K140548 · Corcam Technologia, SA · DSI · May 23, 2014 · Cardiovascular
Device Facts
Record ID
K140548
Device Name
CORCAM HEART MONITOR
Applicant
Corcam Technologia, SA
Product Code
DSI · Cardiovascular
Decision Date
May 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140548 · May 23, 2014
CORCAM HEART MONITOR
Corcam Technologia, SA
Medical records; Clinical patient cohort
A cross-sectional observational study was conducted to assess the accuracy of the Corcam Heart Monitor in detecting arrhythmias and myocardial ischemia in a real-world clinical population compared to conventional ECG.
Cross-sectional observational study; Cross-sectional observational study; Study Period: May to August 2012
152 consecutive inpatients and outpatients with established clinical diagnoses (atrial fibrillation, acute coronary syndromes, ventricular arrhythmias, or bradyarrhythmias); Sample Size: 152
Conventional ECG
Accuracy of arrhythmia and myocardial ischemia detection; transmission time to Central Monitoring Station
Indications for Use
The Corcam Heart Monitor is a portable device intended to continuously monitor a patient's cardiac condition for early signs of arrhythmia. 1. Adult patients who have demonstrated a need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease. 2. Adult patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: dizziness or lightheadedness; syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and dyspnea (shortness of breath). 3. Adult patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms. 4. Adult patients who require outpatient monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation).
Device Story
Portable, battery-operated, non-invasive cardiac monitor; connects to patient via four electrodes. Inputs: ECG signals, accelerometer data (fall detection), GPS location. Operation: continuous monitoring; detects irregular heartbeats; generates ECG; transmits data via GSM/GPRS to central monitoring station. Features: SOS/help button, speaker for two-way communication, vibrating/visual alerts. Used in outpatient settings; operated by patient. Clinician at monitoring center reviews transmitted ECGs; may request additional exams or initiate emergency services. Benefits: enables remote, autonomous tracking of cardiac events and timely clinical intervention.
Clinical Evidence
Cross-sectional observational study (n=152) comparing Corcam device to conventional ECG (gold standard). Patients included those with established arrhythmias or myocardial ischemia. Primary endpoint: ability to detect ECG changes and transmit data to central station without human intervention. Results: 92% of ECGs transmitted within 3 minutes; remaining 10% within 5 minutes during peak network traffic. No adverse events reported.
Technological Characteristics
Battery-operated, portable, non-invasive. Connectivity: GSM/GPRS for remote data transmission. Sensors: 4-electrode ECG, accelerometer, GPS. User interface: speaker, vibrating mechanism, indicator lights, SOS button. Standards: NBR IEC 60601-1, 60601-1-2, 60601-2-27, AAMI/ANSI EC38, AAMI/ANSI EC57, IEC 62209-01/02, IEEE 1528, ETSI TS 151 010-1.
Indications for Use
Indicated for adult patients requiring continuous cardiac monitoring for non-life-threatening arrhythmias (e.g., AFib, flutter, PACs, PSVT, ventricular ectopy), bradyarrhythmias, bundle branch block, or symptoms (dizziness, syncope, dyspnea, palpitations). Also indicated for monitoring drug effects on ventricular rate in atrial arrhythmias.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
CardioNet ECG Monitor (DSI)
Body Guardian Monitor (MHX)
Applied Cardiac Systems CORE™ ECG Monitor (DSH)
Submission Summary (Full Text)
{0}------------------------------------------------
# Section 5: 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for Corcam's 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
The safety and effectiveness of the Corcam Heart Monitor is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate device(s).
