← Product Code [BWK](/productcode/BWK) · K140530

# Electro Auricular Device (K140530)

_Navigant Consulting, Inc. · BWK · Oct 2, 2014 · NE · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140530

## Device Facts

- **Applicant:** Navigant Consulting, Inc.
- **Product Code:** [BWK](/productcode/BWK.md)
- **Decision Date:** Oct 2, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** NE
- **Attributes:** Therapeutic

## Indications for Use

The EAD is an electro acupuncture device for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

## Device Story

Battery-operated, single-use electro-acupuncture device; provides auricular point nerve stimulation for pain therapy. Input: preprogrammed electrical pulse parameters. Transformation: microprocessor-controlled generation of low-frequency electrical pulses transmitted via stainless steel wires to needle arrays (titanium needles) placed behind the ear. Output: periodic electrical stimulation (2 hours on/2 hours rest) for up to 120 hours. Used in clinics or at home; applied by qualified acupuncture practitioners. Output informs pain management therapy; provides continuous stimulation over several days to potentially benefit patients with acute or chronic pain.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993), electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), and software verification/validation.

## Technological Characteristics

Battery-powered (3V CR1220 Lithium) electro-acupuncture stimulator. Output: max 3.2V, 1-10Hz modulating frequency, 1ms pulse width, bipolar. Dimensions: 36x16x7mm; weight 4g. Needles: 0.5mm width x 2mm length. Materials: stainless steel wires, titanium needles. Connectivity: standalone. Sterilization: VDMAX25 method (ISO 11137-2). Software: microprocessor-controlled, minor level of concern.

## Predicate Devices

- P.Stim System ([K050123](/device/K050123.md))

## Reference Devices

- E-Pulse ([K091875](/device/K091875.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are stacked slightly on top of each other, creating a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 2, 2014

Navigant Consulting, Inc. Colleen Hittle Managing Director 30 S. Wacker Drive, Suite 3100 Chicago, Illinois 60606

Re: K140530

Trade/Device Name: Electro Auricular Device Regulation Name: Electro Acupuncture Stimulator Regulatory Class: Unclassified Product Code: BWK Dated: August 29, 2014 Received: September 2, 2014

Dear Ms. Colleen Hittle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K140530

Device Name EAD (electro auricular device)

Indications for Use (Describe)

The EAD is an electro acupuncture device for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Felipe Aquel Date: 2014.10.02 20:47:21 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### Submitter & Contact Information

Colleen Hittle, RAC Managing Director Navigant Consulting, Inc. 30 S. Wacker Drive Suite 3100 Chicago IL 60606 Phone: (317) 228-8730 Fax: (317) 228-8701 colleen.hittle@navigant.com

Date: June 23, 2014

Manufacturer: Key Electronics 2533 Centennial Blvd Jeffersonville, IN 47130

Trade Name: EAD (electro auricular device)

Common Name: electro acupuncture device

Classification Name(s): BWK - stimulator, electro-acupuncture

Classification Number: Unclassified

#### Predicate Device(s)

|           | 510(k) Number | Device Name   | Submitter Name       |
|-----------|---------------|---------------|----------------------|
| Primary   | K050123       | P.Stim System | Neuroscience Therapy |
| Predicate |               |               | Corporation          |
| Reference | K091875       | E-Pulse       | Medevice Corporation |
| Device    |               |               |                      |

{4}------------------------------------------------

# Device Description

The EAD system is a battery-operated, single-use device that has a preprogrammed frequency, pulse and duration for the stimulation of auricular point nerve stimulation for pain. The device power supply connects via four (4) stainless steel wires, sheathed in a plastic over-molding, to three (3) needle arrays comprised of four (4) needles each and one (1) needle array comprised of only 1 needle.

The device is powered by one 3 Volt lithium ion battery.

The device modulates a duty cycle between 2 hours on and 2 hours at rest. The maximum performance time frame is 5 days or 120 hours (5 days X 24 hours). Weight is 4 grams including batteries. The power supply dimensions are 36 mm x 17 mm x 7 mm. The needle dimensions are 0.5 mm width x 2 mm length.

# Intended Use(s)

The EAD is an electro acupuncture device for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

## Technological Characteristics

Appliance: Electro-acupuncture Device Type description: EAD Power supply: 1 x 3V battery (Type CR1220 Lithium) Output: (Load impedance range 1k-10kQ) max. 3.2V, Impulse interval 1000ms, impulse width 1ms, 'ims I bipolar), max possible total duration of treatment 5x 24 hours Protection level: IP20 Type: B Duty type: approx. 2h duty I 2h at rest (periodic duty) Weight incl. battery: 4g Dimensions: 36 x 16 x 7 mm Needle Dimensions: 0.5 mm width x 2 mm length

