← Product Code [GEI](/productcode/GEI) · K140521

# TOPAZ EZ MICRODEBRIDER COBLATION WAND WITH INTEGRATED FINGER SWITCH (K140521)

_ArthroCare Corporation · GEI · Jun 10, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140521

## Device Facts

- **Applicant:** ArthroCare Corporation
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jun 10, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch is indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: Procedure: Fasciotomy, Body Structure: Foot; Procedure: Synovectomy, Body Structure: Foot; Procedure: Tendonotomy, Body Structure: Knee, Wrist, Elbow, Ankle, Shoulder, Foot; Procedure: Rotator Cuff Tendonotomy, Body Structure: Shoulder; Procedure: Capsulotomy, Body Structure: Foot.

## Device Story

The TOPAZ® EZ Microdebrider Coblation® Wand is a bipolar, high-frequency electrosurgical device used in arthroscopic and orthopedic procedures. It consists of a tungsten electrode tip, alumina ceramic spacer, stainless steel shaft, saline irrigation tubing, and a handle with an integrated finger switch. The device connects to an ArthroCare Quantum Controller. During operation, the device delivers radiofrequency energy to saline to create a plasma field, enabling precise soft tissue ablation, resection, and hemostasis. It is operated by a physician in a clinical or surgical setting. The integrated finger switch allows the surgeon to activate the wand directly. By providing controlled tissue removal and coagulation, the device assists in procedures such as fasciotomy, synovectomy, and tendonotomy, potentially reducing surgical time and improving procedural outcomes.

## Clinical Evidence

No clinical data are included in this submission. Substantial equivalence is supported by non-clinical performance testing, including visual and dimensional inspection, Hi-Pot testing, electrical impedance, saline flow rate, ablation set points, finger switch actuation, shaft stiffness, axial compression force, biocompatibility (cytotoxicity, sensitization, maximization), and electrical safety testing per IEC 60601-2-2:2009.

## Technological Characteristics

Bipolar high-frequency electrosurgical device. Materials: tungsten electrode, alumina ceramic spacer, stainless steel shaft, polyester shaft insulation, Pebax saline delivery and electrode insulation, polycarbonate handle. Dimensions: 1.57-inch shaft length, 0.042-inch OD. Energy source: radiofrequency via ArthroCare Quantum Controller. Connectivity: wired connection to controller. Sterilization: sterile. Software: none.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- TOPAZ® Microdebrider Coblation® Wand with Integrated Finger Switch ([K080282](/device/K080282.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN | 0 2014

# K140521

# 510(k) Summary

# ArthroCare® Corporation TOPAZ® EZ Microdebrider Coblation® Wand With Integrated Finger Switch

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

# General Information

| Submitter Name: | ArthroCare Corporation                                                                                |
|-----------------|-------------------------------------------------------------------------------------------------------|
| Address         | 7000 West William Cannon Drive<br>Austin, TX 78735                                                    |
| Contact Person: | Mitchell A. Dhority<br>Vice President, Regulatory Affairs<br>Phone: 512-358-5995<br>Fax: 512-895-1489 |
| Date Prepared:  | February 27, 2014                                                                                     |

## Device Name

| Proprietary Name:  | TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch |
|--------------------|-----------------------------------------------------------------------|
| Regulation Name:   | Electrosurgical cutting and coagulation device and accessories        |
| Regulatory Class:  | I I                                                                   |
| Product Code:      | GEI                                                                   |
| Regulation Number: | 21 CFR 878.4400                                                       |

#### Predicate Device

TOPAZ® Microdebrider Coblation® Wand with Integrated Finger Switch included in: ArthroCare Topaz Wand, K080282 (cleared February 15, 2008)

#### Description

The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch (Topaz EZ IFS) is a bipolar, sterile, high frequency electrosurgical device, which consists of a distal electrode tip composed of tungsten, an alumina ceramic spacer, stainless steel shaft, saline irrigation tubing and a molded handle with an integrated single finger switch used to activate the Wand. The handle connects proximally to the radiofrequency Controller via an electrical cable. The Topaz EZ IFS is only compatible with the ArthroCare Quantum Controllers.

{1}------------------------------------------------

## .. ......... Intended Use/Indications for Use

The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch is indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

| Procedure                | Body Structure as Described Below         |
|--------------------------|-------------------------------------------|
| Fasciotomy               | Foot                                      |
| Synovectomy              | Foot                                      |
| Tendonotomy              | Knee, Wrist, Elbow, Ankle, Shoulder, Foot |
| Rotator Cuff Tendonotomy | Shoulder                                  |
| Capsulotomy              | Foot                                      |

## Summary of Technological Characteristics

This Special 510(k) proposes modifications to the materials, dimensional specifications, and performance specifications of the ArthroCare Topaz Wand cleared under K080282. The indications for use, fundamental scientific technology, principle of operation, and sterilization methodology remain the same as in the previously cleared 510(k).

