CEDIC ENTERAL DISTAL END ENFIT TRANSITION CONNECTOR, CEDIC ENTERAL ENFIT TRANSITION CONNECTOR FOR MEDICATION PORT
K140581 · Cedic S.R.L. · PIO · Oct 17, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K140581
Device Name
CEDIC ENTERAL DISTAL END ENFIT TRANSITION CONNECTOR, CEDIC ENTERAL ENFIT TRANSITION CONNECTOR FOR MEDICATION PORT
Applicant
Cedic S.R.L.
Product Code
PIO · Gastroenterology, Urology
Decision Date
Oct 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The Cedic Enteral Distal End ENFit Transition Connector is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel. The Cedic Enteral ENFit Transition Connector for Medication Port is intended for connecting a male oral tip syringe to an enteral giving set or enteral catheter with an ENFit medication port. The Cedic Enteral Funnel ENFit Transition Connector is intended for connecting an enteral giving set equipped with a stepped enteral distal end to an enteral catheter equipped with an ENFit connector.
Device Story
Transition connectors facilitate physical interconnection between legacy enteral feeding devices and new ISO 80369-3 (ENFit) compliant devices. Three variants: Distal End ENFit (ENFit set to funnel catheter); Medication Port ENFit (male oral syringe to ENFit medication port); Funnel ENFit (stepped distal end set to ENFit catheter). Used in clinical or home settings by healthcare providers or patients. Devices provide a central fluid path for feeding fluids. Connectors mitigate misconnection risks by ensuring compatibility between incompatible end fittings. No patient contact intended; passive mechanical function only.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included enteral connector misconnection assessment, human factors, fluid leakage, stress cracking, resistance to separation from axial load, resistance to separation from unscrewing, resistance to overriding, and disconnection by unscrewing.
Technological Characteristics
Injection-molded transition connectors made of ABS HF 380, LDPE (Riblene MM20), or soft PVC (Nakan FMA919N). Rigid/semi-rigid materials selected per AAMI/ANSI/ISO 80369-1 requirements. Passive mechanical design; no energy source or software. Dimensions/form factor vary by connector type to facilitate specific legacy-to-ENFit transitions.
Indications for Use
Indicated for patients requiring enteral feeding who need to connect legacy enteral feeding devices (funnel or stepped distal ends) to new generation ENFit-compliant enteral feeding systems.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Compat Dualflo Enteral Delivery Pump Set (K080340)
Corflo Anti-IV Enteral Feeding Extension Set (K083786)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, connected by a flowing line. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 17, 2014
Cedic S.r.1. % Roger Gray VP. Ouality and Regulatory Donawa Lifescience Consulting S.r.l. Piazza Albania, 10 Rome 00153 Italy
Re: K140581 Trade/Device Name: Cedic Enteral Distal End ENFit Transition Connector Cedic Enteral ENFit Transition Connector for Medication Port Cedic Enteral Funnel ENFit Transition Connector Regulation Number: 21 CFR& 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIO Dated: September 16, 2014 Received: September 19, 2014
Dear Roger Gray,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -A
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140581
Device Name
Cedic Enteral Distal End ENFit Transition Connector
Indications for Use (Describe)
The Cedic Enteral Distal End ENFit Transition Connecting an enteral giving set with an ENF'it connector to an enteral catheter with funnel.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K140581
Device Name
Cedic Enteral ENFit Transition Connector for Medication Port
Indications for Use (Describe)
The Cedic Enteral ENFit Transition Connector for Medication Port is intended for connecting a male oral tip syringe to an enteral giving set or enteral catheter with an ENFit medication port.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda hhs gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140581
Device Name
Cedic Enteral Funnel ENFit Transition Connector
Indications for Use (Describe)
The Cedic Enteral Furnel ENFit Transition Connecting an enteral giving set equipped with a stepped enteral distal end to an enteral catheter equipped with an ENFit connector.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Cedic" in a stylized, cursive font. The word is white and has a drop shadow, giving it a three-dimensional effect. The background is a teal color. There is a small circle above the "i" in "Cedic".
# 510(k) Summary
| Device Name: | Cedic ENFit Transition Connectors for Enteral Applications |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Traditional |
| Date of Submission: | 28 February 2014 |
| Manufacturer: | Cedic S.r.l.<br>Biomedical Division<br>Via Liberazione, 63/9<br>20068 Peschiera Borromeo (MI)<br>Italy |
| 510(k) Owner and Contact: | Ms Tiziana Melis<br>Regulatory Affairs Manager<br>Cedic S.r.l.<br>Biomedical Division<br>Via Liberazione, 63/9<br>20068 Peschiera Borromeo (MI)<br>Italy |
| Phone:<br>Fax:<br>Email: | +39 02 5530 0174<br>+39 02 5530 1487<br>tiziana.melis@cedicbio.com |
| FDA Registration Number: | 3003593728 |
| Owner/Operator Number: | 9063446 |
| 510(k) Submitter and Contact: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy |
| Phone:<br>Fax:<br>Email: | +39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com |
| FDA Product Code: | PIO |
| FDA Regulation Number: | 876.5980 |
| FDA Classification Name: | Enteral Specific Transition Connectors |
| Classification Panel: | Gastroenterology and Urology |
| Common Name: | Transition Connectors for Enteral Applications |
| FDA Classification: | Class II |
| FDA Identification: | Facilitates enteral specific connections between ENFit<br>connectors and non 80369-1 compliant enteral connectors |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the word "Cedic" in a stylized, cursive font. The word is white and appears to be glowing slightly, set against a teal background that transitions from a darker shade at the top to a lighter shade at the bottom. The letter "e" in "Cedic" is stylized with a swirl above it.