{5}------------------------------------------------

| Device Name        | EAD                                                                                                                                                                                                                                                              | E-Pulse<br>(Reference Device)                                                                                                                                                                                                                                                                           | P.Stim System<br>(Primary Predicate)                                                                                                                                                                                                                                                                          |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 501(k) Number      | This Submission                                                                                                                                                                                                                                                  | K091875                                                                                                                                                                                                                                                                                                 | K050123                                                                                                                                                                                                                                                                                                       |
| Indication for Use | The EAD is an<br>electro acupuncture<br>device for use in the<br>practice of<br>acupuncture by<br>qualified<br>practitioners of<br>acupuncture as<br>determined by the<br>states.                                                                                | The E-Pulse is an<br>electro acupuncture<br>device for use in the<br>practice of<br>acupuncture by<br>qualified<br>practitioners of<br>acupuncture as<br>determined by the<br>states.                                                                                                                   | The P-Stim is an<br>electro acupuncture<br>device for use in the<br>practice of<br>acupuncture by<br>qualified<br>practitioners of<br>acupuncture as<br>determined by the<br>states.                                                                                                                          |
| Device Description | Battery powered<br>device generates<br>low frequency and<br>continual electrical<br>pulse which are<br>transmitted to new<br>endings of the ear. It<br>allows continued<br>therapy over several<br>days.<br>The device is<br>controlled by a<br>micro processor. | Battery powered<br>unit designed to<br>administer auricular<br>point nerve<br>stimulation<br>treatment for pain<br>therapy over a 96-<br>hour period via<br>electrical pulsing.<br>The device is on for<br>3 hours and then off<br>for 3 hours.<br>The device is<br>controlled by a<br>micro processor. | A micro stimulation<br>appliance for pain<br>therapy.<br>The device<br>generates low<br>frequency and<br>continual electrical<br>pulse which are<br>transmitted to new<br>endings of the ear. It<br>allows continued<br>therapy over several<br>days.<br>The device is<br>controlled by a<br>micro processor. |
| Target Population  | Patients with acute<br>and chronic pain                                                                                                                                                                                                                          | Patients with acute<br>and chronic pain                                                                                                                                                                                                                                                                 | Patients with acute<br>and chronic pain                                                                                                                                                                                                                                                                       |
| Human Factors      | To be applied by a<br>qualified<br>practitioner of<br>acupuncture                                                                                                                                                                                                | To be applied by a<br>qualified<br>practitioner of<br>acupuncture                                                                                                                                                                                                                                       | To be applied by a<br>qualified<br>practitioner of<br>acupuncture                                                                                                                                                                                                                                             |
| Where Used         | At the clinic and at<br>home                                                                                                                                                                                                                                     | At the clinic and at<br>home                                                                                                                                                                                                                                                                            | At the clinic and at<br>home                                                                                                                                                                                                                                                                                  |
| Software Based     | Yes                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                           |
| Performance        | 2 hours on/2 hours<br>off; pulses with<br>modulating<br>frequency (1 to 10<br>Hz)                                                                                                                                                                                | 3 hours on/3 hours<br>off; pulse<br>monophasic at 1 Hz                                                                                                                                                                                                                                                  | 3 hours on/3 hours<br>off; pulse<br>monophasic at 1 Hz                                                                                                                                                                                                                                                        |
| Power Source       | Lithium ion battery                                                                                                                                                                                                                                              | Zinc air battery                                                                                                                                                                                                                                                                                        | Zinc air battery                                                                                                                                                                                                                                                                                              |
| Duration           | Up to 120 hours                                                                                                                                                                                                                                                  | 96 hours                                                                                                                                                                                                                                                                                                | 96 hours                                                                                                                                                                                                                                                                                                      |

## Comparison Table: Applicant Device vs. Predicate Devices

{6}------------------------------------------------

| Generator | Attached behind patient ear | Attached behind patient ear            | Attached behind patient ear            |
|-----------|-----------------------------|----------------------------------------|----------------------------------------|
| Leads     | Four electrode needle leads | Three leads that can attach to needles | Three leads that can attach to needles |
| Needles   | Titanium straight shaft     | Titanium hook shaft                    | Titanium straight shaft                |
| Shape     | Elliptical                  | Round                                  | Elliptical                             |

# Non-Clinical Performance Data

## Biocompatibility Testing

The biocompatibility evaluation for the adhesive for the EAD device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, " May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

- . Cytotoxicity
- Sensitization
- Irritation

The EAD adhesive is considered tissue contacting for a duration of less than 30 days. The other patient contacting materials contained in the subject device are unchanged from the P-Stim predicate device.

### Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the EAD device. The system complies with the IEC 60601-1 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.

### Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "minor" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or operator.

### Sterility/Shelf Life

Sterilization Validation was conducted on the EAD device needles and wiring harness using the VDMAX25 method according to ISO 11137-2 and ISO 11737-2. Performance and stability of the EAD packaging was validated in accordance with ISO 11607-1 and Accelerated aging of the EAD was performed in accordance with ASTM F1980-07. Packaging qualifications are according to ISO 11607-1 and machine qualifications for the sealing process are according to ISO 11607-2.

{7}------------------------------------------------

# Conclusion

The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the EAD device should perform as well as the predicate device in the specified use conditions.

---

**Source:** [https://fda-staging.innolitics.com/device/K140530](https://fda-staging.innolitics.com/device/K140530)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).