← Product Code [GXY](/productcode/GXY) · K140586

# ASCEND ELECTRODE (K140586)

_Neurometrix, Inc. · GXY · Nov 17, 2014 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140586

## Device Facts

- **Applicant:** Neurometrix, Inc.
- **Product Code:** [GXY](/productcode/GXY.md)
- **Decision Date:** Nov 17, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1320
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ASCEND Electrode is intended for use as disposable, conductive, adhesive interface between the user's skin and a transcutaneous electrical nerve stimulator.

## Device Story

The ASCEND Electrode serves as a disposable, conductive, adhesive interface between a user's skin and a transcutaneous electrical nerve stimulator (TENS). It facilitates the delivery of electrical stimulation from the TENS device to the patient. The electrode is designed for single-person use and is provided non-sterile. It is intended for over-the-counter use by the patient. The device functions as a passive component in the stimulation circuit, ensuring electrical contact between the stimulator and the skin.

## Clinical Evidence

No clinical data. Bench testing only. Verification and performance testing confirmed the electrode met functional specifications and user needs. The device conforms to ISO 10993-1 for biological evaluation.

## Technological Characteristics

Disposable, conductive, adhesive cutaneous electrode. Non-sterile. Conforms to ISO 10993-1 for biological evaluation. Passive electrical interface component.

## Regulatory Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

## Predicate Devices

- NeuroMetrix SENSUS Electrode ([K121816](/device/K121816.md))

## Reference Devices

- Axelgaard ValuTrode Neurostimulation Electrodes ([K130987](/device/K130987.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines connecting them, creating a sense of unity and movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 17, 2014

NeuroMetrix, Inc. Mr. Rainer Maas Director of QA/RA 62 Fourth Ave. Waltham, MA 02451

Re: K140586

Trade/Device Name: ASCEND Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: August 18, 2014 Received: August 19, 2014

Dear Mr. Maas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aquel -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K140586

Device Name ASCEND Electrode

Indications for Use (Describe)

The ASCEND Electrode is intended for use as disposable, conductive, adhesive interface between the user's skin and a transcutaneous electrical nerve stimulator.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary for ASCEND Electrode

#### SPONSOR

NeuroMetrix, Inc. 62 Fourth Avenue Waltham, MA 02451 USA

| Contact Person: | Rainer Maas    |
|-----------------|----------------|
| Telephone:      | (781) 314-2781 |
| Date Prepared:  | March 6, 2014  |

#### DEVICE NAME

| Proprietary Name:    | ASCEND Electrode                                     |
|----------------------|------------------------------------------------------|
| Common/Usual Name:   | Transcutaneous Electrical Nerve Stimulator Electrode |
| Classification Name: | 882.1320<br>GXY<br>Cutaneous Electrode               |

#### PREDICATE DEVICE

NeuroMetrix SENSUS Electrode (K121816)

#### INTENDED USE

The ASCEND Electrode is intended for use as disposable, conductive, adhesive interface between the user's skin and a transcutaneous electrical nerve stimulator.

#### DEVICE DESCRIPTION

The ASCEND Electrode provides an electrically conductive interface between a transcutaneous electrical nerve stimulator and a user's skin. It is intended to be disposable and for single person use. It is provided non-sterile.

#### COMPARISON TO PREDICATE

The ASCEND Electrode has identical indications for use and technological characteristics as the cleared SENSUS Electrode (K121816). The only difference is that the ASCEND Electrode is labeled for overthe-counter use while SENSUS Electrode is labeled for prescription use. Many cutaneous electrodes intended for use with transcutaneous electrical nerve stimulators have been cleared by the FDA for overthe-counter use, e.g., the Axelgaard ValuTrode Neurostimulation Electrodes (K130987). Because the ASCEND Electrode has the same intended use and fundamental technology as these other electrodes, it is substantially equivalent to them.

{4}------------------------------------------------

#### GUIDANCE DOCUMENT

The recommendations of the FDA's "Draft Guidance for Industry and Staff: Class II Special Controls Guidance Document: Cutaneous Electrode (April 5, 2010)" were taken into account in preparing this 510(k) submission. The draft guidance addresses requirements for cutaneous electrodes which are defined as "an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation." This definition includes electrodes intended for use with transcutaneous electrical nerve stimulators. NeuroMetrix believes that the ASCEND Electrode complies with the special controls as outlined in the draft guidance, thereby providing additional assurance of safe and effective use of the ASCEND Electrode.

#### NON-CLINICAL TESTING

The ASCEND Electrode has the same technological characteristics as the predicate SENSUS Electrode that was validated with electrical and mechanical tests to show that the electrode met its target specifications over a range of operating and storage conditions. Verification and performance testing demonstrated that the electrode met user needs as reflected in the functional specification.

The ASCEND Electrode conforms to the following voluntary standards:

- · ISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing.
### CLINICAL TESTING

NeuroMetrix determined clinical testing was not required.

#### CONCLUSION

The ASCEND Electrode is substantially equivalent to the predicate SENSUS Electrode.

---

**Source:** [https://fda-staging.innolitics.com/device/K140586](https://fda-staging.innolitics.com/device/K140586)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