| Sponsor: | Corcam Technologica S/A<br>Rua Sampaio Viana, n. 202, conjs. 104 / 105, Paraíso,<br>São Paulo, Brazil |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jennifer A. Daudelin<br>M Squared Associates<br>575 Eighth Avenue, Suite 1212<br>New York, NY 10018<br>Ph: 703-562-9800 x251<br>Fax: 703-562-9797<br>Email: jdaudelin@msquaredassociates.com |
| Date Prepared: | March 3, 2014 |
| Proposed Class: | II |
| Proprietary Name: | Corcam Heart Monitor |
| Common Name: | Arrhythmia Detector and Alarm |
| Classification Name: | Arrhythmia Detector and Alarm |
| Regulation Number: | 21 CFR 870.1025 |
| Product Codes: | DSI |
Class Manufacturer Device Name Common Name Procode II CardioNet ECG DSI Arrhythmia Detector and CardioNet, Inc. Alarm Monitor with Arrhythmia Detection Patient Physiological II MHX, Preventice, Inc. Body Guardian Monitor, Arrhythmia DSI Detector and Alarm II DSI, Arrhythmia Detector and Applied Cardiac CORE™ ECG Monitor Systems DSH, Alarm DPS
{1}------------------------------------------------
## Indication for Use
、i
The Corcam Heart Monitor is a portable device intended to continuously monitor a patient's cardiac condition for early signs of arrhythmia.
- 1. Adult patients who have demonstrated a need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease.
- 2. Adult patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: dizziness or lightheadedness; syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and dyspnea (shortness of breath).
- 3. Adult patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms.
- 4. Adult patients who require outpatient monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation).
## Device Description
The Corcam Heart Monitor is a battery operated, non-invasive, portable device intended to continuously monitor a patient's cardiac condition for early signs of arrhythmia. The device allows for autonomous remote monitoring via a Global System for Mobile Communications/General Packet Radio Service (GSM/ GPRS). The device is connected to the patient via four commercially available electrodes (Derivations I. II. III. and ground) and is equipped with an accelerometer to detect patient falls, GPS for geographical and accurate location of the user, and a speaker system to allow communication to or with the Monitoring Center. A vibrating mechanism and lights, when activated, provide physical and visual signals to the user. Indicator lights also show the operating status of the unit. An SOS/ Help button can be pressed by the user to send a help signal to the monitoring center. In addition to the transmission of patient data to the monitoring center, the monitoring center or clinician at any given moment can conditionally request a new exam on demand allowing for actual and accurate tracking of a cardiac event. Once the monitor detects an irregular heartbeat, the device generates an electrocardiogram (ECG) and sends it to the monitoring center. A clinician in the monitoring center carries out an analysis on the received ECG. Depending upon the analysis of the clinician can check the symptoms of the patient, request another ECG transmission, or initiate immediate assistance for the form of an ambulance/ emergency services.
{2}------------------------------------------------
## Performance Data (Non-Clinical)
The Corcam Heart Monitor meets the requirements of the following standards:
- NBR IEC 60601-1:1994, Electromedical Equipment Part 1 General Safety Prescriptions ●
- NBR IEC 60601-1, Amendment 1: 1997 .
- ABNT NBR IEC 60601-1-2:2006, Electromedical Equipment Part 1-2: Collateral . Standard: Electromagnetic Compatibility - Requirements and tests.
