K140543 · Merit Medical Systems, Inc. · DYB · Oct 16, 2014 · Cardiovascular
Device Facts
Record ID
K140543
Device Name
PRELUDEEASE HYDROPHILIC SHEATH INTRODUCER
Applicant
Merit Medical Systems, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Oct 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Merit PreludeEASE™ Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic or therapeutic procedures. The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires.
Device Story
Sheath introducer system facilitates percutaneous vascular access; provides conduit for diagnostic/therapeutic devices into veins/arteries (e.g., radial artery). Components include sheath with hydrophilic coating, vessel dilator, sidearm with 3-way stopcock, integral hemostasis valve, and suture ring. Accessory access needle (inner metal needle/outer plastic cannula) used for initial vessel entry and guide wire placement. Kit may include guide wire, syringe, scalpel, and BowTie™ insertion device. Used in clinical settings by medical professionals. Device maintains hemostasis during procedure; hydrophilic coating reduces friction during insertion. Benefits include reliable vascular access and simplified device introduction.
Clinical Evidence
Bench testing only. No clinical data. Performance testing conducted per ISO 11070:1998 and other standards, including mechanical testing (leak, pull force, kink resistance, lubricity, particulate evaluation) and biocompatibility (cytotoxicity, sensitization, systemic toxicity, genotoxicity, hemolysis, pyrogenicity).
Technological Characteristics
Hydrophilic-coated sheath introducer with vessel dilator. Features integral hemostasis valve, sidearm with 3-way stopcock, and suture ring. Materials evaluated per ISO 10993 series. Sterilized via ethylene oxide (ISO 11135-1). Packaging validated per ISO 11607-1 and ASTM D4169-09. Mechanical testing per ISO 11070:1998.
Indications for Use
Indicated for patients requiring percutaneous vascular access (veins/arteries, including radial artery) for diagnostic or therapeutic procedures.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. The faces are facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2014
Merit Medical Systems, Inc. Ms. Michaela Rivkowich Principal Regulatory Affairs Specialist 1600 West Merit Pkwv. South Jordan, Utah 84095
Re: K140543
> Trade/Device Name: PreludeEase Hydrophilic Sheath Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: September 3, 2014 Received: September 4, 2014
Dear Ms. Rivkowich,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140543
Device Name PreludeEASE Hydrophilic Sheath Introducer
Indications for Use (Describe)
The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries. including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.
The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4196<br>(801) 253-6932<br>Michaela Rivkowich<br>October 16, 2014<br>1721504 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | PreludeEASE™ Hydrophilic Sheath Introducer<br>Sheath Introducer<br>Catheter Introducer |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Prelude® PSI Sheath Introducer<br>Vessel Dilator for Percutaneous<br>Catheterization<br>K070159<br>Merit Medical Systems, Inc. |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Glidesheath<br>Catheter Introducer<br>K102008<br>Terumo Medical Corporation |
| Classification | Class II<br>21 CFR § 870.1340<br>FDA Product Code: DYB<br>Review Panel: Cardiovascular | |
| Intended Use | The Merit PreludeEASE™ Hydrophilic Sheath Introducer is intended to<br>provide access and facilitate the percutaneous introduction of various<br>devices into veins and/or arteries, including but not limited to the radial<br>artery, while maintaining hemostasis for a variety of diagnostic or<br>therapeutic procedures.<br>The access needle with inner metal needle and outer plastic cannula is<br>used to gain access to the vein or artery for placement of guide wires. | |
| Device<br>Description | The PreludeEASE™ Hydrophilic Sheath Introducer consists of a sheath<br>introducer with compatible vessel dilator that snaps securely into the<br>sheath introducer hub. The sheath is coated with a hydrophilic coating<br>and is equipped with a sidearm terminating in a 3-way stopcock. The<br>sheath hub contains an integral hemostasis valve and suture ring. The<br>access needle with inner metal needle and outer plastic cannula is an<br>accessory device which is used to gain access to the vein or artery for<br>placement of guide wires. | |
| | The PreludeEASE Hydrophilic Sheath Introducer is marketed with any<br>of the following components, depending on the product configuration:<br>guide wire, metal access needle, access needle with inner metal needle<br>and outer plastic cannula, syringe, scalpel and BowTie™ guide wire<br>insertion device. | |
