relating to PFOA, the U.S. Environmental Protection Agency has instructed that all processors of PTFE no longer make, buy or use PFOA(2010/2015 PFOA Stewardship Program). PFOA material is used in the manufacture of both the intermediate layer and top layer of the PTFE(Polytetrafluoroethylene) coating. The removal of PFOA from Merit's non Heparin-coated InQwire® Diagnostic Guide Wires was previously assessed under Special 510(k) Premarket Notification, K133230. The only difference between the Heparin-coated InQwire® Guidewires and the non-Heparin coated InQwire® Guidewires, K133230, is the addition of a Benzalkonium Heparin coating applied over the PTFE coating. No changes have been made to the actual heparin coating. The same design and material characteristics that were qualified in K133230, also apply to the heparin-coated InQwire® Diagnostic Guidewires.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Merit Medical. The logo consists of a red square with a stylized white "M" inside, followed by the words "MERIT MEDICAL" in red, with the "MEDICAL" portion slightly smaller and to the right of "MERIT". The logo has a clean and professional design.
Technological characteristics of the subject Merit Medical Heparin-coated InQwire® Diagnostic Guide Wire are substantially equivalent to Predicate Device#1, the Merit Medical Heparin-coated InQwire® Diagnostic Guidewire [K822586] and Predicate Device#2, the Merit Medical non Heparin-coated InQwire® Diagnostic Guidewire [K133230]. The difference between the subject device and Predicate Device# 1 relates to the removal of perfluorooctanoic acid (PFOA), used in the manufacture of the PTFE (Polytetrafluoroethylene) guide wire coating. The guide wire Comparison design and indications remain unchanged. Other than the addition of a to Predicate Benzalkonium Heparin coating applied over the PTFE coating, the subject device and Predicate Device#2 are identical. The guide wire design and indications remain unchanged. Predicate Device#2, the Merit Medical non Heparin-coated InQwire® Guidewire is the predicate for the performance testing parameters.
> No performance standards have been established under section 514 of the Food, Drug and Cosmetic Act for these devices. Performance testing of the subject Merit Heparin-coated InQwire® Diagnostic Guide Wire was conducted based on risk analysis. A battery of testing was conducted in accordance with protocols based on requirements outlined in guidance's and industry standards and these were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence.
> Where appropriate, the tests were based on the requirements of the following documents:
- FDA guidance Coronary and Cerebrovascular Guide Wire . Guidance January 1995.
- ISO 11070:1998, Sterile Single-Use Intravascular Catheter ● Introducers.
- ISO 11135-1:2007 Sterilization of health care products-Ethylene . oxide- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
- ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile ● Barrier Systems for Medical Devices
- ISO 10993-1:2009. Biological Evaluation of Medical Devices Part ● 1: Evaluation and Testing within a risk management process, and the FDA Modified ISO 10993 Test Profile FDA Memo G95-1.
The Merit Heparin-coated InQwire® Diagnostic Guide Wire was compared to the predicate device(s) for various performance attributes that support substantial equivalence of the device. The difference in coating between the modified device and the cleared device, the Merit Medical Heparin-coated InQwire® Diagnostic Guidewire, [K822586] has raised no new issues. Performance testing was based on Predicate Device#2, the Merit Medical non Heparin-coated InQwire® Diagnostic Guidewire, [K133230].
The following is a list of all significant testing that was successfully
Safety & Performance Tests
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Merit Medical. The logo consists of a red square with a white stylized "M" inside, followed by the words "MERIT MEDICAL" in red. The word "MERIT" is solid red, while the word "MEDICAL" is outlined in red.
completed:
- Surface .
- Coating Adherence/Integrity .
- o Fracture Test
- o Flexing Test
- Coil Lubricity Test o
- Catheter Compatibility ●
- Biocompatibility .
All test results were comparable to the predicate devices and the subject Merit Heparin-coated InQwire® Diagnostic Guide Wire met the predeterminded acceptance criteria applicable to the safety and effectiveness of the device. This has demonstrated the subject device is substantially equivalent to predicate devices.
## Summary of Substantial Equivalence
Based on the Indications for Use, design, safety and performance testing, the subject Merit Medical Heparin-coated InQwire® Diagnostic Guide Wire is substantially equivalent to the predicate devices, the cleared Merit Medical Heparin-coated InQwire® Guidewire, K822586, and the Merit Medical non Heparin-coated InQwire® Diagnostic Guide Wire, K133230.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.