The following table summarizes the technological differences between the subject and predicate devices and describes the rationale for the change.

| Parameter                                               | Predicate Device:<br>Topaz Wand<br>(K080282)                                  | Subject Device:<br>TOPAZ EZ IFS<br>Wand                         | Rationale for Change                                                                                                            |
|---------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Spacer Material                                         | Thermal-Set Silicone<br>Elastomer (Silicone)                                  | Alumina                                                         | Facilitates consistency in<br>manufacturability                                                                                 |
| Shaft Length                                            | 2.91 inches (74mm)                                                            | 1.57 inches (40mm)                                              | Length shortened based on user<br>feedback (decreases the<br>working distance between the<br>user and the surgical site)        |
| External Shaft<br>Insulation<br>Material                | Nylon                                                                         | Polyester                                                       | Material now used to insulate<br>the shaft as opposed to also<br>serving as the saline sheath                                   |
| Saline Delivery<br>(irrigant<br>contacting<br>material) | Nylon and stainless<br>steel                                                  | Pebax and stainless<br>steel                                    | Provides saline delivery close to<br>the tip of the wand and greater<br>visibility for the surgeon                              |
| Shaft OD                                                | $0.032 \pm .005"$<br>(.812± .13mm)                                            | $0.042 \pm .005"$ (1.067<br>±.13mm)                             | Increased stiffness of the shaft                                                                                                |
| Active Electrode<br>Insulation<br>Material              | Polyimide                                                                     | Pebax                                                           | Manufacturability                                                                                                               |
| Shaft Y -<br>Connector and<br>Bushing                   | Separate components<br>comprised of PVC and<br>polycarbonate,<br>respectively | Single integrated<br>component<br>comprised of<br>polycarbonate | Cost savings and process<br>improvement. Adhesive was<br>increased to prevent the shaft<br>from sliding within Y-<br>connector. |
| Active/Return<br>Electrode<br>Insulator                 | N/A                                                                           | Clear Pebax                                                     | Insulates electrodes; centers<br>electrode components during<br>manufacturing                                                   |
| Internal                                                | N/A                                                                           | Nylon                                                           | Holds cured adhesive to                                                                                                         |

{2}------------------------------------------------

| Parameter               | Predicate Device:<br>Topaz Wand<br>(K080282) | Subject Device:<br>TOPAZ EZ IFS<br>Wand | Rationale for Change                                     |
|-------------------------|----------------------------------------------|-----------------------------------------|----------------------------------------------------------|
| Reinforcement<br>Tubing |                                              |                                         | facilitate bonding of electrodes<br>during manufacturing |
| Labeling                | Topaz IFS                                    | Topaz EZ IFS                            | Proprietary name change.<br>Instructions strengthened.   |

#### Summary of Non-Clinical Performance Data

Non-clinical performance testing verifies that the Topaz EZ IFS performs as intended and that no additional risks or hazards have been identified as a result of the changes described in this Special 510(k). As such, results from this testing support the determination that the Topaz EZ IFS is substantially equivalent to the predicate device. The following table lists the non-clinical testing performed and the results obtained in support of substantial equivalence.

| Testing Type                 | Test Description                                                                                                                                                                                                                                                             | Result Supporting Substantial<br>Equivalence                                                                                                                                                                                                                       |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Verification Testing         | • Visual Inspection<br>• Dimensional Inspection<br>• Hi-Pot Testing<br>• Electrical Impedance<br>• Saline Flow Rate<br>• Controller Ablation Set Points<br>• Finger Switch Actuation Testing<br>• Ablation Testing<br>• Shaft Stiffness<br>• Axial Compression Force Testing | Both the Topaz EZ IFS and<br>predicate device have substantially<br>equivalent testing specifications<br>and both performed within<br>acceptance criteria. These results<br>support that the Topaz EZ IFS and<br>predicate device are substantially<br>equivalent. |
| Biocompatibility<br>Testing  | • Cytotoxicity<br>• Sensitization<br>• Maximization                                                                                                                                                                                                                          | The Topaz EZ IFS and the<br>predicate device are biocompatible.<br>These results support that the<br>Topaz EZ IFS and predicate device<br>are substantially equivalent.                                                                                            |
| Electrical Safety<br>Testing | Testing in accordance with IEC<br>60601-2-2:2009.                                                                                                                                                                                                                            | The Topaz EZ IFS and the<br>predicate met all acceptance<br>criteria in accordance with IEC<br>60601-2-2: 2009. These results<br>support that the Topaz EZ IFS and<br>predicate device are substantially<br>equivalent.                                            |

## Clinical Data

No clinical data are included in this submission

#### Summary

The Topaz EZ IFS is substantially equivalent to the predicate Topaz Wand (K080282). Both the subject and predicate devices have the same indications for use and utilize the same fundamental scientific technology. All product performance testing demonstrates that the Topaz EZ IFS performs as intended and has acceptable mechanical properties when used in accordance with its labeling. The minor differences between the Topaz EZ IFS and the predicate device do not raise any new questions of safety or effectiveness.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 10, 2014

ArthroCare Coporation Mr. Mitchell A. Dhority Vice President, Regulatory Affairs 7000 West William Cannon Drive Austin, Texas 78735

Re: K140521 Trade/Device Name: TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: May 13, 2014 Received: May 14, 2014

Dear Mr. Dhority:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

Image /page/3/Picture/10 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right.

{4}------------------------------------------------

Page 2 - Mr. Mitchell A. Dhority

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below,

510(k) Number (if known)

K140521

Device Name

TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch

T

Indications for Use (Describe)

The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch is indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic procedures:

Body Structure Procedure Fasciotomy Foot Synovectomy Foot Knee, Wrist, Elbow, Ankle, Shoulder, Foot Tendonotomy Rotator Cuff Tendonotomy Shoulder Capsulotomy Foot

#### Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

CONSTITUTION CONTROLLER COLLECTION FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR MONDENT OF CONSULTION

Concurrence of Center for Davices and Radiological Health (CDRH) (Signalure)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### ·DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Offica of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

---

**Source:** [https://fda-staging.innolitics.com/device/K140521](https://fda-staging.innolitics.com/device/K140521)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).