#### Device Description
The introduction of new connectors to devices and accessories for enteral feeding applications, in order to help avoid misconnections with devices intended for other clinical applications, has resulted in the short-term need to connectors that will allow devices with existing ('legacy') end connectors to be connected with newer devices having end connectors meeting the relevant requirements of the ISO/IEC 80369 series of standards.
Cedic Srl has designed three transition connectors that will allow specific connections between PGLock end connectors being used on new devices to some of the previously used end connectors that may still be in use in healthcare or home settings. The three transition connectors are:
- . Enteral Distal End ENFit Transition Connector, for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel. Available with and without end cap.
- . Enteral ENFit Transition Connector for Medication Port, for connecting a male oral tip syringe to an enteral giving set or enteral catheter with an ENFit medication port. Available with and without end cap.
- . Enteral Funnel ENFit Transition Connector, for connecting an enteral giving set equipped with a stepped enteral distal end to an enteral catheter equipped with an ENFit connector. Available with and without end cap.
All three Cedic ENFit Transition Connectors are intended for prescription use only.
#### Indications for Use/Intended Use
The three ENFit Transition Connectors have the following indications for use/intended use:
The Cedic Enteral Distal End ENFit Transition Connector is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel.
The Cedic Enteral ENFit Transition Connector for Medication Port is intended for connecting a male oral tip syringe to an enteral giving set or enteral catheter with an ENFit medication port.
The Cedic Enteral Funnel ENFit Transition Connector is intended for connecting an enteral giving set equipped with a stepped enteral distal end to an enteral catheter equipped with an ENFit connector.
#### Principle of operation, mechanism of action, and interaction with the patient:
All three Cedic ENFit Transition Connectors operate by providing a means of interconnecting incompatible enteral feeding device end fittings together, so that patient enteral feeding can take place when 'new generation' ENFit end fittings need to be connected to previous designs of end fitting.
The ENFit Transition Connectors provide the mating components at each end of the connector that allow connection with the 'new' end fitting at one end of the connection with the 'old' end fitting at the other.
The ENFit Transition Connectors are not intended to come into contact with the patiental contact may occur. The ENFit Transition Connectors have a central fluid path through which feeding fluids flow during the feeding process.
#### Device Specifications
- . PGLock female to funnel connector: designed to connect an enteral giving set equipped with an ENFit female connector to an enteral catheter equipped with soft funnel port.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Cedic" in a stylized, cursive font. The word is white and has a slight shadow effect, giving it a three-dimensional appearance. The background is a solid teal color, which provides a contrasting backdrop for the white text.
- . Male oral tip syringe to PGLock male connector: designed to connect a syringe equipped with male oral tip to a medication port equipped with an ENFit male connector.
- o Stepped connector to PGLock male connector: designed to connect an enteral giving set equipped with stepped connector to an enteral catheter equipped with an ENFit male connector.
#### Manufacture
All three transition connectors are manufactured by injection molding from ABS HF 380, LDPE (Riblene MM20), or soft PVC (Nakan FMA919N). One component, an end cap, is colored with Remafin Violet PE43076356-ZT (2%). These materials and methods of manufacture have been used previously for the Cedic Enteral Feeding Spike Adapter, cleared by FDA under 510(k) reference K072652.
#### Performance Data:
In relation to performance data for such ENFit Transition Connectors, according to the FDA publication 'Draft Guidance for Industry and Food and Drug Administration Staff: Safety Considerations for 510(k) Submissions to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications', July 27, 2012:
"Adapters are used to connect enteral devices and provide additional connection points where misconnection events can occur. To mitigate misconnection via adapters, the FDA recommends that adapters be treated similarly to enteral connectors, as described in Section VI.A, B, and C within this guidance. This means that they should be made of rigid or semi-rigid materials, and mechanical testing should be performed according to recommendations described in AAMI/ANSI/ISO 80369-1, or an equivalent alternative, to demonstrate that adapters are specific and compatible for enteral applications only and are non-interconnectable with the connectors of non-enteral devices."
