tenure 4G is intended to be used as a resin tooth bonding agent. It is used for composite bonding of porcelain and metals, as well as for desensitizing root, dentin or crown prior to impressions or temporaries.
Device Story
tenure 4G is a multi-purpose, self-cure dental adhesive system; consists of polymerizable dental monomer resins; two-part system mixed by clinician; forms leak and stain-resistant bonds between dental surfaces and restorations; used in dental clinics by dentists; applied to intraoral surfaces including porcelain, ceramics, metals, and tooth structure; provides bonding for direct/indirect restorations and desensitization of dentin/root surfaces; benefits patient by securing restorations and reducing tooth sensitivity.
Clinical Evidence
Bench testing only. Laboratory shear bond testing performed on porcelain, enamel, and dentin substrates. Results showed mean bond strengths (MPa) of 13.149 (porcelain), 12.768 (enamel), and 26.035 (dentin), demonstrating effectiveness comparable to or exceeding predicate devices.
Indicated for routine direct and indirect resin composite bonding; porcelain, ceramic veneer, amalgam, precious and semi-precious metal bonding; indirect gold, porcelain, and ceramic inlay/onlay bonding; and desensitization of root, dentin, or crown prior to impressions or temporaries.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with wavy lines representing hair or flowing fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 20, 2014
DenMat Holdings, LLC C/O Ms. Helen Ragus Regulatory Specialist 1017 W. Central Avenue Lompoc, California, 93436
Re: K140537
Trade/Device Name: tenure®4G Regulation Number: 21 CFR 872.3200 Regulation Name: Resin tooth bonding agent Regulatory Class: II Product Code: KLE Dated: May 27, 2014 Received: May 27, 2014
Dear Ms. Ragus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Ms. Ragus
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510{k) Number (if known)
K140537
Device Name tenure@4G
Indications for Use (Describe)
tenure@4G is recommended for the following types of applications:
1) All routine direct and indirect resin composite bonding
2) Porcelain, ceramic veneers, amalgams, precious and semi-precious metals bonding
3) Indirect gold, porcelain and ceramic inlays and onlays bonding
4) Desensitization of root or dentin prior to impressions or temporaries
5) Preparation desensitization of crown prior to impressions or temporaries
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
B
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for "DenMat". The logo consists of the letters "dm" in a stylized font, with "DenMat" written below the letters. The letters "dm" are connected at the top, and the "d" is slightly smaller than the "m". There is a trademark symbol to the right of the word "DenMat".
1017 West Central Avenue
805-346-3700 www.denmat.com
K140537
### V. 510(k) SUMMARY
#### Submitter:
| Owner's Name: | DenMat Holdings, LLC |
|---------------------------------|----------------------------------------------------------------------------------|
| Address: | 1017 W. Central Avenue<br>Lompoc, CA 93436<br>U.S.A. |
| Phone Number: | 805 346 3700 |
| Fax Number: | 805 347 7940 |
| Contact Person: | Helen Ragus<br>Regulatory Specialist<br>805 346 3700, X2932<br>hragus@denmat.com |
| Date of Summary<br>Preparation: | February 24, 2014 |
| Device Name: | |
| Trade Name: | tenure* 4G |
| Common Name: | Bonding Agent |
| Classification Name: | Agent, Tooth Bonding |
| Predicate Devices: | |
| ALL-BOND 2 by Bisco | K910860 |
| Tenure MPB<br>(Den-Mat Tenure) | K872510 |
:
{4}------------------------------------------------
#### Description of the Device
tenure * 4G is DenMat's next generation of its popular Tenure MPB System with greater bond strength and sensitivity control. It is a 40 generation multi-purpose, self-cure adhesive system for bonding any resin restorative to all intraoral surfaces. They are polymerizable dental monomer resins that are chemically-cured with the reaction initiated when the two parts are mixed together. These polymers form strong leak and stain resistant bonds between the dental surface and restorations placed over them.
#### Intended Use of the Device
tenure #G is intended to be used as a resin tooth bonding agent. It is used for composite bonding of porcelain and metals, as well as for desensitizing root, dentin or crown prior to impressions or temporaries.
#### Indications of Use of the Device
tenure 4G is recommended for the following types of applications:
- 1) All routine direct and indirect resin composite bonding.
- 2) Porcelain, ceramic veneers, amalgams, precious and semi-precious metals bonding.
- 3) Indirect gold, porcelain and ceramic inlays and onlays bonding.
- 4) Desensitization of root or dentin prior to impressions and temporaries.
- 5) Preparation desensitization of crown prior to impression or temporaries.
#### Substantial Equivalence Discussion
- 1) Intended Uses/Indications for Use
tenure 4G and the predicate devices are intended to be used as resin tooth bonding agents. They are used for composite bonding of porcelain and metals, as well as for desensitizing root, dentin or crown prior to impressions or temporaries.
tenure 4G and Tenure MPB are recommended for the following types of applications: all routine direct and indirect resin composite bonding; porcelain, ceramic veneers, amalgams, precious and semi-precious metals bonding; indirect gold, porcelain and ceramic inlays and onlays bonding; desensitization of root or dentin prior to impressions and temporaries; and preparation desensitization of crown prior to impression or temporaries.
