K140592 · Ceragem International, Inc. · JFB · Dec 17, 2014 · Physical Medicine
Device Facts
Record ID
K140592
Device Name
CERAGEM AUTOMATIC THERMAL MASSAGER
Applicant
Ceragem International, Inc.
Product Code
JFB · Physical Medicine
Decision Date
Dec 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Ceragem® Thermal Massager Model CGM-MB-1101 Automatic Thermal Massager is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for: - Temporary relief of minor muscle and joint pain stiffness - Temporary relief of minor joint pain associated with arthritis - Temporary increase in local circulation where applied - Relaxation of muscles
Device Story
Electrically powered, motorized physical therapy table; provides muscle relaxation and topical radiant infrared heat. Device consists of main mat with internal jade roller projector, lower mat, and optional external 3-sphere/9-sphere jade projectors. User lies on mat; motorized carriage moves jade rollers vertically/horizontally along torso to apply massage pressure. Internal lamps heat jade rollers; additional epoxy carbon panels and infrared heating elements provide supplemental heat. Operated via remote control; intended for home use. Heat and massage therapy aim to relieve minor muscle/joint pain, stiffness, and arthritis-associated pain, and increase local circulation. Safety features include limit switches for carriage movement and thermistors for temperature regulation.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified through compliance with UL 60601-1, IEC 60601-1, IEC 60601-1-2, ISO 10993-10, and ISO 14971 standards.
Technological Characteristics
Motorized physical therapy table; steel/ABS frame; polyester/cotton/rayon/polyurethane massager cover; Oxford cover sheet. Sensing: Thermistor-based temperature control, limit switches for carriage. Energy: 100-127Vac, 360VA. Heating: Jade rollers (12V/24Vdc lamps), epoxy carbon panels, infrared heating elements. Emission: 5-20 micron infrared spectrum. Standards: UL 60601-1, CAN/CSA-C22.2 No. 601.1, IEC 60601-1, EN 60601-1, IEC 60601-1-2, IEC 62304, ISO 10993-10, ISO 14971.
Indications for Use
Indicated for adults seeking muscle relaxation therapy, temporary relief of minor muscle/joint pain and stiffness, relief of minor joint pain associated with arthritis, and temporary increase in local circulation.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
Ceragem International, Inc. % E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114
Re: K140592
Trade/Device Name: Ceragem Automatic Thermal Massager, Model CGM-MB-1101 Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-Function Physical Therapy Table Regulatory Class: Class II Product Code: JFB Dated: November 17, 2014 Received: November 17, 2014
Dear E.J. Smith,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
Page 2 - E.J. Smith
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
for
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140592
#### Device Name
Ceragem Automatic Thermal Massager, Model CGM-MB-1101
#### Indications for Use (Describe)
The intended use of the Ceragem® Thermal Massager Model CGM-MB-1101 Automatic Thermal Massager is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:
- Temporary relief of minor muscle and joint pain stiffness
- Temporary relief of minor joint pain associated with arthritis
- Temporary increase in local circulation where applied
- Relaxation of muscles
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 807.92(c)
510(k) Summary
| Sponsor: | Ceragem International, Inc. 807.92(a)(1) |
|------------------------------------|---------------------------------------------------------|
| Address: | 3699 Wilshire Blvd., 930<br>Los Angeles, CA 90010 |
| Contact Person: | Raymond Chung |
| Phone: | 213-480-7070 |
| Fax: | 213-341-2338 |
| Summary Prepared: | November 11, 2014 |
| Device Name: | 807.92(a)(2 |
| Trade Name: | Ceragem Automatic Thermal Massager<br>Model CGM-MB-1101 |
| Common/Usual Name: | Massage Table |
| Classification Name: | Table, Physical Therapy, Multi-Function |
| Establishment Registration Number: | 2087361 |
| Product Code: | JFB |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 890.5880 |
| Manufacturer | Brand Name | 510(k) Number |
|-----------------------------|----------------------------|---------------|
| Ceragem International, Inc. | CERAGEM-C Thermal Massager | K040031 |
## Device Description
This device is comprised of the main body and optional components. The main body components are the main mat which includes the internal projector and the up/down movement motor, and the lower mat. The optional components are the external projectors (3-sphere). This device is an electrically powered, motorized, multi-functional physical therapy table intended to use for temporary relief of muscle and joint pain and stiffness by applying massage pressure from the vertical and horizontal movements, and heat from the internal and external projectors.