| Comparison<br>to Predicate<br>Device | The technological characteristics of the subject PreludeEASE<br>Hydrophilic Sheath Introducer are substantially equivalent to those of<br>the predicate devices. The subject device has the same basic design<br>as the predicate devices in that it consists of a sheath tubing, hub,<br>sidearm and stopcock and is provided with a vessel dilator. The main<br>difference is in the device materials. Just as the predicate devices, the<br>PreludeEASE is provided with kit components typically used to obtain<br>vascular access. | |
| Safety &<br>Performance<br>Tests | No performance standards have been established under Section 514 of<br>the Food, Drug and Cosmetic Act for these devices. Performance<br>testing of the subject PreludeEASE Hydrophilic Sheath Introducer was<br>conducted based on the risk analysis and based on the requirements of<br>the following international standard:<br>ISO 11070:1998, Sterile, single-use intravascular catheter<br>introducers ISO 11135-1:2007, Sterilization of health care products – Ethylene<br>oxide - Part 1: Requirements for development, validation and<br>routine control of a sterilization process for medical devices ISO 11607-1:2010, Packaging for terminally sterilized medical<br>devices Part 1: Requirements for materials, sterile barrier systems<br>and packaging systems ASTM D4169-09, Standard Practice for performance Testing of<br>Shipping Containers and Systems ISO 2233:2000, Packaging – Complete, filled transport packages<br>and unit loads – Conditioning for testing ISO 10993-1:2009, Biological Evaluation of Medical Devices Part<br>1: Evaluation and Testing within a risk management process, and<br>FDA guidance Required Biocompatibility Training and Toxicology<br>Profiles for Evaluation of Medical Devices, May 1, 1995 ANSI/AAMI/ISO 10993-3:2003, Biological Evaluation of Medical<br>Devices – Part 3: Tests for Genotoxicity, Carcinogenicity and<br>Reproductive Toxicity ISO 10993-4:2002 (Amd.1:2006), Biological evaluation of medical<br>devices - Part 4: Selection of tests for interactions with blood ISO 10993-5:2009, Biological evaluation of medical devices – Part<br>5: Tests for in vitro cytotoxicity ISO 10993-7:2008, Biological evaluation of medical devices – Part<br>7: Ethylene oxide sterilization residuals ISO 10993-10:2010, Biological evaluation of medical devices –<br>Part 10: Tests for irritation and skin sensitization ISO 10993-11:2006, Biological evaluation of medical devices –<br>Part 11: Tests for systemic toxicity ASTM F756-08, Standard practice for assessment of hemolytic<br>properties of materials United States Pharmacopeia 35, National Formulary 30, 2012<br><151> Pyrogen Test | |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
### Performance Testing-Bench
- Surface
- Radiodetectability ●
- Size Designation ●
- Sheath Introducer Assembly Leak ●
- Sheath Introducer Hemostasis Valve 5.8 psi Leak ●
- Sheath Introducer Tubing Pull (Force at Break) ●
- Sheath Introducer Hub to Tubing Pull (Force at Break) ●
- Dilator Hub to Tubing Pull (Force at Break)
- Sheath / Dilator Tip Peel Back
- Hemostasis Valve Dislodgement
- Hemostasis Valve Vacuum
- Hemostasis Valve Low Pressure with Dilator Dwell ●
- Sidearm to Sheath Hub Pull (Force at Break)
- Sheath Hub to Cap Pull (Force at Break) ●
- Sheath Introducer Stiffness and Flex Modulus ●
- Sheath Introducer Kink Resistance
- Suture Ring Pull Test
- Dilator Stiffness and Flex Modulus
- Hydrophilic Coating Lubricity
- Hydrophilic Coating Coverage/Adherence and Length
- Two-Part Access Needle Metal Cannula to Hub Pull ●
- Two-Part Access Needle Plastic Cannula to Hub Pull
- Two-Part Access Needle Insertion and Withdrawal Test
- Particulate Evaluation ●
- . Kink Resistance
- Torque Strength .
- Packaging Testing ●
- Scalpel Flex Test ●
## Biocompatibility
- Cytotoxicity ●
- Sensitization ●
- Intracutaneous Injection ●
- Acute Systemic Toxicity
- Pyrogenicity ●
- Genotoxicity ●
- Hemolysis ●
- Thrombogenicity ●
- Complement Activation ●
- Chemical Characterization ●
## Safety & Performance Tests cont.
{7}------------------------------------------------
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, safety and performance<br>testing. the subject PreludeEASE Hydrophilic Sheath Introducer meets<br>the requirements that are considered essential for its intended use and<br>is substantially equivalent to the predicate devices, the Prelude PSI<br>Sheath Introducer, K070159 and the Glidesheath, K102008. |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.