ALL-BOND 2 applications include bonding to dentin, enamel, new or old composite, precious and non-precious casting alloys, silane treated porcelain, and new or old amalgam; and treatment of tooth sensitivity.
{5}------------------------------------------------
Scientific literature have been evaluated to determine safety and efficacy of similar products used for the same indications. The intended uses and indications for use of the subject device are substantially equivalent to those of the predicate devices.
- 2) Chemical Components/Safety
Chemical components in tenure 4G have been used in the predicate devices. Scientific literature have been evaluated to determine safety and efficacy of similar products used for the same indications. The predicate devices have not been a focus of any advisory notice or recalls according to the post-market adverse event reporting requirements in the United States. The conclusion can be made that the safety of the subject device is substantially cquivalent to those of the predicate devices.
- 3) Technological Characteristics/Effectiveness and Performance
There are no international standards concerning performance for these types of devices. Scientific literature have been evaluated to determine safety and efficacy of similar products used for the same indications. The physical/mechanical properties of tenure 4G were tested in the lab using R&D test protocols. Results of shear bond testing indicate that tenure 4G was as effective and performs as good as or even better than the predicate devices. The conclusion can be made that the effectiveness and performance of the subject device is substantially equivalent to those of the predicate devices.
#### Biocompatibility
The subject device is categorized as an external communicating device with contact to tissue/bone/dentin and is a permanent contact device. tenure "4G's chemical ingredients are equivalent to those of the predicate devices. All devices are made of materials with a long history of safe use. The predicate devices have not been a focus of any advisory notice or recalls according to the post-market adverse event reporting requirements in the United States. Accordingly. conclusion can be made that the subject device is substantially equivalent in safety to the predicate devices.
| tenure* 4G | | | | | | |
|------------|-----------|--------|---------|--------|---------|--------|
| | Porcelain | | Enamel | | Dentin | |
| | psi | MPa | psi | MPa | psi | MPa |
| Mean | 1907.14 | 13.149 | 1851.84 | 12.768 | 3776.07 | 26.035 |
| Stdev | 294.42 | 2.030 | 512.14 | 3.531 | 478.64 | 3.300 |
#### Comparative Performance Data
{6}------------------------------------------------
| | Porcelain | | | Enamel | | | Dentin | |
|-------|-----------|--------|-------|---------|-------|-------|---------|--------|
| | psi | MPa | | psi | MPa | | psi | MPa |
| Mean | 1627.21 | 11.219 | Mean | 1147.21 | 7.910 | Mean | 2830.62 | 19.516 |
| Stdev | 656.00 | 4.523 | Stdev | 549.74 | 3.790 | Stdev | 217.68 | 1.501 |
ALL-BOND 2
| Tenure MPB | | | | | | |
|------------|---------|--------|--------------|--------|--------------|--------|
| Porcelain | | Enamel | | Dentin | | |
| | psi | MPa | psi | MPa | Psi | MPa |
| Mean | 2041.29 | 14.074 | Mean 1485.50 | 10.242 | Mean 3027.60 | 20.875 |
| Stdev | 177.37 | 1.223 | Stdev 659.45 | 4.547 | Stdev 410.47 | 2.830 |
## Summary of Features and Characteristics of the Device Compared to the Predicate Device:
| Product | 510(k) | Applications | Chemical<br>Compositon | Technique<br>Application |
|----------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------|
| tenure®4G | | - All routine direct and indirect<br>resin composite bonding<br>- Porcelain, ceramic veneers,<br>amalgams, precious and semi-<br>precious metals bonding<br>- Indirect gold, porcelain and<br>ceramic inlays and onlays<br>bonding<br>- Desensitization of root or<br>dentin prior to impressions and<br>temporaries<br>- Preparation desensitization of<br>crown prior to impression or<br>temporaries | - Self-cure<br>- Methacrylate<br>resin based<br>- Contains<br>acetone solvent<br>- Contains<br>ethanol solvent | Total-Etch |
| ALL-<br>BOND 2 | K910860 | - All dental surfaces including<br>precious and non precious<br>casting alloys, and amalgam<br>bonding<br>- Root sensitivity treatment | - Self-Cure<br>- Methacrylate<br>resin based<br>- Contains<br>acetone solvent<br>- Contains<br>ethanol solvent | Total-Etch |
{7}------------------------------------------------
| Tenure<br>Multi-<br>purpose<br>Bonding | K872510 | - All routine direct and indirect<br>resin composite bonding<br>- Porcelain, ceramic veneers,<br>amalgams, precious and semi-<br>precious metals bonding<br>- Indirect gold, porcelain and<br>ceramic inlays and onlays<br>bonding<br>- Desensitization of root or<br>dentin prior to impressions and<br>temporaries<br>- Preparation desensitization of<br>crown prior to impression or<br>temporaries | - Self-Cure<br>- Methacrylate<br>resin based<br>- Contains<br>acetone solvent | Total -Etch |
|----------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------|
|----------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------|
#### Conclusion
The information provided in this 510(k) submission demonstrates that tenure *40 is substantially equivalent to the predicate devices All-Bond 2 and Tenure MPB in terms of intended use, indications for use, chemical composition and physical properties.