## Indications for Use
The intended use of the Ceragem® Thermal Massager Model CGM-MB-1101 Automatic Thermal Massager is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:
- Temporary relief of minor muscle and joint pain stiffness -
- । Temporary relief of minor joint pain associated with arthritis
- -Temporary increase in local circulation where applied
- -Relaxation of muscles
## 807.92(a)(4)
807.92(a)(5)
{5}------------------------------------------------
#### 12.3 Predicate Product Comparison Chart
807.92(a)(6)
| | | Predicate Device | Subject Device |
|------------------------------|------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Device Brand Name and | Manufacturer | CERAGEM-C Thermal Massager<br>Ceragem International, Inc. | CGM MB-1101<br>Automatic Thermal Massager<br>Ceragem International, Inc. |
| 510(k) Number | | K040031 | |
| Rated Voltage | | AC 120V 60Hz | 100-127Vac 50/60Hz |
| Operation Power | | Maximum 280 Watts | 360 VA<br>240 Watts] |
| Home Use | | Yes | Yes |
| Style | | Flood Model (Massage Bed) | Flood Model (Massage Bed) |
| Remote Control | | Yes | Yes |
| Operation Method | | Auto / Manual | Auto / Manual |
| Emission Source | | Jade Projectors | Jade Projectors |
| Infrared Emission Spectrum | | Jade: 5 ~ 20<br>Epoxy Carbon Panel: 5 ~ 20 | Jade: 5 ~ 20<br>Epoxy Carbon Panel: 5 ~ 20 |
| Heating Lamp | Voltage | 12V | 24Vdc |
| | Power | 10W | 15W |
| Operating Temperature Range | | 30°C - 60°C (86°F - 140°F) | 1. Internal : 30°C – 65°C<br>(86°F - 149°F)<br>2. External, main, Auxiliary :<br>30°C - 60°C (86°F – 140°F) |
| Extra Overheating Protection | | Yes | Yes |
| Moving Device | | Geared DC Motor | Geared DC Motor |
| Limit Detector | | Limit Switch | Limit Switch |
| Temperature Sensor | | Temperature Coefficient<br>Variable Resistor | Thermistor |
| Control Method | | Microchip's Microcontroller | Microchip's Microcontroller |
| Tugging Method | | Wire-Chain | Wire-Chain |
| Material | Support Frame | Steel | Steel, ABS |
| | Massager | Vinyl | Polyester, Cotton, Rayon, Poly<br>urethane |
| | Cover | | |
| | Cover Sheet | Cotton | Oxford |
| | Rails | Ergonomically Molded<br>Polyethylene | Ergonomically Molded<br>Polyethylene |
| Internal Projector | | Ergonomic Tilting Jade Rollers | Ergonomic Tilting Jade Rollers |
| Jade Heads | Internal<br>Projector | 1 Jade Ball in the Middle 4 Jade<br>Rollers (Total of 5) | 1 Jade Ball in the Middle 4 Jade<br>Rollers (Total of 5) |
| | External<br>Projectors | Three (3) Jade Heads Projector<br>Nine (9) Jade Heads Projector | Three (3) Jade Heads Projector<br>Nine (9) Jade Heads Projector |
| | | | |
| Mat Dimensions | | 1920mm x 550mm x 130mm<br>76 in x 22 in x 5 in<br>(L x W x H) | 1. When spread out<br>: 700mm×2016mm×450mm(±5mm)<br>2. When folded<br>:700mm×1258mm×450mm(±5mm) |
| Weight | | Approximately 38 Kg (84 lb.) | Main body: 23kg(±2kg) |
| | | Sliding support: 14kg(±2kg)<br>Frame: 16kg(±2kg) | |
| Patient Population | Adult | Adult | |
{6}------------------------------------------------
#### 12.4 Differences Table
| | | Predicate Device | Subject Device | Difference |
|-----------------------------|---------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Rated Voltage | | AC 120V 60Hz | 100-127Vac 50/60Hz | Changed SMPS<br>(switched-mode<br>power supply) |
| Operation Power | | Maximum 280 Watts | 360 VA<br>[240 Watts] | Changed SMPS<br>(switched-mode<br>power supply) |
| Heating Lamp | Voltage | 12V | 24Vdc | Changed the heating<br>element |
| | Power | 10W | 15W | |