- ABNT NBR IEC 60601-2-27:1997, Medical Electrical Equipment Part 2: Particular . Requirements For the Safety of Electrocardiographic Monitoring Equipment
- · NBR IEC 60601-2-27:1997, Part 2: Specific Prescriptions for the Equipment Safety for the Electrocardiogram Software
- IEC 62209-01: 2005, Human exposure to radio frequency fields from hand-held and body-mounted . wireless communication devices- Human models, instrumentation, and procedures - Part 1: Procedure to determine the specific absorption rate (SAR) for hand-held devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz
- IEC 62209 02: 2010, Procedure to determine the Specific Absorption Rate (SAR) in the head and . body for 30 MHz to 6 GHz Handheld and Body-Mounted Devices used in close proximity to the Bodv
- IEEE Std 1528, IEEE Recommended Practice for Determining the Peak Spatial- Average Specific . Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement Techniques
- ETSI TS 151 010-1 V6.5.0 (2005-11) (3GPP TS 51.010-1 version 6.5.0 Release 6), Digital . cellular telecommunications system (Phase 2+); Mobile Station (MS) conformance specification; Part 1: Conformance specification
- AAMI / ANSI EC38: 2007, Medical Electrical Equipment Part 2-47: Particular Requirements for . the safety, including essential performance, of ambulatory electrocardiographic systems. (Cardiovascular)
- AAMI/ANSI EC 57 1998/(R ) 2003 Testing and Reporting Performance Results of Cardiac . Rhythm and ST Segment Measurement Algorithms
## Performance Data (Clinical)
A cross-sectional observational study was designed to assess the accuracy of this monitor as compared to that of conventional ECG (gold standard) as a screening tool for the detection of arrhythmias and myocardial ischemia. From May to August 2012, 152 consecutive patients were included in the study. Patients were inpatients or outpatients with established clinical diagnoses (as reported in medical records) and ECGs meeting the eligibility criteria of the study. Inclusion criteria were patients over 18 years of age with cardiac arrhythmias or myocardial ischemia and an established diagnosis of attial fibrillation, acute coronary syndromes, or ventricular arrhythmias or bradyarrhythmias. All patients underwent conventional electrocardiography and had an electrocardiogram (ECG) generated by the Corcam device. For the Corcam Heart Monitor ECGs, the device was installed onto the patient and automatically generated and sent an ECG to the Central Monitoring Station without human intervention, either from the patient or the investigator. Both tests were performed within a 10-minute window. There were no adverse
{3}------------------------------------------------
events. Results showed that in 92% of the cases evaluated, the Central Monitoring Station received the ECGs automatically generated by the Corcam device within 3 minutes. ECGs that took between 4 and 5 minutes to send were generated at peak times, when there is greater data traffic over the cellular network. The study concluded that the Corcam Heart Monitor has the ability to detect relevant ECG changes and generate an electrocardiogram and transmit it automatically (without human intervention) within minutes to a Central Monitoring Station.
# Technological Characteristics and Substantial Equivalence
The Corcam Heart Monitor has the same indications for use and similar design features as compared with the predicate arrhythmia detectors. Additionally, both the Corcam device and the predicate CORE™ ECG Monitor use an accelerometer to detect accelerated motion. The bench testing along with the clinical data demonstrate that the performance characteristics of the Corcam Heart Monitor is equivalent to those of other legally marketed arrhythmia detectors, and therefore supports a determination of Substantial Equivalence for the proposed indications for use.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure embracing the world.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002
May 23, 2014
Corcam Technologia, SA % Ms. Jennifer Daudelin Consultant for Corcam M Squared Associates, Inc. 575 Eighth Ave, Ste 1212 New York, New York 10018
K140548 Re:
Trade/Device Name: Corcam Heart Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm Regulatory Class: Class II Product Code: DSI Dated: March 25, 2014 Received: March 26, 2014
Dear Ms. Jennifer Daudelin:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave feviewed your bection by the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce pror to May 20, 1977 in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, access, mains of the Act include requirements for annual registration, listing of general controls provisions gractice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is classified (500 actrols. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
### Page 2 - Ms. Jennifer Daudelin
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/5/Picture/8 description: The image shows a logo with the letters FDA. The letters are stylized and have a bold, geometric design. The logo appears to be a black and white image. The logo is likely associated with the Food and Drug Administration.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
K140548 510(k) Number:
Corcam Heart Monitor Device Name:
### Indications for Use:
The Corcam Heart Monitor is a portable device intended to continuously monitor a patient's cardiac condition for early signs of arrhythmia.
- 1. Adult patients who have demonstrated a need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease.
- 2. Adult patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: dizziness or lightheadedness; syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and dyspnea (shortness of breath).
- 3. Adult patients with palpitations without known arrhythmias to obtain correlation of rhythm with symptoms.
- 4. Adult patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation).
AND/OR
Prescription Use ___X_ (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of __ 1 __
FDA 2014.05.23
0651-04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.