It is concluded that the information supplied in this submission has proven the safety and efficacy of this product.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for "DenMat". The logo consists of the letters "dm" in a stylized font, with "DenMat" written below in a smaller font. The letters "dm" are connected, with the "d" on the left and the "m" on the right. The logo is in black and white.
K140537/52
FDA CDRH DMC MAY 2 7 2014 Received
1-15
#### 510(k) COVER LETTER
U.S. Food and Drug Administration Center for Devices and Radiological Heath Document Control Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
Reference: K140537/S001
| Type of Submission: | Traditional |
|-------------------------------|----------------------|
| Trade Name: | tenure® 4G |
| Common Name: | Dental Bonding Agent |
| Classification Regulation: | 872.3200 |
| Device Class: | II |
| Panel: | Dental |
| Product Code: | KLE |
| 510(k) Submitter: | DenMat Holdings, LLC |
| Establishment Registration #: | 2018957 |
| Contact Person: | Helen Ragus |
Phone Number:
Dear Sir or Madam,
In Reference to 510(k) number K140537/S001, please find the attached response addressing the elements identified as missing or inconsistent in the provided checklist attached to Acceptance Review Notification - Refuse To Accept (RTA) dated May 21, 2014.
Regulatory Specialist
805 346-3700, X2932
In addition, 2 pages from the response to the FDA Acceptance Review Notification - Refuse to Accept dated March 18, 2014 and received by the FDA on 5/8/14 (eCopy 5/13/14) are also provided with corrected information.
- 1) In Response to Section C. Substantial Equivalence Discussion, the 510(k) predicate device number for Tenure MPB (K801216) was listed in error instead of the correct 510(k) K872510.
- 2) In Response to Section G. Biocompatibility, water was listed with the incorrect chemical name. The corrected information simply listed it as water.
Corrections were also made to the initial submission received by the FDA on 3/4/14 where the incorrect predicate device for Tenure MPB (K801216) was listed instead of the correct 510(k) K872510.
- 1) I. 510(k) Cover Letter (page 2)
1017 W. Central Avenue, Lompoc, CA 93436 denmat.com
P: 805-346-3700 F: 805-347-7933
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo for "DenMat". The logo consists of the letters "d" and "m" in a stylized font, with the "d" slightly overlapping the "m". Below the letters "dm" is the word "DenMat" in a smaller, sans-serif font. The logo is black and white.
- 2) V. 510(k) Summary (Summary of Features and Characteristics of the Device Compared to the Predicate Device) (page 16)
- 3) X. Executive Summary (Features and Characteristics of the Device Compared to the Predicate Device) (page 23)
- 4) XII. Substantial Equivalence Discussion (Comparison of Features and Characteristics) (page 30)
In accordance to Section 745A (b) of the Federal Food, Drug, and Cosmetic Act (FD&C Act), added by section 1136 of the Food and Drug Administration Safety and Innovation Act (FDASIA) (Pub. L. 112-144), an eCopy in the form of a CD is submitted with the paper copy,
If you have any questions and/or concerns, please call or e-mail me using the information below.
Sincerely,
Helen Ragn
Helen Ragus Regulatory Specialist (805) 346-3700, X2932 hragus@denmat.com
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a logo for DenMat. The logo consists of the letters 'dm' in a stylized font, with the 'd' being smaller and to the left of the 'm'. Below the letters 'dm' is the word 'DenMat' in a smaller, sans-serif font. The logo is black and white.
K140537/5001
#### 510(k) COVER LETTER
ಲ್ಲಿ
U.S. Food and Drug Administration Center for Devices and Radiological Heath Document Control Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
I·DA CDRH DMC MAY 1 3 2014 Received
Reference: K140537/S001
Type of Submission: Trade Name: Common Name: Classification Regulation: Device Class: Panel: Product Code:
510(k) Submitter: Establishment Registration #: Contact Person:
Phone Number:
DenMat Holdings, LLC 2018957 Helen Ragus Regulatory Specialist 805 346-3700, X2932
Dental Bonding Agent
Dear Sir or Madam,
In Reference to 510(k) number K140537/S001, please find the revised eCopy with the corrected PDF naming convention.
The eCopy is an exact duplicate of the paper copy except for the revised cover letter.
of the states and the became comments of any the section of the count
Traditional
tenure@4G .
872.3200
Dental KLE
II
If you have any questions and/or concerns, please call or e-mail me using the information below.
Sincerely,
Helen Rogers
Helen Ragus Regulatory Specialist (805) 346-3700, X2932 hragus@denmat.com
1017 W. Central Avenue, Lompoc, CA 93436 denmat.com
P: 805-346-3700 F: 805-347-7933
and and and and the States of Cameral
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.