| Operating Temperature Range | | 30°C – 60°C (86°F –<br>140°F) | 1. Internal : 30°C−65°C<br>(86°F - 149°F)<br>2. External, main,<br>Auxiliary :<br>30°C – 60°C (86°F – 140°F) | Changed internal<br>temperature range |
| Temperature Sensor | | Temperature Coefficient<br>Variable Resistor | Thermistor | No changes |
| Material | Support Frame | Steel | Steel, ABS | Added new<br>components |
| | Massager<br>Cover | Vinyl | Polyester, Cotton, Rayon,<br>Poly urethane | Changed components |
| | Cover Sheet | Cotton | Oxford | Changed components |
| | Rails | Ergonomically Molded<br>Polyethylene | Ergonomically Molded<br>Polyethylene | |
| Mat Dimensions | | 1920mm x 550mm x<br>130mm<br>76 in x 22 in x 5 in<br>(L x W x H) | 1. When spread out<br>: 700mm×2016mm×450mm<br>(±5mm)<br>2. When folded<br>: 700mm×1258mm×450mm | New design (Sliding<br>in/out structure) |
| | | (±5mm) | | |
| Static Weight | 300 | 298 pounds | Substantially<br>Equivalent | |
| Maximum Permissible Weight | 300 | 298 pounds | Substantially<br>Equivalent | |
| Weight | Approximately 38 Kg<br>(84 lb.) | Main body: 23kg(±2kg)<br>Sliding support:<br>14kg(±2kg)<br>Frame: 16kg(±2kg) | Changed design | |
{7}------------------------------------------------
#### 12.4 Discussion of Technological Characteristics
## 807.92(b)(1)
Both CGM MB-1101 and the predicate device use massage rollers and heat to provide muscle relaxation therapy. The CGM MB-1101 emits topical radiant infrared heat in the range of 5-20 microns from the jade massage rollers, heated by lamps located inside the rollers. The predicate device also emits topical heat through lamps located in the jade massage rollers in the range of 5-20 microns. Both the CGM MB-1101 device and the predicate provide the massage component by attaching the jade rollers to a carriage mounted under a padded cushion that traverses the torso. The CGM MB-1101 device also transmits heat through the two Epoxy Carbon Panels and an infrared-rays heating element. While the predicate device does not include Epoxy Carbon Panels, the maximal heat of the CGM MB-1101 remains lower than that of the predicate. Additionally, both CGM MB-1101 and the predicate device have been tested to and meet the following standards:
- 1. UL 60601-1, 1st Edition, 2006-04-26 (include National Differences for USA)
- CAN/CSA-C22.2 No. 601.1-M90 (R2005) (includes National Differences for Canada) 2.
- 3. IEC 60601-1:2005 + CORR.1 (2006) + CORR.2 (2007)
- 4. EN 60601-1:2006 + CORR: 2010 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance)
- 5. IEC 60601-1-2: 2007 Medical Electrical Equipment Part 1-2: General Requirements for Electromagnetic Compatibility
- IEC 62304 Medical device Software Software Life Cycle Processes 6.
- 7. ISO 10993-10 Biological Evaluation of medical Devices Part 10: Tests for Irritation and Skin Sensitization
- 8. ISO 14971 Medical Devices Application of Risk Management to Medical Devices
#### 12.5 Conclusions
# 807.92(b)(3)
Based on comparison of the CGM Mb-1101 to the predicate device, we conclude that the CGM MB-1101 has the same intended use, indications for use, and intended population, and similar
{8}------------------------------------------------
functional and performance characteristics. The addition of the two Epoxy panels and one infrared-rays heating element does not raise new safety or effectiveness issues since the temperature never exceeds the maximal temperature of the predicate device. Other visual distinctions do not impact safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.