The FDA draft guidance includes in Section VI. A, B and C the following recommendations:
#### A. Connector materials
"FDA recommends that enteral connectors be made of rigid materials, as described in AAMI/ANS//ISO 80369-1, Clause 4, with testing according to ASTM D747 or ASTM D790, or equivalent. Use of rigid or semi-rigid materials will reduce the likelihood of forced fits between flexible connectors that are not intended to connect with each other."
In this respect, the Cedic ENFit Transition Connectors are made of the appropriate rigid or semi-rigid materials.
#### B. Mechanical testing of enteral connectors to assess incompatibility
"FDA recommends mechanical force testing of enteral connectors following AAMI/ANSI/ISO 80369-1. Clause 5.8, Annex B methods, or an equivalent alternative, to demonstrate that enteral connectors are non-interconnectable with connectors from other health care applications."
Bench tests have been carried out on samples of the Cedic ENFit Transition Connectors for Enteral Applications. The tests carried out include:
- Enteral connector misconnection assessment
- Human factors .
- Fluid leakage ●
- Stress cracking
- . Resistance to separation from axial load
- Resistance to separation from unscrewing ●
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image features the word "Cedic" in a stylized, cursive font. The word is white and has a slight shadow effect, giving it a three-dimensional appearance. Above the "ic" in "Cedic" is a small, stylized "e" symbol, also in white. The background is a gradient of teal, with the lighter shade at the top and a darker shade at the bottom.
- Resistance to overriding ●
- . Disconnection by unscrewing
#### C. Enteral connector risk assessment
"When an applicant submits a new 510(k) application, they should provide a risk assessment to demonstrate they have effectively mitigated the risk of misconnection with their new product. There should be objective evidence that risks have been reduced to acceptable levels according to ISO 14971:2007 or equivalent. For example, the applicant may provide evidence of selection of appropriate material (Section VI.A, above) and quantitative mechanical testing data to demonstrate that the proposed enteral connector has a reduced risk of forming stable attachments to connectors routing into non-enteral devices (Section VI.B, above)."
In this respect, this 510(k) submission includes a copy of the risk analysis for the three subject transition connectors.
#### Predicate devices
The predicate device selected for comparison with the Cedic Enteral Distal End ENFit Transition Connector is:
| Predicate Device: | Compat Dualflo Enteral Delivery Pump Set With<br>Spikeright Piercing Spike and 1000ml water bag |
|----------------------|-------------------------------------------------------------------------------------------------|
| 510(k) Sponsor: | Novartis Nutrition Corp. |
| 510(k) Number: | K080340 |
| Clearance Date: | 5 March 2008 |
| FDA Product Code: | KNT |
| Classification Name: | Gastrointestinal tube and accessories |
| Regulation No: | 876.5980 |
The predicate device selected for comparison with the Cedic Enteral ENFit Transition Connector for Medication Port is:
| Predicate Device: | Corflo Anti-IV Enteral Feeding Extension Set,<br>Models 20-1006AIV-LB and 20-1030AIV-LB |
|----------------------|-----------------------------------------------------------------------------------------|
| 510(k) Sponsor: | Corpak Medsystems |
| 510(k) Number: | K083786 |
| Clearance Date: | 20 February 2009 |
| FDA Product Code: | FPD |
| Classification Name: | Tube, Feeding |
| Regulation No: | 876.5980 |
The predicate device selected for comparison with the Cedic Enteral Funnel ENFit Transition Connector is:
| Predicate Device: | Corflo Anti IV Enteral Feeding Tube |
|----------------------|-------------------------------------|
| 510(k) Sponsor: | Corpak Medsystems |
| 510(k) Number: | K083210 |
| Clearance Date: | 29 December 2008 |
| FDA Product Code: | FPD |
| Classification Name: | Tube, Feeding |
| Regulation No: | 876.5980 |
The Subject Devices and the end connectors of the Predicate Devices both share many identical or similar properties and features, and none of the differences have any significant effect on safety or effectiveness of the Subject Devices.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Cedic" in a stylized, cursive font. The word is white and has a slight shadow effect, giving it a three-dimensional appearance. Above the "e" in "Cedic" is a small, circular design that resembles the at symbol (@).
A fourth predicate device is included with this submission to demonstrate substantial equivalence with a further Cedic interconnecting connector, this being:
| Predicate Device: | Enteral Feeding Spike Adaptor |
|----------------------|---------------------------------------|
| 510(k) Sponsor: | Cedic |
| 510(k) Number: | K072652 |
| Clearance Date: | 18 January 2008 |
| FDA Product Code: | KNT |
| Classification Name: | Gastrointestinal tube and accessories |
| Regulation No: | 876.5980 |
The subject devices share many aspects with this predicate device, including materials, manufacturing processes and methods, packaging components and processes, and biocompatibility data.
#### Conclusion:
Based on the information contained within this submission, it is concluded that the Cedic ENFit Transition Connectors for Enteral Applications are substantially equivalent to the identified predicate devices already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.